STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 39,301–39,400 of 47,868
Gullberg, R. G., & McElroy, A. J. (1992) · Forensic Science International · doi:10.1016/0379-0738(92)90013-m · Cited by 3
A406 MEASUREMENT OF BREATH ETHANOL CONCENTRATION IN PATIENTS WITH GENERAL ANESTHESIA
Gehring, H., Schmitz, A., Nahm, W., Fornara, P., Dörges, V., & Schmucker, P. (1997) · Anesthesiology · doi:10.1097/00000542-199709001-00406 · Cited by 3
Müller, M., Pan, X., Walter, P., & Schimrigk, К. (1998) · European Journal of Ultrasound · doi:10.1016/s0929-8266(98)00044-5 · Cited by 3
Tsikas, D., & Hohlfeld, J. M. (2009) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.01.026 · Cited by 3
FRUIN, M. L., & Rankin, J. W. (2003) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200305001-01587 · Cited by 4
Cardiac output: measurement of the future or relic of the past?
Lamantia, K., & Barash, P. G. (1988) · Journal of Cardiothoracic Anesthesia · doi:10.1016/0888-6296(88)90048-8 · Cited by 5
Taimela, E., Koskinen, P. J., & Irjala, K. M. A. (1994) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365519409087528 · Cited by 3
Evaluation of an expired fraction carbon dioxide monitor
Ratnasabapathy, U., Allam, S., & Souter, M. J. (2002) · Anaesthesia · doi:10.1046/j.1365-2044.2002.02751.x · Cited by 3
de Greeff, A., Ahmed, N., Sheppard, E., & Shennan, A. H. (2008) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3282f7692d · Cited by 3
Has LVEF changed beyond chance? Limits of agreement of radiotracer-derived LVEF
Wackers, F. J. T. (2015) · Journal of Nuclear Cardiology · doi:10.1007/s12350-015-0090-y · Cited by 3
Salmi, T. K., & Brander, P. E. (1994) · International journal of clinical monitoring and computing · doi:10.1007/bf01132846 · Cited by 3
Cardiac output estimation by visual inspection vs thermodilution during cardiac surgery
Aye, T., Milne, B., & Ballantyne, M. (1997) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03012999 · Cited by 3
The Agreement and Repeatability of Computer-Based Wear Measurement of Total Hip Arthroplasties
Valle, A. G. D., Comba, F. M., Ellis, R., Peterson, M. G. E., & Salvati, E. A. (2007) · The Journal of Arthroplasty · doi:10.1016/j.arth.2006.12.106 · Cited by 3
Evaluation of an alternative method of blood sampling from arterial cannulae
Alford, A. M., Kirton, C. B., & Peacock, J. E. (1993) · Anaesthesia · doi:10.1111/j.1365-2044.1993.tb07079.x · Cited by 3
The Effect of Illumination-Microscope Angle on Slit Lamp Estimate of the Anterior Chamber Depth
Osuobeni, E. P., & ODUWAIYE, K. A. (2003) · Optometry and Vision Science · doi:10.1002/j.1538-9235.2003.tb01624.x · Cited by 3
Grochowski, S. J., Amrami, K. K., & Kaufman, K. R. (2005) · Skeletal Radiology · doi:10.1007/s00256-005-0944-5 · Cited by 3
Comparison of Methods for Cardiac Output Measurement
Kööbi, T., Kaukinen, S., & Kauppinen, P. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00052 · Cited by 3
Hadziyannis, E. S., Hadziyannis, A., Yen‐Lieberman, B., Kiwi, M. L., Hodnick, S., Spanou, F., Starkey, C., & Younossi, Z. M. (2001) · Journal of Clinical Gastroenterology · doi:10.1097/00004836-200107000-00011 · Cited by 3
Systolic arterial pressure determination by a new pulse monitor technique
Wong, D. T., VOLGYESI, G. A., & Bissonnette, B. (1992) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03008326 · Cited by 3
