Thin-Film Microbiosensors for Glucose−Lactate Monitoring
The DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
A miniaturized device for simultaneous measurement of glucose and lactate levels was produced by means of photopatterning of enzyme-containing photosensitive membrane precursors. This device shows no cross-talk and a lifetime for both the glucose and the lactate sensors of more than 2 weeks when continuously operated in undiluted bovine serum. Linear response ranges of up to 40 mM for glucose and 25 mM for L-lactate, in combination with 95% response times of < 30 s, were realized. The devices are mass produced by means of thin-film technology on flexible carriers to give catheter-type multisensing devices for in vivo applications. Ex vivo experiments, performed with human volunteers, where the device was continuously operated in an extracorporeal, undiluted, heparinized blood stream for 6 h, gave a correlation of r > 0.98 with respect to laboratory techniques. Subcutaneous measurements of glucose levels in pigs were close to the corresponding blood levels obtained without in vivo calibration.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Figure 2g shows that the responses of glucose and lactate sensors increase rapidly upon elevation of the solution temperature from 22 °C to 40 °C, reflecting the effect of increased enzyme activities26.”
From Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis (2016) · cites this paper as [26]
Citations
Cites
Cortical evoked potential and extracellular K+ and H+ at critical levels of brain ischemia.
Astrup, J., Symon, L., Branston, N. M., & Lassen, N. A. (1977) · Stroke · doi:10.1161/01.str.8.1.51 · Cited by 1,120
PEN-SIZED DIGITAL 30-SECOND BLOOD GLUCOSE METER
Matthews, D. R., Bown, E., Watson, A., Holman, R. R., Steemson, J., Hughes, S., & Scott, D. (1987) · The Lancet · doi:10.1016/s0140-6736(87)92802-9 · Cited by 300
Evaluation of a novel point-of-care system, the i-STAT portable clinical analyzer
Erickson, K. A., & Wilding, P. (1993) · Clinical Chemistry · doi:10.1093/clinchem/39.2.283 · Cited by 248
Electropolymerized films to prevent interferences and electrode fouling in biosensors
Geise, R. J., Adams, J. M., Barone, N. J., & Yacynych, A. M. (1991) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(91)87039-e · Cited by 169
Moussy, F., Harrison, D. J., O'Brien, D. W., & Rajotte, R. V. (1993) · Analytical Chemistry · doi:10.1021/ac00063a023 · Cited by 158
In vivo biocompatibility of collagenpoly(hydroxyethyl methacrylate) hydrogels
Jeyanthi, R., & Rao, K. P. (1990) · Biomaterials · doi:10.1016/0142-9612(90)90004-a · Cited by 130
Moatti-Sirat, D., Capron, F., Poitout, V., Reach, G., Bindra, D. S., Zhang, Y., Wilson, G. S., & Th�venot, D. R. (1992) · Diabetologia · doi:10.1007/bf00400921 · Cited by 123
A Continuous, Implantable Lactate Sensor
Baker, D. A., & Gough, D. A. (1995) · Analytical Chemistry · doi:10.1021/ac00105a010 · Cited by 83
Urban, G. A., Jobst, G., Keplinger, F., Aschauer, E., Tilado, O., Fasching, R. J., & Köhl, F. (1992) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(92)85056-g · Cited by 74
Self-citationIntegrated biosensor for glucose and galactose
Yokoyama, K., Sode, K., Tamiya, E., & Karube, I. (1989) · Analytica Chimica Acta · doi:10.1016/s0003-2670(00)80291-3 · Cited by 68
Research works printed but not linked: 9
- Kimura J., 1988, Biosensors, 4, 52
- Tamiya E., 1989, Sens. Actuators B, 18, 307
- Mastrototaro J., 1991, Sens. Actuators B, 1, 145
- Wilke, D.; Müller, H. InTrends in electrochemical Biosensors;Costa, G., Miertus, S., Eds.; World Scientific: Singapore, 1992; pp 155−162.
