A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Open accessThe DOI matches its Crossref record; 42 of the 43 references it cites were found in a registry or an index.
Abstract
We report a new method for the real-time quantitative analysis of sodium in human sweat, consolidating sweat collection and analysis in a single, integrated, wearable platform. This temporal data opens up new possibilities in the study of human physiology, broadly applicable from assessing high performance athletes to monitoring Cystic Fibrosis (CF) sufferers. Our compact Sodium Sensor Belt (SSB) consists of a sodium selective Ion Selective Electrode (ISE) integrated into a platform that can be interfaced with the human body during exercise. No skin cleaning regime or sweat storage technology is required as the sweat is continually wicked from the skin to a sensing surface and from there to a storage area via a fabric pump. Our results suggest that after an initial equilibration period, a steady-state sodium plateau concentration was reached. Atomic Absorption Spectroscopy (AAS) was used as a reference method, and this has confirmed the accuracy of the new continuous monitoring approach. The steady-state concentrations observed were found to fall within ranges previously found in the literature, which further validates the approach. Daily calibration repeatability (n = 4) was ±3.0% RSD and over a three month period reproducibility was ±12.1% RSD (n = 56). As a further application, we attempted to monitor the sweat of Cystic Fibrosis (CF) sufferers using the same device. We observed high sodium concentrations symptomatic of CF (∼60 mM Na+) for two CF patients, with no conclusive results for the remaining patients due to their limited exercising capability, and high viscosity/low volume of sweat produced.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state14–18.”
From Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis (2016) · cites this paper as [14]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Exercise and Fluid Replacement
(2007) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e31802ca597 · Cited by 1,497
Gibson, L. E., & Cooke, R. E. (1959) · Pediatrics · doi:10.1542/peds.23.3.545 · Cited by 1,434
Variations in Regional Sweat Composition in Normal Human Males
Patterson, M. J., Galloway, S. D. R., & Nimmo, M. A. (2000) · Experimental Physiology · doi:10.1111/j.1469-445x.2000.02058.x · Cited by 438
Shirreffs, S. M., & Maughan, R. J. (1997) · Journal of Applied Physiology · doi:10.1152/jappl.1997.82.1.336 · Cited by 251
Calixarene-based sensing agents
Diamond, D., & McKervey, M. A. (1996) · Chemical Society Reviews · doi:10.1039/cs9962500015 · Cited by 247
Self-citationBio-sensing textile based patch with integrated optical detection system for sweat monitoring
Morris, D., Coyle, S., Wu, Y., Lau, K. T., Wallace, G., & Diamond, D. (2009) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2009.02.032 · Cited by 235
Self-citationSolid-contact polymeric membrane electrodes with detection limits in the subnanomolar range
Sutter, J., Radu, A., Peper, S., Bakker, E., & Pretsch, E. (2004) · Analytica Chimica Acta · doi:10.1016/j.aca.2004.07.016 · Cited by 211
Solid Contact Potentiometric Sensors for Trace Level Measurements
Chumbimuni-Torres, K. Y., Rubinova, N., Radu, A., Kubota, L. T., & Bakker, E. (2006) · Analytical Chemistry · doi:10.1021/ac050749y · Cited by 206
Fluids and hydration in prolonged endurance performance
von Duvillard, S. P., Braun, W. A., Markofski, M., Beneke, R., & Leithäuser, R. (2004) · Nutrition · doi:10.1016/j.nut.2004.04.011 · Cited by 186
