Wearable device signals and home blood pressure data across age, sex, race, ethnicity, and clinical phenotypes in the Michigan Predictive Activity & Clinical Trajectories in Health (MIPACT) study: a prospective, community-based observational study
Open accessThe DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Wearable technology has rapidly entered consumer markets and has health-care potential; however, wearable device data for diverse populations are scarce. We therefore aimed to describe and compare key wearable signals (ie, heart rate, step count, and home blood pressure measurements) across age, sex, race, ethnicity, and clinical phenotypes. METHODS: In the Michigan Predictive Activity & Clinical Trajectories in Health (MIPACT) prospective observational study, we enrolled participants from Michigan Medicine, Ann Abor, MI, USA, and followed them up for at least 90 days. Patients were included if they were aged 18 years or older, were fluent in English, owned an iPhone 6 or newer model with a supported iOS version, and had regular access to the internet throughout the study period. All participants were provided with an Apple Watch Series 3 or 4, an Omron Evolv Wireless Blood Pressure Monitor, and the MyDataHelps study smartphone application. Participants were asked to wear their watch for 12 h per day or longer and to do daily or weekly tasks, including home blood pressure measurements and breathing tasks. Heart rate, blood pressure, step counts, and distance walked were collected. The study was divided into two phases: an intensive 45-day collection phase (phase 1); and a 3-year longitudinal monitoring phase (phase 2). Here we report the first 90 days of data for all participants, which includes all of phase 1 and the first 45 days of phase 2. Participants' electronic health records were used to establish clinical diagnoses for analysis. FINDINGS: We enrolled 6765 eligible participants between Aug 14, 2018, and Dec 19, 2019, of whom 6454 participants from Michigan Medicine completed the phase 1 study protocol and were included in this analysis (3482 [54%] women and 2972 [46%] men; 3657 [57%] participants were White, with 1094 [17%] Asian and 1090 [17%] Black participants). On days when participants wore their smart watches, median daily watch wear time was 15·5 h (IQR 14-17). Participants contributed a total of 1 107 320 blood pressure and 202 198 347 heart rate measurements over 90 days, with 172 (SD 50) blood pressure and 31 329 (SD 24 620) heart rate measurements per participant. Mean systolic blood pressure was 122 mm Hg (SD 10) and mean diastolic blood pressure was 77 mm Hg (SD 8), with 167 312 (15%) measurements having a systolic blood pressure higher than 140 mm Hg or diastolic blood pressure higher than 90 mm Hg. Mean resting heart rate was 64 beats per min (SD 8). Blood pressure and resting heart rate varied by sex, age, race, and ethnicity, with higher blood pressures in males and lower heart rate in participants aged 65 years or older (p<0·0001). Participants took 7511 steps per day (SD 2805) and walked 6009 metres per day (SD 2608), varying across demographic and clinical subgroups. INTERPRETATION: These data could inform clinical trial design, interpretation of wearable data in clinical practice, and health-care interventions. FUNDING: Apple, University of Michigan.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“multi-modal sensor signals are not yet available for population-scale studies with longitudinal follow-up beyond a few years20.”
