Utilizing Microchip Capillary Electrophoresis Electrospray Ionization for Hydrogen Exchange Mass Spectrometry
Open accessAbstract
Hydrogen exchange (HX) mass spectrometry (MS) of complex mixtures requires a fast, reproducible, and high peak capacity separation prior to MS detection. The current paradigm relies on liquid chromatography (LC) with fast gradients performed at low temperatures to minimize back exchange. Unfortunately, under these conditions, the efficiency of LC is limited due to resistance to mass transfer, reducing the capability to analyze complex samples. Capillary electrophoresis (CE), on the other hand, is not limited by resistance to mass transfer, enabling very rapid separations that are not adversely affected by low temperature. Previously, we have demonstrated an integrated microfluidic device coupling CE with electrospray ionization (ESI) capable of very rapid and high efficiency separations. In this work, we demonstrate the utility of this microchip CE-ESI device for HX MS. High speed CE-ESI of a bovine hemoglobin pepsin digestion was performed in 1 min with a peak capacity of 62 versus a similar LC separation performed in 7 min with peak capacity of 31. A room temperature CE method performed in 1.25 min provided similar deuterium retention as an 8.5 min LC method conducted at 0 °C. Separation of a complex mixture with CE was done with considerably better speed and nearly triple the peak capacity than the equivalent separation by LC. Overall, the results indicate the potential utility of microchip CE-ESI for HX MS.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Microchip CE-MS analysis of a bovine hemoglobin pepsin digest demonstrated high peak capacity and speed compared to conventional LC separation of the same sample.
“For instance, microchip CE-MS analysis of a bovine hemoglobin pepsin digest has demonstrated high peak capacity and speed when compared to a conventional LC separation of the same sample136.”
From Recent trends of capillary electrophoresis‐mass spectrometry in proteomics research (2019) · cites this paper as [136]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Primary structure effects on peptide group hydrogen exchange
Bai, Y., Milne, J. S., Mayne, L. C., & Englander, S. W. (1993) · Proteins Structure Function and Bioinformatics · doi:10.1002/prot.340170110 · Cited by 1,904 · see citing works
The Cell as a Collection of Protein Machines: Preparing the Next Generation of Molecular Biologists
Alberts, B. (1998) · Cell · doi:10.1016/s0092-8674(00)80922-8 · Cited by 1,411 · see citing works
Zhang, Z., & Smith, D. L. (1993) · Protein Science · doi:10.1002/pro.5560020404 · Cited by 1,017 · see citing works
Hydrogen exchange mass spectrometry for the analysis of protein dynamics
Wales, T. E., & Engen, J. R. (2005) · Mass Spectrometry Reviews · doi:10.1002/mas.20064 · Cited by 864 · see citing works
Self-citationHydrogen exchange mass spectrometry for studying protein structure and dynamics
Konermann, L., Pan, J., & Liu, Y. (2010) · Chemical Society Reviews · doi:10.1039/c0cs00113a · Cited by 842 · see citing works
Estimation of diffusion coefficients of proteins
Young, M. E., CARROAD, P. A., & Bell, R. L. (1980) · Biotechnology and Bioengineering · doi:10.1002/bit.260220504 · Cited by 529 · see citing works
Plumb, R. S., Johnson, K. A., Rainville, P. D., Smith, B. W., Wilson, I. D., Castro‐Perez, J. M., & Nicholson, J. K. (2006) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.2550 · Cited by 486 · see citing works
High-Speed Separations on a Microchip
Jacobson, S. C., Hergenröder, R., Koutny, L. B., & Ramsey, J. M. (1994) · Analytical Chemistry · doi:10.1021/ac00079a029 · Cited by 450 · see citing works
Self-citationHigh-Speed and High-Resolution UPLC Separation at Zero Degrees Celsius
Wales, T. E., Fadgen, K. E., Gerhardt, G. C., & Engen, J. R. (2008) · Analytical Chemistry · doi:10.1021/ac8008862 · Cited by 353 · see citing works
Self-citationMellors, J. S., Jorabchi, K., Smith, L. M., & Ramsey, J. M. (2010) · Analytical Chemistry · doi:10.1021/ac902218y · Cited by 235 · see citing works
Self-citationFree-zone electrophoresis in glass capillaries.
Jorgenson, J. W., & Lukacs, K. D. (1981) · Clinical Chemistry · doi:10.1093/clinchem/27.9.1551 · Cited by 234 · see citing works
Peer Reviewed: Investigating Protein Structure and Dynamics by Hydrogen Exchange MS
Engen, J. R., & Smith, D. L. (2001) · Analytical Chemistry · doi:10.1021/ac012452f · Cited by 219 · see citing works
Self-citationMellors, J. S., Gorbounov, V., Ramsey, R. S., & Ramsey, J. M. (2008) · Analytical Chemistry · doi:10.1021/ac800428w · Cited by 199 · see citing works
Self-citationHydrogen exchange mass spectrometry: what is it and what can it tell us?
