Wearable biosensors for healthcare monitoring
Open accessAbstract
Wearable biosensors are garnering substantial interest due to their potential to provide continuous, real-time physiological information via dynamic, noninvasive measurements of biochemical markers in biofluids, such as sweat, tears, saliva and interstitial fluid. Recent developments have focused on electrochemical and optical biosensors, together with advances in the noninvasive monitoring of biomarkers including metabolites, bacteria and hormones. A combination of multiplexed biosensing, microfluidic sampling and transport systems have been integrated, miniaturized and combined with flexible materials for improved wearability and ease of operation. Although wearable biosensors hold promise, a better understanding of the correlations between analyte concentrations in the blood and noninvasive biofluids is needed to improve reliability. An expanded set of on-body bioaffinity assays and more sensing strategies are needed to make more biomarkers accessible to monitoring. Large-cohort validation studies of wearable biosensor performance will be needed to underpin clinical acceptance. Accurate and reliable real-time sensing of physiological information using wearable biosensor technologies would have a broad impact on our daily lives.
AI snapshot
Written from the paper’s full text.
TL;DR
This review surveys recent advances in wearable biosensors for non-invasive healthcare monitoring, focusing on electrochemical and optical platforms that measure biomarkers in sweat, tears, saliva, and interstitial fluid. It concludes that despite rapid proof-of-concept progress, clinical and commercial translation requires better blood–biofluid correlation, large-population validation, and improved accuracy and stability under uncontrolled conditions.
Key findings
- Wearable biosensors have expanded from physical activity and vital-sign tracking to non-invasive chemical monitoring in sweat, tears, saliva, and ISF, with representative platforms including multiplexed sweat patches for glucose, lactate, Na⁺, K⁺, and temperature; contact-lens tear glucose sensors; and mouthguard sensors for salivary lactate, uric acid, and glucose.
- Reported quantitative examples include sweat glands at more than 100 glands/cm² of skin; the GlucoWatch Biographer as an FDA-approved non-invasive glucose monitor over 12 h with 6 measurements per hour, later withdrawn due to skin irritation, a 2–3 h warm-up, and calibration needs; salivary glucose–blood correlation of R = 0.64 in healthy subjects and R = 0.95 in diabetic subjects; iontophoretic tattoo sweat alcohol measurement within 10 min; artificial-saliva glucose range of 5–1000 μM; cortisol/IL-6 immunosensor antibody stability up to 96 h; RI glucose monitoring over 6 h; and iontophoretic sweat generation limited to 60 min.
- Major technical advances include multiplexed sensing, microfluidic sampling/transport, flexible and stretchable materials, wireless electronics, self-powered or energy-harvesting designs, and closed-loop “sense-act” systems such as sweat glucose monitoring coupled with microneedle metformin delivery in a mouse model.
- The global wearable sensors market was valued at roughly $150 million in 2016 and projected to reach $2.86 billion by 2025, with non-invasive glucose monitoring expected to make up a large portion of future wearable biosensor market share.
Methods
- Review article synthesizing key advances from the past 2 years, with emphasis on electrochemical and optical transduction, across epidermal (sweat and ISF), ocular (tears), and oral-cavity (saliva) wearable biosensing platforms.
- Discusses representative proof-of-concept and clinical/commercial examples, including multiplexed patches, contact lenses, mouthguards, iontophoresis/reverse iontophoresis, microfluidics, and integrated wireless electronics; no new experimental data or systematic meta-analysis are reported.
Limitations
- The paper states that a limited understanding of correlations between analyte concentrations in blood and non-invasive biofluids remains a major obstacle, and that extensive large-population validation through multidisciplinary collaboration is required for medical and commercial acceptance.
- The paper states that accuracy and stability under uncontrolled conditions, reproducible sample transport, biofouling, bioreceptor stability, calibration, and immunosensor regeneration remain unresolved challenges; many devices rely on exercise for sweat generation or have limited continuous operation, and saliva and tear sampling face low analyte concentrations, contamination, evaporation, and variability.
- Evident from the text: as a review, it does not provide primary experimental validation or a systematic quantitative synthesis, and commercial translation remains limited despite numerous proof-of-concept demonstrations.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 184 cited sources checked · 0 supported · 0 flagged
- Findingnone given
Reported epidermal biosensor devices have been limited to analysis of a single sampled biofluid.
“Despite the tremendous advances in epidermal biosensors, the reported devices have been limited to analysis of a single sampled biofluid.”
Quote in the paperSubstantial claim - Findingcitation
The system allows on-demand, controlled simultaneous sampling of two biofluids at physically separated locations on a single wearable tattoo platform.
“The system allows on-demand, controlled simultaneous sampling of two biofluids at physically separated locations on a single wearable tattoo platform.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [93] Kim, J. et al. Simultaneous monitoring of sweat and interstitial fluid using a single wearable biosensor platform. Adv. Sci. (Weinh.) 5, 1800880 (2018).
Not checked in the free run.
- Findingcitation
The authors' group demonstrated a new concept for simultaneous sampling and analysis of two different epidermal biofluids with a single wearable platform through combined iontophoresis.
“However, our group (J.K., A.S.C., J.W. et al.)93 recently demonstrated a new concept for the simultaneous sampling and analysis of two different epidermal biofluids with a single wearable platform through combined iontophoresis (Fig. 4f).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [93] Kim, J. et al. Simultaneous monitoring of sweat and interstitial fluid using a single wearable biosensor platform. Adv. Sci. (Weinh.) 5, 1800880 (2018).
Not checked in the free run.
- Findingcitation
The biosensing performance of the dual-biofluid platform has been demonstrated by measuring sweat alcohol and ISF glucose as model analytes with human subjects consuming food and alcoholic drink.
“The biosensing performance has been demonstrated by measuring sweat alcohol and ISF glucose as model analytes with human subjects consuming food and alcoholic drink.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [93] Kim, J. et al. Simultaneous monitoring of sweat and interstitial fluid using a single wearable biosensor platform. Adv. Sci. (Weinh.) 5, 1800880 (2018).
Not checked in the free run.
- Findingcitation
The dual-biofluid platform was accomplished by coupling sweat stimulation (iontophoretic drug delivery) with ISF extraction (via RI), enabling concurrent analysis of biomarkers in each biofluid.
“This was accomplished by coupling sweat stimulation (iontophoretic drug delivery) with ISF extraction (via RI), enabling concurrent analysis of biomarkers in each biofluid.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [93] Kim, J. et al. Simultaneous monitoring of sweat and interstitial fluid using a single wearable biosensor platform. Adv. Sci. (Weinh.) 5, 1800880 (2018).
Not checked in the free run.
- Findingcitation
The authors' group demonstrated a mouthguard-based uric acid biosensor that incorporated anatomically miniaturized instrumentation electronics featuring a potentiostat, microcontroller, and a Bluetooth low energy (BLE) transceiver for monitoring salivary uric acid levels toward clinical applications.
“We (J.K., J.W. et al.)36 have further continued development of oral-cavity sensors by demonstrating a mouthguard-based uric acid biosensor that incorporated anatomically miniaturized instrumentation electronics featuring a potentiostat, microcontroller, and a Bluetooth low energy (BLE) transceiver for monitoring salivary uric acid levels toward clinical applications (Fig. 6a).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Kim, J. et al. Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics. Biosens. Bioelectron. 74, 1061–1068 (2015). source
Not checked in the free run.
- Findingcitation
The authors' group was the first to develop mouthguard-based salivary metabolite electrochemical biosensors through integration of screen-printed enzymatic electrodes.
“Our group (J.K., J.W. et al.)145 was the first to develop mouthguard-based salivary metabolite electrochemical biosensors through integration of screen-printed enzymatic electrodes.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [145] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632–1636 (2014). source
Not checked in the free run.
- Findingnone given
The high concentrations of protein in saliva, including mucins and proteolytic enzymes, along with food debris, can lead to rapid biofouling of the oral cavity sensor through non-specific absorption at the transducer surface.
“The high concentrations of protein in saliva, including mucins and proteolytic enzymes, along with food debris, can lead to rapid biofouling of the oral cavity sensor through non-specific absorption at the transducer surface.”
Quote in the paperSubstantial claim - Findingreasoning
Accuracy is often compromised by surface-fouling effects, which represent a major challenge to the continuous operation of on-body biosensors.
“Accuracy is often compromised by surface-fouling effects, which represent a major challenge to the continuous operation of on-body biosensors.”
Quote in the paperSubstantial claim - Findingreasoning
Such rapid adsorption impedes the diffusion of the target analyte to the sensor surface, leading to a gradual decrease of the sensing signal over the time.
“Such rapid adsorption impedes the diffusion of the target analyte to the sensor surface, leading to a gradual decrease of the sensing signal over the time.”
Quote in the paperSubstantial claim - Premisenone given
Wearable biosensors can provide continuous, real-time physiological information in healthcare-related applications via non-invasive measurements of chemical markers in biofluids such as sweat, tears, saliva and interstitial fluid.
“Wearable biosensors are garnering substantial interest due to their potential to provide continuous, real-time physiological information in an array of healthcare-related applications via dynamic non-invasive measurements of chemical markers in biofluids, such as sweat, tears, saliva and interstitial fluid (ISF).”
Quote in the paperSubstantial claim - Interpretationreasoning
Commercialization of wearable bioanalyte sensors is substantially more challenging than that of activity-tracking counterparts or common lab-based biosensors because such devices must be capable of continuous on-body biochemical sensing and reliable measurement of a fragile biorecognition element.
“The commercialization of wearable bioanalyte sensors is substantially more challenging than that of activity-tracking counterparts or common lab-based biosensors because such devices must be capable not only of continuous on-body biochemical sensing but also of reliable measurement of a biorecognition element (or elements) that is highly specific yet fragile.”
Quote in the paperSubstantial claim - Premisecitation
The degree of analyte dilution during sweat excretion is affected by the relationship between sweat rate and analyte partitioning rate.
“Although variations in sweat rate can be monitored using multiplexed analysis (i.e., simultaneous monitoring of analytes with concentration profiles that are independent of sweat rate) or skin impedance measurements, the degree of analyte dilution during sweat excretion is affected by the relationship between sweat rate and analyte partitioning rate64.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [64] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015). source
Not checked in the free run.
- Premisecitation
Tattoo-based epidermal biosensors have been shown to allow real-time, non-invasive measurements of key sweat electrolytes, heavy metal, and metabolites.
“Tattoo-based epidermal biosensors have thus been shown to allow real-time, non-invasive measurements of key sweat electrolytes (pH, ammonium or sodium), heavy metal (Zn) and metabolites (lactate or ethanol) (J.K., J.W. et al.)58,70-74.”
Quote in the paperSubstantial claim0 of 6 sources checked- cites [58] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013). source
Not checked in the free run.
- cites [70] Bandodkar, A. J. et al. Tattoo-based potentiometric ion-selective sensors for epidermal pH monitoring. Analyst 138, 123–128 (2013). source
Not checked in the free run.
- cites [71] Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603–609 (2014). source
Not checked in the free run.
- cites [72] Guinovart, T., Bandodkar, A. J., Windmiller, J. R., Andrade, F. J. & Wang, J. A potentiometric tattoo sensor for monitoring ammonium in sweat. Analyst 138, 7031–7038 (2013). source
Not checked in the free run.
- cites [73] Kim, J. et al. Wearable temporary tattoo sensor for real-time trace metal monitoring in human sweat. Electrochem. Commun. 51, 41–45 (2015). source
Not checked in the free run.
- cites [74] Kim, J. et al. Noninvasive alcohol monitoring using a wearable tattoo-based iontophoretic-biosensing system. ACS Sens. 1, 1011–1019 (2016). source
Not checked in the free run.
- Interpretationcitation
Accurate measurement of physiologically relevant sweat glucose concentrations with epidermal biosensors faces major challenges related to uncontrolled operational conditions, glucose contamination, irregular sampling rates, and low sampling volumes.
“The accurate measurement of physiologically relevant sweat glucose concentrations with epidermal biosensors faces several major challenges related to uncontrolled operational conditions (e.g., varying temperature and pH), glucose contamination from various sources, irregular sampling rates, and low sampling volumes.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [64] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015). source
Not checked in the free run.
- cites [76] Moyer, J., Wilson, D., Finkelshtein, I., Wong, B. & Potts, R. Correlation between sweat glucose and blood glucose in subjects with diabetes. Diabetes Technol. Ther. 14, 398–402 (2012). source
Not checked in the free run.
- cites [77] Sakaguchi, K. et al. Evaluation of a minimally invasive system for measuring glucose area under the curve during oral glucose tolerance tests: usefulness of sweat monitoring for precise measurement. J. Diabetes Sci. Technol. 7, 678–688 (2013). source
Not checked in the free run.
- Premisecitation
ISF composition leads to reliable correlations between blood and ISF concentrations of many analytes, including electrolytes, metabolites, and proteins.
“This function, and the associated ISF composition, lead to reliable correlations between the blood and ISF concentrations of many analytes, including electrolytes (e.g., sodium, phosphate, magnesium, potassium or calcium), metabolites (e.g., glucose, alcohol, lactate or cortisol) and proteins65-69.”
Quote in the paperSubstantial claim0 of 5 sources checked- cites [65] Thennadil, S. N. et al. Comparison of glucose concentration in interstitial fluid, and capillary and venous blood during rapid changes in blood glucose levels. Diabetes Technol. Ther. 3, 357–365 (2001). source
Not checked in the free run.
- cites [66] Campbell, A. S., Kim, J. & Wang, J. Wearable electrochemical alcohol biosensors. Curr. Opin. Electrochem. 10, 126–135 (2018). source
Not checked in the free run.
- cites [67] Fogh-Andersen, N., Altura, B. M., Altura, B. T. & Siggaard-Andersen, O. Composition of interstitial fluid. Clin. Chem. 41, 1522–1525 (1995). source
Not checked in the free run.
- cites [68] Venugopal, M. et al. Clinical evaluation of a novel interstitial fluid sensor system for remote continuous alcohol monitoring. IEEE Sens. J. 8, 71–80 (2008). source
Not checked in the free run.
- cites [69] Venugopal, M., Arya, S. K., Chornokur, G. & Bhansali, S. A realtime and continuous assessment of cortisol in ISF using electrochemical impedance spectroscopy. Sens. Actuators A Phys. 172, 154–160 (2011). source
Not checked in the free run.