From Self-supervised learning of accelerometer data provides new insights for sleep and its association with mortality (2024) · cites this paper as [20]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation
Perez, M. V., Mahaffey, K. W., Hedlin, H., Rumsfeld, J. S., Garcia, A., Ferris, T., Balasubramanian, V., Russo, A. M., Rajmane, A., Cheung, L., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1901183 · Cited by 1,917
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715
(2019) · Cited by 633
Association of Step Volume and Intensity With All-Cause Mortality in Older Women
Lee, I.-M., Shiroma, E. J., Kamada, M., Bassett, D. R., Matthews, C. E., & Buring, J. E. (2019) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2019.0899 · Cited by 604
Wearable sensor data and self-reported symptoms for COVID-19 detection
Quer, G., Radin, J. M., Gadaleta, M., Baca-Motes, K., Ariniello, L., Ramos, E., Kheterpal, V., Topol, E. J., & Steinhubl, S. R. (2021) · Nature Medicine · doi:10.1038/s41591-020-1123-x · Cited by 447
Wearable-device-measured physical activity and future health risk
Strain, T., Wijndaele, K., Dempsey, P. C., Sharp, S. J., Pearce, M., Jeon, J., Lindsay, T., Wareham, N., & Brage, S. (2020) · Nature Medicine · doi:10.1038/s41591-020-1012-3 · Cited by 355
Radin, J. M., Wineinger, N. E., Topol, E. J., & Steinhubl, S. R. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30222-5 · Cited by 300
McConnell, M. V., Shcherbina, A., Pavlovic, A., Homburger, J. R., Goldfeder, R. L., Waggot, D., Cho, M. K., Rosenberger, M. E., Haskell, W. L., Myers, J., et al. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.4395 · Cited by 239
Real-world heart rate norms in the Health eHeart study
Avram, R., Tison, G. H., Aschbacher, K., Kuhar, P., Vittinghoff, E., Butzner, M., Runge, R., Wu, N., Pletcher, M. J., Marcus, G. M., et al. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0134-9 · Cited by 207
Quer, G., Gouda, P., Galarnyk, M., Topol, E. J., & Steinhubl, S. R. (2020) · PLOS ONE · doi:10.1371/journal.pone.0227709 · Cited by 185
Cited by
Acosta, J. N., Falcone, G. J., Rajpurkar, P., & Topol, E. J. (2022) · Nature Medicine · doi:10.1038/s41591-022-01981-2 · Cited by 1,365
“The Michigan Predictive Activity & Clinical Trajectories in Health (MIPACT) study constitutes another example, with participants contributing data from wearables, physiological data (blood pressure), clinical information (EHR and surveys) and laboratory data 101 .”
BackgroundDigital Phenotyping in Health Using Machine Learning Approaches: Scoping Review
Dlima, S. D., Shevade, S., Menezes, S. R., & Ganju, A. (2022) · JMIR Bioinformatics and Biotechnology · doi:10.2196/39618 · Cited by 61
Li, S., Jiang, C., Wang, Y., Lai, Y., Zhao, M., Li, Q., Bai, Y., Dai, W., Guo, Q., He, L., et al. (2023) · Hypertension · doi:10.1161/hypertensionaha.122.20711 · Cited by 41
Yuan, H., Plekhanova, T., Walmsley, R., Reynolds, A. C., Maddison, K. J., Bucan, M., Gehrman, P., Rowlands, A., Ray, D. W., Bennett, D., et al. (2024) · npj Digital Medicine · doi:10.1038/s41746-024-01065-0 · Cited by 40
“Detailed population characteristics and inclusion criteria are listed in Supplementary Section 1.1.”
BackgroundInfluentialWang, J., Jiang, C., Li, S., Wang, Z., Wang, Y., Lai, Y., Wang, Z., Lv, W., Bai, Y., Yang, Z., et al. (2023) · Hypertension · doi:10.1161/hypertensionaha.123.21651 · Cited by 40
Golbus, J. R., GOSCH, K. L., Birmingham, M. C., Butler, J., Lingvay, I., Lanfear, D. E., Abbate, A., Kosiborod, M., Damaraju, C. V., Januzzi, J. L., et al. (2023) · JACC Heart Failure · doi:10.1016/j.jchf.2023.05.033 · Cited by 46
Self-citationModeling personalized heart rate response to exercise and environmental factors with wearables data
Nazaret, A., Tonekaboni, S., Darnell, G., Ren, S., Sapiro, G., & Miller, A. C. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00926-4 · Cited by 48
“In the public health domain, wearables data and ML algorithms have been successfully deployed in disease surveillance 8 such as detecting in fl uenza-like illness 9 , monitoring disease progression in the population 10,11 , and informing the design of new epidemiological studies 12 .”
