STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 4,301–4,400 of 47,868
Deng, Z., Hasegawa, M., AOKI, K., Matayoshi, S., Kiyuna, A., Yamashita, Y., Uehara, T., Agena, S., Maeda, H., Xie, M., et al. (2014) · International Journal of Oncology · doi:10.3892/ijo.2014.2440 · Cited by 98
Estimating energy expenditure by heart-rate monitoring without individual calibration
RENNIE, K. L., HENNINGS, S. J., MITCHELL, J., & WAREHAM, N. J. (2001) · Medicine and Science in Sports and Exercise · doi:10.1097/00005768-200106000-00013 · Cited by 98
Intercontinental hearing assessment – a study in tele-audiology
Swanepoel, D. W., Koekemoer, D., & Clark, J. L. (2010) · Journal of Telemedicine and Telecare · doi:10.1258/jtt.2010.090906 · Cited by 98
Isakow, W., & Schuster, D. P. (2006) · American Journal of Physiology-Lung Cellular and Molecular Physiology · doi:10.1152/ajplung.00277.2006 · Cited by 100
Prous, M. J. G. D. Y., Salvanés, F. R., & Ortells, L. C. (2009) · Reumatología Clínica (English Edition) · doi:10.1016/s2173-5743(09)70115-7 · Cited by 128
Song, J., Jeong, S.-M., Jun, I., Lee, H.-M., & Hwang, G.-S. (2013) · British Journal of Anaesthesia · doi:10.1093/bja/aet325 · Cited by 98
RELIABILITY AND VALIDITY OF AN ACCELEROMETRIC SYSTEM FOR ASSESSING VERTICAL JUMPING PERFORMANCE
Choukou, M.-A., Laffaye, G., & Taïar, R. (2014) · Biology of Sport · doi:10.5604/20831862.1086733 · Cited by 98
Hariri, A. H., Zhang, H., Ho, A., Francis, P. J., Weleber, R. G., Birch, D. G., Ferris, F. L., Sadda, S. R., & for the Trial of Oral Valproic Acid for Retinitis Pigmentosa Group (2016) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2016.0502 · Cited by 101
A real-time approach for heart rate monitoring using a Hilbert transform in seismocardiograms
Tadi, M. J., Lehtonen, E., Hurnanen, T., Koskinen, J., Eriksson, J., Pänkäälä, M., Teräs, M., & Koivisto, T. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/11/1885 · Cited by 114
Meyer, J. J., Gokul, A., Vellara, H. R., Prime, Z. J., & McGhee, C. N. J. (2016) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2016.12.003 · Cited by 98
Attia, A., Dhahbi, W., Chaouachi, A., Padulo, J., Wong, D. P., & Chamari, K. (2017) · Biology of Sport · doi:10.5114/biolsport.2017.63735 · Cited by 112
Improved Binocular Outcomes Following Binocular Treatment for Childhood Amblyopia
Kelly, K. R., Jost, R. M., Wang, Y., Dao, L. M., Beauchamp, C. L., Leffler, J. N., & Birch, E. E. (2018) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.17-23235 · Cited by 98
Movement of anterior teeth using clear aligners: a three-dimensional, retrospective evaluation
Tepedino, M., Paoloni, V., Cozza, P., & Chimenti, C. (2018) · Progress in Orthodontics · doi:10.1186/s40510-018-0207-3 · Cited by 98
Oustric, P., Thivel, D., Dalton, M., Beaulieu, K., Gibbons, C., Hopkins, M., Blundell, J., & Finlayson, G. (2020) · Food Quality and Preference · doi:10.1016/j.foodqual.2019.103824 · Cited by 97
Use of the iPhone for Cobb angle measurement in scoliosis
Shaw, M., Adam, C. J., Izatt, M. T., Licina, P., & Askin, G. N. (2011) · European Spine Journal · doi:10.1007/s00586-011-2059-0 · Cited by 98
Kondili, Ε., Prinianakis, G., Alexopoulou, C., Vakouti, E., Klimathianaki, M., & Georgopoulos, D. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0110-0 · Cited by 104
