An Interlaboratory Evaluation of Drift Tube Ion Mobility–Mass Spectrometry Collision Cross Section Measurements
Open accessAbstract
Collision cross section (CCS) measurements resulting from ion mobility–mass spectrometry (IM-MS) experiments provide a promising orthogonal dimension of structural information in MS-based analytical separations. As with any molecular identifier, interlaboratory standardization must precede broad range integration into analytical workflows. In this study, we present a reference drift tube ion mobility mass spectrometer (DTIM-MS) where improvements on the measurement accuracy of experimental parameters influencing IM separations provide standardized drift tube, nitrogen CCS values ( DT CCS N2 ) for over 120 unique ion species with the lowest measurement uncertainty to date. The reproducibility of these DT CCS N2 values are evaluated across three additional laboratories on a commercially available DTIM-MS instrument. The traditional stepped field CCS method performs with a relative standard deviation (RSD) of 0.29% for all ion species across the three additional laboratories. The calibrated single field CCS method, which is compatible with a wide range of chromatographic inlet systems, performs with an average, absolute bias of 0.54% to the standardized stepped field DT CCS N2 values on the reference system. The low RSD and biases observed in this interlaboratory study illustrate the potential of DTIM-MS for providing a molecular identifier for a broad range of discovery based analyses.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Addona, T. A., Abbatiello, S. E., Schilling, B., Skates, S. J., Mani, D. R., Bunk, D. M., Spiegelman, C. H., Zimmerman, L. J., Ham, A.-J. L., Keshishian, H., et al. (2009) · Nature Biotechnology · doi:10.1038/nbt.1546 · Cited by 1,037
Cited by
Meier, F., Brunner, A.-D., Koch, S., Koch, H., Lubeck, M., Krause, M., Goedecke, N., Decker, J., Kosinski, T., Park, M. A., et al. (2018) · Molecular & Cellular Proteomics · doi:10.1074/mcp.tir118.000900 · Cited by 965
Ion Mobility Spectrometry: Fundamental Concepts, Instrumentation, Applications, and the Road Ahead
Dodds, J. N., & Baker, E. S. (2019) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-019-02288-2 · Cited by 574
Self-citationRecommendations for reporting ion mobility Mass Spectrometry measurements
Gabelica, V., Shvartsburg, A. A., Afonso, C., Barran, P., Benesch, J. L. P., Bleiholder, C., Bowers, M. T., Bilbao, A., Bush, M. F., Campbell, J. L., et al. (2019) · Mass Spectrometry Reviews · doi:10.1002/mas.21585 · Cited by 493
Self-citationIon mobility in the pharmaceutical industry: an established biophysical technique or still niche?
Campuzano, I. D., & Lippens, J. L. (2018) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2017.11.008 · Cited by 42
Jooß, K., Meckelmann, S. W., Klein, J., Schmitz, O. J., & Neusüß, C. (2019) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1515-7 · Cited by 41
Metadata
- Authors
- Stow, S. M., Causon, T. J., Zheng, X., Kurulugama, R. T., Mairinger, T., May, J. C., Rennie, E. E., Baker, E. S., Smith, R. D., McLean, J. A., Hann, S., & Fjeldsted, J. C.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2017-08-16
- Volume
- 89
- Issue
- 17
- Pages
- 9048-9055
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2740977550
- Cited by
- 534
- References
- 1
- FWCI
- 19.19
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An Interlaboratory Evaluation of Drift Tube Ion Mobility–Mass Spectrometry Collision Cross Section Measurements. https://app.cochat.ai/papers/10.1021/acs.analchem.7b01729
- The paper
- Stow, S. M., Causon, T. J., Zheng, X., Kurulugama, R. T., Mairinger, T., May, J. C., Rennie, E. E., Baker, E. S., Smith, R. D., McLean, J. A., Hann, S., & Fjeldsted, J. C. (2017). An Interlaboratory Evaluation of Drift Tube Ion Mobility–Mass Spectrometry Collision Cross Section Measurements. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.7b01729
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.