Analysis of using EMG and mechanical sensors to enhance intent recognition in powered lower limb prostheses
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The purpose of this study was to determine the contribution of electromyography (EMG) data, in combination with a diverse array of mechanical sensors, to locomotion mode intent recognition in transfemoral amputees using powered prostheses. Additionally, we determined the effect of adding time history information using a dynamic Bayesian network (DBN) for both the mechanical and EMG sensors. APPROACH: EMG signals from the residual limbs of amputees have been proposed to enhance pattern recognition-based intent recognition systems for powered lower limb prostheses, but mechanical sensors on the prosthesis-such as inertial measurement units, position and velocity sensors, and load cells-may be just as useful. EMG and mechanical sensor data were collected from 8 transfemoral amputees using a powered knee/ankle prosthesis over basic locomotion modes such as walking, slopes and stairs. An offline study was conducted to determine the benefit of different sensor sets for predicting intent. MAIN RESULTS: EMG information was not as accurate alone as mechanical sensor information (p < 0.05) for any classification strategy. However, EMG in combination with the mechanical sensor data did significantly reduce intent recognition errors (p < 0.05) both for transitions between locomotion modes and steady-state locomotion. The sensor time history (DBN) classifier significantly reduced error rates compared to a linear discriminant classifier for steady-state steps, without increasing the transitional error, for both EMG and mechanical sensors. Combining EMG and mechanical sensor data with sensor time history reduced the average transitional error from 18.4% to 12.2% and the average steady-state error from 3.8% to 1.0% when classifying level-ground walking, ramps, and stairs in eight transfemoral amputee subjects. SIGNIFICANCE: These results suggest that a neural interface in combination with time history methods for locomotion mode classification can enhance intent recognition performance; this strategy should be considered for future real-time experiments.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A new strategy for multifunction myoelectric control
Hudgins, B., Parker, P., & Scott, R. N. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.204774 · Cited by 2,149 · cited 1×
“Non-weight-bearing control, similar to myoelectric control of upper limb prostheses [9, 10], has been accomplished through signal processing techniques applied to EMG from isometric contractions to reposition a knee or ankle [11–13].”
MethodShirazi, S. F. S., Gharehkhani, S., Mehrali, M., Yarmand, H., Metselaar, H. S. C., Adib Kadri, N., & Osman, N. A. A. (2015) · Science and Technology of Advanced Materials · doi:10.1088/1468-6996/16/3/033502 · Cited by 755
Huang, Y., Englehart, K. B., Hudgins, B., & Chan, A. D. C. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.856295 · Cited by 643
Huang, H., Zhang, F., Hargrove, L. J., Dou, Z., Rogers, D. R., & Englehart, K. B. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2161671 · Cited by 542 · cited 10×
“Huang 2011 [31] used EMG from the residual limb together with mechanical sensors on a passive prosthesis and found that the EMG added considerable information to that provided by the mechanical sensors.”
BackgroundMethodResultInfluentialSelf-citationA Strategy for Identifying Locomotion Modes Using Surface Electromyography
Huang, H., Kuiken, T. A., & Lipschutz, R. D. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2003293 · Cited by 512 · cited 4×
“Typically, features are calculated over the previous 100–500 ms data window [21, 28, 31].”
BackgroundMethodResultInfluentialSelf-citationPowered ankle-foot prosthesis to assist level-ground and stair-descent gaits
Au, S., Berniker, M., & Herr, H. (2008) · Neural Networks · doi:10.1016/j.neunet.2008.03.006 · Cited by 487 · cited 1×
“Direct proportional control using EMG information has been applied to lower limb prostheses in various ways including controlling the amount of plantarflexion torque at pushoff [14, 15], for powered ankle orthosis assistance [16–18], and for control of knee torque [19] and knee equilibrium set points [20].”
MethodMulticlass Real-Time Intent Recognition of a Powered Lower Limb Prosthesis
Varol, H. A., Sup, F., & Goldfarb, M. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2034734 · Cited by 425 · cited 7×
“Similarly, previous studies such as Varol 2010 [28] have used only a set of mechanical sensors for intent recognition purposes.”