Normalizing Intermethod Free Triiodothyronine Patient Results: Need for Commutable Materials
Cattozzo, G., Scarsi, G., & Franzini, C. (1998) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.1998.138 · Cited by 3
Reduced k‐space encoding for dual‐contrast TrueFISP imaging
Merrifield, R. D., Keegan, J., Firmin, D. N., & Yang, G. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10374 · Cited by 3
Fumagalli, M., Mosca, F., Knudsen, G. M., & Greisen, G. O. (2003) · Neonatology · doi:10.1159/000074967 · Cited by 3
Hypertensive heart disease: Quantitative evaluation of response to therapy with cine MR imaging
Albrecht, T., Dulce, M.-C., Dulce, K. K., Gotzen, R., & Wolf, K.-J. R. (1997) · Academic Radiology · doi:10.1016/s1076-6332(05)80266-1 · Cited by 3
Kern, S. E., Johnson, J. O., Westenskow, D. R., & Orr, J. A. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01627415 · Cited by 3
Sidery, M. B., & Macdonald, I. A. (1991) · Clinical Autonomic Research · doi:10.1007/bf01826214 · Cited by 3
Perret, C., Poiraudeau, S., Fermanian, J., & Revel, M. (2000) · Annales de Réadaptation et de Médecine Physique · doi:10.1016/s0168-6054(00)88785-1 · Cited by 3
Padulo, J., Oliva, F., & Ardigò, L. P. (2014) · European Spine Journal · doi:10.1007/s00586-014-3407-7 · Cited by 3
Hernández-González, M. A., Espinola-Zavaleta, N. G., Solorio, S., Malacara, J. M., Pérez, V. M. J., de León, V. D., & López-Valenzuela, P. (2007) · Revista Española de Cardiología · doi:10.1157/13109643 · Cited by 3
Doriot, P. A., Dorsaz, P. ‐., Dorsaz, L., Chatelain, P., & Rutishauser, W. (1996) · International journal of cardiac imaging · doi:10.1007/bf01797742 · Cited by 3
The pruning of Doctor McDonough and the “humanizing” of statistics
Norman, G. R. (1997) · Fertility and Sterility · doi:10.1016/s0015-0282(97)90198-9 · Cited by 3
Whole blood assay of theophylline concentrations using immunochromatographic stick.
McKenzie, S. A. (1988) · Archives of Disease in Childhood · doi:10.1136/adc.63.5.571-a · Cited by 3
Booth, J. H., Quinn, T. A., Richmond, M. E., Cabreriza, S. E., Weinberg, A. D., Johnston, T. A., & Spotnitz, H. M. (2009) · ASAIO Journal · doi:10.1097/mat.0b013e3181bbafd5 · Cited by 3
Postoperative evaluation of congenital heart disease by magnetic resonance imaging
Roest, A. A. W., Helbing, W. A., van der Wall, E. E., & Roos, A. (1999) · Journal of Magnetic Resonance Imaging · doi:10.1002/(sici)1522-2586(199911)10:5<821::aid-jmri29>3.0.co;2-7 · Cited by 31
Chemoreflex thresholds to CO2 in decerebrate cats
Iscoe, S., Beaton, M., & Duffin, J. (1998) · Respiration Physiology · doi:10.1016/s0034-5687(98)00052-8 · Cited by 3
de Mey, C., & Erb, K. A. (1997) · The Journal of Clinical Pharmacology · doi:10.1177/009127009703700117 · Cited by 3
Estimating the reliability of clinical tests
Reeves, B. C., & Hill, A. R. (1992) · Ophthalmic and Physiological Optics · doi:10.1016/0275-5408(92)90098-h · Cited by 3
Evaluation of a mathematical model to predict intrapulmonary shunt non-invasively
Seear, M. D., Malagon, I. L., Hui, H. K., Alexander, J., Daoust, C., & Skippen, P. W. (1999) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03012950 · Cited by 3
Capillary testing in an anticoagulant clinic: comparison of two methods: Fig 1
Donohue, S. M., Dufty, J. M., Kafkarkou, N. A., Kirwan, R., & Taberner, D. A. (1996) · BMJ · doi:10.1136/bmj.313.7063.982 · Cited by 3
Coronary Risk Versus Cardiovascular Risk for Treatment Decisions in Mild Hypertension