Peer Reviewed: Calixarenes: Designer Ligands for Chemical Sensors
Diamond, D., & Nolan, K. (2001) · Analytical Chemistry · doi:10.1021/ac012376g · Cited by 172
Self-citation
Research works printed but not linked: 2
- Northall, 1995, British Journal of Biomedical Science, 52, 68
- Legrys, 1994, Archives of Pathology & Laboratory Medicine, 118, 865
Cited by
Items 201–289 of 289
Thirsty for a Noninvasive Wearable to Detect Dehydration: A Review
Chiao, J. ‐., Bing, S., Chawang, K., & Crowe, B. (2024) · IEEE Antennas and Propagation Magazine · doi:10.1109/map.2024.3442823 · Cited by 14
Wearable Device for Continuous and Real-Time Monitoring of Human Sweat Sodium
Noor, A. M., Farisi, M. S. A., Mazalan, M., Ibrahim, N. F. A., Wahab, A. A., Zakaria, Z., Rusli, N. I., Sabani, N. B., & Manaf, A. A. (2025) · Sensors · doi:10.3390/s25113467 · Cited by 13
“In comparison to previous wearable systems such as those developed by Schazmann et al. [42] and Pirovano et al. [43], which employed similar offline data processing approaches, the current system advances the field by integrating a more compact, low-power, and accurate electrochemical acquisition…”
What Is a Biosensor?—A Terminological Guide From Biomolecular Recognition to Bioindicators
Weber, T. E., Fritschen, A., Prins, M. W. J., & Blaeser, A. (2026) · Engineering in Life Sciences · doi:10.1002/elsc.70068 · Cited by 12
Zhang, H., Wu, Z., Zhi, Z., Gao, W., Sun, W., Hua, Z., & Wu, Y. (2021) · ACS Omega · doi:10.1021/acsomega.1c04229 · Cited by 12
Innovations in wearable biosensors: A pathway to 24/7 personalized healthcare
Singh, I., Kumar, S., & Koh, J. (2025) · Measurement · doi:10.1016/j.measurement.2025.117938 · Cited by 12
Wearable electrochemical sensors for monitoring of inorganic ions and pH in sweat
Zhao, K., Kang, B., & Zhou, B. (2022) · International Journal of Electrochemical Science · doi:10.20964/2022.04.35 · Cited by 10
Al-omari, M., Sel, K., Mueller, A., Mellinger, A., & Kaya, T. (2013) · Journal of Applied Physics · doi:10.1063/1.4805359 · Cited by 10
Lab‐on‐a‐Chip Platforms for Disease Detection and Diagnosis
Işıksaçan, Z., Guler, M. T., Kalantarifard, A., Asghari, M. H., & Elbüken, Ç. (2017) · Wiley eBooks · doi:10.1002/9781119065036.ch8 · Cited by 10
An Artificial Sweating System for Sweat Sensor Testing Applications
Brueck, A., Bates, K. D., Wood, T., House, W., Martinez, Z., Peters, S., Root, B., Yelamarthi, K., & Kaya, T. (2019) · Electronics · doi:10.3390/electronics8060606 · Cited by 10
“Therefore, significant efforts were spent on first modifying the traditional sweat collection methods to fit into a smaller form-factor for portability, such as Macroduct [16,20] (a commercial sweat collection device that was developed for Cystic Fibrosis diagnosis on infants [21]), fabric wicking [6], and modified ab…”
BackgroundBiomimetic materials based on hydroxyapatite patterns for studying extracellular cell communication
Zyrianova, P. I., Eltantawy, M. M., Silin, D. V., Korolev, I., Nikolaev, K. G., Kozodaev, D. A., Slautina, A., Анатольевич, С. Р., Kholkin, A. L., Ulasevich, S. A., et al. (2024) · Materials & Design · doi:10.1016/j.matdes.2024.112718 · Cited by 9
Monoolein-Based Wireless Capacitive Sensor for Probing Skin Hydration
Chaturvedi, V., Falk, M., Björklund, S., Gonzalez-Martinez, J. F., & Shleev, S. (2024) · Sensors · doi:10.3390/s24144449 · Cited by 9
“Athletes and fitness enthusiasts may use these sensors to track sweat levels and hydration status during physical activities [38–43].”
