A C60 Primary Ion Beam System for Time of Flight Secondary Ion Mass Spectrometry: Its Development and Secondary Ion Yield Characteristics
Abstract
A buckminsterfullerene (C60)-based primary ion beam system has been developed for routine application in TOF-SIMS analysis of organic materials. The ion beam system is described, and its performance is characterized. Nanoamp beam currents of C60+ are obtainable in continuous current mode. C60(2+) can be obtained in pulsed mode. At 10 keV, the beam can be focused to less than 3 microm with 0.1 nA currents. TOF-SIMS studies of a series of molecular solids and a number of polymer systems in monolayer and thick film forms are reported. Very significant enhancement of secondary ion yields, particularly at higher mass, were observed using 10-keV C60+ for all samples other than PTFE, as compared to those observed from 10 keV Ga+ primary ions. Three materials (PS2000, Irganox 1010, PET) were studied in detail to investigate primary ion-induced disappearance (damage) cross sections to determine the increase in secondary ion formation efficiency. The C60 disappearance cross sections observed from monolayer film PS2000 and self-supporting PET film are close to those observed from Ga+. The resulting C60 efficiencies are 30-100 times those observed from gallium. The cross sections observed from C60 bombardment of multilayer molecular solids are approximately 100 times less, such that essentially zero damage sputtering is possible. The resulting efficiencies are > 10(3) greater than from gallium. It is also shown that C60 primary ions do not generate any more low-mass fragments than any other ion beam system does. C60 is shown to be a very favorable ion beam system for TOF-SIMS, delivering high yield, close to 10% total yield, favoring high-mass ions, and on thick samples, offering the possibility of analysis well beyond the static limit.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Matrix-assisted laser desorption/ionization mass spectrometry of biopolymers.
Hillenkamp, F., Karas, M., Beavis, R. C., & Chait, B. T. (1991) · Analytical Chemistry · doi:10.1021/ac00024a716 · Cited by 1,930
Benninghoven, A. (1994) · Angewandte Chemie International Edition in English · doi:10.1002/anie.199410231 · Cited by 535
Atomic and Molecular Imaging at the Single-Cell Level with TOF-SIMS
Colliver, T. L., Brummel, C. L., Pacholski, M. L., Swanek, F. D., Ewing, A. G., & Winograd, N. (1997) · Analytical Chemistry · doi:10.1021/ac9701748 · Cited by 260
Development and experimental application of a gold liquid metal ion source
Davies, N., Weibel, D. E., Blenkinsopp, P., Lockyer, N., Hill, R., & Vickerman, J. C. (2003) · Applied Surface Science · doi:10.1016/s0169-4332(02)00631-1 · Cited by 223
Self-citationBrummel, C. L., Vickerman, J. C., Carr, S. A., Hemling, M. E., Roberts, G. D., Johnson, W., Weinstock, J., Gaitanopoulos, D. E., Benkovic, S. J., & Winograd, N. (1996) · Analytical Chemistry · doi:10.1021/ac950998u · Cited by 86
Self-citationSingle Photon IonisationMass Spectrometry Using Laser‐Generated Vacuum Ultraviolet Photons
Lockyer, N. P., & Vickerman, J. C. (1997) · Laser Chemistry · doi:10.1155/1997/53174 · Cited by 57
Self-citationDesorption of molecules by cluster impact: A preliminary molecular dynamics study
Kerford, M., & Webb, R. P. (2001) · Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms · doi:10.1016/s0168-583x(01)00395-0 · Cited by 42
Cited by
McDonnell, L. A., & Heeren, R. M. A. (2007) · Mass Spectrometry Reviews · doi:10.1002/mas.20124 · Cited by 1,082
Cluster secondary ion mass spectrometry of polymers and related materials
Mahoney, C. M. (2010) · Mass Spectrometry Reviews · doi:10.1002/mas.20233 · Cited by 293
Mass spectrometry imaging for drug distribution studies
Prideaux, B., & Stoeckli, M. (2012) · Journal of Proteomics · doi:10.1016/j.jprot.2012.07.028 · Cited by 279
Körber, A., Anthony, I. G. M., & Heeren, R. M. A. (2025) · Analytical Chemistry · doi:10.1021/acs.analchem.4c05249 · Cited by 88
Metadata
- Authors
- Weibel, D., Wong, S., Lockyer, N., Blenkinsopp, P., Hill, R., & Vickerman, J. C.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2003-03-04
- Volume
- 75
- Issue
- 7
- Pages
- 1754-1764
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2065657816
- Cited by
- 543
- References
- 8
- FWCI
- 47.58
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A C60 Primary Ion Beam System for Time of Flight Secondary Ion Mass Spectrometry: Its Development and Secondary Ion Yield Characteristics. https://app.cochat.ai/papers/10.1021/ac026338o
- The paper
- Weibel, D., Wong, S., Lockyer, N., Blenkinsopp, P., Hill, R., & Vickerman, J. C. (2003). A C60 Primary Ion Beam System for Time of Flight Secondary Ion Mass Spectrometry: Its Development and Secondary Ion Yield Characteristics. Analytical Chemistry. https://doi.org/10.1021/ac026338o
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.