- Interpretationreasoning
Using reverse iontophoresis or sonophoresis to extract ISF analytes non-invasively, variations in extraction efficiency and skin surface contamination can impact accuracy.
“Using these methods, variations in the extraction efficiency and skin surface contamination can impact the accuracy, similar to sweat-based platforms.”
Quote in the paperSubstantial claim - Premisecitation
The Berkeley team demonstrated simultaneous multiplexed detection of sweat metabolites (glucose and lactate) and electrolytes (sodium and potassium ions), along with skin temperature, by integrating a multi-sensing array.
“In this work, the Berkeley team demonstrated the simultaneous multiplexed detection of sweat metabolites (glucose and lactate) and electrolytes (sodium and potassium ions), along with skin temperature by integrating a multi-sensing array.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [35] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Premisecitation
The caffeine sensing platform relied on direct anodic detection of caffeine at a carbon nanotube-based working electrode with a voltammetric scan.
“Rather than using biological recognition, this sensing platform relied on direct anodic detection of caffeine at a carbon nanotube-based working electrode with a voltammetric scan.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] Tai, L. C. et al. Methylxanthine drug monitoring with wearable sweat sensors. Adv. Mater. 30, e1707442 (2018). source
Not checked in the free run.
- Interpretationreasoning
Due to potential variation in sweat rates, active calibration is required to ensure reliable measurement.
“Yet, due to potential variation in the sweat rates active calibration is required to ensure reliable measurement.”
Quote in the paperSubstantial claim - Premisecitation
A wearable sweat-based sensor has been developed for detecting caffeine using pilocarpine-based iontophoretic sweat stimulation or exercise-induced sweat.
“For example, a wearable sweat-based sensor has been developed for detecting caffeine (a methylxanthine drug) using pilocarpine-based iontophoretic sweat stimulation or exercise-induced sweat92.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] Tai, L. C. et al. Methylxanthine drug monitoring with wearable sweat sensors. Adv. Mater. 30, e1707442 (2018). source
Not checked in the free run.
- Premisecitation
A wearable immunosensor for detecting cortisol and interleukin 6 in sweat has diagnostic potential.
“For instance, a wearable immunosensor for detecting cortisol and interleukin 6 (IL-6) in sweat has diagnostic potential (Fig. 3g)43.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [43] Munje, R. D., Muthukumar, S., Jagannath, B. & Prasad, S. A new paradigm in sweat based wearable diagnostics biosensors using room temperature ionic liquids (RTILs). Sci. Rep. 7, 1950 (2017). source
Not checked in the free run.
- Interpretationcitation
Sweat alcohol represents a useful indicator for blood alcohol levels without time lag and errors common to transdermal devices and breathalyzers.
“Sweat alcohol represents a useful indicator for blood alcohol levels without time lag and errors common to transdermal devices and breathalyzers.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [75] Wang, L. et al. Weaving sensing fibers into electrochemical fabric for real‐time health monitoring. Adv. Funct. Mater. 28, 1804456 (2018). source
Not checked in the free run.
- Premisecitation
Tear analysis presents opportunities for the diagnosis of ocular disease.
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
Quote in the paperSubstantial claim0 of 5 sources checked- cites [94] Farandos, N. M., Yetisen, A. K., Monteiro, M. J., Lowe, C. R. & Yun, S. H. Contact lens sensors in ocular diagnostics. Adv. Healthc. Mater. 4, 792–810 (2015). source
Not checked in the free run.
- cites [95] Baca, J. T., Finegold, D. N. & Asher, S. A. Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus. Ocul. Surf. 5, 280–293 (2007). source
Not checked in the free run.
- cites [96] Thaysen, J. H. & Thorn, N. A. Excretion of urea, sodium, potassium and chloride in human tears. Am. J. Physiol. 178, 160–164 (1954). source
Not checked in the free run.
- cites [97] Mitsubayashi, K. & Arakawa, T. Cavitas sensors: contact lens type sensors & mouthguard sensors. Electroanalysis 28, 1170–1187 (2016). source
Not checked in the free run.
- cites [98] Pankratov, D., Gonzalez-Arribas, E., Blum, Z. & Shleev, S. Tear based bioelectronics. Electroanalysis 28, 1250–1266 (2016). source
Not checked in the free run.
- Premisenone given
Reflex tears, generated during emotional or mechanical stimulation, have different compositions than basal tears, which make up the protective tear film covering the eye surface at all times.
“Reflex tears, generated during emotional or mechanical stimulation, have different compositions than basal tears, which make up the protective tear film covering the eye surface at all times.”
Quote in the paperNot a substantial claim - Premisecitation
The accuracy of in vitro tears diagnostic assays depends strongly on the collection method, with the most common strategies being via glass capillary tube or the Schirmer's strip.
“The accuracy of such in vitro tears diagnostic assays thus depends strongly on the collection method, with the most common strategies being via glass capillary tube or the Schirmer’s strip106.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [106] Green-Church, K. B., Nichols, K. K., Kleinholz, N. M., Zhang, L. & Nichols, J. J. Investigation of the human tear film proteome using multiple proteomic approaches. Mol. Vis. 14, 456–470 (2008).
Not checked in the free run.
- Premisecitation
The possibility of quantifying glucose in human tear fluid at physiological conditions using holographic contact lenses was presented around the same time based on advantages such as no need of battery, easy reading, and continuous glucose signal monitoring.
“The possibility of quantifying glucose in human tear fluid at physiological conditions using holographic contact lenses was also presented around the same time based on its advantages such as a no need of battery, easy reading, and continuous glucose signal monitoring109.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [109] Domschke, A., March, W. F., Kabilan, S. & Lowe, C. Initial clinical testing of a holographic non-invasive contact lens glucose sensor. Diabetes Technol. Ther. 8, 89–93 (2006). source
Not checked in the free run.
- Premisecitation
The Parviz team resolved interference issues by introducing the 'dual sensor setup' that implements additional control (GOx-free) working and counter electrodes.
“Also, they were able to resolve interference issues by introducing the ‘dual sensor setup’ that implements additional control (GOx-free) working and counter electrodes114,115.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [114] Yao, H., Afanasiev, A., Lahdesmaki, I. & Parviz, B. A. A dual microscale glucose sensor on a contact lens, tested in conditions mimicking the eye. Proc. IEEE Micro. Electro. Mech. Syst. 2011, 25–28 (2011).
Not checked in the free run.
- cites [115] Yao, H. et al. A contact lens with integrated telecommunication circuit and sensors for wireless and continuous tear glucose monitoring. J. Micromech. Microeng. 22, 075007 (2012). source
Not checked in the free run.
- Premisecitation
Another recent advance in wearable contact lens biosensors involved the use of smartphones for optical continuous glucose monitoring.
“Another recent advance in wearable contact lens biosensors involved the use of smartphones for optical continuous glucose monitoring112.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
- Premisenone given
A small spring-like electrochemical sensor, consisting of multiple coiled wire electrodes and coated with a protective polysaccharide-based hydrogel material, was designed by Noviosense for placement in the inferior conjunctival fornix toward constant access to tear fluid.
“In addition to contact lens platforms, a small spring-like electrochemical sensor, consisting of multiple coiled wire electrodes, and coated with a protective polysaccharide-based hydrogel material, was designed by Noviosense (http://noviosense.com) for placement in the inferior conjunctival fornix toward constant access to tear fluid (Fig. 5e).”
Quote in the paperNot a substantial claim - Premisecitation
The work was further expanded for integration of wireless power transfer circuits and displays on contact lens biosensors, visualizing the in vivo rabbit tear glucose response in real-time.
“The work was further expanded for integration of wireless power transfer circuits and displays on contact lens biosensors, visualizing the in vivo rabbit tear glucose response in real-time (Fig. 5c)118.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [118] Park, J. et al. Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays. Sci. Adv. 4, eaap9841 (2018). source
Not checked in the free run.
- Premisecitation
A recent clinical trial demonstrated great correlation between tear and blood glucose concentrations in animals and humans (including patients with type 1 diabetes).
“A recent clinical trial demonstrated great correlation between tear and blood glucose concentrations in animals and humans (including patients with type 1 diabetes)119.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [119] Kownacka, A. E. et al. Clinical evidence for use of a noninvasive biosensor for tear glucose as an alternative to painful finger-prick for diabetes management utilizing biopolymer coating. Biomacromolecules 19, 4504–4511 (2018). source
Not checked in the free run.
- Premisecitation
The developed prototype lens consisted of a wireless control chip, miniature electrochemical transducer, and antenna on a soft contact lens platform with the glucose sensor embedded within a hydrogel matrix to non-invasively measure glucose concentrations in the surrounding tears.
“The developed prototype lens consisted of a wireless control chip, miniature electrochemical transducer, and antenna on a soft contact lens platform with the glucose sensor embedded within a hydrogel matrix to non-invasively measure glucose concentrations in the surrounding tears (Fig. 5a).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [33] Senior, M. Novartis signs up for Google smart lens. Nat. Biotechnol. 32, 856 (2014). source
Not checked in the free run.
- Premisecitation
Smart contact lenses for wireless ocular diagnostics have been developed by combining glucose and ocular pressure contact lens sensors for wireless in vivo glucose monitoring in the eye of a rabbit, together with in vitro monitoring of ocular pressure using a bovine eyeball.
“Recently, ‘smart’ contact lenses for wireless ocular diagnostics have been further developed by combining glucose and ocular pressure contact lens sensors for wireless in vivo glucose monitoring in the eye of a rabbit, together with in vitro monitoring of ocular pressure using a bovine eyeball (Fig. 5b)117.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [117] Kim, J. et al. Wearable smart sensor systems integrated on soft contact lenses for wireless ocular diagnostics. Nat. Commun. 8, 14997 (2017). source
Not checked in the free run.
- Interpretationcitation
There have been delays to the clinical trials and subsequent commercial release of the Google/Novartis contact lens product, indicative of technological challenges of successfully achieving a high-performing contact-lens-based sensing platform.
“However, there have been delays to the clinical trials and subsequent commercial release of this product, indicative of the technological challenges of successfully achieving a high-performing contact-lens-based sensing platform.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [33] Senior, M. Novartis signs up for Google smart lens. Nat. Biotechnol. 32, 856 (2014). source
Not checked in the free run.
- Premisecitation
Google, in partnership with Novartis, developed a contact lens sensing platform for tears glucose monitoring.
“As a result of these consistent endeavors, Google, in partnership with Novartis, made the most noticeable progress by applying their expertise in electronics miniaturization and applied medical technologies, respectively, toward development of a contact lens sensing platform for tears glucose monitoring33.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [33] Senior, M. Novartis signs up for Google smart lens. Nat. Biotechnol. 32, 856 (2014). source
Not checked in the free run.
- Premisecitation
A hydrogel-based sensor with photonic microstructure was attached on top of a commercial contact lens and the reflective power was recorded using a smartphone in response to changes in the tears glucose levels.
“A hydrogel-based sensor with photonic microstructure was attached on top of a commercial contact lens and the reflective power was recorded using a smartphone in response to changes in the tears glucose levels (Fig. 5d).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
- Premisecitation
Because saliva can be readily collected, it has been used in connection to in vitro diagnostic biosensors on strips or portable device platforms.
“Because saliva can be readily collected123-125, it has been used in connection to in vitro diagnostic biosensors on strips or portable device platforms,126-134.”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [123] Javaid, M. A., Ahmed, A. S., Durand, R. & Tran, S. D. Saliva as a diagnostic tool for oral and systemic diseases. J. Oral Biol. Craniofac. Res. 6, 66–75 (2016). source
Not checked in the free run.
- cites [124] Goswami, Y., Mishra, R., Agrawal, A. P. & Agrawal, L. A. Salivary biomarkers: a review of powerful diagnostic tool. IOSR J. Dent. Med. Sci. 14, 80–87 (2015).
Not checked in the free run.
- cites [125] Campuzano, S., Yanez-Sedeno, P. & Pingarron, J. M. Electrochemical bioaffinity sensors for salivary biomarkers detection. Trends Analyt. Chem. 86, 14–24 (2017). source
Not checked in the free run.
- cites [126] Soni, A. & Jha, S. K. A paper strip based non-invasive glucose biosensor for salivary analysis. Biosens. Bioelectron. 67, 763–768 (2015). source
Not checked in the free run.
- cites [127] Stevens, R. C., Soelberg, S. D., Near, S. & Furlong C.E. Detection of cortisol in saliva with a flow-filtered, portable surface plasmon resonance biosensor system. Anal. Chem. 80, 6747–6751 (2008). source
Not checked in the free run.
- cites [128] Yamaguchi, M. et al. Hand-held monitor of sympathetic nervous system using salivary amylase activity and its validation by driver fatigue assessment. Biosens. Bioelectron. 21, 1007–1014 (2006). source
Not checked in the free run.
- cites [129] Ballesta Claver, J., Valencia Mirón, M. C. & Capitán-Vallvey, L. F. Disposable electrochemiluminescent biosensor for lactate determination in saliva. Analyst 134, 1423–1432 (2009). source
Not checked in the free run.
- cites [130] Zangheri, M. et al. A simple and compact smartphone accessory for quantitative chemiluminescence-based lateral flow immunoassay for salivary cortisol detection. Biosens. Bioelectron. 64, 63–68 (2015). source
Not checked in the free run.
- cites [131] Mahosenaho, M. et al. A disposable biosensor for the determination of alpha-amylase in human saliva. Mikrochim. Acta 170, 243–249 (2010). source
Not checked in the free run.
- cites [132] Soni, A. & Jha, S. K. Smartphone based non-invasive salivary glucose biosensor. Anal. Chim. Acta 996, 54–63 (2017). source
Not checked in the free run.
- Premisecitation
Salivary lactate has shown good correlation with blood lactate levels, toward assessing physical stress and performance.
“Salivary lactate has shown good correlation with blood lactate levels, toward assessing physical stress and performance146,147.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [146] Segura, R. et al. A new approach to the assessment of anaerobic metabolism: measurement of lactate in saliva. Br. J. Sports Med. 30, 305–309 (1996). source
Not checked in the free run.
- cites [147] Santos, R. V. T., Almeida, A. L. R., Caperuto, E. C., Martins, E. Jr. & Costa Rosa, L. F. B. P. Effects of a 30-km race upon salivary lactate correlation with blood lactate. Comp. Biochem. Physiol. B Biochem. Mol. Biol. 145, 114–117 (2006). source
Not checked in the free run.
