Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review

Open access
Irving, Greg, Holden, John, Stevens, Richard, McManus, Richard J·2016·BMJ Open

Abstract

Objective: To determine the diagnostic accuracy of different methods of blood pressure (BP) measurement compared with reference standards for the diagnosis of hypertension in patients with obesity with a large arm circumference. Design: Systematic review with meta-analysis with hierarchical summary receiver operating characteristic models. Bland-Altman analyses where individual patient data were available. Methodological quality appraised using Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS2) criteria. Data sources: MEDLINE, EMBASE, Cochrane, DARE, Medion and Trip databases were searched. Eligibility criteria: Cross-sectional, randomised and cohort studies of diagnostic test accuracy that compared any non-invasive BP tests (upper arm, forearm, wrist, finger) with an appropriate reference standard (invasive BP, correctly fitting upper arm cuff, ambulatory BP monitoring) in primary care were included. Results: 4037 potentially relevant papers were identified. 20 studies involving 26 different comparisons met the inclusion criteria. Individual patient data were available from 4 studies. No studies satisfied all QUADAS2 criteria. Compared with the reference test of invasive BP, a correctly fitting upper arm BP cuff had a sensitivity of 0.87 (0.79 to 0.93) and a specificity of 0.85 (0.64 to 0.95); insufficient evidence was available for other comparisons to invasive BP. Compared with the reference test of a correctly fitting upper arm cuff, BP measurement at the wrist had a sensitivity of 0.92 (0.64 to 0.99) and a specificity of 0.92 (0.85 to 0.87). Measurement with an incorrectly fitting standard cuff had a sensitivity of 0.73 (0.67 to 0.78) and a specificity of 0.76 (0.69 to 0.82). Measurement at the forearm had a sensitivity of 0.84 (0.71 to 0.92) and a specificity 0.75 of (0.66 to 0.83). Bland-Altman analysis of individual patient data from 3 studies comparing wrist and upper arm BP showed a mean difference of 0.46 mm Hg for systolic BP measurement and 2.2 mm Hg for diastolic BP measurement. Conclusions: BP measurement with a correctly fitting upper arm cuff is sufficiently sensitive and specific to diagnose hypertension in patients with obesity with a large upper arm circumference. If a correctly fitting upper arm cuff cannot be applied, an incorrectly fitting standard size cuff should not be used and BP measurement at the wrist should be considered.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/bmjopen-2016-012429