MethodLeveraging Digital Health to Improve the Cardiovascular Health of Women
Azizi, Z., Adedinsewo, D. A., Rodríguez, F., Lewey, J., Merchant, R. M., & Brewer, L. C. (2023) · Current Cardiovascular Risk Reports · doi:10.1007/s12170-023-00728-z · Cited by 26
“In the MIPACT study by Golbus et al. [31], the authors aimed to monitor CVH behaviors obtained from wearable device signals in a large diverse population of men and women.”
BackgroundFactors associated with long-term use of digital devices in the electronic Framingham Heart Study
Pathiravasan, C. H., Zhang, Y., Wang, X., Trinquart, L., Benjamin, E. J., Borrelli, B., McManus, D. D., Kheterpal, V., Lin, H., Spartano, N. L., et al. (2022) · npj Digital Medicine · doi:10.1038/s41746-022-00735-1 · Cited by 22
Bonilla, D. A., Peralta-Alzate, J. O., Bonilla-Henao, J. A., Cannataro, R., Cardozo, L. A., Vargas-Molina, S., Stout, J. R., Kreider, R. B., & Petro, J. L. (2023) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk8020044 · Cited by 18
“The Michigan Predictive Activity & Clinical Trajectories in Health (MIPACT) study is a good example of how data collected from wearable devices can inform clinical trial design, interpretation in clinical practice, and health-care interventions [72].”
BackgroundMobile health technologies in the prevention and management of hypertension: A scoping review
Babatunde, A. O., Ogundijo, D., Afolayan, A.-G. O., Awosiku, O. V., Aderohunmu, Z., Oguntade, M. S., Alao, U. H., Oseni, A. O., Akintola, A. A., & Amusat, O. (2024) · Digital Health · doi:10.1177/20552076241277172 · Cited by 27
Qi, W., Shen, S., Dong, C., Zhao, M., Zang, S., Zhu, X., Li, J., Wang, B., Shi, Y., Dong, Y., et al. (2025) · Journal of Medical Internet Research · doi:10.2196/71560 · Cited by 22
“Ongoing advancements in digital devices, such as wearable devices and sensors, provide higher-quality and more granular data sources [103], enabling a more accurate portrayal of behaviors of interest in both clinical and research settings [104].”
Zhang, Y., Wang, X., Pathiravasan, C. H., Spartano, N. L., Lin, H., Borrelli, B., Benjamin, E. J., McManus, D. D., Larson, M. G., Vasan, R. S., et al. (2024) · Journal of Medical Internet Research · doi:10.2196/56676 · Cited by 13
Frontiers of medical decision-making in the modern age of data analytics
Denton, B. T. (2022) · IISE Transactions · doi:10.1080/24725854.2022.2092918 · Cited by 17
“For example, Golbus et al. (2021) recently published preliminary research based on a cohort of patients who were recruited to wear Apple watches and use additional devices, including a home blood pressure monitoring device, to collect observational data daily for an extended period to monitor…”
Smart wearable devices for real-time health monitoring
Dhar, R., Kumar, A., & Karmakar, S. (2023) · Asian Journal of Medical Sciences · doi:10.3126/ajms.v14i12.58664 · Cited by 12
Development of an AI based automated analysis of pediatric Apple Watch iECGs
Teich, L., Franke, D. J., Michaelis, A., Dähnert, I., Gebauer, R., Markel, F., & Paech, C. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2023.1185629 · Cited by 11
“Along with the progressive emergence of smart devices like the Apple Watch, people can monitor their health semi-professionally (13 – 16).”
