Comparison of two spatially resolved near-infrared photometers in the detection of tissue oxygen saturation: poor reliability at very low oxygen saturation
The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
Two spatially resolved oximeters, NIRO-300 and OM-200, were compared with regard to the measurement of oxygen saturation values in two forearm muscle groups at rest and during arterial occlusion in nine healthy volunteers. There was a significant correlation between the muscle oxygen saturation values obtained at rest using the two oximeters (n=33, r(2)=0.43, P<0.0001), whereas these values were significantly different during arterial occlusion. Thus, although there was good agreement between muscle oxygen saturation values measured using the two oximeters, the operating range of the tissue oximeters should be recognized and indicated.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Statistical normalization approaches, such as a physiological calibration, are then often used to address these problems.7,9,12”
From Near-infrared spectroscopy-derived muscle oxygen saturation on a 0% to 100% scale: reliability and validity of the Moxy Monitor (2019) · cites this paper as [12]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Tissue oxygenation monitor using NIR spatially resolved spectroscopy:
Suzuki, S., Takasaki, S., Ozaki, T., & Kobayashi, Y. (1999) · SPIE Proceedings · doi:10.1117/12.356862 · Cited by 484
Comparison of time-resolved and -unresolved measurements of deoxyhemoglobin in brain.
Chance, B., Leigh, J. S., Miyake, H., Smith, D. S., Nioka, S., Greenfeld, R., Finander, M., Kaufmann, K., Levy, W., & Young, M. (1988) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.85.14.4971 · Cited by 464
Adipose tissue thickness affects in vivo quantitative near-IR spectroscopy in human skeletal muscle
VAN BEEKVELT, M. C. P., BORGHUIS, M. S., VAN ENGELEN, B. G. M., WEVERS, R. A., & COLIER, W. N. J. M. (2001) · Clinical Science · doi:10.1042/cs1010021 · Cited by 369
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 333
Near-infrared oximetry of the brain
Madsen, P. L., & Secher, N. H. (1999) · Progress in Neurobiology · doi:10.1016/s0301-0082(98)00093-8 · Cited by 315
Near‐infrared spectroscopy for monitoring muscle oxygenation
Boushel, R., & Piantadosi, C. A. (2000) · Acta Physiologica Scandinavica · doi:10.1046/j.1365-201x.2000.00713.x · Cited by 299
Absolute quantification methods in tissue near-infrared spectroscopy:
Matcher, S. J., Kirkpatrick, P. J., Nahid, K., Cope, M., & Delpy, D. T. (1995) · SPIE Proceedings · doi:10.1117/12.209997 · Cited by 262
Accuracy of a Cerebral Oximeter in Healthy Volunteers under Conditions of Isocapnic Hypoxia
Henson, L. C., Calalang, C., Temp, J. A., & Ward, D. S. (1998) · Anesthesiology · doi:10.1097/00000542-199801000-00011 · Cited by 122
Quaresima, V., Sacco, S., Totaro, R., & Ferrari, M. (2000) · Journal of Biomedical Optics · doi:10.1117/1.429987 · Cited by 100
Self-citation
Cited by
Principles, Techniques, and Limitations of Near Infrared Spectroscopy
Ferrari, M., Mottola, L., & Quaresima, V. (2004) · Canadian Journal of Applied Physiology · doi:10.1139/h04-031 · Cited by 997
“Conversely, at rest the intramuscular O 2 stores, as measured by NIRS during suprasystolic cuff occlusion, begin to decrease immediately after the beginning of the occlusion and the maximal desaturation is achieved after 5–6 min (Komiyama et al., 2001).”
BackgroundMethodInfluentialSelf-citationHamaoka, T., McCully, K. K., Quaresima, V., Yamamoto, K., & Chance, B. (2007) · Journal of Biomedical Optics · doi:10.1117/1.2805437 · Cited by 329
Self-citationFeldmann, A., Schmitz, R. W., & Erlacher, D. (2019) · Journal of Biomedical Optics · doi:10.1117/1.jbo.24.11.115001 · Cited by 203
“Statistical normalization approaches, such as a physiological calibration, are then often used to address these problems.(7,9,12) The situation becomes even more complicated as no consensus “gold standard” for NIRS-derived values exists to compare against.”
