STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 39,201–39,300 of 47,868
Gerke, O., Vilstrup, M. H., Halekoh, U., Hildebrandt, M. G., & Høilund‐Carlsen, P. F. (2017) · EJNMMI Research · doi:10.1186/s13550-017-0328-6 · Cited by 4
Packham, T., MacDermid, J. C., Michlovitz, S. L., Cup, E. H. C., & van de Ven-Stevens, L. A. W. (2017) · Journal of Hand Therapy · doi:10.1016/j.jht.2017.03.003 · Cited by 4
Comparative Diagnostic Performance of the Granulocyte and Neutrophil Counts
Pether, N. S., Brothwood, J. L., van Berkel, C., Dunwoodie, E., Blake, R. L., Price, C. P., Jones, R., Baker, K. S., & Hall, G. D. (2017) · Practical Laboratory Medicine · doi:10.1016/j.plabm.2017.10.001 · Cited by 5
Kakinuma, R., Muramatsu, Y., Yamamichi, J., Gomi, S., Oubel, E., & Moriyama, N. (2018) · Journal of Thoracic Disease · doi:10.21037/jtd.2017.11.142 · Cited by 3
Engineering of a miniaturized, robotic clinical laboratory
Nourse, M. B., Engel, K., Anekal, S. G., Bailey, J. A., Bhatta, P., Bhave, D. P., Chandrasekaran, S., Chen, Y., Chow, S., Das, U., et al. (2017) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10084 · Cited by 4
Teigen, L. M., DiCecco, S. R., Vock, D. M., Vierow, K. L., Andrews, L., Hammel, K. D., Sfeir, J. G., Miles, J. M., & Hurley, D. L. (2018) · Clinical Nutrition ESPEN · doi:10.1016/j.clnesp.2018.01.001 · Cited by 3
Panaretos, D., Tzavelas, G., Vamvakari, M. G., & Panagiotakos, D. B. (2018) · Communications in Statistics - Simulation and Computation · doi:10.1080/03610918.2018.1435803 · Cited by 3
Kyriakou, E. S., Nearchakos, N., Bonovas, S., Makri, E., Pantavou, K. G., Nikolopoulos, G. K., Kottaridi, C., Gialeraki, A. E., Douramani, P., Taichert, M., et al. (2018) · Transfusion and Apheresis Science · doi:10.1016/j.transci.2018.06.002 · Cited by 3
Comparison of Non-contact Infrared Temperature with Digital Axillary Temperature
Abraham, V., Johnson, E., & Deep, K. (2018) · Indian Pediatrics · doi:10.1007/s13312-018-1308-2 · Cited by 3
Foot and Soccer Referees’: A Pilot Study Searching “Performance” Throughout Prevention
Gianturco, L., Bodini, B. D., Gianturco, V., Pregliasco, F. E., Cascio, M., Serafin, A., & Turiel, M. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01009 · Cited by 4
Zhu, Y., Hernandez, L. M., Dong, Y., Himes, J. H., Caulfield, L. E., Kerver, J. M., Arab, L., Voss, P., Hirschfeld, S., & Forman, M. R. (2018) · Public Health Nutrition · doi:10.1017/s1368980018002549 · Cited by 3
Kerschberger, B., Mpala, Q., Uribe, P. A. D., Maphalala, G. P., de la Tour, R., Kalombola, S., Bekele, A., Chawinga, T., Mliba, M., Ntshalintshali, N., et al. (2018) · BMC Infectious Diseases · doi:10.1186/s12879-018-3474-1 · Cited by 3
Guerville, F., Baudron, C. R., Duc, S., Salles, N., Rainfray, M., & Bourdel‐Marchasson, I. (2018) · Drugs & Aging · doi:10.1007/s40266-018-0618-3 · Cited by 4
Prognostic Comparison of the Estimations of Renal Function in Patients With Acute Heart Failure
Cheng, Y.-L., Sung, S., Cheng, H., Huang, J.-T., Guo, C., Hsu, P., Yu, W., & Chen, C. (2019) · Circulation Journal · doi:10.1253/circj.cj-18-1013 · Cited by 3
Inns, T., Curtis, D., Crook, P., Vivancos, R., Gardiner, D., McCarthy, N. D., & Mook, P. (2019) · Epidemiology and Infection · doi:10.1017/s0950268819000219 · Cited by 3
Objective measurement of forward-scattered light in the human eye: An electrophysiological approach
Solf, B., Schramm, S., Link, D., & Klee, S. (2019) · PLoS ONE · doi:10.1371/journal.pone.0214850 · Cited by 4
Reliability of Maximal Voluntary Isometric Contraction of Ankle Dorsiflexion in Male Subjects