BackgroundPopham, S. F., Burq, M., Rainaldi, E., Shin, S., Dunn, J., & Kapur, R. (2023) · JMIR Biomedical Engineering · doi:10.2196/43726 · Cited by 10
Ratitch, B., Kunkoski, E., Fontanari, A., Onorati, F., Dabral, A., Rodriguez-Chavez, I. R., Vandendriessche, B., Vega, J., Morton, S., & Damestani, Y. (2022) · Digital Biomarkers · doi:10.1159/000525897 · Cited by 10
Golbus, J. R., Gupta, K., Stevens, R., Jeganathan, V. S. E., Luff, E., Boyden, T. F., Mukherjee, B., Klasnja, P., Kheterpal, S., Kohnstamm, S., et al. (2022) · Circulation Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.122.009182 · Cited by 12
Self-citationKim, D. S., Schütz, N., Johnson, A., Tolas, A., Mantena, S., O’Sullivan, J. W., Hershman, S. G., MYERS, J. N., Christle, J. W., Oppezzo, M., et al. (2025) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2025.01.010 · Cited by 11
Marino, S., Cassidy, R. B., Nanni, J., Wang, Y., Liu, Y., Tang, M., Yuan, Y., Chen, T. D., Sinha, A. K., Pandian, B., et al. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01935-1 · Cited by 14
“Wearable data was collected using Apple Watch Series3or4(Apple,Cupertino,CA,USA),anOmronEvolvWirelessBlood PressureMonitor(Omron,Kyoto,Japan),andtheMyDataHelpsstudy smartphoneapplication(CareEvolution,AnnArbor,MI,USA),asdescri-bedpreviously 29,31 .”
Standardizing Physical Activity Monitoring in Patients With Degenerative Lumbar Disorders
Maldaner, N., Tang, M., Fatemi, P., Leung, C., Desai, A. M., Tomkins‐Lane, C., & Zygourakis, C. C. (2023) · Neurosurgery · doi:10.1227/neu.0000000000002755 · Cited by 9
“Age-and sex-speci fi c reference population mean and SD step count values from the MIPACT study that were used to calculate standard z-scores have recently been published by Golbus et al 2 and are not presented here.”
MethodSeffah, K. D., Zaman, M. A., Awais, N., Satnarine, T., Haq, A., Hernandez, G. N., & Khan, S. (2023) · Cureus · doi:10.7759/cureus.36754 · Cited by 9
Xiao, H., Ding, K., Li, X., Zhou, Z., Ma, Y., Dai, X., Liu, Y., & Chen, D. (2025) · Sleep Medicine · doi:10.1016/j.sleep.2025.01.034 · Cited by 7
Linking coastal environmental and health observations for human wellbeing
Sandifer, P. A. (2023) · Frontiers in Public Health · doi:10.3389/fpubh.2023.1202118 · Cited by 11
Luo, X., Tahabi, F. M., Rollins, D. M., & Sawchuk, A. P. (2024) · JVS Vascular Science · doi:10.1016/j.jvssci.2024.100192 · Cited by 6
Ziga, M., Bättig, L., Gmeiner, R., Yurtsever, E., Stienen, M. N., Netuka, D., Schneider, U. C., Maldaner, N., & Petr, O. (2026) · Brain and Spine · doi:10.1016/j.bas.2026.106207 · Cited by 5
How Could Sensor-Based Measurement of Physical Activity Be Used in Cardiovascular Healthcare?
Hughes, M. E., & Chico, T. (2023) · Sensors · doi:10.3390/s23198154 · Cited by 5
“Providing support and training could increase acceptability [63].”
BackgroundStandardizing Continuous Physical Activity Monitoring in Patients with Cervical Spondylosis
Maldaner, N., Gonzalez-Suarez, A. D., Tang, M., Fatemi, P., Leung, C., Desai, A. M., Tomkins‐Lane, C., & Zygourakis, C. C. (2024) · Spine · doi:10.1097/brs.0000000000004940 · Cited by 6
“6 Understanding the impact of CS on patients ’ baseline physical activity is critical in two ways.”
Rong, J., Pathiravasan, C. H., Zhang, Y., Faro, J. M., Wang, X., Schramm, E., Borrelli, B., Benjamin, E. J., Liu, C., & Murabito, J. M. (2024) · JMIR Aging · doi:10.2196/64636 · Cited by 5
“However, older adults have been underrepresented in electronic cohort (eCohort) [10-12] with few digital health studies including larger numbers of older adults [13].”