MethodBuchheit, M., Abbiss, C. R., Peiffer, J. J., & Laursen, P. B. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-2021-1 · Cited by 135
“2007) or post 2-min upper-body limb occlusion (Komiyama et al. 2001)).”
BackgroundHamaoka, T., & McCully, K. K. (2019) · The Journal of Physiological Sciences · doi:10.1007/s12576-019-00697-2 · Cited by 92
Hyttel-Sorensen, S., Sorensen, L. C., Riera, J., & Greisen, G. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.003047 · Cited by 92
Lawaetz, T. W. H., Hyttel-Sørensen, S., & Greisen, G. O. (2014) · Acta Paediatrica · doi:10.1111/apa.12567 · Cited by 75
Hiroyuki, H., Hamaoka, T., Sako, T., Nishio, S., Kime, R., Murakami, M., & Katsumura, T. (2002) · European Journal of Applied Physiology · doi:10.1007/s00421-002-0644-y · Cited by 69
“However, a physiological validation of the value obtained by NIR spatially resolved spectroscopy was not confirmed in viewing a recent paper by Komiyama et al. (2001) .”
BackgroundQuaresima, V., Ferrari, M., Franceschini, M. A., Hoimes, M. L., & Fantini, S. (2004) · Journal of Biomedical Optics · doi:10.1117/1.1646417 · Cited by 45
Self-citationYang, Y., Soyemi, O. O., Scott, P. J., Landry, M. R., Lee, S. M. C., Stroud, L., & Soller, B. R. (2007) · Optics Express · doi:10.1364/oe.15.013715 · Cited by 49
Near infrared spectroscopy in brain injury: today’s perspective
Al-Rawi, P. G. (2005) · Acta neurochirurgica. Supplementum · doi:10.1007/3-211-32318-x_93 · Cited by 40
“to their accuracy remain controversial [33, 49].”
Skeletal Muscle Deoxygenation Abnormalities in Early Post-Myocardial Infarction
Takagi, S., Murase, N., Kime, R., Niwayama, M., Osada, T., & Katsumura, T. (2014) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000334 · Cited by 28
Limitation of muscle deoxygenation in the triceps during incremental arm cranking in women
Muraki, S., Tsunawake, N., & Yamasaki, M. (2004) · European Journal of Applied Physiology · doi:10.1007/s00421-003-0962-8 · Cited by 27
“In a recent study, however, Komiyama et al. (2001) reported that spatially-resolved NIRS has poor reliability at very low oxygen saturation, indicating a limitation of the quantification of absolute SO2.”
BackgroundKime, R., Fujioka, M., Osawa, T., Takagi, S., Niwayama, M., Kaneko, Y., Osada, T., Murase, N., & Katsumura, T. (2013) · Advances in experimental medicine and biology · doi:10.1007/978-1-4614-7411-1_23 · Cited by 28
“A previous study reported that fat layer thickness affects SmO 2 [ 5 ].”
BackgroundOxygen Saturation Measurement of Calf Muscle During Exercise in Intermittent Claudication
Komiyama, T., Onozuka, A., Miyata, T., & Shigematsu, H. (2002) · European Journal of Vascular and Endovascular Surgery · doi:10.1053/ejvs.2002.1645 · Cited by 29
Self-citationAdvanced Chronic Venous Insufficiency is Associated with Increased Calf Muscle Deoxygenation
Yamaki, T., Nozaki, M., Sakurai, H., Soejima, K., Kono, T., & Hamahata, A. (2010) · European Journal of Vascular and Endovascular Surgery · doi:10.1016/j.ejvs.2010.01.031 · Cited by 27
Kime, R., Niwayama, M., Fujioka, M., Shiroishi, K., Osawa, T., Shimomura, K., Osada, T., Murase, N., & Katsumura, T. (2009) · Advances in experimental medicine and biology · doi:10.1007/978-1-4419-1241-1_51 · Cited by 18
Muscle Oxygen Saturation Heterogeneity Among Leg Muscles During Ramp Exercise
Takagi, S., Kime, R., Niwayama, M., Murase, N., & Katsumura, T. (2012) · Advances in experimental medicine and biology · doi:10.1007/978-1-4614-4989-8_38 · Cited by 13
Yamaki, T., Nozaki, M., Sakurai, H., KIKUCHI, Y., Soejima, K., Kono, T., Hamahata, A., & Kim, K. (2009) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03528.x · Cited by 15
Soga, T., Sakatani, K., Yagi, T., Kawamorita, T., & Yoshino, A. (2013) · Emergency Medicine Journal · doi:10.1136/emermed-2013-202703 · Cited by 15
Pollonini, L. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2794969 · Cited by 10
“tissue oxygenation during hypoxia, [14].”