- Trajanoski Z., Proceedings of the 16th Annual International Conference of the IEEE Engineering in Medicine and Biology Society; IEEE: Piscataway, NJ
- Pfeiffer E. F., 1989, Verh. Dtsch. Ges. Inn. Med., 95, 736
- Clarke W. L., 1984, Diabetes Care, 5, 627
- Varahram M., 1995, Biomed. Tech., 40, 101
- Ammon H. P. T., 1995, Anal. Chem., 67, 471
Cited by
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Hydrogels in Biology and Medicine: From Molecular Principles to Bionanotechnology
Peppas, N. A., Hilt, J. Z., Khademhosseini, A., & Langer, R. (2006) · Advanced Materials · doi:10.1002/adma.200501612 · Cited by 4,109
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 712
Current research activity in biosensors
Nakamura, H., & Karube, I. (2003) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-003-1947-5 · Cited by 299
Ag/AgCl microelectrodes with improved stability for microfluidics
Polk, B. J., Stelzenmuller, A., Mijares, G., MacCrehan, W., & Gaitan, M. (2006) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2005.03.121 · Cited by 246
Microsensor systems for cell metabolism – from 2D culture to organ-on-chip
Kieninger, J., Weltin, A., Flamm, H., & Urban, G. A. (2018) · Lab on a Chip · doi:10.1039/c7lc00942a · Cited by 198
Self-citationBiosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine
Moser, I. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00298-6 · Cited by 181
Self-citationXu, G., Cheng, C., Liu, Z., Yuan, W., Wu, X., Lu, Y., Low, S. S., Liu, J., Zhu, L., Ji, D., et al. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201800658 · Cited by 172
Schuhmann, W. (2002) · Reviews in Molecular Biotechnology · doi:10.1016/s1389-0352(01)00058-7 · Cited by 166
Bollella, P., Sharma, S., Cass, A. E. G., & Antiochia, R. (2019) · Electroanalysis · doi:10.1002/elan.201800630 · Cited by 151
Electroanalysis of Glucose in Transcutaneously Extracted Samples
Tierney, M. J., Kim, H. L., Burns, M. D., Tamada, & Potts, R. O. (2000) · Electroanalysis · doi:10.1002/1521-4109(200005)12:9<666::aid-elan666>3.0.co;2-c · Cited by 68
Nikolaus, N., & Strehlitz, B. (2007) · Microchimica Acta · doi:10.1007/s00604-007-0834-8 · Cited by 142
Palmisano, F., Rizzi, R., Centonze, D., & Zambonin, P. G. (2000) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(00)00107-x · Cited by 140
Pettersen, E. O., Ebbesen, P., Gieling, R. G., Williams, K. J., Dubois, L., Lambin, P., Ward, C., Meehan, J., Kunkler, I. H., Langdon, S. P., et al. (2015) · Journal of Enzyme Inhibition and Medicinal Chemistry · doi:10.3109/14756366.2014.966704 · Cited by 115
Self-citationPolymer-based, flexible glutamate and lactate microsensors for in vivo applications
Weltin, A., Kieninger, J., Enderle, B., Gellner, A., Fritsch, B., & Urban, G. A. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.05.014 · Cited by 117
Self-citationMicrofabricated, amperometric, enzyme-based biosensors for in vivo applications
Weltin, A., Kieninger, J., & Urban, G. A. (2016) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-016-9420-4 · Cited by 115
Self-citationTransparent and Flexible Triboelectric Sensing Array for Touch Security Applications
Yuan, Z., Zhou, T., Yin, Y., Cao, R., Li, C., & Wang, Z. L. (2017) · ACS Nano · doi:10.1021/acsnano.7b03680 · Cited by 108
Sartain, F. K., Yang, X., & Lowe, C. R. (2006) · Analytical Chemistry · doi:10.1021/ac060416g · Cited by 101
The enzyme thermistor—A realistic biosensor concept. A critical review
Yakovleva, M., Bhand, S., & Danielsson, B. R. G. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2012.12.004 · Cited by 103
Thin-Film Glucose Biosensor Based on Plasma-Polymerized Film: Simple Design for Mass Production
Muguruma, H., Hiratsuka, A., & Karube, I. (2000) · Analytical Chemistry · doi:10.1021/ac000014n · Cited by 101
Lactate and Sequential Lactate−Glucose Sensing Using Surface-Enhanced Raman Spectroscopy