Marcsisin, S. R., & Engen, J. R. (2010) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-010-3556-4 · Cited by 190 · see citing works
Self-citationHigh-Efficiency, Two-Dimensional Separations of Protein Digests on Microfluidic Devices
Ramsey, J. D., Jacobson, S. C., Culbertson, C. T., & Ramsey, J. M. (2003) · Analytical Chemistry · doi:10.1021/ac0264574 · Cited by 190 · see citing works
Self-citationKnox, J. H., & McCormack, K. A. (1994) · Chromatographia · doi:10.1007/bf02290338 · Cited by 181 · see citing works
Nelson, R. J., Paulus, A., Cohen, A. S., Guttman, A., & Karger, B. L. (1989) · Journal of Chromatography A · doi:10.1016/s0021-9673(01)84282-5 · Cited by 175 · see citing works
Rosa, J. J., & Richards, F. M. (1979) · Journal of Molecular Biology · doi:10.1016/0022-2836(79)90400-5 · Cited by 167 · see citing works
Gilár, M., Daly, A. E., Kele, M., Neue, U. D., & Gebler, J. C. (2004) · Journal of Chromatography A · doi:10.1016/j.chroma.2004.10.092 · Cited by 165 · see citing works
Busnel, J., Schoenmaker, B., Ramautar, R., Carrasco‐Pancorbo, A., Ratnayake, C. K., Feitelson, J. S., Chapman, J. D., Deelder, A. M., & Mayboroda, O. A. (2010) · Analytical Chemistry · doi:10.1021/ac102159d · Cited by 165 · see citing works
Protein hydrogen exchange studied by the fragment separation method
Englander, J. J., Rogero, J. R., & Englander, S. W. (1985) · Analytical Biochemistry · doi:10.1016/0003-2697(85)90033-8 · Cited by 155 · see citing works
Multidimensional LC-LC and LC-CE for high-resolution separations of biological molecules
Evans, C. R., & Jorgenson, J. W. (2004) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-004-2516-2 · Cited by 140 · see citing works
Redman, E. A., Batz, N. G., Mellors, J. S., & Ramsey, J. M. (2015) · Analytical Chemistry · doi:10.1021/ac503964j · Cited by 136 · see citing works
Self-citationCapillary Liquid Chromatography at Ultrahigh Pressures
Jorgenson, J. W. (2010) · Annual Review of Analytical Chemistry · doi:10.1146/annurev.anchem.1.031207.113014 · Cited by 113 · see citing works
Accurate Mass Measurements in Proteomics
Liu, T., Belov, M. E., Jaitly, N., Qian, W., & Smith, R. D. (2007) · Chemical Reviews · doi:10.1021/cr068288j · Cited by 105 · see citing works
Bonn, B., Leandersson, C., Fontaine, F., & Zamora, I. (2010) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.4753 · Cited by 94 · see citing works
Chambers, A. G., Mellors, J. S., Henley, W. H., & Ramsey, J. M. (2011) · Analytical Chemistry · doi:10.1021/ac102437z · Cited by 93 · see citing works
Self-citationAccurate mass measurement for the determination of elemental formula—A tutorial
Bristow, A. W. T. (2005) · Mass Spectrometry Reviews · doi:10.1002/mas.20058 · Cited by 88 · see citing works
Ahn, J., Jung, M. C., Wyndham, K. D., Yu, Y. Q., & Engen, J. R. (2012) · Analytical Chemistry · doi:10.1021/ac301749h · Cited by 76 · see citing works
Self-citationPuig, P., Borrull, F., Calull, M., & Aguilar, C. (2008) · Analytica Chimica Acta · doi:10.1016/j.aca.2008.03.062 · Cited by 72 · see citing works
Batz, N. G., Mellors, J. S., Alarie, J. P., & Ramsey, J. M. (2014) · Analytical Chemistry · doi:10.1021/ac404106u · Cited by 65 · see citing works
Self-citationDevelopments in coupled solid‐phase extraction–capillary electrophoresis 2009–2011
Ramautar, R., de Jong, G. J., & Somsen, G. W. (2011) · Electrophoresis · doi:10.1002/elps.201100453 · Cited by 60 · see citing works
Jiang, L., & Moini, M. (1999) · Analytical Chemistry · doi:10.1021/ac990777e · Cited by 52 · see citing works
Rob, T., Gill, P. K., Golemi, D., & Wilson, D. J. (2013) · Lab on a Chip · doi:10.1039/c3lc00007a · Cited by 48 · see citing works
Microfluidic Dual Emitter Electrospray Ionization Source for Accurate Mass Measurements
Chambers, A. G., & Ramsey, J. M. (2012) · Analytical Chemistry · doi:10.1021/ac202603s · Cited by 43 · see citing works
Self-citationCharles, L. (2003) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.1060 · Cited by 23 · see citing works
CE hydrogen deuterium exchange‐MS in peptide analysis
Lau, S. S. M., Stainforth, N. M., Watson, D. G., Skellern, G. G., Wren, S. A. C., & Tettey, J. N. A. (2007) · Electrophoresis · doi:10.1002/elps.200700368 · Cited by 15 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- [15]Landers JP Handbook of capillary electrophoresis 1997 894 CRC Press Boca Raton 2nd ed.
