Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement

Open access
Halfon, Matthieu, Wuerzner, Gregoire, Marques-Vidal, Pedro, Taffe, Patrick, Vaucher, Julien, Waeber, Bernard, Liaudet, Lucas, Ltaief, Zied, Popov, Milen, Waeber, Gerard · 2018 · Blood Pressure
Cited by 25

The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.

Abstract

BACKGROUND: The use of automated (oscillometric) blood pressure (BP) devices is not validated in atrial fibrillation (AF) patients. OBJECTIVES: To assess the reliability of three oscillometric BP devices, and the agreement with invasive arterial blood pressure(IBP) in AF patients. METHODS: 48 AF patients with randomized sequences of 10 consecutive BP measurements with two pairs of devices: (1) OmronR7™(wrist) and OmronHEM907™(arm); (2) OmronR7™ and Microlife WatchBPhome(arm). Reliability and agreement of each device were assessed by the intra-class correlation coefficient (ICC) for the continuous BP measurements and Bland & Altman methodology, respectively. In 10 additional AF patients, 10 consecutive measurements with IBP and OmronHEM907™, and IBP and Microlife WatchBPhome were performed. RESULTS: The OmronR7™ was not able to obtain any BP Readings. Arm devices presented better ICC for systolicBP(SBP) than for diastolicBP(DBP) (Omron HEM907™:0.94 [0.90; 0.97] vs. 0.77 [0.67; 0.89]; Microlife WatchBPhome:0.92 [0.88; 0.96] vs.0.79 [0.69; 0.89]).The correlation coefficient between Microlife WatchBPhome and IBP computed using the average of repeated measurements from two to ten measurements improved up to the third and remained stable afterwards. The agreement between IBP and SBP, and IBP and DBP, was moderate as illustrated by a wide limit of agreement [-24; 26](SBP) and [-15;17](DBP) for Microlife WatchBPHome, respectively and [-30; 13](SBP) and [-7; 15](DBP) for OmronHEM907. CONCLUSIONS: BP measurement using the two arm oscillometric devices achieved a high reliability for SBP. The agreement between IBP and arm devices was low but using the average of three consecutive measurements improved the results substantially.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1080/08037051.2017.1383852

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 5 references

Cited by

Metadata

Authors
Halfon, M., Wuerzner, G., Marques-Vidal, P., Taffe, P., Vaucher, J., Waeber, B., Liaudet, L., Ltaief, Z., Popov, M., & Waeber, G.
Venue
Blood Pressure
Published
2017-09-28
Volume
27
Issue
1
Pages
48-55
ISSN
  • 0803-7051 (ISSN-L)
  • 1651-1999
OpenAlex
W2760793119
Cited by
25
References
5
FWCI
1.09
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.