Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-hour blood pressure measurements
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
This randomized, single-center, open, within-subject study evaluated the performance of the Braun PrecisionSensor 2000 blood pressure (BP) measurement wrist device (BP 2000, Braun GmbH, Germany) with and without the position sensor, and compared the results with data obtained from 24-h ambulatory BP measurement (ABPM, A&D TM 2430). In a crossover design, 43 subjects performed BP measurements with the BP 2000 during two 7-day periods, one with the position sensor and the other without the position sensor. The correlation coefficients between all wrist self-measurements and ABPM were 0.73 for systolic and 0.65 for diastolic BP (with position sensor) compared to 0.70 and 0.60 for readings without position sensor, respectively. The categorization of subjects as normotensive or hypertensive, using the wrist device in subjects without specific training, achieved a correct classification of 84% (with position sensor) and 81% (without position sensor) in comparison to ABPM. When the measurements were performed in the clinic by an experienced observer, the correct classification was 79% in comparison to ABPM. The crossover study design did not show any significant variation between the two devices with regard to reproducibility of readings at the wrist. In conclusion, self-BP measurements with BP 2000, by untrained subjects, produced results consistent with those found recently with self-BP measurements with upper arm devices, when both data sets are compared with ABPM. The rates of false classification of normotension/hypertension with the wrist device were small and at least as reliable as office measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Mengden, T., Binswanger, B., Weisser, B., & Vetter, W. (1992) · American Journal of Hypertension · doi:10.1093/ajh/5.3.154 · Cited by 34
Self-citationZweiker, R., Schumacher, M., Fruhwald, F. M., Watzinger, N., & Klein, W. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018080-00004 · Cited by 29
Cited by
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationParati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Self-citationAgarwal, R., Bills, J. E., Hecht, T. J. W., & Light, R. P. (2011) · Hypertension · doi:10.1161/hypertensionaha.110.160911 · Cited by 430
Constanti, M., Boffa, R., Floyd, C. N., Wierzbicki, A. S., McManus, R. J., & Glover, M. (2021) · Journal of Human Hypertension · doi:10.1038/s41371-020-0357-x · Cited by 21
Metadata
- Authors
- UEN, S., WEISSER, B., WIENEKE, P., VETTER, H., & MENGDEN, T.
- Venue
- American Journal of Hypertension
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-09
- Volume
- 15
- Issue
- 9
- Pages
- 787-792
- ISSN
- 0895-7061 (ISSN-L)
- OpenAlex
- W1967548446
- Cited by
- 24
- References
- 2
- FWCI
- 1.66
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-hour blood pressure measurements. https://app.cochat.ai/papers/10.1016/s0895-7061(02)02979-5
- The paper
- UEN, S., WEISSER, B., WIENEKE, P., VETTER, H., & MENGDEN, T. (2002). Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-hour blood pressure measurements. American Journal of Hypertension. https://doi.org/10.1016/s0895-7061(02)02979-5
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.