Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia
Open accessAbstract
BACKGROUND: The threshold-suspend feature of sensor-augmented insulin pumps is designed to minimize the risk of hypoglycemia by interrupting insulin delivery at a preset sensor glucose value. We evaluated sensor-augmented insulin-pump therapy with and without the threshold-suspend feature in patients with nocturnal hypoglycemia. METHODS: We randomly assigned patients with type 1 diabetes and documented nocturnal hypoglycemia to receive sensor-augmented insulin-pump therapy with or without the threshold-suspend feature for 3 months. The primary safety outcome was the change in the glycated hemoglobin level. The primary efficacy outcome was the area under the curve (AUC) for nocturnal hypoglycemic events. Two-hour threshold-suspend events were analyzed with respect to subsequent sensor glucose values. RESULTS: A total of 247 patients were randomly assigned to receive sensor-augmented insulin-pump therapy with the threshold-suspend feature (threshold-suspend group, 121 patients) or standard sensor-augmented insulin-pump therapy (control group, 126 patients). The changes in glycated hemoglobin values were similar in the two groups. The mean AUC for nocturnal hypoglycemic events was 37.5% lower in the threshold-suspend group than in the control group (980 ± 1200 mg per deciliter [54.4 ± 66.6 mmol per liter] × minutes vs. 1568 ± 1995 mg per deciliter [87.0 ± 110.7 mmol per liter] × minutes, P<0.001). Nocturnal hypoglycemic events occurred 31.8% less frequently in the threshold-suspend group than in the control group (1.5 ± 1.0 vs. 2.2 ± 1.3 per patient-week, P<0.001). The percentages of nocturnal sensor glucose values of less than 50 mg per deciliter (2.8 mmol per liter), 50 to less than 60 mg per deciliter (3.3 mmol per liter), and 60 to less than 70 mg per deciliter (3.9 mmol per liter) were significantly reduced in the threshold-suspend group (P<0.001 for each range). After 1438 instances at night in which the pump was stopped for 2 hours, the mean sensor glucose value was 92.6 ± 40.7 mg per deciliter (5.1 ± 2.3 mmol per liter). Four patients (all in the control group) had a severe hypoglycemic event; no patients had diabetic ketoacidosis. CONCLUSIONS: This study showed that over a 3-month period the use of sensor-augmented insulin-pump therapy with the threshold-suspend feature reduced nocturnal hypoglycemia, without increasing glycated hemoglobin values. (Funded by Medtronic MiniMed; ASPIRE ClinicalTrials.gov number, NCT01497938.).
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“People with diabetes who continue to have frequent and severe hypoglycemia and/or impaired awareness of hypoglycemia, despite educational interventions, may benefit from continuous subcutaneous insulin infusion (CSII) therapy or continuous glucose monitoring (CGM) or both (i.e. a sensor augmented pump), to reduce the risk of severe hypoglycemia (44)(45)(46)(47).”
From Hypoglycemia (2018) · cites this paper as [46]
In the citing paper
“Insulin pumps with automated low-glucose suspend and automated insulin delivery systems have been shown to be effective in reducing hypoglycemia in type 1 diabetes (126).”
From 6. Glycemic Goals, Hypoglycemia, and Hyperglycemic Crises: Standards of Care in Diabetes-2026. (2026) · cites this paper as [126]
In the citing paper
“Reduction of nocturnal hypoglycemia in individuals with type 1 diabetes using insulin pumps with CGM is improved by automatic suspension of insulin delivery at a preset glucose level, with further improvements when using devices with predictive low-glucose insulin delivery suspension (24,25).”
From 9. Pharmacologic Approaches to Glycemic Treatment: Standards of Care in Diabetes—2025 (2025) · cites this paper as [24]
In the citing paper
“CGM with automated low glucose suspend and hybrid closed-loop systems have been shown to be effective in reducing hypoglycemia in type 1 diabetes (95).”
From 6. Glycemic Targets: Standards of Care in Diabetes—2023 (2023) · cites this paper as [95]
Citations
Cites
EuroQol: a new facility for the measurement of health-related quality of life
(1990) · Health Policy · doi:10.1016/0168-8510(90)90421-9 · Cited by 15,736
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894
Self-citationHypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes.