Yoong, S. L., Carey, M. L., D’Este, C. A., & Sanson‐Fisher, R. (2013) · BMC Medical Research Methodology · doi:10.1186/1471-2288-13-38 · Cited by 100
Tsai, W., Liu, Y.-W., Huang, Y.-Y., Lin, C., Lee, C., & Tsai, L.-M. (2010) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2010.08.011 · Cited by 98
Machet, L., Belot, V., Naouri, M., Boka, M., Mourtada, Y., Giraudeau, B., Laure, B., Perrinaud, A., Machet, M., & Vaillant, L. (2009) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2009.03.018 · Cited by 98
Dalaveris, E., Kerenidi, T., Katsabeki-Katsafli, A., Kiropoulos, T. S., Tanou, K., Gourgoulianis, K. I., & Κostikas, Κ. (2008) · Lung Cancer · doi:10.1016/j.lungcan.2008.08.015 · Cited by 100
Saad, H. B., Elattar, M. N., Mabrouk, K. H., Abdelaziz, A. B., Abdelghani, A., Boussarsar, M., Limam, K., Maatoug, C., Bouslah, H., Charrada, A., et al. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.10.015 · Cited by 101
Utomo, E., Blok, B. F. M., Steensma, A. B., & Korfage, I. J. (2014) · International Urogynecology Journal · doi:10.1007/s00192-013-2263-z · Cited by 113
Hoge, R. D., Atkinson, J., Gill, B., Crelier, G. R., Marrett, S., & Pike, G. B. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<849::aid-mrm4>3.0.co;2-z · Cited by 576
Cattaneo, C. G., Frank, S. M., Hesel, T. W., El-Rahmany, H. K., Kim, L. J., & Tran, K. M. (2000) · Anesthesia & Analgesia · doi:10.1097/00000539-200004000-00030 · Cited by 98
Duncker, T., Greenberg, J. P., Ramachandran, R., Hood, D. C., Smith, R. T., Hirose, T., Woods, R. L., Tsang, S. H., Delori, F. C., & Sparrow, J. R. (2014) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.13-13834 · Cited by 97
Analysis of Tidal Breathing Parameters in Infancy: How Variable Is TPTEF:TE?
Stocks, J., Dezateux, C. A., Jackson, E. A., Hoo, A., Costeloe, K., & Wade, A. M. (1994) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.150.5.7952563 · Cited by 98
Müller, M., Dahmen, G., Pörksen, E., Geerling, G., Laqua, H., Ziegler, A., & Hoerauf, H. (2006) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2006.07.014 · Cited by 102
Prediction of preterm birth using the cervical consistency index
Parra-Saavedra, M. A., Gómez, L., Barrero, A. V., Parra, G., Vergara, F., & Navarro, É. (2011) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.9010 · Cited by 111
Re‐emergence of iodine deficiency in Australia
Li, M., Ma, G., Guttikonda, K., BOYAGES, S. C., & Eastman, C. J. (2001) · Asia Pacific Journal of Clinical Nutrition · doi:10.1046/j.1440-6047.2001.00254.x · Cited by 100
Comparison of methods for thyroid volume estimation in patients with Graves' disease
van Isselt, J. W., de Klerk, J. M. H., van Rijk, P. P., Gils, A. P. G., Polman, L. J., Kamphuis, C., Meijer, R., & Beekman, F. J. (2003) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-1101-1 · Cited by 98
Hoffman, E. A., Jiang, R., Baumhauer, H., Brooks, M., Carr, J. J., Detrano, R., Reinhardt, J. M., Rodriguez, J., Stukovsky, K. H., Wong, N. D., et al. (2009) · Academic Radiology · doi:10.1016/j.acra.2008.12.024 · Cited by 104
Pulsed Tissue Doppler Imaging in Normal Cats and Cats with Hypertrophic Cardiomyopathy
Koffas, H., Dukes‐McEwan, J., CORCORAN, B. M., MORAN, C. M., French, A. T., Sboros, V., Simpson, K. E., & McDicken, W. N. (2006) · Journal of Veterinary Internal Medicine · doi:10.1111/j.1939-1676.2006.tb02825.x · Cited by 98