Andreis, F. R., Favretto, M. A., Cossul, S., Barbetta, P. A., & Marques, J. L. B. (2019) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-13-2119-1_55 · Cited by 4
Hope, D., French, J., Pizzari, T., Hoy, G., & Barwood, S. A. (2019) · Orthopaedic Journal of Sports Medicine · doi:10.1177/2325967119851084 · Cited by 3
Noh, T., Griffith, B. D., Snyder, J., Zhou, Y., Poisson, L., & Lee, I. Y. (2019) · Lasers in Surgery and Medicine · doi:10.1002/lsm.23111 · Cited by 3
Comparison of Mass Size Measurements: Synthesized Mammography Versus Full-Field Digital Mammography
Şendur, H. N., Cerit, M. N., Gültekin, S., Cındıl, E., & Kılıç, P. (2019) · Academic Radiology · doi:10.1016/j.acra.2019.08.016 · Cited by 3
Evaluation of a developed multichannel R-R interval telemeter and garment-type electrode
Chihara, Y., & Yamakawa, T. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857402 · Cited by 3
Myhr, K. A., Pedersen, F. H. G., Kristensen, C. B., Køber, L., Hassager, C., & Møgelvang, R. (2019) · Echocardiography · doi:10.1111/echo.14529 · Cited by 3
Geometric validation of a computer simulator used in radiography education
Cosson, P., & Lu, Z. (2020) · BJR|Open · doi:10.1259/bjro.20190027 · Cited by 6
’n Model om Nylkrokodileiers se volumes mee te skat nadat hulle uitgebroei het
Nöthling, J. O., Nöthling, J. A., & Myburgh, J. G. (2019) · Suid-Afrikaanse Tydskrif vir Natuurwetenskap en Tegnologie · doi:10.36303/satnt.2019.38.1.680 · Cited by 3
Predictive anthropometric models of total and truncal body fat in Chilean children
Vásquez, F., Corvalán, C., Gahagan, S., Uauy, R. D., & Kain, J. (2020) · Nutrition · doi:10.1016/j.nut.2020.110803 · Cited by 3
Kawai, H. F., Kawakami, T., Tsujimoto, M., Fukushima, A., Isogai, S., Ishizawa, H., Nagano, H., Negi, T., Tochii, D., Tochii, S., et al. (2020) · Institutional Repositories DataBase (IRDB) · doi:10.20407/fmj.2019-012 · Cited by 5
Mitrev, L. V., van Helmond, N., Kaddissi, G., Awad, A. S., Patel, K., Aji, J. I., Ogbara, J., Rafeq, Z., Nagubandi, V., Orr, D., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-64753-3 · Cited by 3
Comparing Image Segmentation Techniques for Determining 3D Orbital Cavernous Hemangioma Size on MRI
Boparai, R. S., Maeng, M. M., Dunbar, K. E., Godfrey, K. J., Tooley, A. A., Maher, M. D., & Kazim, M. (2020) · Ophthalmic Plastic and Reconstructive Surgery · doi:10.1097/iop.0000000000001651 · Cited by 3
Erdélyi, T., Lázár, Z., Odler, B., Tamási, L., & Müller, V. (2020) · Journal of Aerosol Medicine and Pulmonary Drug Delivery · doi:10.1089/jamp.2020.1594 · Cited by 3
Russell, C., Carthy, M. M., Cappelleri, J. C., & Wong, S. (2020) · Therapeutic Innovation & Regulatory Science · doi:10.1007/s43441-020-00188-2 · Cited by 3
Ayabe, M., & Kumahara, H. (2020) · Gait & Posture · doi:10.1016/j.gaitpost.2020.06.032 · Cited by 3
Validation of a Bilateral Simultaneous Computer-Based Tympanometer
Ramatsoma, H., & Koekemoer, D. (2020) · American Journal of Audiology · doi:10.1044/2020_aja-20-00013 · Cited by 3
Chiu, E., Lee, S., Lai, K.-Y., & Gu, F.-Y. (2020) · American Journal of Occupational Therapy · doi:10.5014/ajot.2020.040154 · Cited by 3
Physical Activity Evaluation Using a Voice Recognition App: Development and Validation Study
Namba, H. (2021) · JMIR Biomedical Engineering · doi:10.2196/19088 · Cited by 3
Franz, S., Skopp, G., Dame, T., & Mußhoff, F. (2021) · Drug Testing and Analysis · doi:10.1002/dta.3003 · Cited by 3
Campbell, P., Edgar, D. F., & Shah, R. (2021) · Clinical and Experimental Optometry · doi:10.1080/08164622.2021.1878831 · Cited by 3
