Comparison of Glucose Concentration in Interstitial Fluid, and Capillary and Venous Blood During Rapid Changes in Blood Glucose Levels

Thennadil, Suresh N., Rennert, Jessica L., Wenzel, Brian J., Hazen, Kevin H., Ruchti, Timothy L., Block, Marshall B.·2001·Diabetes Technology & Therapeutics

Abstract

The relationship between glucose concentrations in interstitial fluid (ISF) and blood has generated great interest due to its importance in minimally invasive and noninvasive techniques for measuring blood glucose. The relationship between glucose levels in dermal ISF, and capillary and venous blood was studied with the dermal ISF samples obtained using the suction blister technique. The study was conducted with intensely managed diabetics whose blood glucose levels were manipulated so as to induce rapid changes in blood glucose levels. Glucose levels in the three compartments exhibited high correlations both when individual subjects were considered separately and when data from all subjects were combined. No significant time lag during glucose excursions was observed among the ISF, and capillary and venous glucose levels.

Integrity signals

Registry

Crossref · 10.1089/15209150152607132

Cites

20 of 20 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • ISF composition leads to reliable correlations between blood and ISF concentrations of many analytes, including electrolytes, metabolites, and proteins.

    “This function, and the associated ISF composition, lead to reliable correlations between the blood and ISF concentrations of many analytes, including electrolytes (e.g., sodium, phosphate, magnesium, potassium or calcium), metabolites (e.g., glucose, alcohol, lactate or cortisol) and proteins65-69.”

    From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [65]

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Thennadil, S. N., Rennert, J. L., Wenzel, B. J., Hazen, K. H., Ruchti, T. L., & Block, M. B.
Affiliations
  • Instrumentation Metrics, Inc., Chandler, Arizona; sthennadil@imetricsinc.com
  • Instrumentation Metrics, Inc., Chandler, Arizona
  • Radiant Research (United States) (US)
Venue
Diabetes Technology & Therapeutics
Published
2001-09
Volume
3
Issue
3
Pages
357-365
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2145770174
Cited by
166
References
19
Citations by year
2026: 7 · 2025: 9 · 2024: 7 · 2023: 12 · 2022: 11 · 2021: 13 · 2020: 11 · 2019: 10 · 2018: 11 · 2017: 7 · 2016: 8 · 2015: 4 · 2014: 4 · 2013: 6 · 2012: 5
FWCI
3.68
Type
journal article
Language
en
Open access
closed
Locations
  • Diabetes Technology & Therapeutics (publishedVersion)
  • PubMed (publishedVersion)
Funders
Israel Science Foundation
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
interstitial glucose, capillary blood glucose, venous blood glucose, glycemic excursions, non-invasive glucose monitoring, diabetes mellitus, time delay
MeSH terms
Adult, Blood Glucose, Capillaries, Extracellular Space, Female, Glucose, Humans, Male, Regional Blood Flow, Skin, Veins, Spectroscopy, Near-Infrared

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review