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 49

  1. [1]Ng M, Fleming T, Robinson M Global, regional, and national prevalence of overweight and obesity in children and adults during 1980–2013: a systematic analysis for the Global Burden of Disease Study 2013. Lancet 2014; 384: 766– 81. 10.1016/S0140-6736(14)60460-8 24880830 PMC4624264 · DOI · Paper page
  2. [2]Palatini P, Parati G Blood pressure measurement in very obese patients: a challenging problem. J Hypertens 2011; 29: 425– 9. 10.1097/HJH.0b013e3283435b65 21317721 · DOI
  3. [3]Doshi H, Weder AB, Bard RL Does “hidden undercuffing” occur among obese patients? Effect of arm sizes and other predictors of the difference between wrist and upper arm blood pressures. J Clin Hypertens (Greenwich) 2010; 12: 82– 8. 10.1111/j.1751-7176.2009.00222.x 20167030 PMC8673187 · DOI
  4. [4]Graves JW, Bailey KR, Sheps SG The changing distribution of arm circumferences in NHANES III and NHANES 2000 and its impact on the utility of the ‘standard adult’ blood pressure cuff. Blood Press Monit 2003; 8: 223– 7. 10.1097/01.mbp.0000105483.36014.0d 14688551 · DOI
  5. [5]Graves JW Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives. Blood Press Monit 2001; 6: 17– 20. 10.1097/00126097-200102000-00003 11248756 · DOI
  6. [7]O'Brien E What to do when faced with an unmeasurable ambulatory blood pressure? J Hypertens 2011; 29: 451– 3. 10.1097/HJH.0b013e32834192ef 21119530 · DOI
  7. [9]Mancia G, Fagard R, Narkiewicz K 2013 ESH/ESC guidelines for the management of arterial hypertension: the Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). Blood Pressure 2013; 34: 2159– 219. 10.1093/eurheartj/eht151 17846925 · DOI
  8. [10]Mendis S, Lindholm LH, Mancia G World Health Organization (WHO) and International Society of Hypertension (ISH) risk prediction charts: assessment of cardiovascular risk for prevention and control of cardiovascular disease in low and middle-income countries. J Hypertens 2007; 25: 1578– 82. 10.1097/HJH.0b013e3282861fd3 17620952 · DOI
  9. [11]Lewington S, Clarke R, Qizilbash N, Prospective Studies Collaboration. Age-specific relevance of usual blood pressure to vascular mortality: a meta-analysis of individual data for one million adults in 61 prospective studies. Lancet 2002; 360: 1903– 13. 10.1016/S0140-6736(02)11911-8 12493255 · DOI · Paper page
  10. [12]Leeflang MM, Deeks JJ, Gatsonis C Systematic reviews of diagnostic test accuracy. Ann Intern Med 2008; 149: 889– 97. 10.7326/0003-4819-149-12-200812160-00008 19075208 PMC2956514 · DOI · Paper page
  11. [13]Whiting PF, Rutjes AW, Westwood ME QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 2011; 155: 529– 36. 10.7326/0003-4819-155-8-201110180-00009 22007046 · DOI · Paper page
  12. [14]Rutter CM, Gatsonis CA A hierarchical regression approach to meta-analysis of diagnostic test accuracy evaluations. Stat Med 2001; 20: 2865– 84. 10.1002/sim.942 11568945 · DOI
  13. [15]Altman DG, Bland JM Commentary on quantifying agreement between two methods of measurement. Clin Chem 2002; 48: 801– 2.
  14. [16]Deeks JJ, Macaskill P, Irwig L The performance of tests of publication bias and other sample size effects in systematic reviews of diagnostic test accuracy was assessed. J Clin Epidemiol 2005; 58: 882– 93. 10.1016/j.jclinepi.2005.01.016 16085191 · DOI · Paper page
  15. [17]Stergiou GS, Christodoulakis GR, Nasothimiou EG Can validated wrist devices with position sensors replace arm devices for self-home blood pressure monitoring? A randomized crossover trial using ambulatory monitoring as reference. Am J Hypertens 2008; 21: 753– 8. 10.1038/ajh.2008.176 18443566 · DOI
  16. [18]Bertrand CA, Trout KW, Williams MH Measurement of blood pressure in obese persons. J Am Med Assoc 1956; 162: 970– 1. 10.1001/jama.1956.72970270001010 13366691 · DOI
  17. [19]Bennett JM, van den Bogaerde E, van Rensburg JT [Evaluation of a new blood pressure cuff for patients with large and normal arm circumferences]. S Afr Med J 1987; 72: 610– 1. 3686298
  18. [20]Berntsen R, Romo O, Nielsen EW [Blood pressure measurements using three different cuff sizes]. Tidsskr Nor Laegeforen 1998; 118: 3112– 15. 9760852
  19. [21]Chiolero A, Gervasoni JP, Rwebogora A Difference in blood pressure readings with mercury and automated devices: impact on hypertension prevalence estimates in Dar es Salaam, Tanzania. Eur J Epidemiol 2006; 21: 427– 33. 10.1007/s10654-006-9015-z 16826452 · DOI
  20. [22]Cuckson AC, Moran P, Seed P Clinical evaluation of an automated oscillometric blood pressure wrist device. Blood Press Monit 2004; 9: 31– 7. 10.1097/00126097-200402000-00007 15021076 · DOI
  21. [23]de Senarclens O, Feihl F, Giusti V Brachial or wrist blood pressure in obese patients: which is the best? Blood Press Monit 2008; 13: 149– 51. 10.1097/MBP.0b013e3282fd16e3 18496290 · DOI
  22. [24]Domiano KL, Hinck SM, Savinske DL Comparison of upper arm and forearm blood pressure. Clin Nurs Res 2008; 17: 241– 50. 10.1177/1054773808324651 18927258 · DOI
