A cadaver-based biomechanical model of acetabulum reaming for surgical virtual reality training simulators
Open accessThe DOI matches its Crossref record; 25 of the 27 references it cites were found in a registry or an index.
Abstract
Total hip arthroplasty (THA) is a highly successful surgical procedure, but complications remain, including aseptic loosening, early dislocation and misalignment. These may partly be related to lacking training opportunities for novices or those performing THA less frequently. A standardized training setting with realistic haptic feedback for THA does not exist to date. Virtual Reality (VR) may help establish THA training scenarios under standardized settings, morphology and material properties. This work summarizes the development and acquisition of mechanical properties on hip reaming, resulting in a tissue-based material model of the acetabulum for force feedback VR hip reaming simulators. With the given forces and torques occurring during the reaming, Cubic Hermite Spline interpolation seemed the most suitable approach to represent the nonlinear force–displacement behavior of the acetabular tissues over Cubic Splines. Further, Cubic Hermite Splines allowed for a rapid force feedback computation below the 1 ms hallmark. The Cubic Hermite Spline material model was implemented using a three-dimensional-sphere packing model. The resulting forces were delivered via a human–machine-interaction certified KUKA iiwa robotic arm used as a force feedback device. Consequently, this novel approach presents a concept to obtain mechanical data from high-force surgical interventions as baseline data for material models and biomechanical considerations; this will allow THA surgeons to train with a variety of machining hardness levels of acetabula for haptic VR acetabulum reaming.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“The gaze tracking features of the HTC VIVE Pro Eye was used for identifying distracted learners [60] and for foveated rendering in procedural task training [80]”
From A systematic review of immersive virtual reality for industrial skills training (2021) · cites this paper as [80]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
de Boor, C. (1978) · Applied Mathematical Sciences · doi:10.1007/978-1-4612-6333-3 · Cited by 11,944
The operation of the century: total hip replacement
Learmonth, I. D., Young, C., & Rorabeck, C. (2007) · The Lancet · doi:10.1016/s0140-6736(07)60457-7 · Cited by 2,848
Singh, J. A., Yu, S., Chen, L., & Cleveland, J. D. (2019) · The Journal of Rheumatology · doi:10.3899/jrheum.170990 · Cited by 1,190
Factors affecting the Z-Width of a haptic display
Colgate, J. E., & Brown, J. M. (2002) · Proceedings of the 1994 IEEE International Conference on Robotics and Automation · doi:10.1109/robot.1994.351077 · Cited by 635
Simulation in surgical training: educational issues and practical implications
Kneebone, R. (2003) · Medical Education · doi:10.1046/j.1365-2923.2003.01440.x · Cited by 619
The Role of Haptics in Medical Training Simulators: A Survey of the State of the Art
Coles, T. R., Meglan, D., & John, N. W. (2011) · IEEE Transactions on Haptics · doi:10.1109/toh.2010.19 · Cited by 484
Athanasiou, K. A., Agarwal, A., & Dzida, F. J. (1994) · Journal of Orthopaedic Research · doi:10.1002/jor.1100120306 · Cited by 343
A Review of Simulators with Haptic Devices for Medical Training
Escobar-Castillejos, D., Noguez, J., Neri, L., Magana, A., & Benes, B. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0459-8 · Cited by 245
Dawe, S. R., Windsor, J. A., Broeders, J. A. J. L., Cregan, P. C., Hewett, P. J., & Maddern, G. J. (2014) · Annals of Surgery · doi:10.1097/sla.0000000000000245 · Cited by 219
Epperson, J. F. (1987) · The American Mathematical Monthly · doi:10.1080/00029890.1987.12000642 · Cited by 180
Research works printed but not linked: 3
- Erdogan K Spline interpolation techniques J. Tech. Sci. Technol. 2014 2 47 52
- Kaluschke, M., Weller, R., Zachmann, G. & Lorenz, M. A continuous material cutting model with haptic feedback for medical simulations. In 2019 IEEE Conference on Virtual Reality and 3D User Interfaces (VR) 1002–1003 (IEEE, New York, 2019).
- Knopp, S., Lorenz, M., Pelliccia, L. & Klimant, P. Using industrial robots as haptic devices for VR-training. In 2018 IEEE Conference on Virtual Reality and 3D User Interfaces (VR) 607–608 (IEEE, New York, 2018).