Shah, N. C., Lyandres, O., Walsh, J. T., Glucksberg, M. R., & Duyne, R. P. V. (2007) · Analytical Chemistry · doi:10.1021/ac0704107 · Cited by 92
Patel, N. G., Erlenkötter, A., Cammann, K., & Chemnitius, G. C. (2000) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(00)00410-x · Cited by 87
Killoran, S. J., & O'Neill, R. D. (2008) · Electrochimica Acta · doi:10.1016/j.electacta.2008.03.076 · Cited by 85
Recycling Systems Based on Screen-Printed Electrodes
Lisdat, F., Ho, W. O., Wollenberger, U., Scheller, F. W., Richter, T., & Bilitewski, U. (1998) · Electroanalysis · doi:10.1002/(sici)1521-4109(199809)10:12<808::aid-elan808>3.0.co;2-k · Cited by 29
Matsumoto, T., Ohashi, A., & Ito, N. (2002) · Analytica Chimica Acta · doi:10.1016/s0003-2670(02)00334-3 · Cited by 78
Yan, J., Pedrosa, V. A., Simonian, A. L., & Revzin, A. (2010) · ACS Applied Materials & Interfaces · doi:10.1021/am9007819 · Cited by 77
Miyashita, M., Ito, N., Ikeda, S., Murayama, T., Oguma, K., & Kimura, J. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2008.07.072 · Cited by 75
Micro- and nanobiosensors—state of the art and trends
Urban, G. A. (2008) · Measurement Science and Technology · doi:10.1088/0957-0233/20/1/012001 · Cited by 95
Self-citationPoly(pyrrole–biotin): a new polymer for biomolecule grafting on electrode surfaces
Cosnier, S., & Lepellec, A. (1999) · Electrochimica Acta · doi:10.1016/s0013-4686(98)00294-1 · Cited by 71
Microscale hydrogels for medicine and biology: synthesis, characteristics and applications
Rivest, C., Morrison, D. W. G., Ni, B., Rubin, J., Yadav, V. G., Mahdavi, A., Karp, J. M., & Khademhosseini, A. (2007) · Journal of mechanics of materials and structures · doi:10.2140/jomms.2007.2.1103 · Cited by 77
Biosensors for L-malate and L-lactate based on solid binding matrix
Katrlík, J., Pizzariello, A., Mastihuba, V., Švorc, J., Streďanský, M., & Miertuš, S. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(98)00610-2 · Cited by 65
Hossain, M. F., & Slaughter, G. E. (2020) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2020.113990 · Cited by 64
Micromachined sensor for lactate monitoring in saliva
Schabmueller, C. G. J., Loppow, D., Piechotta, G., Schütze, B., Albers, J., & Hintsche, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2005.09.015 · Cited by 59
Wáng, Y., He, Q., Dong, Y., & Chen, H. (2009) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2009.11.068 · Cited by 58
Intelligent Soft Sweat Sensors for the Simultaneous Healthcare Monitoring and Safety Warning
Huang, X., Liu, Y., Park, W., Li, J., Ma, J., Yiu, C. K., Zhang, Q., Li, J., Wu, P., Zhou, J., et al. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202202846 · Cited by 57
Vilkanauskyte, A., Erichsen, T. J., Marcinkevičienė, L., Laurinavičius, V., & Schuhmann, W. W. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00095-7 · Cited by 56
Pontius, K., Semenova, D., Silina, Y. E., Gernaey, K. V., & Junicke, H. (2020) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.00436 · Cited by 53
Mass producible miniaturized flow through a device with a biosensor array
Jobst, G., Moser, I., Svasek, P., Varahram, M., Trajanoski, Z., Wach, P., Kotanko, P., Skrabal, F., & Urban, G. (1997) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(97)00080-4 · Cited by 51
Self-citationNon-enzymatic glucose sensing based on hierarchical platinum micro-/nanostructures
Unmüssig, T., Weltin, A., Urban, S., Daubinger, P., Urban, G. A., & Kieninger, J. (2018) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2018.03.061 · Cited by 49
Self-citationSenillou, A., Jaffrezic-Renault, N. J., Martelet, C., & Cosnier, S. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(99)00520-6 · Cited by 49
Petrou, P. S., Moser, I., & Jobst, G. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00038-1 · Cited by 46