- Premisecitation
Oral biosensing devices have been further miniaturized to a detachable 'cavitas sensors' device to measure salivary glucose levels on a mouthguard platform fabricated to fit over the wearer's teeth.
“Oral biosensing devices have been further miniaturized to a detachable ‘cavitas sensors’ device to measure salivary glucose levels on a mouthguard platform fabricated to fit over the wearer’s teeth (Fig. 6b)148.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [148] Arakawa, T. et al. Mouthguard biosensor with telemetry system for monitoring of saliva glucose: a novel cavitas sensor. Biosens. Bioelectron. 84, 106–111 (2016). source
Not checked in the free run.
- Premisecitation
Mannoor et al. reported graphene-based nanosensors printed onto water-soluble silk and transferred directly onto tooth enamel for passive, wireless bacteria detection.
“The field of oral biosensing was expanded by Mannoor et al.32, who reported graphene-based nanosensors printed onto water-soluble silk and transferred directly onto tooth enamel for passive, wireless bacteria detection.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Mannoor, M. S. et al. Graphene-based wireless bacteria detection on tooth enamel. Nat. Commun. 3, 763 (2012). source
Not checked in the free run.
- Interpretationreasoning
Salivary glucose can offer an alternative painless screening route for diabetic patients.
“Salivary glucose can thus offer an alternative painless screening route for diabetic patients149-154.”
Quote in the paperSubstantial claim0 of 6 sources checked- cites [149] Yamaguchi, M., Mitsumori, M. & Kano, Y. Noninvasively measuring blood glucose using saliva. IEEE Eng. Med. Biol. Mag. 17, 59–63 (1998). source
Not checked in the free run.
- cites [150] Agrawal, R. et al. Noninvasive method for glucose level estimation by saliva. J. Diabetes Metab. 4, 2–5 (2013).
Not checked in the free run.
- cites [151] Zhang, W., Du, Y. & Wang, M. L. Noninvasive glucose monitoring using saliva nano-biosensor. Sens. Biosensing Res. 4, 23–29 (2015). source
Not checked in the free run.
- cites [152] Jurysta, C. et al. Salivary glucose concentration and excretion in normal and diabetic subjects. J. Biomed. Biotechnol. 2009, 430426 (2009). source
Not checked in the free run.
- cites [153] Campbell, M. J. A. Glucose in the saliva of the non-diabetic and the diabetic patient. Arch. Oral Biol. 10, 197–205 (1965). source
Not checked in the free run.
- cites [154] Abikshyeet, P., Ramesh, V. & Oza, N. Glucose estimation in the salivary secretion of diabetes mellitus patients. Diabetes Metab. Syndr. Obes. 5, 149–154 (2012).
Not checked in the free run.
- Premisecitation
Another wearable sensor based on an oral-cavity platform recently has been demonstrated using in-mouth operation with a human subject.
“Another wearable sensor based on an oral-cavity platform recently has been demonstrated using in-mouth operation with a human subject (Fig. 6c)155.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [155] Tseng, P., Napier, B., Garbarini, L., Kaplan, D. L. & Omenetto, F. G. Functional, RF-trilayer sensors for tooth-mounted, wireless monitoring of the oral cavity and food consumption. Adv. Mater. 30, e1703257 (2018). source
Not checked in the free run.
- Interpretationreasoning
The direct relationship between blood glucose and salivary glucose provides a highly advantageous, easily accessible route for glucose sampling.
“The direct relationship between blood glucose and salivary glucose provides a highly advantageous, easily accessible route for glucose sampling.”
Quote in the paperSubstantial claim - Premisecitation
In vivo use was demonstrated through the measurement of different fluid properties, such as alcohol content, salinity, sugars, pH and temperature using RF sensors.
“In vivo use was demonstrated through the measurement of different fluid properties, such as alcohol content, salinity, sugars, pH and temperature using RF sensors.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [155] Tseng, P., Napier, B., Garbarini, L., Kaplan, D. L. & Omenetto, F. G. Functional, RF-trilayer sensors for tooth-mounted, wireless monitoring of the oral cavity and food consumption. Adv. Mater. 30, e1703257 (2018). source
Not checked in the free run.
- Premisecitation
Salivary glucose has been highlighted as having a positive correlation with blood glucose levels.
“As demonstrated in the above studies, salivary glucose has been highlighted as having a positive correlation with blood glucose levels.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [144] Viswanath, B., Choi, C. S., Lee, K. & Kim, S. Recent trends in the development of diagnostic tools for diabetes mellitus using patient saliva. Trends Analyt. Chem. 89, 60–67 (2017). source
Not checked in the free run.
- Premisecitation
Such oral sensing was realized by introducing biocompatible materials, such as porous silk and hydrogels, on a tooth-mounted oral cavity sensor capable of wireless monitoring of foods during ingestion.
“Such oral sensing was realized by introducing biocompatible materials, such as porous silk and hydrogels, on a tooth-mounted oral cavity sensor capable of wireless monitoring of foods during ingestion.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [155] Tseng, P., Napier, B., Garbarini, L., Kaplan, D. L. & Omenetto, F. G. Functional, RF-trilayer sensors for tooth-mounted, wireless monitoring of the oral cavity and food consumption. Adv. Mater. 30, e1703257 (2018). source
Not checked in the free run.
- Premisecitation
The correlation between salivary and blood glucose reflects the diffusion and active transport of blood components to the salivary gland.
“This correlation reflects the diffusion and active transport of blood components to the salivary gland144.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [144] Viswanath, B., Choi, C. S., Lee, K. & Kim, S. Recent trends in the development of diagnostic tools for diabetes mellitus using patient saliva. Trends Analyt. Chem. 89, 60–67 (2017). source
Not checked in the free run.
- Premisecitation
This oral sensing platform relies on a user-comfortable system using ultrathin stretchable electronics along with miniaturized sensors.
“This oral sensing platform relies on a user-comfortable system using ultrathin stretchable electronics along with miniaturized sensors.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [156] Lee, Y. et al. Wireless, intraoral hybrid electronics for real-time quantification of sodium intake toward hypertension management. Proc. Natl Acad. Sci. USA 115, 5377–5382 (2018). source
Not checked in the free run.
- Premisecitation
An alternative in vivo oral monitoring device has been developed for sodium intake via long-range wireless telemetry.
“An alternative in vivo oral monitoring device has also been developed for sodium intake via long-range wireless telemetry (Fig. 6d)156.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [156] Lee, Y. et al. Wireless, intraoral hybrid electronics for real-time quantification of sodium intake toward hypertension management. Proc. Natl Acad. Sci. USA 115, 5377–5382 (2018). source
Not checked in the free run.
- Premisecitation
The performance of the sodium monitoring device has been demonstrated with human subjects, proving its feasibility for real-time monitoring of sodium consumption, which is desired for managing hypertension.
“The performance of the device has been demonstrated with human subjects, proving its feasibility for real-time monitoring of sodium consumption, which is desired for managing hypertension.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [156] Lee, Y. et al. Wireless, intraoral hybrid electronics for real-time quantification of sodium intake toward hypertension management. Proc. Natl Acad. Sci. USA 115, 5377–5382 (2018). source
Not checked in the free run.
- Premisecitation
Wearable supercapacitor materials have shown fast charge and discharge capacity.
“These materials have shown fast charge and discharge capacity, but also possess low gravimetric and volumetric energy densities180-182.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [180] Qiu, Y. C. et al. Vertically aligned carbon nanotubes on carbon nanofibers: a hierarchical three-dimensional carbon nanostructure for high-energy flexible supercapacitors. Chem. Mater. 27, 1194–1200 (2015). source
Not checked in the free run.
- cites [181] Li, Q., Lu, X. F., Xu, H., Tong, Y. X. & Li, G. R. Carbon/MnO2 double-walled nanotube arrays with fast ion and electron transmission for high-performance supercapacitors. ACS Appl. Mater. Interfaces 6, 2726–2733 (2014). source
Not checked in the free run.
- cites [182] Wang, Q., Yan, J. & Fan, Z. J. Carbon materials for high volumetric performance supercapacitors: design, progress, challenges and opportunities. Energy Environ. Sci. 9, 729–762 (2016). source
Not checked in the free run.
- Premisecitation
Wearable power sources have been reported that harvest energy during operation.
“Wearable power sources have also been reported that harvest energy during operation, depending on the type of wearable platform being powered.”
Quote in the paperNot a substantial claim0 of 6 sources checked- cites [183] O’Connor, T. F. et al. Wearable organic solar cells with high cyclic bending stability: materials selection criteria. Sol. Energy Mater. Sol. Cells 144, 438–444 (2016). source
Not checked in the free run.
- cites [184] Kim, B. J. et al. Highly efficient and bending durable perovskite solar cells: toward a wearable power source. Energy Environ. Sci. 8, 916–921 (2015). source
Not checked in the free run.
- cites [185] Yang, J.H. et al. Effect of garment design on piezoelectricity harvesting from joint movement. Smart Mater. Struct. 25, 1–15 (2016). source
Not checked in the free run.
- cites [186] Pu, X. et al. Wearable self-charging power textile based on flexible yarn supercapacitors and fabric nanogenerators. Adv. Mater. 28, 98–105 (2016). source
Not checked in the free run.
- cites [187] Oh, J. Y. et al. Chemically exfoliated transition metal dichalcogenide nanosheet-based wearable thermoelectric generators. Energy Environ. Sci. 9, 1696–1705 (2016). source
Not checked in the free run.
- cites [188] Lu, Z. S., Zhang, H. H., Mao, C. P. & Li, C. M. Silk fabric-based wearable thermoelectric generator for energy harvesting from the human body. Appl. Energy 164, 57–63 (2016). source
Not checked in the free run.
- Premisecitation
Wearable supercapacitor materials possess low gravimetric and volumetric energy densities.
“These materials have shown fast charge and discharge capacity, but also possess low gravimetric and volumetric energy densities180-182.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [180] Qiu, Y. C. et al. Vertically aligned carbon nanotubes on carbon nanofibers: a hierarchical three-dimensional carbon nanostructure for high-energy flexible supercapacitors. Chem. Mater. 27, 1194–1200 (2015). source
Not checked in the free run.
- cites [181] Li, Q., Lu, X. F., Xu, H., Tong, Y. X. & Li, G. R. Carbon/MnO2 double-walled nanotube arrays with fast ion and electron transmission for high-performance supercapacitors. ACS Appl. Mater. Interfaces 6, 2726–2733 (2014). source
Not checked in the free run.
- cites [182] Wang, Q., Yan, J. & Fan, Z. J. Carbon materials for high volumetric performance supercapacitors: design, progress, challenges and opportunities. Energy Environ. Sci. 9, 729–762 (2016). source
Not checked in the free run.
- Interpretationreasoning
Substantial biofouling is expected for saliva-based oral-cavity biosensors as the complex saliva matrix contains a much higher protein content compared to other non-invasive biofluids (e.g., sweat or tears).
“In contrast, substantial biofouling is expected for saliva-based oral-cavity biosensors as the complex saliva matrix contains a much higher protein content compared to other non-invasive biofluids (e.g., sweat or tears).”
Quote in the paperSubstantial claim - Premisecitation
Energy can be harvested from chemical constituents of the sampled biofluids with wearable biofuel cells.
“chemical constituents of the sampled biofluids with wearable biofuel cells189-192”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [189] Jia, W., Valdés-Ramírez, G., Bandodkar, A. J., Windmiller, J. R. & Wang, J. Epidermal biofuel cells: energy harvesting from human perspiration. Angew. Chem. Int. Ed. Engl. 52, 7233–7236 (2013). source
Not checked in the free run.
- cites [190] Jia, W. Z. et al. Wearable textile biofuel cells for powering electronics. J. Mater. Chem. A Mater. 2, 18184–18189 (2014). source
Not checked in the free run.
- cites [191] Bandodkar, A. J. & Wang, J. Wearable biofuel cells: a review. Electroanalysis 28, 1188–1200 (2016). source
Not checked in the free run.
- cites [192] Ogawa, Y. et al. Stretchable biofuel cell with enzyme-modified conductive textiles. Biosens. Bioelectron. 74, 947–952 (2015). source
Not checked in the free run.
- Premisecitation
Wearable biofuel cells harvest energy from the same biofluids of interest.
“Wearable biofuel cells, in particular, are promising for the powering of non-invasive wearable platforms as they harvest energy from the same biofluids of interest and can operate as self-powered biosensors, but yet still suffer from the same stability issues discussed above42,193,194.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J. M. & Wang, J. Stretchable biofuel cells as wearable textile-based self-powered sensors. J. Mater. Chem. A Mater. Energy Sustain. 4, 18342–18353 (2016). source
Not checked in the free run.
- cites [193] Falk, M., Andoralov, V., Silow, M., Toscano, M. D. & Shleev, S. Miniature biofuel cell as a potential power source for glucose-sensing contact lenses. Anal. Chem. 85, 6342–6348 (2013). source
Not checked in the free run.
- cites [194] Grattieri, M. & Minteer, S. D. Self-powered biosensors. ACS Sens. 3, 44–53 (2018). source
Not checked in the free run.
- Premisecitation
Researchers have focused on combination approaches that involve the development of both the power sources and the design of more energy-efficient devices and adaptive algorithms to decrease energy demands.
“Researchers have focused on a combination approaches that involve the development of both the power sources and the design of more energy-efficient devices and adaptive algorithms to decrease energy demands.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [1] Bandodkar, A. J., Jeerapan, I. & Wang, J. Wearable chemical sensors: present challenges and future prospects. ACS Sens. 1, 464–482 (2016). source
Not checked in the free run.
- Premisecitation
Wearable biofuel cells can operate as self-powered biosensors.
“Wearable biofuel cells, in particular, are promising for the powering of non-invasive wearable platforms as they harvest energy from the same biofluids of interest and can operate as self-powered biosensors, but yet still suffer from the same stability issues discussed above42,193,194.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J. M. & Wang, J. Stretchable biofuel cells as wearable textile-based self-powered sensors. J. Mater. Chem. A Mater. Energy Sustain. 4, 18342–18353 (2016). source
Not checked in the free run.
- cites [193] Falk, M., Andoralov, V., Silow, M., Toscano, M. D. & Shleev, S. Miniature biofuel cell as a potential power source for glucose-sensing contact lenses. Anal. Chem. 85, 6342–6348 (2013). source
Not checked in the free run.