BackgroundRovira, M., Fernández‐Sánchez, C., & Jiménez‐Jorquera, C. (2021) · Biosensors · doi:10.3390/bios11090303 · Cited by 10
“The development of wearable devices based on electrochemical sensors for in situ sweat biomarker detection was first published in 2010 by the group of Diamond from Dublin University that developed a Na ion selective electrode (ISE) wearable device and was applied to compare the change in concentration of this target s…”
MethodBiosensors for animal health and meat safety monitoring: farm-to-slaughterhouse continuum
Nastasijević, I., Mitrović, R., & Janković, S. (2021) · IOP Conference Series Earth and Environmental Science · doi:10.1088/1755-1315/854/1/012063 · Cited by 9
“They have been developed mostly for human health monitoring, such as analysis of sweat for sodium concentration and lactate levels [23].”
BackgroundRojas, D., Cuartero, M., & Crespo, G. A. (2025) · ACS Sensors · doi:10.1021/acssensors.5c00682 · Cited by 8
Principles, applications, and challenges of E-skin: A mini-review
Lu, Q., Lei, T., Xu, J., & Qin, J. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.166936 · Cited by 9
Properties and Customization of Sensor Materials for Biomedical Applications
Zuliani, C., Curto, V. F., Matzeu, G., Fraser, K. J., & Diamond, D. (2014) · Elsevier eBooks · doi:10.1016/b978-0-08-096532-1.01308-x · Cited by 8
Self-citationNanoparticle-based biomedical sensors
Tricoli, A., & Bo, R. (2020) · Frontiers of nanoscience · doi:10.1016/b978-0-08-102515-4.00009-x · Cited by 9
Vending-Machine-Style Skin Excretion Sensing
Yu, K.-C., Hsu, C.-Y., Prabhu, G. R. D., Chiu, H., & Urban, P. L. (2023) · ACS Sensors · doi:10.1021/acssensors.2c02325 · Cited by 8
Microfluidic Nonwoven-Based Device as a Potential Biosensor for Sweat Analysis
Baysal, G., Kök, F. N., Trabzon, L., Kızıl, H., Göcek, İ., & Karaguzel, B. (2014) · Applied Mechanics and Materials · doi:10.4028/www.scientific.net/amm.490-491.274 · Cited by 8
“This application provided diagnosis of Cystic Fibrosis (CF) disease [7].”
MethodDyshko, K., Nicodemus, M. P., Otterstetter, R., Ghadimi, H., Daniels, S., Fulmer, M., Cheney, Z., Ellis, R., Stege, V., & Monty, C. N. (2023) · European Journal of Applied Physiology · doi:10.1007/s00421-023-05364-4 · Cited by 7
Bayansal, F., Yuksel, M. T., MacDougall, S., & Ahmadi, A. (2023) · Ceramics International · doi:10.1016/j.ceramint.2023.12.322 · Cited by 7
Advances in Wearable Sensing Technologies and Their Impact for Personalized and Preventive Medicine
Nasiri, N., & Tricoli, A. (2018) · InTech eBooks · doi:10.5772/intechopen.76916 · Cited by 8
“The information about the composition of a sample is obtained from the potential difference between two electrodes [11, 12].”
BackgroundHerrero, E. J., & Bühlmann, P. (2021) · Analysis & Sensing · doi:10.1002/anse.202100027 · Cited by 7
Zhao, H., Zhang, X., Khan, W. U., Karbarz, M., & Zhang, Q. M. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202502468 · Cited by 10
Integration of PEDOT:PSS into Cotton Yarn as Ion‐Selective Sensors for Smart Clothing Applications
Li, C., Zhang, N., Ingebrandt, S., & Vu, X. T. (2024) · physica status solidi (a) · doi:10.1002/pssa.202400086 · Cited by 6
Non-Invasive Hydration Monitoring with a Graphene Dual Sweat Sensor
Ng, J. J. Q., Tkachev, S. V., Sim, G. C. F., de Lima, L. F., Koon, G. K. W., Lima, A. P., & Neto, A. H. C. (2025) · Applied Sciences · doi:10.3390/app15094970 · Cited by 6
Wearable nanosensor systems and their applications in healthcare
Ramasamy, M., Kumar, P. S., & Varadan, V. K. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2264812 · Cited by 7
“A wearable potentiometric sodium sensor for cystic fibrosis monitoring was demonstrated by Schazmann et al [31] A conductometric sensor that is fabric-based for measuring the level of hydration was demonstrated by Coyle.”