BackgroundMethodResultInfluentialAn improved powered ankle–foot orthosis using proportional myoelectric control
Ferris, D. P., Gordon, K. E., Sawicki, G. S., & Peethambaran, A. (2006) · Gait & Posture · doi:10.1016/j.gaitpost.2005.05.004 · Cited by 342
Dynamic Bayesian Networks for Audio-Visual Speech Recognition
Nefian, A. V., Liang, L., Pi, X., Liu, X., & Murphy, K. (2002) · EURASIP Journal on Advances in Signal Processing · doi:10.1155/s1110865702206083 · Cited by 320
Preliminary Evaluations of a Self-Contained Anthropomorphic Transfemoral Prosthesis
Sup, F., Varol, H. A., Mitchell, J., Withrow, T. J., & Goldfarb, M. (2009) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2009.2032688 · Cited by 298 · cited 2×
“tronics at Vanderbilt University [33], was used for this”
Method
Cited by
A decade retrospective of medical robotics research from 2010 to 2020
Dupont, P. E., Nelson, B. J., Goldfarb, M., Hannaford, B., Menciassi, A., O’Malley, M. K., Simaan, N., Valdastri, P., & Yang, G.-Z. (2021) · Science Robotics · doi:10.1126/scirobotics.abi8017 · Cited by 516
The exoskeleton expansion: improving walking and running economy
Sawicki, G. S., Beck, O. N., Kang, I., & Young, A. J. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00663-9 · Cited by 424
Self-citationProprioception from a neurally controlled lower-extremity prosthesis
Clites, T. R., Carty, M. J., Ullauri, J. B., Carney, M. E., Mooney, L. M., Duval, J.-F., Srinivasan, S. S., & Herr, H. M. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aap8373 · Cited by 233
Fleming, A., Stafford, N., Huang, S., Hu, X., Ferris, D. P., & Huang, H. (. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac1176 · Cited by 216
Improved prosthetic hand control with concurrent use of myoelectric and inertial measurements
Krasoulis, A., Kyranou, I., Erden, M. S., Nazarpour, K., & Vijayakumar, S. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0284-4 · Cited by 214
“Finally, the combination of (EMG) and accelerometry has been also found to improve classification accuracy (CA) in lower-limb movement intent decoding [18].”
BackgroundReal-Time Neural Network-Based Gait Phase Estimation Using a Robotic Hip Exoskeleton
Kang, I., Kunapuli, P., & Young, A. J. (2020) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2019.2961749 · Cited by 168
“In order to further improve the model’s robustness, ML with sensor fusion techniques can be applied [27], [28].”
MethodSelf-citationRecognition of walking environments and gait period by surface electromyography
Kyeong, S., Shin, W., Yang, M., Heo, U., Feng, J.-R., & Kim, J. (2019) · Frontiers of Information Technology & Electronic Engineering · doi:10.1631/fitee.1800601 · Cited by 68
Xu, D., Feng, Y., Mai, J., & Wang, Q. (2018) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2018.2870152 · Cited by 67
“sEMG signals and mechanical signal [7], [26]–[28], mechanical signals and capacitive signals [8], etc.”
BackgroundDaily Locomotion Recognition and Prediction: A Kinematic Data-Based Machine Learning Approach
Figueiredo, J., Carvalho, S. P., Goncalve, D., Moreno, J. C., & Santos, C. P. (2020) · IEEE Access · doi:10.1109/access.2020.2971552 · Cited by 37
Whole body awareness for controlling a robotic transfemoral prosthesis
Parri, A., Martini, E., Geeroms, J., Flynn, L., Pasquini, G., Crea, S., Molino Lova, R., Lefeber, D., Kamnik, R., Munih, M., et al. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00025 · Cited by 29
“…successful approaches (still based on LDA) have proposed integrated neuromechanical sensory fusion (Huang et al., 2011; Tkach and Hargrove, 2013; Young et al., 2014) combining sEMG with mechanical sensors embedded in the prosthetic segments such as, pressure sensors, load cells, or inertial…”
MethodA strain gauge based locomotion mode recognition method using convolutional neural network
Feng, Y., Chen, W., & Wang, Q. (2019) · Advanced Robotics · doi:10.1080/01691864.2018.1563500 · Cited by 28
Metadata
- Authors
- Young, A. J., Kuiken, T. A., & Hargrove, L. J.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2014-09-22
- Volume
- 11
- Issue
- 5
- Pages
- 056021
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2093185395
- Cited by
- 200
- References
- 29
- FWCI
- 5.46
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.