- cites [194] Grattieri, M. & Minteer, S. D. Self-powered biosensors. ACS Sens. 3, 44–53 (2018). source
Not checked in the free run.
- Interpretationreasoning
Wearable biofuel cells are promising for powering non-invasive wearable platforms.
“Wearable biofuel cells, in particular, are promising for the powering of non-invasive wearable platforms as they harvest energy from the same biofluids of interest and can operate as self-powered biosensors, but yet still suffer from the same stability issues discussed above42,193,194.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J. M. & Wang, J. Stretchable biofuel cells as wearable textile-based self-powered sensors. J. Mater. Chem. A Mater. Energy Sustain. 4, 18342–18353 (2016). source
Not checked in the free run.
- cites [193] Falk, M., Andoralov, V., Silow, M., Toscano, M. D. & Shleev, S. Miniature biofuel cell as a potential power source for glucose-sensing contact lenses. Anal. Chem. 85, 6342–6348 (2013). source
Not checked in the free run.
- cites [194] Grattieri, M. & Minteer, S. D. Self-powered biosensors. ACS Sens. 3, 44–53 (2018). source
Not checked in the free run.
- Premisecitation
Wearable biofuel cells still suffer from stability issues.
“Wearable biofuel cells, in particular, are promising for the powering of non-invasive wearable platforms as they harvest energy from the same biofluids of interest and can operate as self-powered biosensors, but yet still suffer from the same stability issues discussed above42,193,194.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [42] Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J. M. & Wang, J. Stretchable biofuel cells as wearable textile-based self-powered sensors. J. Mater. Chem. A Mater. Energy Sustain. 4, 18342–18353 (2016). source
Not checked in the free run.
- cites [193] Falk, M., Andoralov, V., Silow, M., Toscano, M. D. & Shleev, S. Miniature biofuel cell as a potential power source for glucose-sensing contact lenses. Anal. Chem. 85, 6342–6348 (2013). source
Not checked in the free run.
- cites [194] Grattieri, M. & Minteer, S. D. Self-powered biosensors. ACS Sens. 3, 44–53 (2018). source
Not checked in the free run.
- Interpretationreasoning
The successful translation of wearable biosensors and proof-of-concept demonstrations to the commercial market faces several hurdles related to their fundamental operation.
“The successful translation of wearable biosensors and proof-of-concept demonstrations to the commercial market faces several hurdles related to their fundamental operation.”
Quote in the paperSubstantial claim
- May combine two claims. 47
- Classified as probably not a substantial claim. 48
May combine two claims.
FindingcitationAlthough the smart contact lens device is capable of multiplexed sensing, the simultaneous demonstration of two functionalities has not yet been evaluated.
“Although this device is capable of multiplexed sensing, the simultaneous demonstration of two functionalities has not yet been evaluated.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [117] Kim, J. et al. Wearable smart sensor systems integrated on soft contact lenses for wireless ocular diagnostics. Nat. Commun. 8, 14997 (2017). source
Not checked in the free run.
May combine two claims.
FindingcitationThe mouthguard lactate device is capable of detecting salivary lactate electrochemically in connection to selectivity imparted by the lactate oxidase enzyme; protection against biofouling in undiluted human salivary samples was conferred by electro-polymerized ortho-phenylenediamine (OPD) for continuous, non-invasive physiological monitoring of an individual's fitness state.
“This device is capable of detecting salivary lactate electrochemically in connection to selectivity imparted by the lactate oxidase enzyme; protection against biofouling in undiluted human salivary samples was conferred by electro-polymerized ortho-phenylenediamine (OPD) for continuous, non-invasive physiological monitoring of an individual’s fitness state.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [145] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632–1636 (2014). source
Not checked in the free run.
May combine two claims.
FindingcitationThe mouthguard uric acid platform enables non-invasive monitoring of blood uric acid levels, which is a biomarker for various diseases (e.g., hyperuricemia, gout and renal syndrome), and it displays highly sensitive, selective, stable and rapid response toward obtaining dynamic chemical data on salivary biomarkers in the oral cavity.
“This platform enables non-invasive monitoring of blood uric acid levels, which is a biomarker for various diseases (e.g., hyperuricemia, gout and renal syndrome), and it displays highly sensitive, selective, stable and rapid response toward obtaining dynamic chemical data on salivary biomarkers in the oral cavity.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Kim, J. et al. Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics. Biosens. Bioelectron. 74, 1061–1068 (2015). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingreasoningFew studies have focused on developing wearable oral cavity biosensors, likely due to potential biofouling effects caused by the rich salivary protein content and the very low concentration of some target biomarkers.
“However, few studies have focused on developing wearable oral cavity biosensors, likely due to potential biofouling effects caused by the rich salivary protein content and the very low concentration of some target biomarkers.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationThe toxicity of the sodium monitoring device was evaluated without the chemical sensing layer, and hence practical oral-cavity applications would require further critical assessment along with a biocompatible recognition layer.
“However, the toxicity of the device was evaluated without the chemical sensing layer, and hence practical oral-cavity applications would require further critical assessment along with a biocompatible recognition layer.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [156] Lee, Y. et al. Wireless, intraoral hybrid electronics for real-time quantification of sodium intake toward hypertension management. Proc. Natl Acad. Sci. USA 115, 5377–5382 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenCompared with other non-invasively sampled biofluids, saliva can be readily sampled without complicated procedures, but is composed of a rich matrix of constituents that can also be easily contaminated by external factors (e.g., food and drink).
“Compared with other non-invasively sampled biofluids, saliva can be readily sampled without complicated procedures, but is composed of a rich matrix of constituents that can also be easily contaminated by external factors (e.g., food and drink).”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationWearable biosensors face many fundamental challenges and technological gaps related to the scope, validation, stability and accuracy, along with power, communication and security/privacy issues.
“Wearable biosensors face many fundamental challenges and technological gaps related to the scope, validation, stability and accuracy, along with power (Box 1)1,42,157-167, communication (Box 2)1,168 and security/privacy issues (Box 3)1,169-172”
Quote in the paperSubstantial claim0 of 10 sources checked- cites [1] Bandodkar, A. J., Jeerapan, I. & Wang, J. Wearable chemical sensors: present challenges and future prospects. ACS Sens. 1, 464–482 (2016). source
Not checked in the free run.
- cites [42] Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J. M. & Wang, J. Stretchable biofuel cells as wearable textile-based self-powered sensors. J. Mater. Chem. A Mater. Energy Sustain. 4, 18342–18353 (2016). source
Not checked in the free run.
- cites [157] Sarpeshkar, R. Universal principles for ultra low power and energy efficient design. IEEE Trans. Circuits Syst. II 59, 193–198 (2012). source
Not checked in the free run.
- cites [158] De Marcellis, A., Depari, A., Ferri, G., Flammini, A. & Sisinni, E. A CMOS integrated low-voltage low-power time-controlled interface for chemical resistive sensors. Sens. Actuators B Chem. 179, 313–318 (2013). source
Not checked in the free run.
- cites [159] Zhang, Y. et al. Realizing both high energy and high power densities by twisting three carbon-nanotube-based hybrid fibers. Angew. Chem. Int. Ed. Engl. 54, 11177–11182 (2015). source
Not checked in the free run.
- cites [160] Huang, Y. et al. Weavable, conductive yarn-based NiCo//Zn textile battery with high energy density and rate capability. ACS Nano 11, 8953–8961 (2017). source
Not checked in the free run.
- cites [161] Mesin, L. A neural algorithm for the non-uniform and adaptive sampling of biomedical data. Comput. Biol. Med. 71, 223–230 (2016). source
Not checked in the free run.
- cites [162] Zi, Y. et al. Triboelectric-pyroelectric-piezoelectric hybrid cell for high-efficiency energy-harvesting and self-powered sensing. Adv. Mater. 27, 2340–2347 (2015). source
Not checked in the free run.
- cites [163] Zhang, K. & Yang, Y. Thermo-phototronic effect enhanced InP/ZnO nanorod heterojunction solar cells for self-powered wearable electronics. Adv. Funct. Mater. 27, 1703331 (2017). source
Not checked in the free run.
- cites [164] Wen, Z. et al. Self-powered textile for wearable electronics by hybridizing fiber-shaped nanogenerators, solar cells, and supercapacitors. Sci. Adv. 2, e1600097 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Findingnone givenMost current wearable biosensing devices measure a small number of biomarkers.
“Most current wearable biosensing devices measure a small number of biomarkers.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenBiofouling is driven by the accumulation of proteins, cells or macromolecules on the sensor surface through non-specific binding.
“Biofouling is driven by the accumulation of proteins, cells or macromolecules on the sensor surface through non-specific binding.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenA plethora of innovative wearable biosensing devices have already been demonstrated in diverse applications, ranging from the detection of metabolites (e.g., lactate or glucose) to the monitoring of electrolytes (e.g., sodium, potassium or calcium) in fluids, such as sweat, ISF, saliva or tears in connection to enzymatic and ion-recognition reactions.
“A plethora of innovative wearable biosensing devices have already been demonstrated in diverse applications, ranging from the detection of metabolites (e.g., lactate or glucose) to the monitoring of electrolytes (e.g., sodium, potassium or calcium) in fluids, such as sweat, ISF, saliva or tears in connection to enzymatic and ion-recognition reactions.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Findingnone givenThe concentrations of many important biomarkers in saliva are substantially lower than in blood, requiring highly sensitive sensing systems for accurate monitoring.
“The concentrations of many important biomarkers in saliva are substantially lower than in blood, requiring highly sensitive sensing systems for accurate monitoring.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingreasoningDespite tremendous recent progress in wearable biosensors, the state-of-the-art in this field remains at demonstrating proof-of-concept wearable biosensing platforms for detecting several representative biomarkers, and only small steps can be taken to practical applications in the field.
“Despite tremendous recent progress in wearable biosensors, the state-of-the-art in this field remains at demonstrating proof-of-concept wearable biosensing platforms for detecting several representative biomarkers, and only small steps can be taken to practical applications in the field.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationWearable batteries have been developed on stretchable and flexible materials for greater wearability, but have not yet demonstrated sufficient energy density for long-term use.
“Alternatively, wearable batteries have been developed on stretchable and flexible materials for greater wearability, but have not yet demonstrated sufficient energy density for long-term use177-179.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [177] Xu, S. et al. Stretchable batteries with self-similar serpentine interconnects and integrated wireless recharging systems. Nat. Commun. 4, 1543 (2013). source
Not checked in the free run.
- cites [178] Berchmans, S. et al. An epidermal alkaline rechargeable Ag-Zn printable tattoo battery for wearable electronics. J. Mater. Chem. A Mater. 2, 15788–15795 (2014). source
Not checked in the free run.
- cites [179] Kumar, R. et al. All-printed, stretchable Zn-Ag2O rechargeable battery via hyperelastic binder for self-powering wearable electronics. Adv. Energy Mater. 7, 1–8 (2017). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenNew data mining algorithms are being developed for making sense of the large amount of data and predicting patterns in connection to specific applications.
“New data mining algorithms are being developed for making sense of the large amount of data and predicting patterns in connection to specific applications.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenThe provision of power to wearable biosensing platforms can be accomplished in several ways, with the most common approach being lithium ion or alkaline batteries.
“The provision of power to wearable biosensing platforms (see Box 1) can be accomplished in several ways, with the most common approach being lithium ion or alkaline batteries.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationLithium ion or alkaline batteries exhibit long lifetimes for extended biosensor operation; however, they are relatively bulky and can pose toxicity issues, particularly lithium ion-based systems.
“These devices exhibit long lifetimes for extended biosensor operation; however, they are relatively bulky and can pose toxicity issues, particularly lithium ion-based systems175,176.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [175] Hammami, A., Raymond, N. & Armand, M. Lithium-ion batteries: runaway risk of forming toxic compounds. Nature 424, 635–636 (2003). source
Not checked in the free run.
- cites [176] Lisbona, D. & Snee, T. A review of hazards associated with primary lithium and lithium-ion batteries. Process Saf. Environ. Prot. 89, 434–442 (2011). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Findingnone givenUnlike traditional (lab-based) biosensors, wearable biosensors can be exposed to harsh and fluctuating conditions (e.g., temperature) during prolonged outdoor activities in uncontrolled environments.
“Unlike traditional (lab-based) biosensors, wearable biosensors can be exposed to harsh and fluctuating conditions (e.g., temperature) during prolonged outdoor activities in uncontrolled environments.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Interpretationnone givenWearable biosensors can obviate painful and risky blood sampling procedures and can be readily blended with a wearer's daily routine.
“Wearable biosensors can obviate painful and risky blood sampling procedures and can be readily blended with a wearer’s daily routine.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationSweat contains metabolites, electrolytes, trace elements, and small amounts of large molecules.
“Generally, sweat contains metabolites (e.g., lactate, glucose, urea, ethanol or cortisol) along with electrolytes (e.g., sodium, potassium, chloride or ammonium), trace elements (e.g., zinc or copper), and small amounts of large molecules (e.g., proteins, nucleic acids, neuropeptides or cytokines)64.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [64] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015). source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
Interpretationnone givenRobust reliable measurement of wearable bioanalyte sensors must overcome challenges including gradual surface biofouling, inefficient sample transport, limited bioreceptor stability, complexity of multi-step bioaffinity assays and receptor regeneration, and calibration issues.
“Robust, reliable measurement also must overcome such challenges as gradual surface biofouling at the body–sensor interface, inefficient transport of sample over the sensor, limited stability of many bioreceptors, the complexity of multi-step bioaffinity assays and related receptor regeneration, and issues posed by calibration for on-body biosensors.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenLimited understanding of the correlations between analyte concentrations in blood and non-invasive biofluids remains a major obstacle.
“Even so, our limited understanding of the correlations between analyte concentrations in the blood and non-invasive biofluids remains a major obstacle.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisereasoningSweat, tears, saliva and interstitial fluid can be sampled non-invasively without disrupting the stratum corneum and without contacting blood, posing minimal risk of harm or infection and being more user-friendly.
“Sweat, tears, saliva and ISF have been targeted as they can be sampled in a non-invasive manner, meaning that they can be readily accessed without disrupting the outermost protecting layers of the body’s skin (i.e., stratum corneum) and without contacting blood. As such, non-invasive sensing methods pose minimal risk of harm or infection and are generally more user-friendly.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationcitationSweat lactate does not directly reflect concurrent blood levels, but it indicates the level of physical exertion during prolonged exercise and can be used as a marker for athletic efficiency without invasive blood sampling.