Okamura, T., Onuma, Y., García‐García, H. M., van Geuns, R., Wykrzykowska, J. J., Schultz, C. J., van der Giessen, W., Ligthart, J. M. R., Regar, E. S., & Serruys, P. W. J. C. (2011) · EuroIntervention · doi:10.4244/eijv6i9a182 · Cited by 98
Bein, B., Worthmann, F., Tonner, P. H., Paris, A., Steinfath, M., Hedderich, J., & Scholz, J. (2004) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2004.01.025 · Cited by 98
Nebuchennykh, M., Løseth, S., Lindal, S., & Mellgren, S. I. (2009) · Journal of Neurology · doi:10.1007/s00415-009-5065-y · Cited by 109
Leitner, C., Mair, P., Paul, B., Wick, F., Mittermaier, C., Sycha, T., & EBENBICHLER, G. R. (2007) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2007.09.007 · Cited by 112
Fielden, S. W., Fornwalt, B. K., Jerosch‐Herold, M., Eisner, R. L., Stillman, A. E., & Oshinski, J. N. (2008) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21387 · Cited by 103
β-Adrenergic Sweat Secretion as a Diagnostic Test for Cystic Fibrosis
Quinton, P., Molyneux, L., Ip, W., Dupuis, A., Avolio, J., Tullis, E. D., Conrad, D. J., Shamsuddin, A. K. M., Durie, P. R., & Gonska, T. Y. (2012) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201205-0922oc · Cited by 113
Rupp, T. L., & Balkin, T. J. (2011) · Behavior Research Methods · doi:10.3758/s13428-011-0098-4 · Cited by 98
Anatomical and functional changes in the lower urinary tract during pregnancy
Wijma, J., Potters, A. E. W., de Wolf, B. T. H. M., Tinga, D. J., & Aarnoudse, J. G. (2001) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1016/s0306-5456(00)00123-6 · Cited by 98
Mainstream End-tidal Carbon Dioxide Monitoring in the Neonatal Intensive Care Unit
Rozycki, H. J., Sysyn, G. D., Marshall, M. K., Malloy, R., & Wiswell, T. E. (1998) · PEDIATRICS · doi:10.1542/peds.101.4.648 · Cited by 98
Yu, E., Sloggett, C. E., Iwanochko, R. M., Rakowski, H., & Siu, S. C. (2000) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(00)90058-0 · Cited by 99
The effect of body mass index on three methods of fetal weight estimation
Farrell, T. J., Holmes, R., & STONE, P. R. (2002) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.2002.01249.x · Cited by 98
In vitro measurement of kidney size: comparison of ultrasonography and MRI
Bakker, J., Olree, M., Kaatee, R. S. J. P., de Lange, E. E., & Beek, F. J. A. (1998) · Ultrasound in Medicine & Biology · doi:10.1016/s0301-5629(97)00212-3 · Cited by 98
Tompuri, T. T., Lakka, T. A., Hakulinen, M., Lindi, V., Laaksonen, D. E., Kilpeläinen, T. O., Jääskeläinen, J., Lakka, H., & Laitinen, T. (2015) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12118 · Cited by 98
A Single-Visit Field Test of Critical Speed
Galbraith, A., Hopker, J., Lelliott, S., Diddams, L., & Passfield, L. (2014) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2013-0507 · Cited by 98
Pfau, T., Stubbs, N. C., Kaiser, L. J., Brown, L. E. A., & Clayton, H. M. (2012) · American Journal of Veterinary Research · doi:10.2460/ajvr.73.12.1890 · Cited by 98
Ramamurthy, N., Baliga, N., Wahr, J. A., Schaller, U., Yang, V. C.-M., & Meyerhoff, M. E. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.3.606 · Cited by 98
Pringle, A. M., Petrie, A., Cunningham, S. J., & McKnight, M. (2009) · American Journal of Orthodontics and Dentofacial Orthopedics · doi:10.1016/j.ajodo.2007.08.032 · Cited by 113