BackgroundInfluentialMuscle Deoxygenation and Its Heterogeneity Changes After Endurance Training
Kime, R., Niwayama, M., Kaneko, Y., Takagi, S., Fuse, S., Osada, T., Murase, N., & Katsumura, T. (2016) · Advances in experimental medicine and biology · doi:10.1007/978-3-319-38810-6_37 · Cited by 5
Skeletal Muscle Deoxygenation Responses During Treadmill Exercise in Children
Takagi, S., Kime, R., Midorikawa, T., Niwayama, M., Sakamoto, S., & Katsumura, T. (2014) · Advances in experimental medicine and biology · doi:10.1007/978-1-4939-0620-8_45 · Cited by 3
Al-Rawi, P. G., & Kirkpatrick, P. J. (2013) · Elsevier eBooks · doi:10.1016/b978-1-4377-0167-8.00033-9 · Cited by 3
Ogata, H., Akai, M., & Nakazawa, K. (2008) · Journal of PHYSIOLOGICAL ANTHROPOLOGY · doi:10.2114/jpa2.27.83 · Cited by 5
Near infrared spectroscopy in carotid endarterectomy
Al-Rawi, P. G., & Kirkpatrick, P. J. (2006) · Cambridge University Press eBooks · doi:10.1017/cbo9780511544941.028 · Cited by 1
ЦЕРЕБРаЛьНаЯ ИНФР аКРа СНаЯ оКСИМЕТРИЯ пРИ вНу ТРИчЕРЕпНых КРовоИЗЛИЯНИЯх
Square, P., Novgorod, N., & Russian Federation (2014)
Evaluation of muscle oxygenation based on spatial-resolved broad band and time-resolved spectroscopy
Rohm, O., Gürtler, R., & Kohl‐Bareis, M. (2004) · Biomedical Topical Meeting · doi:10.1364/bio.2004.sh2
CereBrAl infrAred o ximetry in intrACrAniAl hemorrhAge
Square, P., Novgorod, N., & Russian Federation (2014)
A review on the non-invasive evaluation of skeletal muscle oxygenation
Halim, A., Laili, M. H., Aziz, N. A., Laili, A. R., Salikin, M. S., & RUSOP, M. (2016) · AIP conference proceedings · doi:10.1063/1.4948829
Metadata
- Authors
- KOMIYAMA, T., QUARESIMA, V., SHIGEMATSU, H., & FERRARI, M.
- Affiliations
- The University of Tokyo (JP)
- University of L'Aquila (IT)
- Venue
- Clinical Science
- Publisher
- Portland Press Ltd.
- Published
- 2001-11-21
- Volume
- 101
- Issue
- 6
- Pages
- 715-718
- ISSN
- 0143-5221 (ISSN-L)
- 1470-8736
- OpenAlex
- W2093580391
- Semantic Scholar
- 1f398fa2e1344fef0eccbe53e199a06ce74d2dd3
- Cited by
- 31
- References
- 18
- Citations by year
- 2019: 2 · 2018: 1 · 2016: 2 · 2014: 5 · 2013: 3 · 2012: 1
- FWCI
- 1.77
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Science (publishedVersion)
- PubMed (publishedVersion)
- Archivio Istituzionale della Ricerca - Università degli Studi dell Aquila (submittedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Keywords
- tissue oxygen saturation, muscle oxygenation, arterial occlusion, measurement reliability, low oxygen saturation, forearm muscles
- MeSH terms
- Adult, Blood Gas Monitoring, Transcutaneous, Constriction, Humans, Male, Oxygen, Photometry, Reproducibility of Results, Muscle, Skeletal, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.