Ma, Y., Norton, D., Hulle, C. A. V., Chappell, R. J., Lazar, K. K., Jonaitis, E. M., Koscik, R. L., Clark, L. R., Krause, R. A., Andréasson, U., et al. (2021) · Alzheimer s & Dementia Diagnosis Assessment & Disease Monitoring · doi:10.1002/dad2.12194 · Cited by 4
Validation and Reliability of the Japanese Version of the Modified Parkinson Activity Scale (M-PAS)
Taniguchi, S., Nakata, Y., Inoue, M. B., & Marumoto, K. (2021) · Progress in Rehabilitation Medicine · doi:10.2490/prm.20210051 · Cited by 3
van den Berg, B., Vanwinsen, L., Pezzali, G., & Buitenweg, J. R. (2022) · Attention Perception & Psychophysics · doi:10.3758/s13414-022-02484-5 · Cited by 3
Oe, B., de Fijter, C. W. H., Geers, A. B., Vos, P. F., Donker, A. J. M., & VRIES, P. M. J. M. D. (2000) · The International Journal of Artificial Organs · doi:10.1177/039139880002300305 · Cited by 3
Vilete, L. M. P. (2002) · Arca - Repositório Institucional da Fiocruz · Cited by 5
Probst, Y. (2006) · Research Online (University of Wollongong) · Cited by 3
Unidimensionality, Agreement and Concordance Probability
Jin, Z., & Mesbah, M. (2013) · Wiley eBooks · doi:10.1002/9781118826805.ch1 · Cited by 3
Physiological and clinical aspects of uterine contractility during the postpartum period in cows
Bajcsy, Á. C. (2005) · Utrecht University Repository (Utrecht University) · Cited by 3
Virtual three-dimensional analysis of digitized dental impressions and stone replicas
Persson, A. (2008) · Figshare · Cited by 3
A Novel Automated Left Ventricle Segmentation Routine
Liu, N., Trakic, A., Liu, F., Appleton, B., Wilson, S. J., Slaughter, R., Strugnell, W., & Crozier, . (2005) · Cited by 3
Exercise and dietary behaviour change in a sample of midlife Australian women
Anderson, R. L. (2008) · Faculty of Health · Cited by 6
Characterization of normal facial features and their association with genes
Toma, A. M. (2014) · ORCA Online Research @Cardiff (Cardiff University) · Cited by 6
Dosimetry and Optimisation in High Dose Fluoroscopic and Fluorographic Procedures
Morrell, R. (2006) · Nottingham ePrints (University of Nottingham) · Cited by 3
Chouliara, N. (2013) · Nottingham ePrints (University of Nottingham) · Cited by 6
Evaluation of Hematology Analyzers
Briggs, C. (2012) · Wiley eBooks · doi:10.1002/9781444398595.ch9 · Cited by 3
Prediction of Pulmonary Artery Pressure
Fisher, M. R., & Hassoun, P. M. (2005) · CHEST Journal · doi:10.1016/s0012-3692(15)37990-3 · Cited by 3
Gene Expression: Microarray Data Analysis
Pevsner, J. (2009) · Wiley eBooks · doi:10.1002/9780470451496.ch9 · Cited by 5
Psychophysical capacity in non-specific chronic low back pain
Hodselmans, A. P. (2009) · University of Groningen research database (University of Groningen / Centre for Information Technology) · Cited by 3
A Time Motion Analysis of Elite Women's Hockey - Implications for Fitness Assessment and Training
Holmes, L. (2011) · Coventry University Open Collections (Coventry university) · Cited by 3
Pre-Performance Practices: Breathing Imagery and Warm-Up For Singers
Moorcroft, L. (2011) · The Sydney eScholarship Repository (The University of Sydney) · Cited by 7
Bonomi, A. (2010) · Research Publications (Maastricht University) · doi:10.26481/dis.20101208abo · Cited by 3
Bone mineral density and body composition of children and adolescents in health and disease
Boot, A. M. (1997) · RePub (Erasmus University Rotterdam) · Cited by 6
Neuerer, F. (2005) · Electronic Theses of LMU Munich (Ludwig-Maximilians-Universität München) · doi:10.5282/edoc.3339 · Cited by 3
Evaluation of physical fitness in relation to performance and injury severity in contemporary dance
Angioi, M. (2010) · Wolverhampton Intellectual Repository and E-Theses (University of Wolverhampton) · Cited by 5