“Although sweat lactate does not directly reflect the concurrent blood levels, it indicates the level of physical exertion experienced during prolonged exercise and can be used as a marker for athletic efficiency without invasive blood sampling.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [58] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013). source
Not checked in the free run.
- Rests on an assumption: Sweat lactate is a byproduct of local sweat gland metabolism and intense physical activity induces higher generation rates.
May combine two claims.
InterpretationcitationThe reported multiplexed sweat biosensor system proved advantageous for monitoring fitness parameters during exercise, but its utility would be limited in continuous monitoring applications due to its reliance on physical exertion for sweat generation.
“The reported system proved advantageous for monitoring fitness parameters during exercise, but its utility would be limited in continuous monitoring applications due to its reliance on physical exertion for sweat generation.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [35] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationThe duration of sweat generation in a patch-type iontophoretic sweat sensor was limited to 60 min, with varying rates over that time, which could hinder continuous monitoring applications.
“However, the duration of sweat generation was limited to 60 min, with varying rates over that time, which could hinder continuous monitoring applications.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [41] Emaminejad, S. et al. Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform. Proc. Natl Acad. Sci. USA 114, 4625–4630 (2017). source
Not checked in the free run.
May combine two claims.
Interpretationnone givenThe efficiency of glucose extraction by reverse iontophoresis is difficult to control, which can lead to inconsistent volumes of sampled ISF and thus variations in glucose concentration.
“However, the efficiency of glucose extraction by RI is difficult to control, which can lead to inconsistent volumes of sampled ISF and thus variations in glucose concentration.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningAntibody-based bioassays hold great promise for expanding the scope of epidermal wearable biosensors, but their successful on-body evaluation has yet to be demonstrated.
“Although such antibody-based bioassays hold great promise for expanding the scope of epidermal wearable biosensors, which focus primarily on enzymatic metabolite detection, their successful on-body evaluation has yet to be demonstrated.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationcitationAlthough several studies have indicated that sweat glucose concentrations can correlate with concurrent blood levels, the limitations of uncontrolled conditions, contamination, irregular sampling, and low volumes can significantly impact the accuracy of collected data.
“Although several studies have indicated that sweat glucose concentrations can correlate with concurrent blood levels, these limitations can significantly impact the accuracy of collected data 64, 76,77.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [64] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015). source
Not checked in the free run.
- cites [76] Moyer, J., Wilson, D., Finkelshtein, I., Wong, B. & Potts, R. Correlation between sweat glucose and blood glucose in subjects with diabetes. Diabetes Technol. Ther. 14, 398–402 (2012). source
Not checked in the free run.
- cites [77] Sakaguchi, K. et al. Evaluation of a minimally invasive system for measuring glucose area under the curve during oral glucose tolerance tests: usefulness of sweat monitoring for precise measurement. J. Diabetes Sci. Technol. 7, 678–688 (2013). source
Not checked in the free run.
May combine two claims.
InterpretationreasoningUnlike epidermal enzyme-based biosensors, immunosensors cannot be readily regenerated toward continuous monitoring applications, and along with other challenges of multi-step affinity bioassays, require further efforts and innovations.
“Unlike epidermal enzyme-based biosensors, such immunosensors cannot be readily regenerated toward continuous monitoring applications, and along with other challenges of multi-step affinity bioassays, these devices require further efforts and innovations.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningPatch-type sweat biosensors require physical exercise for their operation and are thus not compatible with continuous glucose monitoring in daily-life without exercise.
“Yet, these devices require physical exercise for their operation, and are thus not compatible with continuous glucose monitoring in daily-life without exercise.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenIontophoresis-induced sweat generation obviates the need to exercise for sweat sampling.
“Iontophoresis-induced sweat generation obviates the need to exercise for sweat sampling.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe GlucoWatch Biographer was an FDA-approved device capable of non-invasive glucose monitoring over a 12 hour period with 6 measurements per hour.
“This US Food and Drug Administration (FDA)-approved device is capable of non-invasive glucose monitoring over a 12 hour period with 6 measurements per hour.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [30] Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L. & Garg, S. Clinical evaluation of the GlucoWatch biographer: a continual, non-invasive glucose monitor for patients with diabetes. Biosens. Bioelectron. 16, 621–629 (2001). source
Not checked in the free run.
May combine two claims.
PremisecitationThe caffeine sensing device was not integrated with the sweat generation device, but was coupled to a commercially available Macroduct sweat collector using pilocarpine delivery.
“Even so, as described, the sensing device was not integrated with the sweat generation device, but rather coupled to a commercially available Macroduct sweat collector using pilocarpine delivery.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [92] Tai, L. C. et al. Methylxanthine drug monitoring with wearable sweat sensors. Adv. Mater. 30, e1707442 (2018). source
Not checked in the free run.
May combine two claims.
PremisecitationBiomarker molecules in tears diffuse directly from the blood and exhibit close tear-blood concentration correlations.
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
Quote in the paperSubstantial claim0 of 5 sources checked- cites [94] Farandos, N. M., Yetisen, A. K., Monteiro, M. J., Lowe, C. R. & Yun, S. H. Contact lens sensors in ocular diagnostics. Adv. Healthc. Mater. 4, 792–810 (2015). source
Not checked in the free run.
- cites [95] Baca, J. T., Finegold, D. N. & Asher, S. A. Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus. Ocul. Surf. 5, 280–293 (2007). source
Not checked in the free run.
- cites [96] Thaysen, J. H. & Thorn, N. A. Excretion of urea, sodium, potassium and chloride in human tears. Am. J. Physiol. 178, 160–164 (1954). source
Not checked in the free run.
- cites [97] Mitsubayashi, K. & Arakawa, T. Cavitas sensors: contact lens type sensors & mouthguard sensors. Electroanalysis 28, 1170–1187 (2016). source
Not checked in the free run.
- cites [98] Pankratov, D., Gonzalez-Arribas, E., Blum, Z. & Shleev, S. Tear based bioelectronics. Electroanalysis 28, 1250–1266 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationTear glucose concentrations have been demonstrated to exhibit good correlation with blood glucose levels, reflecting diffusion from the lachrymal artery when tears are sampled without eye irritation or stimulation, which can compromise the relationship.
“In particular, tear glucose concentrations have been demonstrated to exhibit good correlation with blood glucose levels, reflecting the diffusion from the lachrymal artery when the tears are sampled without any eye irritation or stimulation, which can compromise the relationship94, 95, 99-101.”
Quote in the paperSubstantial claim0 of 5 sources checked- cites [94] Farandos, N. M., Yetisen, A. K., Monteiro, M. J., Lowe, C. R. & Yun, S. H. Contact lens sensors in ocular diagnostics. Adv. Healthc. Mater. 4, 792–810 (2015). source
Not checked in the free run.
- cites [95] Baca, J. T., Finegold, D. N. & Asher, S. A. Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus. Ocul. Surf. 5, 280–293 (2007). source
Not checked in the free run.
- cites [99] Sen, D. K. & Sarin, G. S. Tear glucose levels in normal people and in diabetic patients. Br. J. Ophthalmol. 64, 693–695 (1980). source
Not checked in the free run.
- cites [100] Chu, M. X. et al. Soft contact lens biosensor for in situ monitoring of tear glucose as non-invasive blood sugar assessment. Talanta 83, 960–965 (2011). source
Not checked in the free run.
- cites [101] Chu, M. et al. Biomedical soft contact-lens sensor for in situ ocular biomonitoring of tear contents. Biomed. Microdevices 13, 603–611 (2011). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationContact lens-based sensors were initially introduced utilizing optical detection of tear glucose levels based on the interaction of glucose with concanavalin A or phenylboronic acid derivatives.
“Contact lens-based sensors were initially introduced utilizing optical detection of tear glucose levels based on the interaction of glucose with concanavalin A or phenylboronic acid derivatives107,108.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [107] Alexeev, V. L., Das, S., Finegold, D. N. & Asher, S. A. Photonic crystal glucose-sensing material for noninvasive monitoring of glucose in tear fluid. Clin. Chem. 50, 2353–2360 (2004). source
Not checked in the free run.
- cites [108] Ben-Moshe, M., Alexeev, V. L. & Asher, S. A. Fast responsive crystalline colloidal array photonic crystal glucose sensors. Anal. Chem. 78, 5149–5157 (2006). source
Not checked in the free run.
May combine two claims.
PremisecitationSampling tears for in vitro diagnoses is associated with errors due to small sample volumes, ease of evaporation during sample collection, variations in tear production among individuals and throughout the day, and challenging collection methods that can affect sampled analyte concentrations.
“Despite the demonstrated correlations, sampling tears for in vitro diagnoses is associated with several errors that are due to the following: first, small sample volumes102; second, ease of evaporation during sample collection; third, variations in tear production among individuals and throughout the day103; and lastly, challenging collection methods that can affect the sampled analyte concentra”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [102] Mishima, S., Gasset, A., Klyce, S. D. Jr. & Baum, J. L. Determination of tear volume and tear flow. Invest. Ophthalmol. 5, 264–276 (1966).
Not checked in the free run.
- cites [103] Fullard, R. J. & Carney, L. G. Diurnal variation in human tear enzymes. Exp. Eye Res. 38, 15–26 (1984). source
Not checked in the free run.
- cites [104] Yan, Q. et al. Measurement of tear glucose levels with amperometric glucose biosensor/capillary tube configuration. Anal. Chem. 83, 8341–8346 (2011). source
Not checked in the free run.
- cites [105] Stuchell, R. N., Feldman, J. J., Farris, R. L. & Mandel, I. D. The effect of collection technique on tear composition. Invest. Ophthalmol. Vis. Sci. 25, 374–377 (1984).
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationOther contact lens-based optical sensors have involved the use of photonic crystal materials in combination with responsive hydrogels or fluorescent dyes for measuring glucose and other target analytes in tears.
“Other contact lens-based optical sensors have involved the use of photonic crystal materials in combination with responsive hydrogels or fluorescent dyes for measuring glucose and other target analytes in tears110-113.”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [110] Khalil, O. S. Noninvasive photonic-crystal material for sensing glucose in tears. Clin. Chem. 50, 2236–2237 (2004). source
Not checked in the free run.
- cites [111] Cai, Z., Smith, N. L., Zhang, J. T. & Asher, S. A. Two-dimensional photonic crystal chemical and biomolecular sensors. Anal. Chem. 87, 5013–5025 (2015). source
Not checked in the free run.
- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
- cites [113] Sung, Y., Campa, F. & Shih, W. C. Open-source do-it-yourself multi-color fluorescence smartphone microscopy. Biomed. Opt. Express 8, 5075–5086 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationThe Parviz team demonstrated electrochemical biosensing and investigated different biosensing strategies to achieve good sensitivity, linearity and accuracy.
“Further advances for the field came with the demonstration of electrochemical biosensing by the Parviz team114-116. This group investigated different biosensing strategies to achieve good sensitivity, linearity and accuracy.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [114] Yao, H., Afanasiev, A., Lahdesmaki, I. & Parviz, B. A. A dual microscale glucose sensor on a contact lens, tested in conditions mimicking the eye. Proc. IEEE Micro. Electro. Mech. Syst. 2011, 25–28 (2011).
Not checked in the free run.
- cites [115] Yao, H. et al. A contact lens with integrated telecommunication circuit and sensors for wireless and continuous tear glucose monitoring. J. Micromech. Microeng. 22, 075007 (2012). source
Not checked in the free run.
- cites [116] Liao, Y. T., Yao, H. F., Lingley, A., Parviz, B. & Otis, B. P. A. 3-µW CMOS glucose sensor for wireless contact-lens tear glucose monitoring. IEEE J. Solid-State Circuits 47, 335–344 (2012). source
Not checked in the free run.
May combine two claims.
InterpretationcitationSuch sensor placement at the base of the eye provides continuously accessible tear glucose measurements when coupled with wireless data transmission, without causing discomfort.
“Such sensor placement at the base of the eye (behind the eyelid) provides continuously accessible tear glucose measurements when coupled with wireless data transmission, without causing discomfort.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [119] Kownacka, A. E. et al. Clinical evidence for use of a noninvasive biosensor for tear glucose as an alternative to painful finger-prick for diabetes management utilizing biopolymer coating. Biomacromolecules 19, 4504–4511 (2018). source
Not checked in the free run.
May combine two claims.
PremisecitationThe hydrogel-based contact lens device offers fast and easy fabrication along with a rapid and sensitive glucose response.
“This device offers fast and easy fabrication along with a rapid and sensitive glucose response.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
May combine two claims.
Interpretationnone givenTear-based sensors have been focused primarily on glucose monitoring but show considerable promise for non-invasive sensing of other physiologically important biomarkers.
“Overall, tear-based sensors have been focused primarily on glucose monitoring but show considerable promise for non-invasive sensing of other physiologically important biomarkers.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationFurther progress was made by embedding a wireless read out chip (2.4 GHz) and by powering with far-field electromagnetic radiation (3μW within 15 cm).
“Further progress has since been made by embedding a wireless read out chip (2.4 GHz) and by powering with far-field electromagnetic radiation (3μW within 15 cm)116.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [116] Liao, Y. T., Yao, H. F., Lingley, A., Parviz, B. & Otis, B. P. A. 3-µW CMOS glucose sensor for wireless contact-lens tear glucose monitoring. IEEE J. Solid-State Circuits 47, 335–344 (2012). source
Not checked in the free run.
May combine two claims.
PremisecitationThe advanced contact lens device focused on ensuring wearer comfort without hindrance to vision through transparent soft materials, while integrating wireless electronics to eliminate the need for an external power source.
“This advanced device focused on ensuring wearer comfort without hindrance to vision through the use of transparent soft materials, while integrating wireless electronics to eliminate the need for an external power source.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [118] Park, J. et al. Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays. Sci. Adv. 4, eaap9841 (2018). source
Not checked in the free run.
May combine two claims.
PremisecitationTear fluid contains thousands of proteins-the most abundant of which are lysozyme, lactoferrin and albumin.
“As tear fluid contains thousands of proteins—the most abundant of which are lysozyme, lactoferrin and albumin—non-invasive tear monitoring could also be used to detect protein biomarkers correlated with disease121.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [121] Zhou, L. et al. In-depth analysis of the human tear proteome. J. Proteomics 75, 3877–3885 (2012). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenThe high protein content of saliva makes it attractive for detecting disease and stress biomarkers toward biomedical and fitness monitoring.