Gala, A. R. P. B. E., Sharp, A. J., Pope, M., Leo, M., Varini, R., Paisey, J., Curzen, N. P., Banerjee, A., & Betts, T. R. (2025) · Heart Rhythm · doi:10.1016/j.hrthm.2025.04.015 · Cited by 7
Golbus, J. R., Basu, T., Luff, E., Hu, Y., Zeng, D., Gesierich, C., Resnicow, K., Klasnja, P., & Nallamothu, B. K. (2025) · Circulation Heart Failure · doi:10.1161/circheartfailure.125.013082 · Cited by 5
Self-citationRelatability of Blood Pressure Monitoring With Wearable Cuffless Devices
Chrysant, S. G. (2022) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2021.12.028 · Cited by 4
Nikolovski, J., Návar, A. M., Steinhubl, S. R., Baca-Motes, K., Curtis, A. B., Peterson, E. D., Lakkireddy, D., Tarino, M., Juan, S., Damaraju, C. V., et al. (2025) · JACC Advances · doi:10.1016/j.jacadv.2025.102072 · Cited by 3
Keshmiri, S., Tomonaga, S., Mizutani, H., & Doya, K. (2024) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2024.108335 · Cited by 4
Yuan, H., Plekhanova, T., Walmsley, R., Reynolds, A. C., Maddison, K. J., Bucan, M., Gehrman, P., Rowlands, A., Ray, D. W., Bennett, D., et al. (2023) · doi:10.1101/2023.07.07.23292251 · Cited by 3
Users’ Expectations of Smart Devices during Physical Activity—A Literature Review
Tóth, K., Takács, P., & Balatoni, I. (2024) · Applied Sciences · doi:10.3390/app14083518 · Cited by 5
Lutz, L., Veltmann, P., Meisterhans, R., & Giurgiu, M. (2025) · Frontiers in Psychology · doi:10.3389/fpsyg.2025.1623886 · Cited by 3
Pastapur, A., Pescatore, N. A., Shah, N., Kheterpal, S., Nallamothu, B., & Golbus, J. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1243574 · Cited by 3
Self-citationGonzalez-Suarez, A. D., Maldaner, N., Tang, M., Fatemi, P., Leung, C., Desai, A. M., Tomkins‐Lane, C., Han, S. S., & Zygourakis, C. C. (2024) · Scientific Reports · doi:10.1038/s41598-024-70912-7 · Cited by 3
Spector‐Bagdady, K. (2023) · The Journal of Law Medicine & Ethics · doi:10.1017/jme.2023.74 · Cited by 3
da Silva, J. H. B., Germano-Soares, A. H., de Oliveira, T. V., Silva, L. C. B., Filho, V. C. B., Silva, T. C. D. A., & Tassitano, R. M. (2026) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2026.106370 · Cited by 1
Xiao, H., Ding, K., Li, X., Ma, Y., Liu, Y., Huang, J., & Chen, D. (2026) · BMC Medicine · doi:10.1186/s12916-026-04830-y · Cited by 3
Digital Phenotyping in Health Research: Scoping Review of Methods, Gaps, and Opportunities
Badier, N., Lafitte, A., Difernand, A., Aguayo, G., Fagherazzi, G., Onnela, J.-P. “., Vittrant, B., Lechat, B., Larochelambert, Q. D., Toussaint, J., et al. (2026) · Journal of Medical Internet Research · doi:10.2196/91747 · Cited by 1
Another step (count) towards leveraging mobile health data for clinical prediction
Murphy, A., & Low, C. A. (2021) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00212-0 · Cited by 1
The Promise and Pitfalls of Big Data Studies in Perioperative Medicine
Mathis, M. R., Janda, A. M., & Kheterpal, S. (2023) · Elsevier eBooks · doi:10.1016/b978-0-323-83399-8.00005-3 · Cited by 1
Self-citationKim, D. S., Schütz, N., Johnson, A., Tolas, A., Mantena, S., O’Sullivan, J. W., Hershman, S. G., King, A. C., Christle, J. W., Oppezzo, M., et al. (2024) · medRxiv · doi:10.1101/2024.06.26.24309535 · Cited by 2
Data continuity and linkage in the healthcare ecosystem
Charles, W. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-824010-6.00020-4 · Cited by 1
Keshmiri, S., Tomonaga, S., Mizutani, H., & Doya, K. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.01.21.576502
Rong, J., Pathiravasan, C. H., Zhang, Y., Faro, J. M., Wang, X., Schramm, E., Borrelli, B., Benjamin, E. J., Liu, C., & Murabito, J. M. (2024) · doi:10.2196/preprints.64636
AN EXAMINATION OF THE IMPLEMENTATION OF INTERNET OF THINGS IN HEALTHCARE UTILIZING SMARTWATCHES
Boshrabadi, F. S., Abolhassani, M., Shafaghi, S., Ghorbani, F., & Shafaghi, M. (2024) · Jurnal Administrasi Kesehatan Indonesia · doi:10.20473/jaki.v12i2.2024.292-304 · Cited by 1
“Using a smartwatch compatible with IoT, caregivers can monitor the patient’s practices remotely and assess the progress using the old data and the new data collected by the gadget (Golbus et al. , 2021).”