Cited by
A systematic review of immersive virtual reality for industrial skills training
Radhakrishnan, U., Koumaditis, K., & Chinello, F. (2021) · Behaviour & Information Technology · doi:10.1080/0144929x.2021.1954693 · Cited by 200
Tokgöz, E. (2022) · Springer eBooks · doi:10.1007/978-3-031-08927-5_8 · Cited by 12
Lorenz, M., Hoffmann, A., Kaluschke, M., Ziadeh, T., Pillen, N., Kusserow, M., Perret, J., Knopp, S., Dettmann, A., Klimant, P., et al. (2023) · Scientific Reports · doi:10.1038/s41598-023-38201-x · Cited by 12
Self-citationLorenz, M., Kaluschke, M., Melzer, A., Pillen, N., Sanrow, M., Hoffmann, A., Schmidt, D., Dettmann, A., Bullinger, A. C., Perret, J., et al. (2025) · IEEE Transactions on Visualization and Computer Graphics · doi:10.1109/tvcg.2025.3549896 · Cited by 11
“…a progression from the conventional Virtuose 6D, characterized by elevated torques across all motors, resulting in a peak force of 70 N in translation (5 Nm in rotation) at the wrist within the entire workspace, which covers a fair spectrum of forces occurring during acetabula reaming [16].”
Self-citationWang, J., Hua, J., Ding, R., Zou, L., Li, H., Zhang, L., Sun, Q., & He, D. (2023) · Quantitative Imaging in Medicine and Surgery · doi:10.21037/qims-22-1239 · Cited by 8
“Generating a smooth curve by discrete data to simulate the actual situation was performed by cubic splint data, mathematical method (23,24).”
Kriener, K., Whiting, H., Storr, N., Homes, R., Lala, R., Gabrielyn, R., Kuang, J., Rubin, B., Frails, E., Sandstrom, H., et al. (2023) · JBI Evidence Synthesis · doi:10.11124/jbies-22-00363 · Cited by 5
Neuwirth, L. S., Ng, S., Devors, S., Lonjon, N., & Ros, M. (2024) · Frontiers in Virtual Reality · doi:10.3389/frvir.2024.1408092 · Cited by 5
Pratap, J. S., Laane, C. L. E., Chen, N. C., & Bhashyam, A. R. (2024) · Frontiers in Virtual Reality · doi:10.3389/frvir.2024.1392825 · Cited by 3
Lorenz, M., Knopp, S., Pillen, N., Sanrow, M., Hoffmann, A., Stoiser, V., Schmidt, D., Atze, J. P. G., Dettmann, A., Bullinger, A. C., et al. (2025) · Frontiers in Virtual Reality · doi:10.3389/frvir.2025.1620135 · Cited by 3
Self-citationInvestigation of bone tissue mechanical properties in the supra-acetabular region
Udintseva, M. Y., Зайцев, Д. В., Волокитина, Е. А., Петровна, А. И., & Kutepov, S. M. (2022) · Genij Ortopedii · doi:10.18019/1028-4427-2022-28-4-559-564 · Cited by 2
Teaching Human Anatomy: A Comparison of In-Person and Online Learning Programs
McKivigan, J. M., Guison, N., & Qureshi, R. (2021) · European Journal of Medical and Health Sciences · doi:10.24018/ejmed.2021.3.6.1011 · Cited by 2
Compensation of acetabular defects in hip arthroplasty
Udintseva, M. Y., Волокитина, Е. А., & Kutepov, S. M. (2022) · Kazan medical journal · doi:10.17816/kmj2022-89 · Cited by 2
Asia-Pacific expert consensus on robotic-assisted hip and knee arthroplasty
Zhang, G., Zhou, Y., 陈云苏, Budhiparama, N. C., Chun Hoi, Y. A. N., Lim, S. J., Liow, M. H. L., Sugano, N., Tang, A. W. W., Cai, H., et al. (2026) · Arthroplasty · doi:10.1186/s42836-026-00427-1 · Cited by 1
“The importance of simulation and cadaver labs in the training process cannot be overstated [52].”
Devinta, S., Wiyono, N., Munawaroh, S., & Hastami, Y. (2024) · Jurnal Pendidikan Kedokteran Indonesia The Indonesian Journal of Medical Education · doi:10.22146/jpki.88670
“24 In addition, virtual reality can be used by students to practice cardiopulmonary resuscitation skills even though they are far from the hospital and get input from instructors.”