“The high protein content of saliva makes it attractive for detecting disease and stress biomarkers toward biomedical and fitness monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenCompared with epidermally focused wearable biosensors, tear-based systems benefit from the advantage of continuous access to the target biofluid without the need for induction or extraction.
“Compared with epidermally focused wearable biosensors, tear-based systems benefit from the advantage of continuous access to the target biofluid without the need for induction or extraction.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone givenDifficult sampling procedures complicate reliable tear-based sensing platforms while contact lens-type systems suffer from design constraints imposed by the nature of their operating environment.
“However, difficult sampling procedures complicate reliable tear-based sensing platforms while contact lens-type systems suffer from design constraints imposed by the nature of their operating environment.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSeveral saliva biomarkers (e.g., drug, hormones, metabolites or antibodies) have been used in clinical settings as they offer meaningful diagnostic information.
“Several of these saliva biomarkers (e.g., drug, hormones, metabolites or antibodies) have been used in clinical settings as they offer meaningful diagnostic information140-144.”
Quote in the paperNot a substantial claim0 of 5 sources checked- cites [140] Malon, R. S. P., Sadir, S., Balakrishnan, M. & Córcoles, E. P. Saliva-based biosensors: noninvasive monitoring tool for clinical diagnostics. Biomed Res. Int. 2014, 962903 (2014). source
Not checked in the free run.
- cites [141] Chicharro, J. L., Lucía, A., Pérez, M., Vaquero, A. F. & Ureña, R. Saliva composition and exercise. Sports Med. 26, 17–27 (1998). source
Not checked in the free run.
- cites [142] Singh, M. Non invasive diagnostic tool for pathological conditions salivary biomarkers. Int. J. Pharm. Biol. Arch. 5, 1–12 (2014).
Not checked in the free run.
- cites [143] Gatti, R. & De Palo, E. F. An update: salivary hormones and physical exercise. Scand. J. Med. Sci. Sports 21, 157–169 (2011). source
Not checked in the free run.
- cites [144] Viswanath, B., Choi, C. S., Lee, K. & Kim, S. Recent trends in the development of diagnostic tools for diabetes mellitus using patient saliva. Trends Analyt. Chem. 89, 60–67 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationA fluidic platform was suggested by the Butt group for integration with an optical monitoring system, but has not been demonstrated for biosensing applications.
“Such a fluidic platform was suggested by the Butt group112 for integration with an optical monitoring system, but has not been demonstrated for biosensing applications.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Interpretationnone givenWearable contact lens tear monitoring platforms do not cause any eye irritation and yield a relatively consistent tear fluid composition.
“Wearable contact lens tear monitoring platforms are highly advantageous as they do not cause any eye irritation and yield a relatively consistent tear fluid composition.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone givenMany biomarkers in saliva pass directly from the bloodstream via transcellular or paracellular paths, making saliva the "mirror of the human body" that reflects the body's physiological state to offer a non-invasive alternative to blood analysis.
“Many of the biomarkers in saliva pass directly from the bloodstream via transcellular or paracellular paths, making saliva the “mirror of the human body” that reflects the body’s physiological state to offer a non-invasive alternative to blood analysis.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationSaliva is a complex oral fluid, produced mainly by the parotid gland and composed of numerous constituents, such as metabolites, enzymes, hormones, proteins, microorganisms and ions.
“Saliva is a complex oral fluid, which is produced mainly by the parotid gland and is composed of numerous constituents, such as metabolites, enzymes, hormones, proteins, microorganisms and ions125,135-140.”
Quote in the paperNot a substantial claim0 of 7 sources checked- cites [125] Campuzano, S., Yanez-Sedeno, P. & Pingarron, J. M. Electrochemical bioaffinity sensors for salivary biomarkers detection. Trends Analyt. Chem. 86, 14–24 (2017). source
Not checked in the free run.
- cites [135] Aguirre, A. et al. Sialochemistry: a diagnostic tool? Crit. Rev. Oral Biol. Med. 4, 343–350 (1993). source
Not checked in the free run.
- cites [136] Kaufman, E. & Lamster, I. B. The diagnostic applications of saliva—a review. Crit. Rev. Oral Biol. Med. 13, 197–212 (2002). source
Not checked in the free run.
- cites [137] Chiappin, S., Antonelli, G., Gatti, R. & De Palo, E. F. Saliva specimen: a new laboratory tool for diagnostic and basic investigation. Clin. Chim. Acta 383, 30–40 (2007). source
Not checked in the free run.
- cites [138] Streckfus, C. F. & Bigler, L. R. Saliva as a diagnostic fluid. Oral Dis. 8, 69–76 (2002). source
Not checked in the free run.
- cites [139] Mandel, I. D. The diagnostic uses of saliva. J. Oral Pathol. Med. 19, 119–125 (1990). source
Not checked in the free run.
- cites [140] Malon, R. S. P., Sadir, S., Balakrishnan, M. & Córcoles, E. P. Saliva-based biosensors: noninvasive monitoring tool for clinical diagnostics. Biomed Res. Int. 2014, 962903 (2014). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenSuch in-mouth operation required the replacement of several teeth by the sensors and was subject to risks associated with potential leakage of the internal sensor solution.
“However, such in-mouth operation required the replacement of several teeth by the sensors and was subject to risks associated with potential leakage of the internal sensor solution.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThese oral-cavity sensors were integrated with a resonant coil for battery-free operation and were capable of detecting salivary bacteria at the single cell level in vitro using a naturally occurring antimicrobial peptide (AMP) biorecognition element along with label-free impedance transduction.
“These oral-cavity sensors were integrated with a resonant coil for battery-free operation and were capable of detecting salivary bacteria at the single cell level in vitro using a naturally occurring antimicrobial peptide (AMP) biorecognition element along with label-free impedance transduction.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Mannoor, M. S. et al. Graphene-based wireless bacteria detection on tooth enamel. Nat. Commun. 3, 763 (2012). source
Not checked in the free run.
Classified as probably not a substantial claim.
Interpretationnone givenIn-mouth biosensing platforms can offer an attractive painless route for obtaining dynamic chemical information from saliva.
“Despite these challenges, in-mouth biosensing platforms can offer an attractive painless route for obtaining dynamic chemical information from saliva.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenThe first example of a wearable oral sensor was demonstrated in the 1960s and was based on a partial denture platform for monitoring mastication, monitoring plaque pH and fluoride concentrations.
“The first example of a wearable oral sensor was demonstrated in the 1960s and was based on a partial denture platform for monitoring mastication monitoring plaque pH and fluoride concentrations.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenIn diabetes, changes in hormonal and neural balance may impact the salivary glands, which act as a filter from the blood, and lead to increased secretion of salivary glucose.
“In the case of diabetes, changes in hormonal and neural balance may impact the salivary glands, which act as a filter from the blood, and lead to increased secretion of salivary glucose.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe developed cavitas sensors are based on a GOX-modified poly(ethylene terephthalate) glycol (PETG) surface, and the device is seamlessly integrated with a wireless transmitter on a custom-fitted monolithic mouthguard platform.
“The developed sensors are based on a GOX-modified poly(ethylene terephthalate) glycol (PETG) surface, and the device is seamlessly integrated with a wireless transmitter on a custom-fitted monolithic mouthguard platform.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [148] Arakawa, T. et al. Mouthguard biosensor with telemetry system for monitoring of saliva glucose: a novel cavitas sensor. Biosens. Bioelectron. 84, 106–111 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationSoni et al. demonstrated the correlation between blood and saliva glucose concentrations to be R = 0.64 in healthy subjects, whereas diabetic subjects show a much closer relationship with R = 0.95.
“Soni et al.126 have demonstrated the correlation between blood and saliva glucose concentrations to be R = 0.64 in healthy subjects, whereas diabetic subjects show a much closer relationship with R = 0.95.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [126] Soni, A. & Jha, S. K. A paper strip based non-invasive glucose biosensor for salivary analysis. Biosens. Bioelectron. 67, 763–768 (2015). source
Not checked in the free run.
May combine two claims.
PremisecitationThis configuration enables the telemetric measurement of salivary glucose in artificial saliva over the relevant physiological range (5-1000 μM) and the device has been further characterized through connection to a phantom jaw that mimics the human oral cavity with a saliva flow system.
“This configuration enables the telemetric measurement of salivary glucose in artificial saliva over the relevant physiological range (5–1000 μM) and the device has been further characterized through connection to a phantom jaw that mimics the human oral cavity with a saliva flow system.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [148] Arakawa, T. et al. Mouthguard biosensor with telemetry system for monitoring of saliva glucose: a novel cavitas sensor. Biosens. Bioelectron. 84, 106–111 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing). May combine two claims.
InterpretationreasoningThese advances have greatly enhanced the reliability of wearable biosensor, the analyte monitoring capabilities and wearability.
“These advances have greatly enhanced the reliability of wearable biosensor, the analyte monitoring capabilities and wearability.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone givenSome of these challenges are specific to the individual platforms or target analytes, whereas others are shared by all wearable biosensing systems.
“Some of these challenges are specific to the individual platforms or target analytes, whereas others are shared by all wearable biosensing systems.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThese demonstrations have shown that wearable biosensors have immense potential for real-world applications.
“These demonstrations have shown that wearable biosensors have immense potential for real-world applications.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenFouling of the surface of wearable biosensors, particularly those operating in sweat or tears over short durations, is expected to be less severe compared with that observed with implantable or minimally-invasive sensors.
“Fouling of the surface of wearable biosensors, particularly those operating in sweat or tears over short durations, is expected to be less severe compared with that observed with implantable or minimally-invasive sensors.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenSuch dramatic progress has benefited from the refinement of multiplexed-sensing platforms, improved biofluid sampling and advances in flexible materials and wireless electronics.
“Such dramatic progress has benefited from the refinement of multiplexed-sensing platforms, improved biofluid sampling and advances in flexible materials and wireless electronics.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenA wireless electronic printed circuit board (PCB) containing a fully functional microcontroller is a widely used platform for wireless electronics due to its flexibility and cost-effectiveness.
“A wireless electronic printed circuit board (PCB) containing a fully functional microcontroller is a widely used platform for wireless electronics due to its flexibility and cost-effectiveness.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
InterpretationreasoningSuch severe conditions may affect the stability of their fragile bioreceptors.
“Such severe conditions may affect the stability of their fragile bioreceptors.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningThese issues can be partially addressed using proper microfluidic sampling systems and optimization of surface coating techniques.
“These issues can be partially addressed using proper microfluidic sampling systems and optimization of surface coating techniques.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningMultiplexed sensing, including both biosensors and physical sensors, could provide active calibration for variations in temperature, pH and humidity.
“Multiplexed sensing, including both biosensors and physical sensors, could provide active calibration for variations in temperature, pH and humidity.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationEnergy can be harvested from motion with piezoelectric or triboelectric devices.
“motion with piezoelectric or triboelectric devices185,186”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [185] Yang, J.H. et al. Effect of garment design on piezoelectricity harvesting from joint movement. Smart Mater. Struct. 25, 1–15 (2016). source
Not checked in the free run.
- cites [186] Pu, X. et al. Wearable self-charging power textile based on flexible yarn supercapacitors and fabric nanogenerators. Adv. Mater. 28, 98–105 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationEnergy can be harvested from heat with thermoelectric materials.
“heat with thermoelectric materials187,188”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [187] Oh, J. Y. et al. Chemically exfoliated transition metal dichalcogenide nanosheet-based wearable thermoelectric generators. Energy Environ. Sci. 9, 1696–1705 (2016). source
Not checked in the free run.
- cites [188] Lu, Z. S., Zhang, H. H., Mao, C. P. & Li, C. M. Silk fabric-based wearable thermoelectric generator for energy harvesting from the human body. Appl. Energy 164, 57–63 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationEnergy can be harvested from light with wearable solar cells.
“Energy can be harvested from light with wearable solar cells183,184”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [183] O’Connor, T. F. et al. Wearable organic solar cells with high cyclic bending stability: materials selection criteria. Sol. Energy Mater. Sol. Cells 144, 438–444 (2016). source
Not checked in the free run.
- cites [184] Kim, B. J. et al. Highly efficient and bending durable perovskite solar cells: toward a wearable power source. Energy Environ. Sci. 8, 916–921 (2015). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
InterpretationreasoningAdvances in wearable power sources are a significant need, particularly with increases in energy demands caused by multiplexed sensing platforms.
“Advances in wearable power sources are a significant need, particularly with increases in energy demands caused by multiplexed sensing platforms.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe global wearable sensors market was valued at roughly $150 million in 2016 and is projected to reach $2.86 billion by 2025.
“The global wearable sensors market was valued at roughly $150 million in 2016 and is projected to reach $2.86 billion by 2025, according to a recent market report (https://www.grandviewresearch.com/industry-analysis/global-wearable-sensor-market).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites https://www.grandviewresearch.com/industry-analysis/global-wearable-sensor-market
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Interpretationnone givenThe acceptance of non-invasive biosensors by the medical community will require extensive and successful validation in human testing and improved understanding of the clinical relevancy of sensor information.
“The acceptance of these non-invasive biosensors by the medical community will require extensive and successful validation in human testing and improved understanding of the clinical relevancy of sensor information.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Interpretationnone givenFuture advances, breakthroughs and growth in wearable biosensors will require close multidisciplinary collaboration among the engineering, scientific and medical communities.
“Such future advances, breakthroughs and growth will require close multidisciplinary collaboration among the engineering, scientific and medical communities.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
InterpretationreasoningSensor preparation must ensure high bioreceptor stability for maintaining accuracy and reliability of the response.
“The sensor preparation thus must ensure high bioreceptor stability for maintaining accuracy and reliability of the response.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenFully integrated wearable biosensing platforms would require incorporation of wireless electronics (integrated with energy source) to facilitate data processing and secured signal transmission.
“Fully integrated wearable biosensing platforms would require incorporation of wireless electronics (integrated with energy source) to facilitate the data processing and secured signal transmission.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenThe regeneration issue of immunosensors is a major challenge to overcome if prolonged use of wearables is to be attained.
“The regeneration issue of immunosensors is another major challenge to overcome if prolonged use of wearables is to be attained.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenFor completely reliable use, wearable devices must overcome stability issues caused by prolonged operation under uncontrolled conditions, biofouling from constituents present in the sampled biofluids, and intrinsic instabilities of the biological recognition components themselves.
