Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation
Open accessAbstract
There is an unmet clinical need for a low cost and easy to use wearable devices for continuous cardiovascular health monitoring. A flexible and wearable wristband, based on microelectromechanical sensor (MEMS) elements array was developed to support this need. The performance of the device in cardiovascular monitoring was investigated by (i) comparing the arterial pressure waveform recordings to the gold standard, invasive catheter recording (n = 18), (ii) analyzing the ability to detect irregularities of the rhythm (n = 7), and (iii) measuring the heartrate monitoring accuracy (n = 31). Arterial waveforms carry important physiological information and the comparison study revealed that the recordings made with the wearable device and with the gold standard device resulted in almost identical (r = 0.9–0.99) pulse waveforms. The device can measure the heart rhythm and possible irregularities in it. A clustering analysis demonstrates a perfect classification accuracy between atrial fibrillation (AF) and sinus rhythm. The heartrate monitoring study showed near perfect beat-to-beat accuracy (sensitivity = 99.1%, precision = 100%) on healthy subjects. In contrast, beat-to-beat detection from coronary artery disease patients was challenging, but the averaged heartrate was extracted successfully (95% CI: −1.2 to 1.1 bpm). In conclusion, the results indicate that the device could be useful in remote monitoring of cardiovascular diseases and personalized medicine.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Survey on Wearable Sensor-Based Systems for Health Monitoring and Prognosis
Pantelopoulos, A., & Bourbakis, N. G. (2010) · IEEE Transactions on Systems Man and Cybernetics Part C (Applications and Reviews) · doi:10.1109/tsmcc.2009.2032660 · Cited by 2,296
A wearable and highly sensitive pressure sensor with ultrathin gold nanowires
Gong, S., Schwalb, W., Wang, Y., Chen, Y., Tang, Y., Si, J., Shirinzadeh, B., & Cheng, W. (2014) · Nature Communications · doi:10.1038/ncomms4132 · Cited by 2,064
Schwartz, G., Tee, B. C.-K., Mei, J., Appleton, A. L., Kim, D. H., Wang, H., & Bao, Z. (2013) · Nature Communications · doi:10.1038/ncomms2832 · Cited by 2,042
Pan, L., Chortos, A., Yu, G., Wang, Y., Isaacson, S., Allen, R., Shi, Y., Dauskardt, R., & Bao, Z. (2014) · Nature Communications · doi:10.1038/ncomms4002 · Cited by 1,519
Chen, C.-H., Nevo, E., Fetics, B., Pak, P. H., Yin, F. C. P., Maughan, W. L., & Kass, D. A. (1997) · Circulation · doi:10.1161/01.cir.95.7.1827 · Cited by 1,307
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206
Wearable health devices-vital sign monitoring, systems and technologies
Dias, D., & Paulo Silva Cunha, J. (2018) · Sensors · doi:10.3390/s18082414 · Cited by 919
Flexible suspended gate organic thin-film transistors for ultra-sensitive pressure detection
Zang, Y., Zhang, F., Huang, D., Gao, X., Di, C.-A., & Zhu, D. (2015) · Nature Communications · doi:10.1038/ncomms7269 · Cited by 548
Cited by
Microelectromechanical Systems (MEMS) for Biomedical Applications
Chircov, C., & Grumezescu, A. M. (2022) · Micromachines · doi:10.3390/mi13020164 · Cited by 243
Jeong, H., Lee, J. Y., Lee, K., Kang, Y. J., Kim, J.-T., Avila, R., Tzavelis, A., Kim, J., Ryu, H., Kwak, S. S., et al. (2021) · Science Advances · doi:10.1126/sciadv.abg3092 · Cited by 121
Metadata
- Authors
- Kaisti, M., Panula, T., Leppänen, J., Punkkinen, R., Jafari Tadi, M., Vasankari, T., Jaakkola, S., Kiviniemi, T., Airaksinen, J., Kostiainen, P., Meriheinä, U., Koivisto, T., & Pänkäälä, M.
- Venue
- npj Digital Medicine
- Published
- 2019-05-14
- Volume
- 2
- Issue
- 1
- Pages
- 39-39
- ISSN
- 2398-6352 (ISSN-L)
- OpenAlex
- W2946036604
- Cited by
- 164
- References
- 8
- FWCI
- 7.05
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation. https://app.cochat.ai/papers/10.1038/s41746-019-0117-x
- The paper
- Kaisti, M., Panula, T., Leppänen, J., Punkkinen, R., Jafari Tadi, M., Vasankari, T., Jaakkola, S., Kiviniemi, T., Airaksinen, J., Kostiainen, P., Meriheinä, U., Koivisto, T., & Pänkäälä, M. (2019). Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation. npj Digital Medicine. https://doi.org/10.1038/s41746-019-0117-x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.