Can Validated Wrist Devices With Position Sensors Replace Arm Devices for Self-Home Blood Pressure Monitoring? A Randomized Crossover Trial Using Ambulatory Monitoring as Reference

Stergiou, G. S., Christodoulakis, G. R., Nasothimiou, E. G., Giovas, P. P., Kalogeropoulos, P. G. · 2008 · American Journal of Hypertension
Cited by 21

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Electronic devices that measure blood pressure (BP) at the arm level are regarded as more accurate than wrist devices and are preferred for home BP (HBP) monitoring. Recently, wrist devices with position sensors have been successfully validated using established protocols. This study assessed whether HBP values measured with validated wrist devices are sufficiently reliable to be used for making patient-related decisions in clinical practice. METHODS: This randomized crossover study compared HBP measurements taken using validated wrist devices (wrist-HBP, Omron R7 with position sensor) with those taken using arm devices (arm-HBP, Omron 705IT), and also with measurements of awake ambulatory BP (ABP, SpaceLabs), in 79 subjects (36 men and 43 women) with hypertension. The mean age of the study population was 56.7 +/- 11.8 years, and 33 of the subjects were not under treatment for hypertension. RESULTS: The average arm-HBP was higher than the average wrist-HBP (mean difference, systolic 5.2 +/- 9.1 mm Hg, P < 0.001, and diastolic 2.2 +/- 6.7, P < 0.01). Twenty-seven subjects (34%) had a > or =10 mm Hg difference between systolic wrist-HBP and arm-HBP and twelve subjects (15%) showed similar levels of disparity in diastolic HBP readings. Strong correlations were found between arm-HBP and wrist-HBP (r 0.74/0.74, systolic/diastolic, P < 0.0001). However, ABP was more strongly correlated with arm-HBP (r 0.73/0.76) than with wrist-HBP (0.55/0.69). The wrist-arm HBP difference was associated with systolic ABP (r 0.34) and pulse pressure (r 0.29), but not with diastolic ABP, sex, age, arm circumference, and wrist circumference. CONCLUSIONS: There might be important differences in HBP measured using validated wrist devices with position sensor vs. arm devices, and these could impact decisions relating to the patient in clinical practice. Measurements taken using arm devices are more closely related to ABP values than those recorded by wrist devices. More research is needed before recommending the widespread use of wrist monitors in clinical practice. American Journal of Hypertension doi:10.1038/ajh.2008.176American Journal of Hypertension (2008); 21, 7, 753-758. doi:10.1038/ajh.2008.176.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1038/ajh.2008.176

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 21

Metadata

Authors
Stergiou, G. S., Christodoulakis, G. R., Nasothimiou, E. G., Giovas, P. P., & Kalogeropoulos, P. G.
Venue
American Journal of Hypertension
Published
2008-07-01
Volume
21
Issue
7
Pages
753-758
ISSN
  • 0895-7061 (ISSN-L)
  • 1941-7225
OpenAlex
W2028267159
Cited by
21
References
3
FWCI
1.54
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Can Validated Wrist Devices With Position Sensors Replace Arm Devices for Self-Home Blood Pressure Monitoring? A Randomized Crossover Trial Using Ambulatory Monitoring as Reference. https://app.cochat.ai/papers/10.1038/ajh.2008.176
The paper
Stergiou, G. S., Christodoulakis, G. R., Nasothimiou, E. G., Giovas, P. P., & Kalogeropoulos, P. G. (2008). Can Validated Wrist Devices With Position Sensors Replace Arm Devices for Self-Home Blood Pressure Monitoring? A Randomized Crossover Trial Using Ambulatory Monitoring as Reference. American Journal of Hypertension. https://doi.org/10.1038/ajh.2008.176

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.