- [16]Nelson RJ Paulus A Cohen AS Guttman A Karger BL J Chromatogr 1989 480 111 127
Cited by
Sorted by citations
James, E. I., Murphree, T. A., Vorauer, C., Engen, J. R., & Guttman, M. (2021) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00279 · Cited by 308 · see citing works
Self-citationContemporary hydrogen deuterium exchange mass spectrometry
Oganesyan, I., Lento, C., & Wilson, D. J. (2018) · Methods · doi:10.1016/j.ymeth.2018.04.023 · Cited by 119 · see citing works
Trabjerg, E., Nazari, Z. E., & Rand, K. D. (2018) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2018.06.008 · Cited by 117 · see citing works
Micro Total Analysis Systems: Fundamental Advances and Applications
Patabadige, D. E. W., Jia, S., Sibbitts, J., Sadeghi, J., Sellens, K. A., & Culbertson, C. T. (2015) · Analytical Chemistry · doi:10.1021/acs.analchem.5b04310 · Cited by 113 · see citing works
Recent trends of capillary electrophoresis‐mass spectrometry in proteomics research
Gomes, F. P., & Yates, J. R. (2019) · Mass Spectrometry Reviews · doi:10.1002/mas.21599 · Cited by 112 · see citing works
Developments in Hydrogen/Deuterium Exchange Mass Spectrometry
Engen, J. R., Botzanowski, T., Peterle, D., Georgescauld, F., & Wales, T. E. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c04281 · Cited by 97 · see citing works
Self-citationOzohanics, O., & Ambrus, A. (2020) · Life · doi:10.3390/life10110286 · Cited by 74 · see citing works
Analysis of proteins and peptides by electromigration methods in microchips
Štěpánová, S., & Kašička, V. (2016) · Journal of Separation Science · doi:10.1002/jssc.201600962 · Cited by 62 · see citing works
Redman, E. A., Ramos‐Payán, M., Mellors, J. S., & Ramsey, J. M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b00622 · Cited by 41 · see citing works
Self-citationHydrogen deuterium exchange and other mass spectrometry- based approaches for epitope mapping
Jethva, P. N., & Gross, M. L. (2023) · Frontiers in Analytical Science · doi:10.3389/frans.2023.1118749 · Cited by 34 · see citing works
Nguyen, T. T. T. N., Petersen, N. J., & Rand, K. D. (2016) · Analytica Chimica Acta · doi:10.1016/j.aca.2016.07.002 · Cited by 33 · see citing works
Recent advances of capillary electrophoresis-mass spectrometry instrumentation and methodology
Jiang, Y., He, M.-Y., Zhang, W.-J., Luo, P., Guo, D., Fang, X., & Xu, W. (2017) · Chinese Chemical Letters · doi:10.1016/j.cclet.2017.05.008 · Cited by 30 · see citing works
Nanospray HX-MS configuration for structural interrogation of large protein systems
Sheff, J. G., Hepburn, M., Yu, Y., Lees‐Miller, S. P., & Schriemer, D. C. (2017) · The Analyst · doi:10.1039/c6an02707e · Cited by 26 · see citing works
Shen, Y., Zhao, X., Wang, G., & Chen, D. D. Y. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b00558 · Cited by 25 · see citing works
Tracking Higher Order Protein Structure by Hydrogen-Deuterium Exchange Mass Spectrometry
Benhaim, M. A., Lee, K. K., & Guttman, M. (2018) · Protein and Peptide Letters · doi:10.2174/0929866526666181212165037 · Cited by 22 · see citing works
Svejdal, R. R., Sticker, D., Sønderby, C., Kutter, J. P., & Rand, K. D. (2020) · Analytica Chimica Acta · doi:10.1016/j.aca.2020.09.062 · Cited by 21 · see citing works
Dada, O. O., Zhao, Y., Jaya, N., & Salas-Solano, O. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b03405 · Cited by 20 · see citing works
Development of a Microchip CE-HPMS Platform for Cell Growth Monitoring
Gilliland, W. M., & Ramsey, J. M. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b03708 · Cited by 20 · see citing works
Self-citationMaaty, W. S. A., & Weis, D. D. (2016) · Journal of the American Chemical Society · doi:10.1021/jacs.5b11742 · Cited by 17 · see citing works
Cong, H., Xu, X., Yu, B., Liu, H., & Yuan, H. T. (2016) · Biomicrofluidics · doi:10.1063/1.4959239 · Cited by 15 · see citing works
Increase the flow rate and improve hydrogen deuterium exchange mass spectrometry