Yeo, W. W., & Yeo, K. R. (2002) · European Journal of Cardiovascular Prevention & Rehabilitation · doi:10.1177/174182670200900508 · Cited by 3
Estimating the risk of driving under the influence of psychoactive substances
Houwing, S. (2013) · University of Groningen research database (University of Groningen / Centre for Information Technology) · Cited by 7
Knee Joint Stability and Functional Ability in Patients with Osteoarthritis of the Knee
van der Esch, M. (2008) · Pure Amsterdam UMC · Cited by 3
van Dalen, B. M., Snelder, S. M., & Geleijnse, M. L. (2018) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.10.097 · Cited by 3
King, A. M. (2011) · University of the Witwatersrand, Johannesburg Institutional Repository on DSpace (University of the Witwatersrand, Johannesburg) · Cited by 3
Assessment of dietary intake in young populations using new approaches and technologies
Svensson, Å. (2014) · DiVA at Umeå University (Umeå University) · Cited by 3
Vessel wall tracking based on the modified autocorrelation estimator
Rabben, S. I., Segers, P., Backer, J. F. A. D., SUTTER, J. D., Verdonck, P., & Gillebert, T. C. (2003) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2002.1192645 · Cited by 3
KONG, C. S., Ryder, I. G., KAHN, R., Gregory, L. C., & Mackenzie, C. F. (1995) · British Journal of Anaesthesia · doi:10.1093/bja/74.2.201 · Cited by 3
Development of sampling and bioselective techniques for on-line clinical biosensors
Kaptein, W. A. (1998) · University of Groningen research database (University of Groningen / Centre for Information Technology) · Cited by 4
Danilouchkine, M. G., van der Geest, R. J., Westenberg, J. J. M., Lelieveldt, B. P. F., & Reiber, J. H. C. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20442 · Cited by 4
Detection of bursts of microneurographic activity and estimation of burst parameters
Baker, L., & Shiavi, R. (1999) · Computers in Biology and Medicine · doi:10.1016/s0010-4825(99)00005-0 · Cited by 3
Lead Screening Among Low-Income Children in Galveston, Texas
Javier, F. C., McCormick, D. P., & Alcock, N. W. (1999) · Clinical Pediatrics · doi:10.1177/000992289903801104 · Cited by 3
Computer-Aided Analysis of Gastrointestinal Pressure Recordings: Comparison with Manual Scoring
Melsen, N. C., Andersen, M. B., Kraglund, K., & Juhl, B. (1992) · Scandinavian Journal of Gastroenterology · doi:10.3109/00365529209011188 · Cited by 3
Alcohol concentrations in post-mortem body fluids
Jones, A. W., & Kugelberg, F. C. (2006) · Human & Experimental Toxicology · doi:10.1177/0960327106070669 · Cited by 3
Identification of ultrasound-contrast-agent dilution systems for cardiac quantification
Mischi, M., Kalker, A. A. C. M., & Korsten, H. H. M. (2005) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2004.1417656 · Cited by 3
Schmidt, C., Theilmeier, G., Aken, H. V., Flottmann, C., Wirtz, S. P., Kehl, H.-G., Hoffmeier, A., & Berendes, E. (2005) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2005.00763.x · Cited by 3
Less Invasive Cardiac Output Monitoring: Characteristics and Limitations
Hofer, C. K., & Zollinger, A. (2006) · Springer eBooks · doi:10.1007/0-387-35096-9_16 · Cited by 3
Hardy, B. G., Rice, M. J., & McDonald, R. W. (1998) · The American Journal of Cardiology · doi:10.1016/s0002-9149(98)00656-0 · Cited by 3
Blood Flow Quantification with Permeable Contrast Agents: A Valid Technique?