BackgroundCarvajal, J. L., Timko, S. E., Kim, S. E., Lewis, D. D., & Lee, H. (2024) · VCOT Open · doi:10.1055/s-0044-1787746
“10,11 12 – 15 CSGs have been used in veterinary medicine and have showed improved surgical accuracy in a wide variety of applications, 16 – 20 but to our knowledge, no study has assessed virtual arthroplasty planning or CSGs for THA in dogs.”
MethodNovel Approach for Personalized Femoral Broach and Acetabular Reamer
Amitasari, N. R., Ayati, N. S., Utomo, M. S., Mayasari, D. A., & Asmaria, T. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-80355-0_4
Yuan, X., Zhou, Y., Zhou, S., & Wang, F. (2025) · Acta Geodynamica et Geomaterialia · doi:10.13168/agg.2025.0027
Physical Human-Robot Collaboration-Assisted Acetabular Preparation for Total Hip Replacement Surgery
Wang, Z., Li, T., Carmichael, M. G., & Huang, S. (2025) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros60139.2025.11245831
“Email: ziqi.r.wang@student.uts.edu.au (b) Manual acetabular reaming with a handheld electric drill and a reamer [6].”
Handpiece Integrating Tactile Feedback for Surgical VR-Training of Total Hips Arthroplasty (THA)
Günther, F., Okoniewski, M., Müller, M., Knopp, S., & Lorenz, M. (2026) · IEEE Conference on Virtual Reality and 3D User Interfaces Abstracts and Workshops (VRW) · doi:10.1109/vrw70859.2026.00200
“140 N was chosen as load as this is the maximum force of the KUKA iiwa 14 R820, which is the most powerful haptic device the handpiece is attached to and because such forces can occur during acetabulum reaming [5].”
Self-citationExtended Reality for Enhancing Surgery: Research Directions to Transform the Operating Environment
Zachmann, G., Haddawy, P., Kikinis, R., Hennemuth, A., Cattin, P. C., Döring, T., Drouin, S., Fichtinger, G., Flügge, T. V., Jorge, J. A., et al. (2026) · Journal of Healthcare Informatics Research · doi:10.1007/s41666-026-00250-y
“Tissue models from cadavers and ex-vivo tissue can be obtained [ 31 ], but their material properties can be grossly different from those of per - fused live tissues.”
Self-citation
Metadata
- Authors
- Pelliccia, L., Lorenz, M., Heyde, C.-E., Kaluschke, M., Klimant, P., Knopp, S., Schleifenbaum, S., Rotsch, C., Weller, R., Werner, M., Zachmann, G., Zajonz, D., & Hammer, N.
- Affiliations
- Chemnitz University of Technology (DE)
- Medical University of Graz (AT)
- University Hospital Leipzig (DE)
- University of Bremen (DE)
- Fraunhofer Institute for Machine Tools and Forming Technology (DE)
- Corresponding author
- Hammer, Niels
- ORCID
- Venue
- Scientific Reports
- Published
- 2020-09-03
- Volume
- 10
- Issue
- 1
- Pages
- 14545-14545
- ISSN
- 2045-2322 (ISSN-L)
- PMC
- PMC7471911
- OpenAlex
- W3081768073
- Cited by
- 20
- References
- 27
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 5 · 2023: 3 · 2022: 3 · 2021: 2
- FWCI
- 1.83
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Scientific Reports (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion, other-oa)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- Qucosa (Saxon State and University Library Dresden) (submittedVersion, cc-by)
- Funders
- Projekt DEAL
- Topics
- Health Sciences › Medicine › Surgery › Surgical Simulation and Training
- Physical Sciences › Engineering › Biomedical Engineering › Anatomy and Medical Technology
- Health Sciences › Medicine › Surgery › Orthopaedic implants and arthroplasty
- Keywords
- total hip arthroplasty, surgical training simulator, virtual surgery, haptic feedback, force feedback, force-displacement behavior
- MeSH terms
- Virtual Reality, Acetabulum, Biomechanical Phenomena, Computer Simulation, Hip Prosthesis, Humans, Arthroplasty, Replacement, Hip
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.