PolLux: A new facility for soft x-ray spectromicroscopy at the Swiss Light Source
Open accessAbstract
We report on the successful installation and operation of a scanning transmission x-ray microspectroscope (STXM) at the PolLux facility at the Swiss Light Source. This integration of an advanced STXM with improved sample handling capabilities and a novel beamline provides unique capabilities. PolLux uses linearly or circularly polarized x-rays from a bending magnet with an extended photon energy range (200–1400 eV). It is therefore well suited to determine a sample’s quantitative chemical composition, molecular orientation, or thickness of organic as well as condensed matter materials. The local magnetic state of magnetic thin films is accessible through fast helicity switching by steering the electron beam off axis through the bending magnet. Ex vacuo girder movers allow fast and highly reproducible (<1 μm) alignment of the instrument with respect to the photon beam. The present spatial resolution is ∼20 nm, limited by the zone plates utilized. The instrument has the stability and positional resolution to operate with much higher resolution optics as it becomes available. In addition to characterization experiments, we present several typical examples from materials research and environmental science to exemplify the capabilities.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Henke, B. L., Gullikson, E. M., & Davis, J. C. (1993) · Atomic Data and Nuclear Data Tables · doi:10.1006/adnd.1993.1013 · Cited by 6,285
Microbubbles in medical imaging: current applications and future directions
Lindner, J. R. (2004) · Nature Reviews Drug Discovery · doi:10.1038/nrd1417 · Cited by 872
Magnetic vortex core reversal by excitation with short bursts of an alternating field
Van Waeyenberge, B., Puzic, A., Stoll, H., Chou, K. W., Tyliszczak, T., Hertel, R., Fähnle, M., Brückl, H., Rott, K., Reiss, G., et al. (2006) · Nature · doi:10.1038/nature05240 · Cited by 861
Self-citationInterferometer-controlled scanning transmission X-ray microscopes at the Advanced Light Source
Kilcoyne, A. L. D., Tyliszczak, T., Steele, W. F., Fakra, S., Hitchcock, P., Franck, K., Anderson, E., Harteneck, B., Rightor, E. G., Mitchell, G. E., et al. (2003) · Journal of Synchrotron Radiation · doi:10.1107/s0909049502017739 · Cited by 692
Self-citationSoft X-ray microscopes and their biological applications
Kirz, J., Jacobsen, C., & Howells, M. (1995) · Quarterly Reviews of Biophysics · doi:10.1017/s0033583500003139 · Cited by 672
Chemical Contrast in X-Ray Microscopy and Spatially Resolved XANES Spectroscopy of Organic Specimens
Ade, H., Zhang, X., Cameron, S., Costello, C., Kirz, J., & Williams, S. (1992) · Science · doi:10.1126/science.1439809 · Cited by 373
Self-citationUrquhart, S. G., & Ade, H. (2002) · The Journal of Physical Chemistry B · doi:10.1021/jp0255379 · Cited by 299
Self-citationCalibrated NEXAFS spectra of some common polymers
Dhez, O., Ade, H., & Urquhart, S. G. (2003) · Journal of Electron Spectroscopy and Related Phenomena · doi:10.1016/s0368-2048(02)00237-2 · Cited by 298
Self-citationAde, H., & Hitchcock, A. P. (2008) · Polymer · doi:10.1016/j.polymer.2007.10.030 · Cited by 286
Self-citationSoft X-ray spectromicroscopy beamline at the CLS: Commissioning results
Kaznatcheev, K. V., Karunakaran, C., Lanke, U. D., Urquhart, S. G., Obst, M., & Hitchcock, A. P. (2007) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2007.08.083 · Cited by 250
Cited by
Liati, A., Schreiber, D., Alpert, P. A., Liao, Y., Brem, B. T., Corral Arroyo, P., Hu, J., Jonsdottir, H. R., Ammann, M., & Dimopoulos Eggenschwiler, P. (2019) · Environmental Pollution · doi:10.1016/j.envpol.2019.01.078 · Cited by 75
Foetisch, A., Filella, M., Watts, B., Vinot, L.-H., & Bigalke, M. (2022) · Journal of Hazardous Materials · doi:10.1016/j.jhazmat.2021.127804 · Cited by 58
He, X., Mukherjee, S., Watkins, S., Chen, M., Qin, T., Thomsen, L., Ade, H., & McNeill, C. R. (2014) · The Journal of Physical Chemistry C · doi:10.1021/jp501222w · Cited by 44
Self-citationFoetisch, A., Filella, M., Watts, B., Bragoni, M., & Bigalke, M. (2023) · Microplastics and Nanoplastics · doi:10.1186/s43591-023-00066-2 · Cited by 21
Metadata
- Authors
- Raabe, J., Tzvetkov, G., Flechsig, U., Böge, M., Jaggi, A., Sarafimov, B., Vernooij, M. G. C., Huthwelker, T., Ade, H., Kilcoyne, D., Tyliszczak, T., Fink, R. H., & Quitmann, C.
- Venue
- Review of Scientific Instruments
- Publisher
- AIP Publishing
- Published
- 2008-11-01
- Volume
- 79
- Issue
- 11
- Pages
- 113704-113704
- ISSN
- 0034-6748 (ISSN-L)
- 1089-7623
- OpenAlex
- W2018951073
- Cited by
- 269
- References
- 23
- FWCI
- 15.13
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: PolLux: A new facility for soft x-ray spectromicroscopy at the Swiss Light Source. https://app.cochat.ai/papers/10.1063/1.3021472
- The paper
- Raabe, J., Tzvetkov, G., Flechsig, U., Böge, M., Jaggi, A., Sarafimov, B., Vernooij, M. G. C., Huthwelker, T., Ade, H., Kilcoyne, D., Tyliszczak, T., Fink, R. H., & Quitmann, C. (2008). PolLux: A new facility for soft x-ray spectromicroscopy at the Swiss Light Source. Review of Scientific Instruments. https://doi.org/10.1063/1.3021472
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.