van den Hoff, J. (2007) · Radiology · doi:10.1148/radiol.2433060992 · Cited by 3
A practical method to statistically classify devices according to their relative accuracy
Minini, P., Gaspoz, J., Pichot, V., Costes, F., Roche, F., Lacour, J. R., Antoniadis, A., & Barthélemy, J. (2001) · Clinical Physiology · doi:10.1046/j.1365-2281.2001.00320.x · Cited by 3
Image Registration Based on Lifting Process: An Application to Digital Subtraction Radiography
Matsopoulos, G. K., Mouravliansky, N. A., Asvestas, P. A., Delibasis, K. K., Gröndahl, K., & Gröndahl, H. (2006) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2006.875683 · Cited by 3
Shoukri, M. M. (2014) · Wiley eBooks · doi:10.1002/9781118445112.stat05301 · Cited by 3
Measuring morbidity following major surgery
Grocott, M. P. W. (2010) · UCL Discovery (University College London) · Cited by 4
Yates, J. O., Nott, D. M., Maltby, P., Billington, D., BAXTER, J. N., Shields, R. C., & Jenkins, S. A. (1995) · HPB Surgery · doi:10.1155/1995/31761 · Cited by 3
In vivo pressure-flow curve in unilateral rat lung ischemia-reperfusion injury
Yu, C., & Lai, Y. (2001) · Journal of Applied Physiology · doi:10.1152/jappl.2001.90.5.1865 · Cited by 3
Retinal Blood Flow and Vascular Reactivity in Diabetic Retinopathy
Gilmore, E. D. (2006) · UWSpace (University of Waterloo) · Cited by 4
Evaluation of a forced oscillation method to measure thoracic gas volume
Peslin, R., & Duvivier, C. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.84.3.862 · Cited by 3
dos Santos, J. A. P. P. (2009) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT) · Cited by 3
Computer Based Interactive Medical Simulation 
Cotin, S. (2008) · HAL · Cited by 6
Derieppe, M., Bos, C., de Greef, M., MOONEN, C. T. W., & de Senneville, B. D. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/2/588 · Cited by 5
Unwin, N. C., SKILLEN, A. W., & Alberti, K. G. M. M. (1995) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329503200312 · Cited by 3
Measuring confidence with manual wheelchair use : a four phase, mixed-methods study
Rushton, P. W. (2011) · cIRcle (University of British Columbia) · doi:10.14288/1.0071504 · Cited by 7
Mahoney, K. B. (2016) · Digital Commons - URI (University of Rhode Island) · doi:10.23860/thesis-mahoney-kayla-2016 · Cited by 3
CAN THEORETICAL VALUES FOR CHEST WALL COMPLIANCE BE USED IN ARDS PATIENTS?
Chen, G., Xu, M., Chen, X., Rittayamai, N., Kim, M., Soliman, I., Yang, Y.-L., Jx, Z., & Brochard, L. J. (2015) · Intensive Care Medicine Experimental · doi:10.1186/2197-425x-3-s1-a999 · Cited by 3
Improvements to the [13C]MTG breath test for measuring fat digestion
Slater, C. (2004) · Enlighten: Theses (The University of Glasgow) · doi:10.5525/gla.thesis.71085 · Cited by 4
Sources of Noise and Signal‐to‐Noise Ratio
Droitcour, A. D., Borić-Lubecke, O., & Yamada, S. (2016) · Wiley eBooks · doi:10.1002/9781119078418.ch6 · Cited by 4
Relative Agreement between Raman and Mass Spectrometry for Measuring End-Tidal Carbon Dioxide
Graybeal, J. M., & Russell, G. B. (1994) · Respiratory Care · doi:10.1177/194336549403900304 · Cited by 3
Factors Influencing Measures of Trochanteric Soft Tissue Thickness
Lafleur, B. (2016) · UWSpace (University of Waterloo) · Cited by 3
Sedrez, J. A., Candotti, C. T., da Rosa, M. I. Z., Medeiros, F. D. S., Marques, M. T., & Loss, J. F. (2022) · Figshare · doi:10.6084/m9.figshare.20015474 · Cited by 3
A Validation of a Simulation Environment for Motion Sensing Electronic Textiles
Einsmann, C. (2006) · VTechWorks (Virginia Tech) · Cited by 3