Self-citationBioMEMS device with integrated microdialysis probe and biosensor array
Petrou, P. S., Moser, I., & Jobst, G. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00055-6 · Cited by 44
Self-citationKieninger, J., Aravindalochanan, K., Sandvik, J. A., Pettersen, E. O., & Urban, G. A. (2014) · Cell Proliferation · doi:10.1111/j.1365-2184.2013.12089.x · Cited by 40
Self-citationHiratsuka, A., Muguruma, H., Nagata, R., Nakamura, R., Sato, K., Uchiyama, S., & Karube, I. (2000) · Journal of Membrane Science · doi:10.1016/s0376-7388(00)00403-8 · Cited by 42
Fiber optic pH/Ca2+ fluorescence microsensor based on spectral processing of sensing signals
Ji, J., & Rosenzweig, Z. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(99)00395-5 · Cited by 41
Kohma, T., Oyamatsu, D., & Kuwabata, S. (2007) · Electrochemistry Communications · doi:10.1016/j.elecom.2006.12.015 · Cited by 39
Rothwell, S. A., McMahon, C. P., & O'Neill, R. D. (2009) · Electrochimica Acta · doi:10.1016/j.electacta.2009.09.069 · Cited by 38
Urban, G. A., & Weiss, T. (2009) · Springer series on chemical sensors and biosensors · doi:10.1007/978-3-540-75645-3_6 · Cited by 35
Self-citationMicroelectrode Biosensors for in vivo Analysis of Brain Interstitial Fluid
Chatard, C., Meiller, A., & Marinesco, S. (2018) · Electroanalysis · doi:10.1002/elan.201700836 · Cited by 35
An Integrated Needle-Type Biosensor for Intravascular Glucose and Lactate Monitoring
Yang, Q., Atanasov, P., & Wilkins, E. S. (1998) · Electroanalysis · doi:10.1002/(sici)1521-4109(199809)10:11<752::aid-elan752>3.0.co;2-t · Cited by 34
Niculescu, M. D., Gáspár, S., Schulte, A., Csöregi, E., & Schuhmann, W. W. (2003) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2003.11.008 · Cited by 32
Hiratsuka, A., Ikebukuro, K., Karube, I., Kojima, K., Suzuki, H. R., & Muguruma, H. (2001) · The Analyst · doi:10.1039/b008036p · Cited by 32
Dong, J., Li, S., Wang, H., Meng, Q., Fan, L.-Y., Xie, H., Cao, C., & Zhang, W. (2013) · Analytical Chemistry · doi:10.1021/ac400642d · Cited by 31
Ellmerer, M., Schaupp, L., Trajanoski, Z., Jobst, G., Moser, I., Urban, G. A., Skrabal, F., & Wach, P. (1998) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(98)00002-5 · Cited by 30
Self-citationSensors for Fetal Hypoxia and Metabolic Acidosis: A Review
Cummins, G., Kremer, J., Bernassau, A. L., Brown, A., Bridle, H. L., Schulze, H., Bachmann, T. T., Crichton, M. L., Denison, F. C., & Desmulliez, M. P. Y. (2018) · Sensors · doi:10.3390/s18082648 · Cited by 26
Schulz, B., Riedel, A., & Abel, P. (1999) · Journal of Molecular Catalysis B Enzymatic · doi:10.1016/s1381-1177(99)00034-x · Cited by 24
Wang, J., Chen, L., Chen, F., Lu, X., Li, X., Bao, Y., Wang, W., Han, D., & Niu, L. (2024) · Chemosensors · doi:10.3390/chemosensors12110222 · Cited by 25
Baghelani, M., Abbasi, Z., Daneshmand, M., & Light, P. E. (2022) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2022.3162315 · Cited by 26
Development of a micro-planar amperometric bile acid biosensor for urinalysis
Koide, S., Ito, N., & Karube, I. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.09.009 · Cited by 23
Impedance characterization of silver/silver chloride micro-electrodes for bio-sensing applications
Tjon, K. C. E., & Yuan, J. G. (2019) · Electrochimica Acta · doi:10.1016/j.electacta.2019.134638 · Cited by 22
Nolan, J. K., Nguyen, T. N. H., Le, K. V. H., DeLong, L. E., & Lee, H. (2019) · SLAS TECHNOLOGY · doi:10.1177/2472630319888442 · Cited by 24
Fibbioli, M., Morf, W. E., Badertscher, M., de Rooij, N. F., & Pretsch, E. (2000) · Electroanalysis · doi:10.1002/1521-4109(200011)12:16<1286::aid-elan1286>3.0.co;2-q · Cited by 443
Determination of Surface pKa of SAM Using an Electrochemical Titration Method
Zhao, J., Luo, L., Yang, X., Wang, E., & Dong, S. (1999) · Electroanalysis · doi:10.1002/(sici)1521-4109(199911)11:15<1098::aid-elan1098>3.0.co;2-n · Cited by 153