Applications of hybrid diffuse optics for clinical management of adults after brain injury
Kim, M. N. (2013) · ScholarlyCommons (University of Pennsylvania) · Cited by 7
Toward Accurate Small Animal Dosimetry and Irradiator Quality Assurance
Rodrigues, A. (2012) · DukeSpace (Duke University) · Cited by 3
Comparison between cell motion analyser, autosperm, and conventional semen analysis
Kamtchouing, P., Vermeulen, L., Schoonjans, F., & Comhaire, F. H. (2009) · Andrologia · doi:10.1111/j.1439-0272.1993.tb02740.x · Cited by 3
Iosa, M., Morone, G., & Paolucci, S. (2011) · Stroke · doi:10.1161/strokeaha.111.629055 · Cited by 3
Choosing the Stimulation Rate in Patients with Intermittent Bradycardia and a Single Lead Pacemaker
Rosenqvist, M., Edhag, O., Lagergren, H., Nordlander, R., & Vallin, H. (1993) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1993.tb01648.x · Cited by 3
Myocardial Systolic Strain in Normal Children Using a Tissue Tracking System
Sakata, M., Mori, K. J., Hayabuchi, Y., Suzuki, M., Nakagawa, R., Kagami, S., & Baba, H. (2006) · Journal of Echocardiography · doi:10.2303/jecho.4.19 · Cited by 3
Lyles, R. H., & Chambless, L. E. (1995) · Statistics in Medicine · doi:10.1002/sim.4780141509 · Cited by 3
Barry, B., Mallick, A., Vucevic, M., & Bodenham, A. (1997) · Critical Care · doi:10.1186/cc92 · Cited by 3
Sierra, À., Bragulat, E., Gómez-Angelats, E., Sierra, C., Coca, A., & Montañá, F. (2002) · Hipertensión y Riesgo Vascular · doi:10.1016/s1889-1837(02)71300-7 · Cited by 3
Which is the Optimal Testing Method for Identifying an AV Delay that Allows Intrinsic Conduction?
Eguía, L. E., Pinski, S. L., Haw, J. M., & Trohman, R. G. (2000) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2000.tb07013.x · Cited by 3
Otsuki, T., Sugawara, J., Tanabe, T., Maeda, S., Ajisaka, R., & Matsuda, M. (2003) · International Journal of Sport and Health Science · doi:10.5432/ijshs.1.142 · Cited by 3
Sheynberg, B. V., Jang, I., Han, R. O., Sabatine, M. S., Brown, D., & Dinsmore, R. E. (2002) · Catheterization and Cardiovascular Interventions · doi:10.1002/ccd.10156 · Cited by 3
Challenges in evaluating summaries of short stories
Kazantseva, A., & Śzpakowicz, S. (2006) · doi:10.3115/1654679.1654682 · Cited by 3
Huang, J., Savini, G., Feng, Y., & Wang, Q. (2012) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.11-9416 · Cited by 3
Krüger, D., Stierle, U., Schmücker, G., Sheikhzadeh, A., & Giannitsis, E. (1999) · Zeitschrift für Kardiologie · doi:10.1007/s003920050330 · Cited by 3
Validierung eines ambulanten Screeninggerätes für schlafbezogene Atmungsstörungen: APNOESCREEN I®
Cohrs, S., Herrendorf, G., Kinkelbur, J., Titkemeyer, M., & Hajak, G. (1998) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-998-0002-3 · Cited by 3
Ethanol elimination in the anesthetized dog: Intragastric versus intravenous administration
Tzamaloukas, A. H., Jackson, J. E., Hermann, J. J., Gallegos, J. C., Long, D. A., & McLane, M. M. (1988) · Alcohol · doi:10.1016/0741-8329(88)90006-7 · Cited by 3
Mohan, J. C., & Bhargava, M. (1996) · International Journal of Cardiology · doi:10.1016/0167-5273(96)02676-9 · Cited by 3
Reproducibility and Reversibility of Tidal Forced Expirations
Lum, S., & Stocks, J. (2003) · Respiration · doi:10.1159/000074219 · Cited by 3
Lopez, M. L., Datiles, M. B., Podgor, M. J., Vivino, M. A., Mahurkarl, A. A., & Lasa, M. S. M. (1994) · Eye · doi:10.1038/eye.1994.162 · Cited by 3
A Gastroesophageal Electrode for Electrophysiological Studies
MCENEANEY, D. J., Escalona, O. J., Anderson, J. A. D., & Adgey, A. A. J. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00477.x · Cited by 3
