Reliability and Validity of the Lactate Pro 2 Analyzer
Open accessThe DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Abstract
This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L−1 for blood lactate (BLa) ≤4.0 mmol.L−1 to −1.6 to +1.4 mmol.L−1 for BLa >8.0 mmol.L−1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa ≤ 8 mmol.L−1, and using the LP2 to identify training intensities at lactate threshold.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The lactate analyzer (Lactate Pro 2) was verified as reliable in research using a graded incremental rowing test, comparing it with a laboratory-based analyzer (Crotty et al., 2021).”
From The relationship of muscle oxygen saturation analyzer with other monitoring and quantification tools in a maximal incremental treadmill test (2023) · cites this paper as Crotty et al., 2021
In the citing paper
“which has been shown to have strong intra-device reliability with intraclass correlation coefficients (ICC) = 0.99, with a coefficient of variation (CV) = 3.3% and 95% limits of agreement (LoA) between −1.0 and 1.0 mmol L-1, making it ideal for testing in large team scenarios (Crotty et al., 2021).”
From Monitoring skeletal muscle oxygen saturation kinetics during graded exercise testing in NCAA division I female rowers (2025) · cites this paper as Crotty et al., 2021
In the citing paper
“Previous studies showed that Lactate Pro 2 presented a good reliability compared with blood samples (Crotty et al., 2021; Raa et al., 2020), with higher agreement if the measurement is performed in the earlobe than in the fingertip (Raa et al., 2020), and with Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L -1 to +1.4 mmol.L -1 for concentrations lower than 4.0 mmol.L -1 and higher”
From A comparative analysis of mathematical methods for detecting lactate thresholds using muscle oxygenation data during a graded cycling test (2023) · cites this paper as Crotty et al., 2021
Citations
Cites
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 31,709
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532 · cited 1×
“The ICC (3,1) was used in the current study according to the guidelines of Shrout and Fleiss (1979) and as described by Koo and Li (2016).”
MethodMeasuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
Atkinson, G., & Nevill, A. M. (1998) · Sports Medicine · doi:10.2165/00007256-199826040-00002 · Cited by 3,779
JOURNAL OF SPORTS SCIENCE & MEDICINE
Kallinen, M. (2005) · Cited by 1,481
Lactate Threshold Concepts: How Valid are They?
Faude, O., Kindermann, W., & Meyer, T. (2009) · Sports Medicine · doi:10.2165/00007256-200939060-00003 · Cited by 882 · cited 2×
“L−1 (Faude et al., 2009; Goodwin et al., 2007).”
(2022) · Journal of Sports Medicine · doi:10.1155/5363 · Cited by 416
Doglikuu (2021) · doi:10.21315/mjms · Cited by 406
Bentley, D. J., Newell, J., & Bishop, D. (2007) · Sports Medicine · doi:10.2165/00007256-200737070-00002 · Cited by 404
A New Approach for the Determination of Ventilatory and Lactate Thresholds
Cheng, B., Kuipers, H., Snyder, A., Keizer, H., Jeukendrup, A., & Hesselink, M. (1992) · International Journal of Sports Medicine · doi:10.1055/s-2007-1021309 · Cited by 371
Research works printed but not linked: 5
- Buckley J. D., 2003, Journal of Science and Medicine, 6, 408
- Ghosh A. K., 2004, Malaysian Journal of Medical Sciences, 11, 24
- Goodwin M. L., 2007, Journal of Diabetes, 1, 558
- Mc Grath E., 2019, International Journal of Exercise Science, 12, 1334
- Rowe G. S., 2016, Gazzetta Medica Italiana Archivio per le Scienze Mediche, 175, 378
Cited by
Mentzoni, F., Skaugen, M., Eythorsdottir, I., Roterud, S., JOHANSEN, E. S., & Losnegard, T. (2024) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05572-6 · Cited by 33
Maré, L. D., Boshuizen, B., de Vega, C. V. M., meeûs, C. D., Plancke, L., Gansemans, Y., Nieuwerburgh, F. V., Deforce, D. L. D., de Oliveira, J. E., Hosotani, G., et al. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.792052 · Cited by 26
Conte, D., Lukonaitienė, I., Matulaitis, K., Sniečkus, A., Kniubaitė, A., Kreivytė, R., & Kamandulis, S. (2023) · Biology of Sport · doi:10.5114/biolsport.2023.122478 · Cited by 24
Spitz, S., Schobesberger, S., Brandauer, K., & Ertl, P. (2023) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10604 · Cited by 20
Takei, N., Inaba, T., Morita, Y., Kakinoki, K., Hatta, H., & Kitaoka, Y. (2024) · Temperature · doi:10.1080/23328940.2024.2375693 · Cited by 19
“%) compared to a reliable laboratory-based analyzer [24].”
Metadata
- Authors
- Crotty, N. M., Boland, M., Mahony, N., Donne, B., & Fleming, N.
- Affiliations
- Trinity College Dublin (IE)
- Corresponding author
- Crotty, Nora May
- Venue
- Measurement in Physical Education and Exercise Science
- Publisher
- Informa UK Limited
- Published
- 2021-01-25
- Volume
- 25
- Issue
- 3
- Pages
- 202-211
- ISSN
- 1091-367X (ISSN-L)
- 1532-7841
- OpenAlex
- W3122543005
- Cited by
- 50
- References
- 19
- Citations by year
- 2026: 8 · 2025: 11 · 2024: 10 · 2023: 15 · 2022: 4 · 2021: 2
- FWCI
- 7.65
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- cc-by-nc-sa
- Locations
- Measurement in Physical Education and Exercise Science (publishedVersion)
- Trinity's Access to Research Output (TARA) (Trinity College Dublin) (submittedVersion, cc-by-nc-sa)
- Funders
- Trinity College Dublin
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Sports Performance and Training
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Health Sciences › Health Professions › Physical Therapy, Sports Therapy and Rehabilitation › Physical Education and Training Studies
- Keywords
- blood lactate measurement, measurement reliability, measurement validity, lactate threshold, oxygen consumption, heart rate, exercise intensity
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Reliability and Validity of the Lactate Pro 2 Analyzer. https://app.cochat.ai/papers/10.1080/1091367x.2020.1865966
- The paper
- Crotty, N. M., Boland, M., Mahony, N., Donne, B., & Fleming, N. (2021). Reliability and Validity of the Lactate Pro 2 Analyzer. Measurement in Physical Education and Exercise Science. https://doi.org/10.1080/1091367x.2020.1865966
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.