Thermal Biosensor and Microbiosensor Techniques
Xie, B., & Danielsson, B. R. G. (2007) · Wiley eBooks · doi:10.1002/9780470061565.hbb065 · Cited by 21
Hughes, G., Pemberton, R. M., Nicholas, P., & Hart, J. P. (2018) · Electroanalysis · doi:10.1002/elan.201800104 · Cited by 20
Continuous-flow multi-analyte biosensor cartridge with controllable linear response range
Frey, O., Talaei, S., van der Wal, P. D., Koudelka‐Hep, M., & de Rooij, N. F. (2010) · Lab on a Chip · doi:10.1039/c004851h · Cited by 19
Microfabricated glucose biosensor for culture welloperation
Pemberton, R. M., Cox, T. I., Tuffin, R. P., Sage, I. C., Drago, G. A., Biddle, N., Griffiths, J., Pittson, R. W., Johnson, G., Xu, J., et al. (2012) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2012.11.032 · Cited by 19
Wearable microdialysis system for continuous in vivo monitoring of glucose
Vering, T., Adam, S., Drewer, H., Dumschat, C., Steinkuhl, R., Knoll, M., Schulze, A., & Siegel, E. G. (1998) · The Analyst · doi:10.1039/a800017d · Cited by 18
Pre‐Calibrated Biosensors for Single‐Use Applications
Moser, I., & Jobst, G. (2012) · Chemie Ingenieur Technik · doi:10.1002/cite.201200129 · Cited by 17
Self-citationGfrerer, R., Brunner, G. A., Trajanoski, Z., Schaupp, L., Sendlhofer, G., Skrabal, F., Jobst, G., Moser, I., Urban, G. A., Pieber, T. R., et al. (1998) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(98)00090-6 · Cited by 16
Self-citationMatsumoto, T., Saito, S., & Ikeda, S. (2005) · Journal of Biotechnology · doi:10.1016/j.jbiotec.2005.09.001 · Cited by 15
Mizuguchi, H., Sasaki, K., Ichinose, H., Seino, S., Sakurai, J., Iiyama, M., Kijima, T., Tachibana, K., Nishina, T., Takayanagi, T., et al. (2017) · Bulletin of the Chemical Society of Japan · doi:10.1246/bcsj.20170193 · Cited by 14
Arrays of Screen‐Printed Graphite Microband Electrodes as a Versatile Electroanalysis Platform
Vagin, M. Y., Sekretareva, A., Reategui, R. S., Lundström, I., Winquist, F., & Eriksson, M. (2014) · ChemElectroChem · doi:10.1002/celc.201300204 · Cited by 14
Microstructuring of organic layers for microsystems
Urban, G. A. (1999) · Sensors and Actuators A Physical · doi:10.1016/s0924-4247(98)00299-4 · Cited by 14
Self-citationFernandes, E., Ledo, A., & Barbosa, R. M. (2022) · Molecules · doi:10.3390/molecules27020514 · Cited by 13
Liu, X., Feng, T., Ji, W., Wang, Z., & Zhang, M. (2019) · The Analyst · doi:10.1039/c9an01946d · Cited by 10
Khan, M. R. R., Khalilian, A., Seo, J., Oh, S., Thakre, A., An, T. K., & Lee, H. S. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3174210 · Cited by 9
Toldrà, A., Ainla, A., Khaliliazar, S., Landin, R., Chondrogiannis, G., Hanze, M., Réu, P., & Hamedi, M. M. (2022) · The Analyst · doi:10.1039/d2an00923d · Cited by 9
Valk, T. W., & McMorrow, C. (2020) · SAGE Open Medicine · doi:10.1177/2050312120974174 · Cited by 9
Enzyme modified membranes protected by plasma polymer layers
Janasek, D., Spohn, U., Kiesow, A., & Heilmann, A. M. F. (2001) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(01)00817-6 · Cited by 8
Urban, G. A., & Jobst, G. (1998) · Topics in current chemistry · doi:10.1007/3-540-69544-3_8 · Cited by 8
Self-citationBioprocess monitoring by biosensor-based technologies
Göker, M., Kurbanoğlu, S., Huri, P. Y., Sezgintürk, M. K., & Özkan, S. A. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-818592-6.00010-4 · Cited by 7
Chip-Based Scanning Nano-Calorimeter for Protein Stability Analysis in Biosensor Membranes
Weiss, T., Igel, G., & Urban, G. A. (2007) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/sensor.2007.4300494 · Cited by 7
Self-citationCórcoles, E. P., & Boutelle, M. G. (2013) · SpringerBriefs in applied sciences and technology · doi:10.1007/978-3-319-00360-3_5 · Cited by 5
Laboratory on Chip for Clinical Applications