MEASUREMENT OF FERRITIN LEVELS: COMPARISON OF A COMMERCIAL IRMA TO AN IN-HOUSE ELISA METHOD
Zemelka, S., & Biesalski, H. K. (2001) · Journal of Immunoassay and Immunochemistry · doi:10.1081/ias-100107401 · Cited by 3
Continuous end‐tidal carbon dioxide monitoring during normofrequent jet‐ventilation
Bach, L. F., Wanner-Olsen, H., Andersen, B., Madsen, I. K., & Kruse, S. H. (1996) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1996.tb05557.x · Cited by 3
Validation of diagnostic tests
Borràs-Bermejo, B., Masclans, J. R., Olona, M., de Latorre, F. J., & Bizouarn, P. (1993) · Intensive Care Medicine · doi:10.1007/bf01690554 · Cited by 3
Letters to the Editor-in-Chief
Doherty, M., & Smith, P. M. (2001) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200110000-00028 · Cited by 3
A modern approach to the determination of expanded uncertainty in noninvasive blood oximetry
Cysewska-Sobusiak, A. R. (1999) · Measurement · doi:10.1016/s0263-2241(98)00073-6 · Cited by 3
Kelb�k, H., Gj�rup, T., Hvid-Jacobsen, K., Skagen, K., Munck, O., GODTFREDSEN, J., Heslet, L., Tjur, T., & Jensen, A. (1992) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/bf00175861 · Cited by 3
Intraocular pressure : clinical aspects and new measurement methods
Jóhannesson, G. (2011) · DiVA at Umeå University (Umeå University) · Cited by 5
Ward, G., Macaulay, F., Jagger, C., & Harper, W. (1998) · British Journal of Occupational Therapy · doi:10.1177/030802269806100308 · Cited by 3
Corsi, C., Lamberti, C., Cerutti, S., Laulom, J.-P., Bailliart, O., Cholley, B., Capderou, A., Vaïda, P., & Caiani, E. G. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02347542 · Cited by 3
Magnetic-resonance pelvimetry in breech presentation
van der Post, J., & Maathuis, J. B. (1998) · The Lancet · doi:10.1016/s0140-6736(05)70333-0 · Cited by 3
Validation of pulse contour derived stroke volume variation
Bein, B., Renner, J., Meybohm, P., & Scholz, J. (2007) · British Journal of Anaesthesia · doi:10.1093/bja/aem284 · Cited by 3
Çiftçi, Ö., Elemen, L., Elemen, F., Yı́ldı́rı́r, A., Çalışkan, M., Erdoğan, D., & Müderrisoğlu, H. (2008) · Surgical Laparoscopy Endoscopy & Percutaneous Techniques · doi:10.1097/sle.0b013e3181654470 · Cited by 3
Carkeet, A. (2015) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12205 · Cited by 3
Experimental and predicted dual oximetry variability
Wong, D. H. W., Weir, P. D., Mahutte, C. K., ONISHI, R., Daves, S., Foran, W., Ferraro, L., & Barker, S. J. (1993) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02886697 · Cited by 3
Quality control in quantitation of components of the plasminogen activator system in tumour extracts
Benraad, T. J., Geurts-Moespot, A. J., Grebenschikov, N., de Witte, J., Heuvel, J. J. T. M., & Sweep, F. C. G. J. (1996) · Fibrinolysis and Proteolysis · doi:10.1016/s0268-9499(96)80073-7 · Cited by 3
Evaluation of the home pad test in the investigation of female urinary incontinence
Khan, K. S., & CHIEN, P. F. W. (1996) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1996.tb09848.x · Cited by 3
Evaluation of a new real continuous cardiac output pulmonary artery catheter
Gardner, A., Daly, D., Screaton, M., Burns, T. I., & Vuylsteke, A. (2000) · Critical Care · doi:10.1186/cc703 · Cited by 3
Ryde, S. J. S., Eston, R. G., Laskey, M. A., Evans, C. J., & Hancock, D. (1998) · Applied Radiation and Isotopes · doi:10.1016/s0969-8043(97)00062-6 · Cited by 3
Validation of the rotation ratios method
Foss, O. A., Klaksvik, J., Benum, P., & Anda, S. (2007) · Acta Radiologica · doi:10.1080/02841850701348705 · Cited by 3
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.