Measurement Accuracy of Heart Rate and Respiratory Rate during Graded Exercise and Sustained Exercise in the Heat Using the Zephyr BioHarnessTM
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
The Zephyr BioHarness was tested to determine the accuracy of heart rate (HR) and respiratory rate (RR) measurements during 2 exercise protocols in conjunction with either a laboratory metabolic cart (Vmax) or a previously validated portable metabolic system (K4b2). In one protocol, HR and RR were measured using the BioHarness and Vmax during a graded exercise up to V˙O2max (n=12). In another protocol, HR and RR were measured using the BH and K4b2 during sustained exercise (30% and 50% V˙O2max for 20 min each) in a hot environment (30 °C, 50% relative humidity) (n=6). During the graded exercise, HR but not RR, obtained from the BioHarness was higher compared to the Vmax at baseline and 30% V˙O2max (p<0.05), but showed no significant difference at other stages with high correlation coefficients for both HR (r=0.87-0.96) and RR (r=0.90-0.99 above 30% V˙O2max). During the exercise in the heat, there were no significant differences between the BioHarness and K4b2 system. Correlation coefficients between the methods were low for HR but moderately to highly correlated (0.49-0.99) for RR. In conclusion, the BioHarness is comparable to Vmax and K4b2 over a wide range of V˙O2 during graded exercise and sustained exercise in the heat.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
Cited by
Evolution of Wearable Devices with Real-Time Disease Monitoring for Personalized Healthcare
Guk, K., Han, G., Lim, J., Jeong, K., Kang, T., Lim, E.-K., & Jung, J. (2019) · Nanomaterials · doi:10.3390/nano9060813 · Cited by 624
The Importance of Respiratory Rate Monitoring: From Healthcare to Sport and Exercise
Nicolò, A., Massaroni, C., Schena, E., & Sacchetti, M. (2020) · Sensors · doi:10.3390/s20216396 · Cited by 440
Wearable sensors for monitoring the internal and external workload of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0149-2 · Cited by 330
Nazari, G., MacDermid, J. C., Sinden, K. E., Richardson, J., & Tang, A. (2019) · Physiotherapy Canada · doi:10.3138/ptc.2018-25 · Cited by 15
Metadata
- Authors
- Kim, J.-H., Roberge, R., Powell, J., Shafer, A., & Jon Williams, W.
- Venue
- International Journal of Sports Medicine
- Publisher
- Georg Thieme Verlag KG
- Published
- 2012-11-22
- Volume
- 34
- Issue
- 06
- Pages
- 497-501
- ISSN
- 0172-4622 (ISSN-L)
- 1439-3964
- OpenAlex
- W2331248963
- Cited by
- 123
- References
- 2
- FWCI
- 9.29
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Measurement Accuracy of Heart Rate and Respiratory Rate during Graded Exercise and Sustained Exercise in the Heat Using the Zephyr BioHarnessTM. https://app.cochat.ai/papers/10.1055/s-0032-1327661
- The paper
- Kim, J.-H., Roberge, R., Powell, J., Shafer, A., & Jon Williams, W. (2012). Measurement Accuracy of Heart Rate and Respiratory Rate during Graded Exercise and Sustained Exercise in the Heat Using the Zephyr BioHarnessTM. International Journal of Sports Medicine. https://doi.org/10.1055/s-0032-1327661
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.