Moser, I., Jobst, G., & Urban, G. A. (1998) · Springer eBooks · doi:10.1007/978-94-011-5286-0_107 · Cited by 5
Self-citationZhang, X., Wang, J., Ogorevc, B., & Spichiger, U. E. (1999) · Electroanalysis · doi:10.1002/(sici)1521-4109(199909)11:13<945::aid-elan945>3.0.co;2-7 · Cited by 92
Tjon, K. C. E., & Yuan, J. G. (2020) · Journal of Visualized Experiments · doi:10.3791/60820 · Cited by 4
Development of an amperometric biosensor for lactate
Lillis, B., Grogan, C., Berney, H., & Lane, B. (2002) · International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine and Biology (MMB) · doi:10.1109/mmb.2000.893813 · Cited by 4
Multi-Parameter Biomems for Clinical Monitoring
Moser, I. (2007) · Springer eBooks · doi:10.1007/978-0-387-28732-4_2 · Cited by 3
Self-citationBiosensor Design and Fabrication
Urban, G. A. (2000) · Wiley eBooks · doi:10.1002/9780470027318.a0505 · Cited by 2
Self-citationMicroelectrode Biosensors for In Vivo Functional Monitoring of Biological Molecules
Marinesco, S. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-409547-2.13879-x · Cited by 2
Jia, N., Zhang, Z., Zhu, J.-Z., & Zhang, G. (2004) · Chinese Journal of Chemistry · doi:10.1002/cjoc.20040220907 · Cited by 1
Schuhmann, W. W., & Bonsen, E. M. (2003) · Wiley eBooks · doi:10.1002/9783527610426.bard030211 · Cited by 1
Implantable microelectrode biosensors for neurochemical monitoring of brain functioning
Vasylieva, N. (2012) · theses.fr (ABES) · doi:10.70675/a77c8d30z8a67z4377z8fbdz1b58f5da801b · Cited by 3
Continuous ex vivo lactate monitoring in whole blood
Gfrerer, R., Wach, P., Brunners, G., Trajanoski, Z., Schaupp, L., Leitner, B., Skrabal, F., Jobst, G., Moser, I., Urban, G. A., et al. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1997.756789
Self-citationIntelligent Hydrogels in Nanoscale Sensing and Drug Delivery Applications
Hilt, J. Z. (2003) · Wiley eBooks · doi:10.1002/9783527610419.ntls0117
Microfabrication of asymmetric planar electrodes for long-term glucose monitoring
Chih-Hsiu, Parng, S.-H., Chen, J.-T., Lee, K.-F., Chen, C.-H., Chou, T.-C., Chuang, Y.-J., Lee, H.-C., & Lin, Y.-J. (2009) · IEEE International Conference on Nano/Molecular Medicine and Engineering (NANOMED) · doi:10.1109/nanomed.2009.5559121
Reply to Osborne re. Ellmerer et al. (1998)
Ellmerer, M. (1999) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(99)00011-1
Neural Stem Cells: From Cell Fate and Metabolic Monitoring Toward Clinical Applications
Pruszak, J., Döbrössy, M. D., Kieninger, J., Aravindalochanan, K., Urban, G. A., & Nikkhah, G. (2010) · Springer eBooks · doi:10.1007/978-3-642-11865-4_20 · Cited by 1
Self-citationMultifunctional Electrode Arrays
Polsky, R., Xiao, X., Wheeler, D. R., & Brozik, S. M. (2014) · Pan Stanford Publishing eBooks · doi:10.1201/b15534-4
Metadata
- Authors
- Jobst, G., Moser, I., Varahram, M., Svasek, P., Aschauer, E., Trajanoski, Z., Wach, P., Kotanko, P., Skrabal, F., & Urban, G.
- Affiliations
- TU Wien (AT)
- ORCID
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 1996-01-01
- Volume
- 68
- Issue
- 18
- Pages
- 3173-3179
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2070038049
- Cited by
- 102
- References
- 20
- Citations by year
- 2024: 1 · 2023: 2 · 2022: 4 · 2020: 5 · 2019: 5 · 2018: 5 · 2017: 4 · 2016: 2 · 2014: 5 · 2013: 2 · 2012: 3
- FWCI
- 8.23
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Funders
- TU Graz, Internationale Beziehungen und Mobilitätsprogramme
- Austrian Science Fund
- Universität Wien
- Karl-Franzens-Universität Graz
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Keywords
- glucose biosensor, lactate biosensor, photopatterning, flexible substrates, sensor lifetime
- MeSH terms
- Adult, Animals, Blood Glucose, Electrochemistry, Female, Humans, Male, Polyhydroxyethyl Methacrylate, Swine, Biosensing Techniques, Lactic Acid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.