Beker, L., Toymuş, A. T., Peker, S., Atik, A. Y., Yener, U. C., Albay, M. M., Yılgör, E., & Yılgör, İ. (2025) · Research Square · doi:10.21203/rs.3.rs-5633578/v1
Yurtsever, E., Bättig, L., Gmeiner, R., Stienen, M. N., Schneider, U. C., Petr, O., Maldaner, N., & Ziga, M. (2025) · The Spine Journal · doi:10.1016/j.spinee.2025.12.017
Toymuş, A. T., Peker, S., Atik, A. Y., Yener, U. C., Albay, M. M., Yılgör, E., YILGÖR, I., & Beker, L. (2026) · Device · doi:10.1016/j.device.2026.101076
Tolas, A., Johnson, A., Schütz, N., Ramírez‐Posada, M., Montalvo, S., Yang, E., Schmiedmayer, P., Rodríguez, F., Lawrie, A., MYERS, J. N., et al. (2026) · doi:10.2196/preprints.102176
Francisco, M. D. Á. (2026) · Revista Master Ensino Pesquisa e Extensão · doi:10.47224/revistamaster.v10i20.806
Metadata
- Authors
- Golbus, J. R., Pescatore, N. A., Nallamothu, B. K., Shah, N., & Kheterpal, S.
- Affiliations
- University of Michigan (US)
- Corresponding author
- Kheterpal, Sachin
- Venue
- The Lancet Digital Health
- Publisher
- Elsevier BV
- Published
- 2021-11
- Volume
- 3
- Issue
- 11
- Pages
- e707-e715
- ISSN
- 2589-7500 (ISSN-L)
- OpenAlex
- W3209205731
- Cited by
- 57
- References
- 18
- Citations by year
- 2026: 7 · 2025: 10 · 2024: 13 · 2023: 19 · 2022: 7 · 2021: 1
- FWCI
- 3.04
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by-nc-nd
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- creativecommons.org/licenses/by-nc-nd/4.0/ (vor)
- Locations
- The Lancet Digital Health (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Merck
- Edwards Lifesciences
- National Institutes of Health (N025926)
- University of Michigan
- BD
- American Heart Association (54423, US 9 962 124)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Keywords
- wearable devices, Apple Watch, home blood pressure monitoring, heart rate, step count, physical activity, racial and ethnic differences, sex differences, age differences, clinical phenotype, smartphone health monitoring, health disparities, diabetes mellitus, chronic kidney disease, cardiovascular disease, respiratory diseases, cancer
- MeSH terms
- Smartphone, Wearable Electronic Devices, Digital Technology, Adult, Aged, Aged, 80 and over, Blood Pressure, Ethnicity, Female, Health Status, Heart Rate, Humans, Male, Michigan, Middle Aged, Phenotype, Prospective Studies, Walking, Telemedicine, Monitoring, Ambulatory, Racial Groups, Young Adult, Electronic Health Records, Mobile Applications
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.