Validity of Six Field and Laboratory Methods for Measurement of Body Composition in Boys
Parker, L., Reilly, J. J., Slater, C., Wells, J. C. K., & Pitsiladis, Y. P. (2003) · Obesity Research · doi:10.1038/oby.2003.117 · Cited by 100
Marx, R. G. (2003) · Arthroscopy The Journal of Arthroscopic and Related Surgery · doi:10.1016/j.arthro.2003.10.029 · Cited by 108
Age-Specific Reliability of Two Grip-Strength Dynamometers When Used by Children
Molenaar, H. M., Zuidam, J. M., Selles, R. W., Stam, H. J., & Hovius, S. E. R. (2008) · Journal of Bone and Joint Surgery · doi:10.2106/jbjs.g.00469 · Cited by 98
Testa, A. C., Ludovisi, M., Manfredi, R. M., Zannoni, G. F., Gui, B., Basso, D., Legge, A. D., Licameli, A., Bidino, R. D., Scambia, G., et al. (2009) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.7325 · Cited by 98
An Automated System for 24-h Monitoring of Cough Frequency: The Leicester Cough Monitor
Matos, S., Birring, S. S., Pavord, I. D., & Evans, D. H. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.900811 · Cited by 98
The reliability, validity, and applicability of an English language version of the Mini-ICF-APP
Molodynski, A., Linden, M., Juckel, G., Yeeles, K., Anderson, C., Vazquez-Montes, M., & Burns, T. (2012) · Social Psychiatry and Psychiatric Epidemiology · doi:10.1007/s00127-012-0604-8 · Cited by 109
Botero, M., Kirby, D. S., Lobato, E. B., Staples, E. D., & Gravenstein, N. L. (2004) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2004.07.005 · Cited by 98
Kelly, A., Powell, C., & WILLIAMS, A. (2002) · Pediatric Emergency Care · doi:10.1097/00006565-200206000-00002 · Cited by 110
Hoffer, K. J., Hoffmann, P., & Savini, G. (2016) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2016.07.013 · Cited by 98
Pirpiris, M., Wilkinson, A. J., Rodda, J., Nguyen, T. C., Baker, R. J., Nattrass, G. R., & Graham, H. K. (2003) · Journal of Pediatric Orthopaedics · doi:10.1097/01241398-200305000-00006 · Cited by 112
Cohen, M. S., & DuBois, R. M. (1999) · Journal of Magnetic Resonance Imaging · doi:10.1002/(sici)1522-2586(199907)10:1<33::aid-jmri5>3.0.co;2-n · Cited by 150
Percutaneous radiofrequency ablation of pulmonary metastases in patients with colorectal cancer
King, J., Glenn, D. W., Clark, W. A., Zhao, J., Steinke, K., Clingan, P. R., & Morris, D. L. (2004) · British journal of surgery · doi:10.1002/bjs.4392 · Cited by 103
Clinical assessment of central venous pressure in the critically ill.
Cook, D. J. (1990) · American Journal of the Medical Sciences · doi:10.1097/00000441-199003000-00006 · Cited by 92
Reproducibility of ambulatory blood pressure monitoring in hemodialysis patients
Peixoto, A. J., Santos, S. F. F., Mendes, R. B., Crowley, S. T., Maldonado, R., Orias, M., Mansoor, G. A., & White, W. B. (2000) · American Journal of Kidney Diseases · doi:10.1053/ajkd.2000.19100 · Cited by 98
Beach, J. R., Young, C., Avery, A. J., Stenton, S., Dennis, J., Walters, E. H., & Hendrick, D. J. (1993) · Thorax · doi:10.1136/thx.48.3.239 · Cited by 98
Friedberg, M. K., Su, X., Tworetzky, W., Soriano, B. D., Powell, A. J., & Marx, G. R. (2010) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.109.928663 · Cited by 97
Wundersitz, D. W. T., Gastin, P. B., Richter, C., Robertson, S. J., & Netto, K. (2014) · European Journal of Sport Science · doi:10.1080/17461391.2014.955131 · Cited by 123