Cryer, P. (2002) · Diabetologia · doi:10.1007/s00125-002-0822-9 · Cited by 794
Fear of hypoglycemia: quantification, validation, and utilization
Cox, D. J., Irvine, A., Gonder-Frederick, L., Nowacek, G., & Butterfield, J. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.617 · Cited by 579
Mechanisms of Hypoglycemia-Associated Autonomic Failure and Its Component Syndromes in Diabetes
Cryer, P. E. (2005) · Diabetes · doi:10.2337/diabetes.54.12.3592 · Cited by 401
Adaptive increase in sample size when interim results are promising: a practical guide with examples
Mehta, C. R., & Pocock, S. J. (2011) · Statistics in Medicine · doi:10.1002/sim.4102 · Cited by 367
Towler, D. A., Havlin, C. E., Craft, S., & Cryer, P. (1993) · Diabetes · doi:10.2337/diab.42.12.1791 · Cited by 258
Tanenberg, R. J., Newton, C. A., & Drake, A. J. (2010) · Endocrine Practice · doi:10.4158/ep09260.cr · Cited by 225
Increasing the sample size when the unblinded interim result is promising
Chen, Y. H. J., DeMets, D. L., & Gordon Lan, K. K. (2004) · Statistics in Medicine · doi:10.1002/sim.1688 · Cited by 207
Hypoglycemia in Type 1 Diabetes Mellitus
Cryer, P. E. (2010) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2010.05.003 · Cited by 205
Cited by
Standards of medical care in diabetes—2014
American Diabetes Association (2014) · Diabetes Care · doi:10.2337/dc14-s014 · Cited by 4,572
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
Self-citation9. Pharmacologic Approaches to Glycemic Treatment: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s009 · Cited by 1,555
6. Glycemic Targets: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s006 · Cited by 1,403
“CGM with automated low glucose suspend and hybrid closed-loop systems have been shown to be effective in reducing hypoglycemia in type 1 diabetes (97).”
MethodType 1 Diabetes Through the Life Span: A Position Statement of the American Diabetes Association
Chiang, J. L., Kirkman, M. S., Laffel, L. M. B., Peters, A. L., & Authors, O. B. O. T. T. 1. D. S. (2014) · Diabetes Care · doi:10.2337/dc14-1140 · Cited by 965
Metadata
- Authors
- Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R.
- Affiliations
- Park Nicollet Health Services (US)
- Mills Peninsula Health Services (US)
- University of Colorado Denver (US)
- Atlanta Diabetes Associates (US)
- University of Wisconsin–Madison (US)
- Oregon Health & Science University (US)
- ORCID
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2013-07-18
- Volume
- 369
- Issue
- 3
- Pages
- 224-232
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2102877966
- Cited by
- 652
- References
- 21
- Citations by year
- 2026: 7 · 2025: 17 · 2024: 18 · 2023: 36 · 2022: 30 · 2021: 41 · 2020: 39 · 2019: 62 · 2018: 75 · 2017: 80 · 2016: 82 · 2015: 74 · 2014: 73 · 2013: 17
- FWCI
- 64.50
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- cc-by-nc
- Locations
- New England Journal of Medicine (publishedVersion)
- PubMed (publishedVersion)
- Repositorio Institucional (Pontificia Universidad Javeriana) (submittedVersion, cc-by-nc)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- sensor-augmented pump, insulin pump therapy, nocturnal hypoglycemia, type 1 diabetes, HbA1c, hypoglycemia prevention, continuous glucose monitoring, glycemic control, hypoglycemic events, hypoglycemia severity, hypoglycemia duration
- MeSH terms
- Adolescent, Adult, Aged, Blood Glucose, Diabetes Mellitus, Type 1, Female, Glycated Hemoglobin, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged, Area Under Curve, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia. https://app.cochat.ai/papers/10.1056/nejmoa1303576
- The paper
- Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013). Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia. New England Journal of Medicine. https://doi.org/10.1056/nejmoa1303576
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.