Peterle, D., DePice, D., Wales, T. E., & Engen, J. R. (2022) · Journal of Chromatography A · doi:10.1016/j.chroma.2022.463742 · Cited by 12 · see citing works
Self-citationHao, Y., Bao, Y., Huang, X., Hu, Y., & Xiong, B. (2018) · RSC Advances · doi:10.1039/c8ra08276f · Cited by 8 · see citing works
Optical Technologies for Single-Cell Analysis on Microchips
Ou, X., Chen, P., & Liu, B. (2023) · Chemosensors · doi:10.3390/chemosensors11010040 · Cited by 8 · see citing works
Hansen, A. B., Svejdal, R. R., Aerts, J. T., Stahlhut, V. B. P., Harmash, D., Hamidi, M., Sticker, D., Kutter, J. P., & Rand, K. D. (2025) · Analytical Chemistry · doi:10.1021/acs.analchem.4c06230 · Cited by 7 · see citing works
Seven perspectives on GPCR H/D-exchange proteomics methods
Zhang, X. (2017) · F1000Research · doi:10.12688/f1000research.10667.1 · Cited by 6 · see citing works
Madren, S. M., & Yi, L. (2022) · Electrophoresis · doi:10.1002/elps.202200129 · Cited by 6 · see citing works
Cong, H., Xu, X., Yu, B., Liu, H., & Yuan, H. T. (2016) · Biomicrofluidics · doi:10.1063/1.4943915 · Cited by 6 · see citing works
Aerts, J. T., Andrén, P. E., & Jansson, E. T. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.2c03893 · Cited by 5 · see citing works
Developments in hydrogen/deuterium exchange mass spectrometry at sub-zero temperatures
Hansen, A. B., Rouhi, O., & Rand, K. D. (2025) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2025.118539 · Cited by 3 · see citing works
Schmidt, A. J., Zamuruyev, K. O., LeVasseur, M. K., Fung, S. L., Anishchenko, I. M., Kenyon, N. J., & Davis, C. E. (2023) · Microelectronic Engineering · doi:10.1016/j.mee.2023.111997 · Cited by 2 · see citing works
Witkiewicz, I. M., & Neering, H. (1981) · PubMed · Cited by 2 · see citing works
Introduction to semiconductor gas sensors
D’Amico, A., & Natale, C. D. (2019) · Elsevier eBooks · doi:10.1016/b978-0-08-102559-8.00004-5 · Cited by 2 · see citing works
Complementary Mass Spectrometry Methods for Characterizing Protein Folding, Structure, and Dynamics
Vahidi, S. (2015) · Scholarship@Western (Western University)
Advances in Structural Proteomics using Mass Spectrometry
Otun, S., Mokhantso, T., & Achilonu, I. (2024) · Applications of modern mass spectrometry · doi:10.2174/9789815050059124020007
Almeida, M. B., da Conceição, R. S. V., Barraca, M., & Carrilho, E. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-34209-7.00016-9
- Authors
- Black, W. A., Stocks, B. B., Mellors, J. S., Engen, J. R., & Ramsey, J. M.
- Affiliations
- University of North Carolina at Chapel Hill (US)
- Northeastern University (US)
- Corresponding author
- Ramsey, J. Michael
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2015-06-02
- Volume
- 87
- Issue
- 12
- Pages
- 6280-6287
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- PMC
- PMC4470751
- OpenAlex
- W281344288
- Cited by
- 35
- References
- 37
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 1 · 2023: 3 · 2022: 3 · 2021: 1 · 2020: 3 · 2019: 3 · 2018: 5 · 2017: 4 · 2016: 6 · 2015: 2
- FWCI
- 2.53
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- UNC Libraries
- Funders
- Northeastern University
- Center for Biomedical Microdevices, University of North Carolina, Chapel Hill
- National Institute of General Medical Sciences (R01-GM101135)
- Topics
- Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Physical Sciences › Chemistry › Spectroscopy › Advanced Proteomics Techniques and Applications
- Keywords
- hydrogen-deuterium exchange mass spectrometry, capillary electrophoresis, electrospray ionization, microfluidics, peak capacity, separation speed, deuterium retention, bovine hemoglobin, pepsin digestion
- MeSH terms
- Animals, Cattle, Chromatography, High Pressure Liquid, Hemoglobins, Mass Spectrometry, Deuterium Exchange Measurement, Electrophoresis, Microchip
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review