Lactose, Lactase, and Bowel Disorders
Peuhkuri, K. (2000) · Helda - University of Helsinki Institutional Repository · Cited by 4
Development of Virtual Toddler Fit Models for Child Safety Restraint Design
Kim, K. H., Jones, M. L. H., Ebert, S. M., Malik, L. A., Manary, M. A., Reed, M. P., & Klinich, K. D. (2015) · Deep Blue (University of Michigan) · Cited by 3
The validation of the sit-to-stand test for COPD patients
Chambellan, A., Nusinovici, S., Vaidya, T., Gourraud, P., & de Bisschop, C. (2017) · European Respiratory Journal · doi:10.1183/13993003.01203-2017 · Cited by 3
Batista, B. N., Baiocchi, J. M. T., Fraga, M. F. P., Sampaio, M., & de Barros, A. C. (2017) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000004223 · Cited by 3
Shoukri, M. M. (2005) · Wiley eBooks · doi:10.1002/0470011815.b2a04002 · Cited by 3
Schmitz, C., Hüttmann, A., Müller, S. P., Hanoun, M., Boellaard, R., Brinkmann, M., Jöckel, K., Dührsen, U., & Rekowski, J. (2019) · Data in Brief · doi:10.1016/j.dib.2019.104976 · Cited by 3
Syncope : an integrative physiological approach
Thijs, R. D. (2008) · Leiden Repository (Leiden University) · Cited by 4
Lorenz, K. A., Yeshurun, S., Aziz, R., Ortiz-Delatorre, J., Bagley, J. R., Mor, M., & Kern, M. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.05.05.078980 · Cited by 3
Is there a place for lower limb lymphedema prevention after cervical cancer treatment?
Baiocchi, J. M. T., Baiocchi, G., & Bergmann, A. K. (2020) · International Journal of Gynecological Cancer · doi:10.1136/ijgc-2020-001532 · Cited by 3
Validation of the Vmax Metabolic Cart in a Simulated Pediatric Model
Smallwood, C. D., Gouldstone, A., & Mehta, N. M. (2015) · Journal of Parenteral and Enteral Nutrition · doi:10.1177/0148607114536444 · Cited by 3
Claessens, J. L. J., van Egmond, J., de Boer, J. H., & Wisse, R. P. L. (2021) · medRxiv · doi:10.1101/2021.04.14.21255457 · Cited by 4
Musical Aesthetic SensitivityPreprint
Clemente, A., Pearce, M. T., & Nadal, M. (2021) · PsyArXiv (OSF Preprints) · doi:10.31234/osf.io/z4dqj · Cited by 3
Lee, H. J., Chae, J. S., An, S.-M., Oh, H., Kim, Y. J., & Woo, J. H. (2022) · Dove Medical Press (Taylor and Francis Group) · Cited by 3
Wang, C., Shiraishi, S., Leung, A., Baravarian, S., Hull, L., Goh, V. H. H., Lee, P. W. N., & Swerdloff, R. S. (2018) · Steroids · doi:10.1016/j.steroids.2018.05.006 · Cited by 3
Genovese, D., Badano, L. P., & Marra, M. P. (2022) · European Heart Journal · doi:10.1093/eurheartj/ehac739 · Cited by 3
MacKay, J. R. D., & Wu, S. C. (2023) · Wiley eBooks · doi:10.1002/9781119852865.ch17 · Cited by 6
Batubo, N. P., Auma, C. I., Moore, J. B., & Zulyniak, M. A. (2024) · medRxiv · doi:10.1101/2024.01.24.24301732 · Cited by 3
Framework for Analytical Validation of DHT-Based Actigraphy and Signal Measures in HF Trials
Köse, O., Sahal, A., Zhang, G. K., Possik, E., Fiuzat, M., Psotka, M. A., Kao, D. P., Shalowitz, E. L., Martyn, T., Kittipibul, V., et al. (2025) · JACC Heart Failure · doi:10.1016/j.jchf.2025.102831 · Cited by 3
Allocating census data to general practice populations
Aveyard, P. N. (1995) · BMJ · doi:10.1136/bmj.311.7009.875b · Cited by 3
Bell, J. A. (2008) · University of Huddersfield Repository (University of Huddersfield) · Cited by 3
Demoulin, C. (2008) · ORBi (University of Liège) · Cited by 3
Knee joint stiffness and function following total knee arthroplasty
Lane, J. (2010) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 5
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.