MethodGao, Y., Zeng, M., Huang, K., & Li, G. (2025) · Microchimica Acta · doi:10.1007/s00604-025-07498-y · Cited by 6
Real-time on-body sweat sensing: Electrochemical insights, progress, and prospects
Saikrithika, S., Zahra, L., Theyagarajan, K., & Kim, Y.-J. (2026) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2026.118882 · Cited by 6
Perumal, K., Sreeja, B. S., Sulthana, S. F., Rajesh, A., & Amirtharajan, R. (2026) · Microfluidics and Nanofluidics · doi:10.1007/s10404-025-02865-w · Cited by 5
Fully Organic Graphene Oxide-based Sensor with Integrated Pump for Sodium Detection
Huang, J., Harvey, J., Hu, C., Faulkner, S. H., King, J. A., Nimmo, M. A., & Tok, A. I. Y. (2013) · International Conference on Sport Sciences Research and Technology Support (icSPORTS) · doi:10.5220/0004636200830088 · Cited by 4
Functionalized Materials for Miniaturized Analytical Devices
Abdelhamid, H. N. (2021) · Wiley eBooks · doi:10.1002/9783527827213.ch9 · Cited by 4
Wearable Biosensors on Sutures and Threads
Sharma, A., Asci, C., Marty, J., & Sonkusale, S. (2024) · Springer eBooks · doi:10.1007/978-981-99-8122-9_13 · Cited by 5
Heikenfeld, J. C. (2024) · Frontiers in Sensors · doi:10.3389/fsens.2024.1408158 · Cited by 4
“In 2013, bolstered by early publications by Dermot Diamond (Schazmann et al., 2010) and Joseph Wang (Jia et al., 2013), the author and the US Air Force began a focused effort in sweat biosensing.”
Harris, J., Bickford, J. R., Cho, P. S., Coppock, M. B., Farrell, M. E., Holthoff, E. L., & Ratcliff, E. L. (2019) · Chemical, Biological, Radiological, Nuclear, and Explosives (CBRNE) Sensing · doi:10.1117/12.2518543 · Cited by 4
Nanosensor and Actuator Technologies for Wearable Mobile Patient Monitoring Systems: A Review
Paliwal, G., & Bunglowala, A. (2021) · Springer proceedings in mathematics & statistics · doi:10.1007/978-981-15-9708-4_7 · Cited by 3
Flexible electrochemical aptasensor for non-invasive potassium detection in human sweat
Ji, S. (2025) · International Journal of Electrochemical Science · doi:10.1016/j.ijoes.2025.101095 · Cited by 3
Holmes, N. P., Elkington, D. C., Walters, J. L. H., Nicholson, L. M., Capozza, M., Magno, M. H. R., Vijayarajan, S., Zhou, X., Belcher, W. J., & Dastoor, P. C. (2018) · Medical Devices & Sensors · doi:10.1002/mds3.10001 · Cited by 3
Development of a miniaturised potentiostat for urea detection using the LMP91000
Nhu, C. T., Dang, P. N., Bui, T. T., & Duc, T. C. (2023) · Ministry of Science and Technology Vietnam · doi:10.31276/vjste.65(1).19-24 · Cited by 3
Biocatalytic Sensors: Potentials, Maxims and Mechanisms for Optimal Performance
Sanni, S. E., Sadiku, E. R., Okoro, E. E., Oni, B. A., & Ndubuisi, A. O. (2023) · Springer eBooks · doi:10.1007/978-981-19-8501-0_6 · Cited by 3
Sanapala, P., & Pola, S. (2023) · Wiley eBooks · doi:10.1002/9783527834266.ch33 · Cited by 3
Fu, Y., Yang, A., & Yan, F. (2017) · Springer eBooks · doi:10.1007/978-3-319-69362-0_12 · Cited by 3
Greene, J. (2021) · Liverpool John Moores University · doi:10.24377/ljmu.t.00014802 · Cited by 5