“For completely reliable use, wearable devices must overcome stability issues caused by prolonged operation under uncontrolled conditions, biofouling from constituents present in the sampled biofluids, and intrinsic instabilities of the biological recognition components themselves.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe use of mobile devices and smartphone-based microscopes, including algorithm-based applications, is expected to facilitate the readout of the response in optical wearable biosensors.
“Furthermore, the use of mobile devices and smartphone-based microscopes, including algorithm-based applications, is expected to facilitate the readout of the response in optical wearable biosensors112,113,195.”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [112] Elsherif, M., Hassan, M. U., Yetisen, A. K. & Butt, H. Wearable contact lens biosensors for continuous glucose monitoring using smartphones. ACS Nano 12, 5452–5462 (2018). source
Not checked in the free run.
- cites [113] Sung, Y., Campa, F. & Shih, W. C. Open-source do-it-yourself multi-color fluorescence smartphone microscopy. Biomed. Opt. Express 8, 5075–5086 (2017). source
Not checked in the free run.
- cites [195] Yetisen, A. K., Martinez-Hurtado, J. L., Garcia-Melendrez, A., Vasconcellos, F. D. & Lowe, C. R. A smartphone algorithm with inter-phone repeatability for the analysis of colorimetric tests. Sens. Actuators B Chem. 196, 156–160 (2014). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA large portion of the future wearable sensors market share is expected to be made up of wearable biosensors, primarily non-invasive glucose monitoring.
“A large portion of this future market share is expected to be made up of wearable biosensors, primarily non-invasive glucose monitoring2.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2] Heikenfeld, J. et al. Wearable sensors: modalities, challenges, and prospects. Lab Chip 18, 217–248 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S4 (based).
InterpretationreasoningWearable devices must be capable of robust operation without the need for constant (re)calibration.
“Furthermore, the devices must be capable of robust operation without the need for constant (re)calibration (common in laboratory settings).”
Quote in the paperNot a substantial claim
The paper relies on this without support.
Sweat glands are distributed across the entire body with more than 100 glands/cm2 of skin.
“Sweat is the most readily obtainable biofluid for chemical sensing applications since sweat glands are distributed across the entire body with more than 100 glands/cm2 of skin.”
The paper relies on this without support.
Sweat lactate is a byproduct of local sweat gland metabolism and intense physical activity induces higher generation rates.
“Sweat lactate is a byproduct of local sweat gland metabolism and intense physical activity induces higher generation rates.”
The paper relies on this without support.
ISF components diffuse directly from the capillary endothelium.
“Within the viable skin tissue, skin cells are surrounded by ISF, which provides nutrients that diffuse directly from the capillary endothelium.”
The paper relies on this without support.
Wearable biosensor technologies can improve health and performance.
“Considering all the above challenges, we are only at the beginning of understanding how wearable biosensor technologies can improve our health and performance.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Biosensing technology is increasingly being used to provide real-time physiological information using dynamic non-invasive measurements of markers including but not limited to heart rate.
“Biosensing technology is increasingly being used to provide real-time physiological information using dynamic non-invasive measurements of markers including but not limited to heart rate (Kim et al. 2019).”
From Effects of a Personalized Stress Management Intervention on Maternal Mental Health: A Randomized Clinical Trial. (2025) · cites this paper as Kim et al. 2019
Despite some efforts, several challenges remain for tear biosensors, such as difficulty in sampling tears, small sample volumes, issues with fast evaporation, dependence of tear composition on emotional or mechanical stimulation, and reliable operation of fully integrated wireless sensing platforms.
“Despite some efforts, several challenges remain, such as difficulty in sampling tears, small sample volumes, issues with fast evaporation, dependence of tear composition on emotional or mechanical stimulation, and reliable operation of fully integrated wireless sensing platforms [126,152].”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [126]
Biochemical sensors can measure body fluid electrolytes with electrochemical transducers, offering valuable information about plasma volume status and analyte concentrations.
“Biochemical sensors can measure body fluid electrolytes with the use of electrochemical transducers, offering valuable information about plasma volume status and analyte concentrations30.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [30]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Iontophoretic alcohol biosensor cites an unrelated paper
MinorReference 75 concerns sensing fibers woven into fabric, whereas the cited passage describes an iontophoretic tattoo biosensor for sweat alcohol.
“A recent paper from our group (J.K., J.W. et al.)75 describes the merging of an iontophoretic sweat-generation system with an amperometric biosensing system in a single wearable tattoo platform (Fig. 4d).”
“75. Wang L Weaving sensing fibers into electrochemical fabric for real-time health monitoring. Adv. Funct. Mater 28, 1804456 (2018).”
Piezoelectric biosensor description cites an optical-sensor paper
MinorThe cited reference 56 describes a textile patch with optical detection, not the piezoelectric electronic-skin biosensor described in the text.
“piezoelectric biosensing systems have also been introduced as new electronic-skin platforms monitoring sweat metabolites (Fig. 3h)56.”
“56. Morris D Bio-sensing textile based patch with integrated optical detection system for sweat monitoring. Sens. Actuators B 139, 231– 236 (2009).”
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
Clark, L. C., & Lyons, C. (1962) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1962.tb13623.x · Cited by 3,945 · see citing works
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498 · see citing works
Self-citationStretchable, Skin-Mountable, and Wearable Strain Sensors and Their Potential Applications: A Review
Amjadi, M., Kyung, K., Park, I., & Sitti, M. (2016) · Advanced Functional Materials · doi:10.1002/adfm.201504755 · Cited by 3,202 · see citing works
Multiplexed electrical detection of cancer markers with nanowire sensor arrays
Zheng, G., Patolsky, F., Cui, Y., Wang, W. U., & Lieber, C. M. (2005) · Nature Biotechnology · doi:10.1038/nbt1138 · Cited by 2,430 · see citing works
Wearable Electronics and Smart Textiles: A Critical Review
Stoppa, M., & Chiolerio, A. (2014) · Sensors · doi:10.3390/s140711957 · Cited by 2,107 · see citing works
Lee, H., Choi, T. K., Lee, Y. B., Cho, H. R., Ghaffari, R., Wang, L., Choi, H. J., Chung, T. D., Lu, N., Hyeon, T., et al. (2016) · Nature Nanotechnology · doi:10.1038/nnano.2016.38 · Cited by 1,802 · see citing works
Lab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring
Liu, Y., Pharr, M., & Salvatore, G. A. (2017) · ACS Nano · doi:10.1021/acsnano.7b04898 · Cited by 1,725 · see citing works
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · see citing works
Gibson, L. E., & Cooke, R. E. (1959) · Pediatrics · doi:10.1542/peds.23.3.545 · Cited by 1,434 · see citing works
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434 · see citing works
Self-citationXu, S., Zhang, Y., Cho, J., Lee, J., Huang, X., Jia, L., Fan, J. A., Su, Y., Su, J., Zhang, H., et al. (2013) · Nature Communications · doi:10.1038/ncomms2553 · Cited by 1,404 · see citing works
Wearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315 · see citing works
A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., et al. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf2593 · Cited by 1,308 · see citing works
Self-citationThe Rise of Consumer Health Wearables: Promises and Barriers
Piwek, L., Ellis, D. A., Andrews, S., & Joinson, A. (2016) · PLoS Medicine · doi:10.1371/journal.pmed.1001953 · Cited by 1,301 · see citing works
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · see citing works
Self-citationWang, Q., Yan, J., & Fan, Z. (2016) · Energy & Environmental Science · doi:10.1039/c5ee03109e · Cited by 1,225 · see citing works
Label‐Free Impedance Biosensors: Opportunities and Challenges
Daniels, J. S., & Pourmand, N. (2007) · Electroanalysis · doi:10.1002/elan.200603855 · Cited by 1,197 · see citing works
Lee, H., Song, C., Hong, Y. S., Kim, M., Cho, H. R., Kang, T., Shin, K., Choi, S. H., Hyeon, T., & Kim, D.-H. (2017) · Science Advances · doi:10.1126/sciadv.1601314 · Cited by 1,118 · see citing works
Infrared Regulating Smart Window Based on Organic Materials
Khandelwal, H., Schenning, A. P. H. J., & Debije, M. G. (2017) · Advanced Energy Materials · doi:10.1002/aenm.201602209 · Cited by 1,031 · see citing works
Graphene-based wireless bacteria detection on tooth enamel
Mannoor, M. S., Tao, H., Clayton, J. D., Sengupta, A., Kaplan, D. L., Naik, R. R., Verma, N., Omenetto, F. G., & McAlpine, M. C. (2012) · Nature Communications · doi:10.1038/ncomms1767 · Cited by 1,014 · see citing works
Wearable smart sensor systems integrated on soft contact lenses for wireless ocular diagnostics
Kim, J., Kim, M., Lee, M.-S., Kim, K., Ji, S., Kim, Y.-T., Park, J., Na, K., Bae, K.-H., Kyun Kim, H., et al. (2017) · Nature Communications · doi:10.1038/ncomms14997 · Cited by 951 · see citing works
Self-citationHome blood glucose biosensors: a commercial perspective
Newman, J. D., & Turner, A. P. F. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.11.012 · Cited by 901 · see citing works
Kaufman, E., & Lamster, I. B. (2002) · Critical Reviews in Oral Biology & Medicine · doi:10.1177/154411130201300209 · Cited by 887 · see citing works
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862 · see citing works
Self-citationImani, S., Bandodkar, A. J., Mohan, A. M. V., Kumar, R., Yu, S., Wang, J., & Mercier, P. P. (2016) · Nature Communications · doi:10.1038/ncomms11650 · Cited by 861 · see citing works
Self-citationWen, Z., Yeh, M.-H., Guo, H., Wang, J., Zi, Y., Xu, W., Deng, J., Zhu, L., Wang, X., Hu, C., et al. (2016) · Science Advances · doi:10.1126/sciadv.1600097 · Cited by 854 · see citing works
Self-citationEmaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839 · see citing works
Saliva specimen: A new laboratory tool for diagnostic and basic investigation
Chiappin, S., Antonelli, G., Gatti, R., & De Palo, E. F. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.04.011 · Cited by 828 · see citing works
Pu, X., Li, L., Liu, M., Jiang, C., Du, C., Zhao, Z., Hu, W., & Wang, Z. L. (2016) · Advanced Materials · doi:10.1002/adma.201504403 · Cited by 820 · see citing works
Label‐Free Electronic Detection of Thrombin in Blood Serum by Using an Aptamer‐Based Sensor
Xiao, Y., Lubin, A. A., Heeger, A. J., & Plaxco, K. W. (2005) · Angewandte Chemie International Edition · doi:10.1002/anie.200500989 · Cited by 814 · see citing works
Diagnostic Potential of Saliva: Current State and Future Applications
Pfaffe, T., Cooper-White, J. J., Beyerlein, P., Kostner, K., & Punyadeera, C. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.153767 · Cited by 803 · see citing works
Wearable Chemical Sensors: Present Challenges and Future Prospects
Bandodkar, A. J., Jeerapan, I., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00250 · Cited by 778 · see citing works
Self-citationSURVEY AND SUMMARY: From DNA biosensors to gene chips
Wang, J. (2000) · Nucleic Acids Research · doi:10.1093/nar/28.16.3011 · Cited by 753 · see citing works
Self-citationContinuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 743 · see citing works
Sonner, Z., Wilder, E., Heikenfeld, J., Kasting, G., Beyette, F., Swaile, D., Sherman, F., Joyce, J., Hagen, J., Kelley-Loughnane, N., et al. (2015) · Biomicrofluidics · doi:10.1063/1.4921039 · Cited by 724 · see citing works
A review of hazards associated with primary lithium and lithium-ion batteries
Lisbona, D., & Snee, T. (2011) · Process Safety and Environmental Protection · doi:10.1016/j.psep.2011.06.022 · Cited by 701 · see citing works
Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
Bandodkar, A. J., Jia, W., Yardımcı, C., Wang, X., Ramirez, J., & Wang, J. (2015) · Analytical Chemistry · doi:10.1021/ac504300n · Cited by 700 · see citing works
Self-citationWearable Electrochemical Sensors and Biosensors: A Review
Windmiller, J. R., & Wang, J. (2012) · Electroanalysis · doi:10.1002/elan.201200349 · Cited by 699 · see citing works
Self-citationHighly efficient and bending durable perovskite solar cells: toward a wearable power source
Kim, B. J., Kim, D. H., Lee, Y.-Y., Shin, H.-W., Han, G. S., Hong, J. S., Mahmood, K., Ahn, T. K., Joo, Y.-C., Hong, K. S., et al. (2015) · Energy & Environmental Science · doi:10.1039/c4ee02441a · Cited by 679 · see citing works
Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics
Kim, J., Imani, S., de Araujo, W. R., Warchall, J., Valdés-Ramírez, G., Paixão, T. R. L. C., Mercier, P. P., & Wang, J. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.039 · Cited by 678 · see citing works
Self-citationSoft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays
Park, J., Kim, J., Kim, S.-Y., Cheong, W. H., Jang, J., Park, Y.-G., Na, K., Kim, Y.-T., Heo, J. H., Lee, C. Y., et al. (2018) · Science Advances · doi:10.1126/sciadv.aap9841 · Cited by 671 · see citing works
Self-citationBiosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
Determination of tear volume and tear flow.