Zahnd, G., Boussel, L., Marion, A., Durand, M., Moulin, P., Sérusclat, A., & Vray, D. (2011) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2011.05.843 · Cited by 110
Tomlinson, M. J., Pooley, K., Simpson, T., Newton, T., Hopkisson, J. F., Jayaprakasan, K., Jayaprakasan, R., Naeem, A., & Pridmore, T. (2009) · Fertility and Sterility · doi:10.1016/j.fertnstert.2008.12.064 · Cited by 97
de Rochefort, L., Vial, L., Fodil, R., Maı̂tre, X., Louis, B., Isabey, D., Caillibotte, G., Thiriet, M., Bittoun, J., Durand, E., et al. (2007) · Journal of Applied Physiology · doi:10.1152/japplphysiol.01610.2005 · Cited by 97
Soliman, O. I. I., Krenning, B. J., Geleijnse, M. L., Nemes, A., van Geuns, R., Baks, T., Anwar, A. M., Galema, T. W., Vletter, W. B., & ten Cate, F. J. (2007) · Echocardiography · doi:10.1111/j.1540-8175.2007.00502.x · Cited by 97
Schwaiger, B. J., Gersing, A. S., Baum, T. H., Noël, P. B., Zimmer, C., & Bauer, J. S. (2014) · American Journal of Neuroradiology · doi:10.3174/ajnr.a3893 · Cited by 106
Smartphone-Derived Heart-Rate Variability and Training Load in a Women’s Soccer Team
Flatt, A. A., & Esco, M. R. (2015) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2014-0556 · Cited by 112
Zhang, L., Kong, H., Chin, C. T., Liu, S., Chen, Z., Wang, T., & Chen, S. (2014) · Computerized Medical Imaging and Graphics · doi:10.1016/j.compmedimag.2014.02.001 · Cited by 106
Moghames, P., Hammami, N., Hwalla, N. C., Yazbeck, N., Shoaib, H., Nasreddine, L., & Naja, F. (2015) · Nutrition Journal · doi:10.1186/s12937-015-0121-1 · Cited by 118
Gąsior, J. S., Sacha, J., Jelen, P., Zieliński, J., & Przybylski, J. (2016) · Frontiers in Physiology · doi:10.3389/fphys.2016.00356 · Cited by 104
Validation of Near-Infrared Spectroscopy for Monitoring Cerebral Autoregulation in Comatose Patients
Rivera‐Lara, L., Geocadin, R. G., Vaca, A. Z., Healy, R., Radzik, B. R., Palmisano, C., Mirski, M. A., Ziai, W., & Hogue, C. W. J. (2017) · Neurocritical Care · doi:10.1007/s12028-017-0421-8 · Cited by 97
Feasibility of simplifying renal dosimetry in 177Lu peptide receptor radionuclide therapy
Sundlöv, A., Gustafsson, J., Brolin, G., Mortensen, N., Hermann, R., Bernhardt, P., Svensson, J. B., Ljungberg, M. H., Tennvall, J., & Gleisner, K. S. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0210-2 · Cited by 104
Using multiple agreement methods for continuous repeated measures data: a tutorial for practitioners
Parker, R. A., Scott, C., Inácio, V., & Stevens, N. T. (2020) · BMC Medical Research Methodology · doi:10.1186/s12874-020-01022-x · Cited by 110
Davies, R. H., Augusto, J. B., Bhuva, A., Xue, H., Treibel, T. A., Ye, Y., Hughes, R. K., Bai, W., Lau, C. W. Z., Shiwani, H., et al. (2022) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-022-00846-4 · Cited by 113
Gwiazda, J. E., Marsh‐Tootle, W. L., Hyman, L., Hussein, M., & Norton, T. T. (2002) · PubMed · Cited by 97
Assessment of Maximum Inspiratory Pressure
Volianitis, S., MCCONNELL, A. K., & Jones, D. A. (2001) · Respiration · doi:10.1159/000050458 · Cited by 97
Front-Crawl Instantaneous Velocity Estimation Using a Wearable Inertial Measurement Unit
Dadashi, F., Crettenand, F., Millet, G. P., & Aminian, K. (2012) · Sensors · doi:10.3390/s121012927 · Cited by 122