Song, T., Wang, Q., Han, T., Dufresne, A., da Costa, F. A. T., & Parra, D. F. (2025) · Cellulose · doi:10.1007/s10570-025-06823-1 · Cited by 3
Sensing the Presence of Inorganic Ions in Water: The Use of Electrochemical Sensors
Otun, K. O., Idris, A. O., Ama, O. M., Anku, W. W., Aigbe, U. O., Ukhurebor, K. E., & Onyancha, R. B. (2021) · Engineering materials · doi:10.1007/978-3-030-85555-0_4 · Cited by 3
Biosensor Based on Chitosan Nanocomposite
Li, B., Cheng, Y., Xu, F., Wang, L., Wei, D., Jia, D., Feng, Y., & Zhou, Y. (2015) · Wiley eBooks · doi:10.1002/9781118998861.ch8 · Cited by 2
Clinical Biosensors: Considerations and Development Process
Ramasamy, M., Kumar, P. S., & Varadan, V. K. (2022) · Springer eBooks · doi:10.1007/978-981-16-9897-2_4 · Cited by 2
Molecular analysis of Sweat for Evidence based Ayurvedic Diagnosis
Soni, P., Singh, S., Singh, U., & Agarwal, A. (2023) · IEEE Applied Sensing Conference (APSCON) · doi:10.1109/apscon56343.2023.10101052 · Cited by 1
Wearable Health Monitoring System Based on Human Sweat Analysis using Electrochemical Sensors
Pagar, V., Bhadane, P. K., & Shaligram, A. D. (2023) · Romanian Journal of Biophysics · doi:10.59277/rjb.2023.4.02 · Cited by 1
Tu, A., Lai, K., Lai, J., He, S., Xiao, L., Aw, K. C., Gao, J., & Jiang, H. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3567150 · Cited by 2
“Diamond et al. [34] demonstrated a wearable belt sodium sensor with a potential repeatability of − 10.8 ± 0.8 mV for four tests within one day and enabled the monitoring of sweat sodium ion concentration of approximately 60 mM in CF patients.”
InfluentialZhang, Y., van Santen, H. M., Musson, R. E. A., Markopoulos, P., & Lu, Y. (2025) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2025.100850 · Cited by 1
Recent advances in wearable sensor technology for sweat analysis
Jacob, J., Niranjan, D. B., Sharma, A., Jayakrishnan, S., Jose, R. C., Medikonda, J., Peter, M., & Namboothiri, P. K. (2026) · Discover Applied Sciences · doi:10.1007/s42452-026-08404-8 · Cited by 3
Umapathi, K. (2026) · Analytical Methods · doi:10.1039/d6ay00459h · Cited by 2
Kızılkurtlu, A. A., & Akpek, A. (2026) · Biosensors · doi:10.3390/bios16080448 · Cited by 1
On-Body Chemo/Bio-Sensing - Opportunities and Challenges
Matzeu, G., Fay, C., O’Quigley, C., Orpen, D., Coleman, S., Kavanagh, A., & Diamond, D. (2014) · Advances in science and technology · doi:10.4028/www.scientific.net/ast.96.78 · Cited by 1
“One major challenge inherent to this area of study was in the low volume production of sweat by CF patients [21].”
BackgroundSelf-citationMaterial Processing for Edible Electronics
Xu, W. (2019) · PhDT · Cited by 1
2D MoS2 for sweat based biosensor application
Lekha, P. C., Satheeshkumar, E., & Chandra, T. S. (2020) · AIP conference proceedings · doi:10.1063/5.0017028 · Cited by 1
Textronic Solutions Used to Produce Layers Sensitive to Chemical Stimuli—Gas Sensors: A Review
Skrzetuska, E., & Szablewska, P. (2023) · Materials · doi:10.3390/ma16175893 · Cited by 1
“This product is used in patients suffering from cystic fibrosis [27].”