  23. [25]Guagnano MT, Pace-Palitti V, Murri R The prevalence of hypertension in gynaecoid and android obese women. J Hum Hypertens 1996; 10: 619– 24. 8953208
  24. [26]Guagnano MT, Palitti VP, Murri R Many factors can affect the prevalence of hypertension in obese patients: role of cuff size and type of obesity. Panminerva Med 1998; 40: 22– 7. 9573749
  25. [27]Leblanc MÈ, Croteau S, Ferland A Blood pressure assessment in severe obesity; validation of a forearm approach. Obesity (Silver Spring) 2013; 21: E533– 41. 10.1002/oby.20458 23512945 · DOI
  26. [28]Linfors EW, Feussner JR, Blessing CL Spurious hypertension in the obese patient. Effect of sphygmomanometer cuff size on prevalence of hypertension. Arch Intern Med 1984; 144: 1482– 5. 10.1001/archinte.1984.00350190188032 6732409 · DOI
  27. [29]Nielsen PE, Larsen B, Holstein P Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects. Hypertension 1983; 5: 122– 7. 10.1161/01.HYP.5.1.122 6848459 · DOI
  28. [30]Pierin AM, Alavarce DC, Gusmão JL Blood pressure measurement in obese patients: comparison between upper arm and forearm measurements. Blood Press Monit 2004; 9: 101– 5. 10.1097/01.mbp.0000132425.25263.ac 15199302 · DOI
  29. [31]Poncelet P, Durand P, Lechantre R [Measurement of blood pressure in obese patients: reliability and value of finger measurement (Finapres)]. Arch Mal Coeur Vaiss 1992; 85: 1165– 8. 1482253
  30. [32]Simpson JA, Jamieson G, Dickhaus DW Effect of size of cuff bladder on accuracy of measurement of indirect blood pressure. Am Heart J 1965; 70: 208– 15. 10.1016/0002-8703(65)90069-4 14328006 · DOI
  31. [33]Stolt M, Sjonell G, Astrom H The reliability of auscultatory measurement of arterial blood pressure. A comparison of the standard and a new methodology. Am J Hypertens 1990; 3: 697– 703. 10.1093/ajh/3.9.697 2222978 · DOI
  32. [34]Stolt M, Sjönell G, Aström H Improved accuracy of indirect blood pressure measurement in patients with obese arms. Am J Hypertens 1993; 6: 66– 71. 8427664
  33. [35]Vinyoles E, Pujol E, de la Figuera M Measuring blood pressure in the forearm of obese patients: concordance with arm measurement. Med Clin (Barc) 2005; 124: 213– 14. 10.1157/13071766 15737302 · DOI
  34. [37]Blackburn H, Kihlberg J, Brozek J Arm versus forearm blood pressure in obesity. 1965; 69: 423– 4. 14280884
  35. [38]Bovet P, Hungerbuhler P, Quilindo J, Grettve M L, Waeber B, Burnand B Systematic difference between blood pressure readings caused by cuff type. Hypertension 1994; 24: 786– 92. 7995638
  36. [39]Genc Y, Altunkan S, Kilinc O, Altunkan E Comparative study on auscultatory and oscillometric methods of ambulatory blood pressure measurements in adult patients. Blood Press Monit 2008; 13: 29– 35. 18199921
  37. [40]Guagnano MT, Merlitti D, Murri R Ambulatory blood pressure monitoring in evaluating the relationship between obesity and blood pressure. J Hum Hypertens 1994; 8: 245– 50 8021904
  38. [41]Julien J, Pagny J Y, Jeunemaitre X, Fouqueray B, Plouin P F, Corvol P Comparison of 3 methods of blood pressure measurement in obesity. Arch Mal Coeur Vaiss 1988; 81 Spec No: 241– 5. 3142414
  39. [42]King G E Errors in clinical measurement of blood pressure in obesity. 1967; 32: 223– 37. 6022817
  40. [44]Kvols L K, Rohlfing B M, Alexander J K A comparison of intra-arterial and cuff blood pressure measurements in very obese subjects. Cardiovasc Res Cent Bull 1969; 7: 118– 23. 5800933
  41. [45]Maxwell Morton H, Schroth Phillip C, Waks Abraham U, Karam Maroun, Dornfeld Leslie P Error in blood-pressure measurement due to incorrect cuff size in obese patients. 1982; 320: 33– 6. 6123760
  42. [46]Maxwell G F, Pruijt J F, Arntzenius A C Comparison of the conical cuff and the standard rectangular cuffs. 1985; 14: 468– 72. 4055213
  43. [47]Pang T C, Brown M A Accuracy of ambulatory blood pressure monitors in routine clinical practice. 2006; 19: 801– 9. 16876678
  44. [48]Schell K, Bradley E, Bucher L, Seckel M, Lyons D, Wakai S Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm. Am J Crit Care 2005; 14: 232– 41. 15840897
  45. [49]Schell K, Lyons D, Bradley E, Bucher L, Seckel M, Wakai S Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm with the patient supine or with the head of the bed raised 45 degrees: a follow-up study. Am J Crit Care 2006; 15: 196– 205. 16501139
  46. [50]Umana E, Ahmed W, Fraley M A, Alpert M A Comparison of oscillometric and intraarterial systolic and diastolic blood pressures in lean, overweight, and obese patients. Angiology 2006; 57: 41– 5. 16444455
  47. [51]van Montfrans G, van der Hoeven G, Karemaker J, Wieling W, Dunning A Accuracy of auscultatory blood pressure measurement with a long cuff. Br Med J (Clin Res Ed) 1987; 295: 354– 5. PMC1247210 3115445
  48. [52]Warembourg A, Poncelet P, Carre A Reliability of the measurement of arterial pressure in the forearm in the obese subject. 1987; 80: 1015– 9. 3116964
  49. [53]Altman D, Bland M Commentary on quantifying agreement between two methods of measurement. Cochrane Statistical Methods Group 2002; 48: 801– 2.