Otsuka, M., Bruining, N., van Pelt, N. C., Mollet, N. R., Ligthart, J. M. R., Vourvouri, E. C., Hamers, R., de Jaegere, P. P. T., Wijns, W. C., van Domburg, R. T., et al. (2008) · Investigative Radiology · doi:10.1097/rli.0b013e31816a88a9 · Cited by 97
Jeong, H., Jeong, S., Lim, H. J., Lee, J., & Yoo, K. Y. (2012) · Anesthesiology · doi:10.1097/aln.0b013e31825154d2 · Cited by 97
Haug, A., Williams, R. B., & Larsen, S. (2006) · Veterinary Parasitology · doi:10.1016/j.vetpar.2005.11.024 · Cited by 116
Capron, F., Fortier, L., Racine, S., & Donati, F. (2006) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000204288.24395.38 · Cited by 97
Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study
Pache, M., Wilmsmeyer, S., Lautebach, S., & Funk, J. (2005) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-005-1124-y · Cited by 97
Körperich, H., Gieseke, J., Barth, P., Hoogeveen, R. M., Esdorn, H., Peterschröder, A., Meyer, H., & Beerbaum, P. (2004) · Circulation · doi:10.1161/01.cir.0000126494.66859.a2 · Cited by 97
Impaired Bronchodilator Response to Albuterol in Healthy Elderly Men and Women
Connolly, M. J., Crowley, J. J., Charan, N. B., Nielson, C., & Vestal, R. E. (1995) · CHEST Journal · doi:10.1378/chest.108.2.401 · Cited by 97
Sharma, B., Boet, S., Grantcharov, T., Shin, E., Barrowman, N. J., & Bould, M. D. (2013) · Surgery · doi:10.1016/j.surg.2012.09.006 · Cited by 101
Heo, M., Murphy, C. F., & Meyers, B. S. (2007) · American Journal of Geriatric Psychiatry · doi:10.1097/jgp.0b013e318098614e · Cited by 109
Yoon, J. H., Lee, J. M., Han, J. K., & Cho, B. I. (2014) · Journal of Ultrasound in Medicine · doi:10.7863/ultra.33.3.437 · Cited by 112
Ono, R., Hirata, S., Yamada, M., Nishiyama, T., Kurosaka, M., & Tamura, Y. (2007) · BMC Musculoskeletal Disorders · doi:10.1186/1471-2474-8-61 · Cited by 123
Little, S. H., Igo, S. R., Pirat, B., McCulloch, M. L., Hartley, C. J., Nosé, Y., & Zoghbi, W. A. (2007) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2006.12.079 · Cited by 97
Joannides, R., Moore, N., Iacob, M., Compagnon, P., Lerebours, G., Menard, J., & Thuillez, C. (2006) · British Journal of Clinical Pharmacology · doi:10.1111/j.1365-2125.2005.02544.x · Cited by 97
van den Bosch, A. E., Robbers‐Visser, D., Krenning, B. J., Voormolen, M. M., McGhie, J. S., Helbing, W. A., Roos‐Hesselink, J. W., Simoons, M. L., & Meijboom, F. J. (2006) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2005.06.009 · Cited by 97
Jimidar, M., Hartmann, C., Cousement, N., & Massart, D. L. (1995) · Journal of Chromatography A · doi:10.1016/0021-9673(94)01290-u · Cited by 97
Link, T. M., Koppers, B., Licht, T., Bauer, J. S., Lü, Y., & Rummeny, E. J. (2004) · Radiology · doi:10.1148/radiol.2313030325 · Cited by 104
Reproducibility of HPV 16 and HPV 18 viral load quantitation using TaqMan real-time PCR assays
Gravitt, P. E., Peyton, C. L., Wheeler, C. M., Apple, R. J., Higuchi, R., & Shah, K. V. (2003) · Journal of Virological Methods · doi:10.1016/s0166-0934(03)00186-1 · Cited by 110
Interpretation of the Gamma Interferon Test for Diagnosis of Subclinical Paratuberculosis in Cattle
Jungersen, G., Huda, A., Hansen, J. J., & Lind, P. (2002) · Clinical and Vaccine Immunology · doi:10.1128/cdli.9.2.453-460.2002 · Cited by 97
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.