BackgroundChallenges in Agri-Food Chain: Biosensors in the Meat Production System
Nastasijević, I., Podunavac, I., Janković, S., Mitrović, R., & Radonić, V. (2023) · Meat Technology · doi:10.18485/meattech.2023.64.2.17 · Cited by 1
Lu, D., Yang, C., Chen, L., Cao, C., Xia, W., Li, G., & Wen, D. (2023) · Electroanalysis · doi:10.1002/elan.202300385 · Cited by 1
Updates of Wearing Devices (WDs) In Healthcare, And Disease Monitoring
Naser, M., Nasr, M., & Shehata, L. H. (2023) · International Journal of Progressive Sciences and Technologies · doi:10.52155/ijpsat.v39.1.5361 · Cited by 1
Wearable Health Monitoring System for Sweat Analysis
Pagar, V., Bhadane, P. K., & Shaligram, A. D. (2024) · Communications in computer and information science · doi:10.1007/978-3-031-54547-4_31 · Cited by 1
Towards Stable Electrochemical Sensing for Wearable Wound Monitoring
RoyChoudhury, S. (2019) · FIU - Digital Commons (Florida International University) · doi:10.25148/etd.fidc007817 · Cited by 1
Electrochemical Wearable Devices
Lee, B. H., & Daniele, M. A.-A. (2021) · Wiley eBooks · doi:10.1002/9783527610426.bard030113 · Cited by 1
Glucose Sensor and Its Potential Directions
Cheng, C., Kuan, C.-M., & Chen, C. (2015) · Springer eBooks · doi:10.1007/978-3-319-19737-1_4
Electrochemistry and the Olympics
Ramani, V. (2016) · The Electrochemical Society Interface · doi:10.1149/2.001163if
WIRELESS SWEAT SIGNAL ANALYSER
Raj, G., & Swetha, R. R. (2016) · Computer Science and Information Technology
Paliwal, G., & Bunglowala, A. (2018) · Advances in medical diagnosis, treatment, and care (AMDTC) book series · doi:10.4018/978-1-5225-7131-5.ch007
Desarrollo de diferentes sensores electroquímicos para la determinación de cloruro
Bujes-Garrido, J. (2017) · Repositorio Institucional de la Universidad de Burgos (University of Burgos) · doi:10.36443/10259/5880
Legner, C. M. (2020) · Iowa State University Digital Repository (Iowa State University) · doi:10.31274/etd-20210114-80
Progress of Flexible Wearable Electrochemical Sensors for Human Sweat Detection
曹, 小. (2022) · Advances in Analytical Chemistry · doi:10.12677/aac.2022.123033
Correlating the Effect of Dynamic Variability in the Sensor Environment on Sensor Design
Bhushan, P. (2021) · FIU - Digital Commons (Florida International University) · doi:10.25148/etd.fidc009698 · Cited by 1
Sempionatto, J. R., Lasalde‐Ramírez, J. A., Yang, Y., & Gao, W. F. (2023) · Elsevier eBooks · doi:10.1016/b978-0-443-15315-0.00013-4
Biomedical Applications of Surfactant Modified Graphene Based Electrochemical Sensing Devices
Adimule, V. M., Sharma, K., & Parashuram, L. (2024) · Springer eBooks · doi:10.1007/978-3-031-60832-2_8
Nanotechnology in Animal and Microenvironment Management
Yadav, A., Kumar, T., Sindhu, N., Kumar, A., Mudgil, N., & Anil Kumar, M. S. (2024) · Livestock diseases and management · doi:10.1007/978-981-16-1610-5_17
Nazari, H., Asbagh, R. A., Abasipanah, P., Nazari, Z., & Warkiani, M. E. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-27345-2.00007-2
Wearable Sweat Sensors: A Review
Chowdhury, I., Ali, M. Y., & Howlader, M. M. R. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-95478-5.00015-7 · Cited by 2
Recent Applications, and Commercialization
