Non-invasive continuous core temperature measurement by zero heat flux
Open accessThe DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
Reliable continuous core temperature measurement is of major importance for monitoring patients. The zero heat flux method (ZHF) can potentially fulfil the requirements of non-invasiveness, reliability and short delay time that current measurement methods lack. The purpose of this study was to determine the performance of a new ZHF device on the forehead regarding these issues. Seven healthy subjects performed a protocol of 10 min rest, 30 min submaximal exercise (average temperature increase about 1.5 °C) and 10 min passive recovery in ambient conditions of 35 °C and 50% relative humidity. ZHF temperature (T(zhf)) was compared to oesophageal (T(es)) and rectal (T(re)) temperature. ΔT(zhf)-T(es) had an average bias ± standard deviation of 0.17 ± 0.19 °C in rest, -0.05 ± 0.18 °C during exercise and -0.01 ± 0.20 °C during recovery, the latter two being not significant. The 95% limits of agreement ranged from -0.40 to 0.40 °C and T(zhf) had hardly any delay compared to T(es). T(re) showed a substantial delay and deviation from T(es) when core temperature changed rapidly. Results indicate that the studied ZHF sensor tracks T(es) very well in hot and stable ambient conditions and may be a promising alternative for reliable non-invasive continuous core temperature measurement in hospital.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Evaluation of wireless determination of skin temperature using iButtons
VANMARKENLICHTENBELT, W., DAANEN, H., WOUTERS, L., FRONCZEK, R., RAYMANN, R., SEVERENS, N., & VANSOMEREN, E. (2006) · Physiology & Behavior · doi:10.1016/j.physbeh.2006.04.026 · Cited by 392
Self-citationSelkirk, G. A., McLellan, T. M., Wright, H. E., & Rhind, S. G. (2008) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00917.2007 · Cited by 170
Bitar, D., Goubar, A., & Desenclos, J. C. (2009) · Eurosurveillance · doi:10.2807/ese.14.06.19115-en · Cited by 150
Exercise induces gastric ischemia in healthy volunteers: a tonometry study
Otte, J. A., Oostveen, E., Geelkerken, R. H., Groeneveld, A. B. J., & Kolkman, J. J. (2001) · Journal of Applied Physiology · doi:10.1152/jappl.2001.91.2.866 · Cited by 138
A non-invasive device to continuously determine heat strain in humans
Gunga, H.-C., Sandsund, M., Reinertsen, R. E., Sattler, F., & Koch, J. (2008) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2008.03.004 · Cited by 129
Gunga, H.-C., Werner, A., Stahn, A., Steinach, M., Schlabs, T., Koralewski, E., Kunz, D., Belavý, D. L., Felsenberg, D., Sattler, F., et al. (2009) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2009.04.005 · Cited by 122
Krack, A., Richartz, B. M., Gastmann, A., Greim, K., Lotze, U., Anker, S. D., & Figulla, H. R. (2004) · European Journal of Heart Failure · doi:10.1016/j.ejheart.2004.03.002 · Cited by 83
Selective brain cooling: a multiple regulatory mechanism
Caputa, M. (2004) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2004.08.079 · Cited by 65
Determining core body temperature via heat flux-a new promising approach
Opatz, O., Stahn, A., Werner, A., & Gunga, H.-C. (2010) · Resuscitation · doi:10.1016/j.resuscitation.2010.05.025 · Cited by 11 · cited 2×
“As recently brought up by Opatz et al (2010), the device has not shown its capabilities under different ambient and physiological conditions.”
Background
Cited by
Considerations for the measurement of core, skin and mean body temperatures
Taylor, N. A. S., Tipton, M. J., & Kenny, G. P. (2014) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2014.10.006 · Cited by 420
Smart health monitoring systems: An overview of design and modeling
Baig, M. M., & Gholamhosseini, H. (2013) · Journal of Medical Systems · doi:10.1007/s10916-012-9898-z · Cited by 396
Mendt, S., Maggioni, M. A., Nordine, M., Steinach, M., Opatz, O., Belavý, D., Felsenberg, D., Koch, J., Shang, P., Gunga, H.-C., et al. (2017) · Chronobiology International · doi:10.1080/07420528.2016.1224241 · Cited by 60
“To date, various studies have demonstrated the reliability and validity of heat-flux approaches recorded under varying environmental conditions (Gunga et al., 2008; Teunissen et al., 2011), as well as in different clinical settings (Kimberger et al.”
BackgroundAccuracy of Zero-Heat-Flux Cutaneous Temperature in Intensive Care Adults
Dahyot-Fizelier, C., Lamarche, S., Kerforne, T., Bénard, T., Giraud, B., Bellier, R., Carise, E., Frasca, D., & Mimoz, O. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002317 · Cited by 34
Clothing for protection against heat and flames
Rossi, R. (2014) · Protective Clothing · doi:10.1533/9781782420408.1.70 · Cited by 17
Metadata
- Authors
- Teunissen, L. P. J., Klewer, J., de Haan, A., de Koning, J. J., & Daanen, H. A. M.
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2011-03-28
- Volume
- 32
- Issue
- 5
- Pages
- 559-570
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2010277406
- Cited by
- 107
- References
- 10
- FWCI
- 4.24
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Non-invasive continuous core temperature measurement by zero heat flux. https://app.cochat.ai/papers/10.1088/0967-3334/32/5/005
- The paper
- Teunissen, L. P. J., Klewer, J., de Haan, A., de Koning, J. J., & Daanen, H. A. M. (2011). Non-invasive continuous core temperature measurement by zero heat flux. Physiological Measurement. https://doi.org/10.1088/0967-3334/32/5/005
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.