Cited by

Sorted by citations

Authors
Irving, G., Holden, J., Stevens, R., & McManus, R. J.
Affiliations
  • University of Cambridge (GB)
  • Department of Garswood Surgery, Garswood, UK
  • University of Oxford (GB)
Corresponding author
Irving, Greg
Venue
BMJ Open
Publisher
BMJ
Published
2016-11
Volume
6
Issue
11
Pages
e012429
ISSN
2044-6055 (ISSN-L)
OpenAlex
W2546973642
Cited by
51
References
2
Citations by year
2025: 3 · 2024: 9 · 2023: 3 · 2022: 6 · 2021: 5 · 2020: 8 · 2019: 5 · 2018: 6 · 2017: 5 · 2014: 1
FWCI
0.94
Type
journal article
Language
en
Open access
Licence
cc-by-nc
Locations
  • BMJ Open (publishedVersion, cc-by-nc)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
  • PubMed Central (publishedVersion, cc-by-nc)
  • PubMed (publishedVersion)
  • Edge Hill University Research Information Repository (Edge Hill University) (submittedVersion, cc-by-nc)
  • HighWire Press Open Archive (submittedVersion)
  • Apollo (University of Cambridge) (submittedVersion, cc-by-nc)
  • Apollo
Funders
  • National Institute for Health and Care Research (CDF/01/017, CL-2013-14-001, NIHR-RP-02-12-015, RP-PG-1209-10051)
  • European Society of Cardiology
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
Keywords
hypertension diagnosis, obesity, blood pressure measurement, cuff size, upper arm cuff, finger blood pressure, diagnostic accuracy, ambulatory blood pressure monitoring, invasive blood pressure, sensitivity and specificity, measurement agreement, primary care, individual patient data, systolic and diastolic blood pressure, cardiovascular risk stratification
MeSH terms
Adult, Aged, Arm, Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Middle Aged, Obesity, Sensitivity and Specificity, Wrist, Blood Pressure Monitoring, Ambulatory, Body Size

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review