(2025) · Royal Society of Chemistry eBooks · doi:10.1039/9781837676774-part3
Supercapacitors in Flexible Electronics
Murugesan, M., Nagavenkatesh, K. R., Sambathkumar, C., Devendran, P., & Nallamuthu, N. (2025) · Royal Society of Chemistry eBooks · doi:10.1039/9781837676774-00662
Integrating Wearable Sensor Data and Neural Networks for Real-Time Electrolyte Prediction
Kumar, D. (2025) · Biomedical Materials & Devices · doi:10.1007/s44174-025-00524-w
Smart Materials and Soldiers: Applications and Challenges—A Mini Review
Lu, Q., Lei, T., Xu, J., Qin, J., & Hu, W. (2026) · Advanced Engineering Materials · doi:10.1002/adem.202502320
Development of Low Cost Paper-based Microfluidic Device for the Diagnosis of Cystic Fibrosis
Fong, Z. (2024) · Otago University Research Archive (University of Otago)
Krishnamoorthy, P., & Rakkesh, R. A. (2026) · RSC Advances · doi:10.1039/d6ra06028e
“217 Due to its simplicity and low power consumption, potentiometry is particularly attractive for continuous wearable monitoring.”
KR2‐Hydrogel Opto‐Iontronic Platform for Na + Sensing Over 0.001–1 M
Li, F., Shen, Z., Hou, D., Cai, J., Wei, L., Hu, T., Li, Y., & Lv, Y. (2026) · Advanced Materials Technologies · doi:10.1002/admt.71338
Smart Textile Systems as Dynamic Platform for Human Hydration Monitoring
Zou, X., Zhang, Y., Forrester, S., & Xu, L. (2026) · Biomedical Materials & Devices · doi:10.1007/s44174-026-00780-4
Review of Recent Advances in Non-invasive, Flexi-ble, Wearable Sweat Monitoring Sensors
Medagedara, M. H., Peiris, T. T. S., & Wanasekara, N. D. (2024) · Instrumentation · Cited by 1
She, Z., Shi, L., Xie, D., Jiang, F., Zhong, C., Cheng, Z., Shuning, J., Xie, H., Li, Y., & Zhou, J. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.183437
Tao, Y., Hao, Z., Zhang, Y., Li, G., & Fang, B. (2026) · Nanomaterials · doi:10.3390/nano16211367
Metadata
- Authors
- Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D.
- Affiliations
- Dublin Institute of Technology (IE)
- Dublin City University (IE)
- Corresponding author
- Schazmann, Benjamin
- Venue
- Analytical Methods
- Publisher
- Royal Society of Chemistry (RSC)
- Published
- 2010
- Volume
- 2
- Issue
- 4
- Pages
- 342
- ISSN
- 1759-9660 (ISSN-L)
- 1759-9679
- OpenAlex
- W2144506268
- Cited by
- 286
- References
- 43
- Citations by year
- 2026: 13 · 2025: 20 · 2024: 23 · 2023: 28 · 2022: 18 · 2021: 28 · 2020: 27 · 2019: 29 · 2018: 20 · 2017: 25 · 2016: 18 · 2015: 13 · 2014: 12 · 2013: 7 · 2012: 3
- FWCI
- 2.85
- Type
- journal article
- Language
- en
- Open access
- green
- orcid.org
- Locations
- Analytical Methods (publishedVersion)
- Dublin City University Open Access Institutional Repository (Dublin City University) (submittedVersion)
- Research Online (University of Wollongong) (submittedVersion)
- Arrow@dit (Dublin Institute of Technology) (submittedVersion)
- Arrow@dit (Dublin Institute of Technology) (submittedVersion, cc-by-nc-sa)
- Arrow - TU Dublin (Technological University Dublin) (submittedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Keywords
- sweat sodium concentration, wearable electrochemical sensors, ion-selective electrodes, real-time physiological monitoring, cystic fibrosis, exercise physiology, sweat collection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.