Mishima, S., Gasset, A. R., Klyce, S. D., & BAUM, J. L. (1966) · PubMed · Cited by 607 · see citing works
Noninvasive Alcohol Monitoring Using a Wearable Tattoo-Based Iontophoretic-Biosensing System
Kim, J., Jeerapan, I., Imani, S., Cho, T. N., Bandodkar, A., Cinti, S., Mercier, P. P., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00356 · Cited by 599 · see citing works
Self-citationData security and privacy in wireless body area networks
Li, M., Lou, W., & Ren, K. (2010) · IEEE Wireless Communications · doi:10.1109/mwc.2010.5416350 · Cited by 589 · see citing works
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2017) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 585 · see citing works
Self-citationCorrelation Between Sweat Glucose and Blood Glucose in Subjects with Diabetes
Moyer, J., Wilson, D., Finkelshtein, I., Wong, B., & Potts, R. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0262 · Cited by 570 · see citing works
Mani, V., Chikkaveeraiah, B. V., Patel, V., Gutkind, J. S., & Rusling, J. F. (2009) · ACS Nano · doi:10.1021/nn800863w · Cited by 534 · see citing works
In Situ Interfacial Mass Detection with Piezoelectric Transducers
Ward, M. D., & Buttry, D. A. (1990) · Science · doi:10.1126/science.249.4972.1000 · Cited by 526 · see citing works
Simultaneous Monitoring of Sweat and Interstitial Fluid Using a Single Wearable Biosensor Platform
Kim, J., Sempionatto, J. R., Imani, S., Hartel, M. C., Barfidokht, A., Tang, G., Campbell, A. S., Mercier, P. P., & Wang, J. (2018) · Advanced Science · doi:10.1002/advs.201800880 · Cited by 525 · see citing works
Self-citationBandodkar, A. J., Molinnus, D., Mirza, O., Guinovart, T., Windmiller, J. R., Valdés-Ramírez, G., Andrade, F. J., Schöning, M. J., & Wang, J. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.11.039 · Cited by 490 · see citing works
Self-citationZi, Y., Lin, L., Wang, J., Wang, S., Chen, J., Fan, X., Yang, P., Yi, F., & Wang, Z. L. (2015) · Advanced Materials · doi:10.1002/adma.201500121 · Cited by 482 · see citing works
Self-citationMartín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · see citing works
Self-citationSkin-like biosensor system via electrochemical channels for noninvasive blood glucose monitoring
Chen, Y., Lu, S., Zhang, S., Li, Y., Qu, Z., Chen, Y., Lu, B., Wang, X., & Feng, X. (2017) · Science Advances · doi:10.1126/sciadv.1701629 · Cited by 470 · see citing works
Contact Lens Sensors in Ocular Diagnostics
Farandos, N. M., Yetisen, A. K., Monteiro, M. J., Lowe, C. R., & Yun, S. H. (2015) · Advanced Healthcare Materials · doi:10.1002/adhm.201400504 · Cited by 466 · see citing works
Advanced Materials for Printed Wearable Electrochemical Devices: A Review
Kim, J., Kumar, R., Bandodkar, A. J., & Wang, J. (2017) · Advanced Electronic Materials · doi:10.1002/aelm.201600260 · Cited by 452 · see citing works
Self-citationSaliva as a diagnostic tool for oral and systemic diseases
Javaid, M. A., Ahmed, A. S., Durand, R., & Tran, S. D. (2016) · Journal of Oral Biology and Craniofacial Research · doi:10.1016/j.jobcr.2015.08.006 · Cited by 431 · see citing works
Bandodkar, A. J., You, J., Kim, N.-H., Gu, Y. Q., Kumar, R., Mohan, A. M. V., Kurniawan, J. F., Imani, S., Nakagawa, T., Parish, B., et al. (2017) · Energy & Environmental Science · doi:10.1039/c7ee00865a · Cited by 413 · see citing works
Self-citationTierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390 · see citing works
Tattoo-based potentiometric ion-selective sensors for epidermal pH monitoring
Bandodkar, A. J., Hung, V. W. S., Jia, W., Valdés-Ramírez, G., Windmiller, J. R., Martinez, A. G., Ramírez, J., Chan, G., Kerman, K., & Wang, J. (2013) · The Analyst · doi:10.1039/c2an36422k · Cited by 379 · see citing works
Self-citationIn-depth analysis of the human tear proteome
Zhou, L., Zhao, S. Z., Koh, S. K., Chen, L., Vaz, C., Tanavde, V., Li, X. R., & Beuerman, R. W. (2012) · Journal of Proteomics · doi:10.1016/j.jprot.2012.04.053 · Cited by 379 · see citing works
Streckfus, C., & Bigler, L. (2002) · Oral Diseases · doi:10.1034/j.1601-0825.2002.1o834.x · Cited by 374 · see citing works
Zangheri, M., Cevenini, L., Anfossi, L., Baggiani, C., Simoni, P., Di Nardo, F., & Roda, A. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.08.048 · Cited by 373 · see citing works
A 3-$\mu\hbox{W}$ CMOS Glucose Sensor for Wireless Contact-Lens Tear Glucose Monitoring
Liao, Y.-T., Yao, H., Lingley, A., Parviz, B., & Otis, B. P. (2012) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2170633 · Cited by 372 · see citing works
Wearable Contact Lens Biosensors for Continuous Glucose Monitoring Using Smartphones
Elsherif, M., Hassan, M. U., Yetisen, A. K., & Butt, H. (2018) · ACS Nano · doi:10.1021/acsnano.8b00829 · Cited by 371 · see citing works
Photonic Crystal Glucose-Sensing Material for Noninvasive Monitoring of Glucose in Tear Fluid
Alexeev, V. L., Das, S., Finegold, D. N., & Asher, S. A. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2004.039701 · Cited by 370 · see citing works
Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites
Kim, J., Valdés-Ramírez, G., Bandodkar, A. J., Jia, W., Martinez, A. G., Ramírez, J., Mercier, P., & Wang, J. (2014) · The Analyst · doi:10.1039/c3an02359a · Cited by 367 · see citing works
Self-citationSmartphone based health accessory for colorimetric detection of biomarkers in sweat and saliva
Oncescu, V., O’Dell, D., & Erickson, D. (2013) · Lab on a Chip · doi:10.1039/c3lc50431j · Cited by 362 · see citing works
Chemical Sensors Based on Fiber Optics
Seitz, W. R. (1984) · Analytical Chemistry · doi:10.1021/ac00265a711 · Cited by 359 · see citing works
Mouthguard biosensor with telemetry system for monitoring of saliva glucose: A novel cavitas sensor
Arakawa, T., Kuroki, Y., Nitta, H., Chouhan, P., Toma, K., Sawada, S.-I., Takeuchi, S., Sekita, T., Akiyoshi, K., Minakuchi, S., et al. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.12.014 · Cited by 356 · see citing works
Weaving Sensing Fibers into Electrochemical Fabric for Real‐Time Health Monitoring
Wang, L., Wang, L., Zhang, Y., Pan, J., Li, S., Sun, X., Zhang, B., & Peng, H. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201804456 · Cited by 353 · see citing works
Huang, Y., Ip, W. S., Lau, Y. Y., Sun, J., Zeng, J., Yeung, N. S. S., Ng, W. S., Li, H., Pei, Z., Xue, Q., et al. (2017) · ACS Nano · doi:10.1021/acsnano.7b03322 · Cited by 352 · see citing works
Piezoelectric crystal biosensor modified with protein A for determination of immunoglobulins
Muramatsu, H., Dicks, J. M., Tamiya, E., & Karube, I. (1987) · Analytical Chemistry · doi:10.1021/ac00150a007 · Cited by 350 · see citing works
Runaway risk of forming toxic compounds
Hammami, A., Raymond, N., & Armand, M. (2003) · Nature · doi:10.1038/424635b · Cited by 346 · see citing works
Saliva Composition and Exercise
Chicharro, J. L., Lucía, A., Pérez, M., Vaquero, A. F., & Ureña, R. (1998) · Sports Medicine · doi:10.2165/00007256-199826010-00002 · Cited by 345 · see citing works
A potentiometric tattoo sensor for monitoring ammonium in sweat
Guinovart, T., Bandodkar, A. J., Windmiller, J. R., Andrade, F. J., & Wang, J. (2013) · The Analyst · doi:10.1039/c3an01672b · Cited by 343 · see citing works
Self-citationLipani, L., Dupont, B. G. R., Doungmene, F., Marken, F., Tyrrell, R. M., Guy, R. H., & Ilie, A. (2018) · Nature Nanotechnology · doi:10.1038/s41565-018-0112-4 · Cited by 340 · see citing works
Silk fabric-based wearable thermoelectric generator for energy harvesting from the human body
Lu, Z., Zhang, H., Mao, C., & Li, C. M. (2016) · Applied Energy · doi:10.1016/j.apenergy.2015.11.038 · Cited by 326 · see citing works
Epidermal Biofuel Cells: Energy Harvesting from Human Perspiration
Jia, W., Valdés‐Ramírez, G., Bandodkar, A. J., Windmiller, J. R., & Wang, J. (2013) · Angewandte Chemie International Edition · doi:10.1002/anie.201302922 · Cited by 324 · see citing works
Self-citationStretchable biofuel cells as wearable textile-based self-powered sensors
Jeerapan, I., Sempionatto, J. R., Pavinatto, A., You, J.-M., & Wang, J. (2016) · Journal of Materials Chemistry A · doi:10.1039/c6ta08358g · Cited by 317 · see citing works
Self-citationAdvanced Materials for Health Monitoring with Skin‐Based Wearable Devices
Jin, H., Abu‐Raya, Y. S., & Haick, H. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700024 · Cited by 317 · see citing works
Tattoo‐Based Wearable Electrochemical Devices: A Review
Bandodkar, A. J., Jia, W., & Wang, J. (2015) · Electroanalysis · doi:10.1002/elan.201400537 · Cited by 316 · see citing works
Self-citationSelf-Powered Wearable Electrocardiography Using a Wearable Thermoelectric Power Generator
Kim, C. S., Yang, H. M., Lee, J., Lee, G. S., Choi, H., Kim, Y. J., Lim, S. H., Cho, S. H., & Cho, B. J. (2018) · ACS Energy Letters · doi:10.1021/acsenergylett.7b01237 · Cited by 316 · see citing works
Biosensing using dynamic-mode cantilever sensors: A review
Johnson, B. N., & Mutharasan, R. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2011.10.054 · Cited by 315 · see citing works
Grattieri, M., & Minteer, S. D. (2018) · ACS Sensors · doi:10.1021/acssensors.7b00818 · Cited by 312 · see citing works
Eyeglasses based wireless electrolyte and metabolite sensor platform
Sempionatto, J. R., Nakagawa, T., Pavinatto, A., Mensah, S. T., Imani, S., Mercier, P., & Wang, J. (2017) · Lab on a Chip · doi:10.1039/c7lc00192d · Cited by 311 · see citing works
Self-citationMethylxanthine Drug Monitoring with Wearable Sweat Sensors
Tai, L., Gao, W., Chao, M., Bariya, M., Ngo, Q. P., Shahpar, Z., Nyein, H. Y. Y., Park, H., Sun, J., Jung, Y., et al. (2018) · Advanced Materials · doi:10.1002/adma.201707442 · Cited by 310 · see citing works
Composition of interstitial fluid
Fogh-Andersen, N., Altura, B. M., Altura, B. T., & Siggaard-Andersen, O. (1995) · Clinical Chemistry · doi:10.1093/clinchem/41.10.1522 · Cited by 309 · see citing works
Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring
Bindra, D. S., Zhang, Y., Wilson, G. S., Sternberg, R., Thevenot, D. R., Moatti, D., & Reach, G. (1991) · Analytical Chemistry · doi:10.1021/ac00017a008 · Cited by 309 · see citing works
Mandel, I. D. (1990) · Journal of Oral Pathology & Medicine · doi:10.1111/j.1600-0714.1990.tb00809.x · Cited by 305 · see citing works
Advances in wearable chemical sensor design for monitoring biological fluids
Matzeu, G., Florea, L., & Diamond, D. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.01.077 · Cited by 304 · see citing works
Fast Responsive Crystalline Colloidal Array Photonic Crystal Glucose Sensors
Ben-Moshe, M., Alexeev, V. L., & Asher, S. A. (2006) · Analytical Chemistry · doi:10.1021/ac060643i · Cited by 292 · see citing works
Oh, J. Y., Lee, J. H., Han, S. W., Chae, S. S., Bae, E. J., Kang, Y. H., Choi, W. J., Cho, S. Y., Lee, J.-O., Baik, H. K., et al. (2016) · Energy & Environmental Science · doi:10.1039/c5ee03813h · Cited by 289 · see citing works
Triboelectric Generators and Sensors for Self-Powered Wearable Electronics
Ha, M., Park, J., Lee, Y., & Ko, H. (2015) · ACS Nano · doi:10.1021/acsnano.5b01478 · Cited by 288 · see citing works
A smartphone algorithm with inter-phone repeatability for the analysis of colorimetric tests
Yetisen, A. K., Martinez-Hurtado, J. L., Garcia-Melendrez, A., da Cruz Vasconcellos, F., & Lowe, C. R. (2014) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2014.01.077 · Cited by 285 · see citing works
Stretchable, Wireless Sensors and Functional Substrates for Epidermal Characterization of Sweat
Huang, X. L., Liu, Y., Chen, K., Shin, W., Lu, C., Kong, G., Patnaik, D., Lee, S., Cortes, J. F., & Rogers, J. A. (2014) · Small · doi:10.1002/smll.201400483 · Cited by 282 · see citing works
Chu, M. X., Miyajima, K., Takahashi, D., Arakawa, T., Sano, K., Sawada, S.-I., Kudo, H., Iwasaki, Y., Akiyoshi, K., Mochizuki, M., et al. (2011) · Talanta · doi:10.1016/j.talanta.2010.10.055 · Cited by 279 · see citing works
Wearable Wireless Tyrosinase Bandage and Microneedle Sensors: Toward Melanoma Screening
Ciui, B., Martín, A., Mishra, R. K., Brunetti, B., Nakagawa, T., Dawkins, T. J., Lyu, M., Cristea, C., Săndulescu, R., & Wang, J. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701264 · Cited by 262 · see citing works
Self-citationContinuous Glucose Monitoring Systems: A Review
Vashist, S. (2013) · Diagnostics · doi:10.3390/diagnostics3040385 · Cited by 261 · see citing works
Kumar, R., Shin, J., Yin, L., You, J., Meng, Y. S., & Wang, J. (2017) · Advanced Energy Materials · doi:10.1002/aenm.201602096 · Cited by 261 · see citing works
Self-citationSaliva-Based Biosensors: Noninvasive Monitoring Tool for Clinical Diagnostics
Malon, R. S. P., Sadir, S., Balakrishnan, M., & Córcoles, E. P. (2014) · BioMed Research International · doi:10.1155/2014/962903 · Cited by 260 · see citing works
An introduction to electrochemical DNA biosensors
Odenthal, K. J., & Gooding, J. J. (2007) · The Analyst · doi:10.1039/b701816a · Cited by 257 · see citing works
Cutting out the middleman: DNA biosensors based on electrochemical oxidation
Thorp, H. H. (1998) · Trends in biotechnology · doi:10.1016/s0167-7799(97)01162-1 · Cited by 255 · see citing works
Frew, J. E., & Hill, H. A. O. (1987) · Analytical Chemistry · doi:10.1021/ac00142a001 · Cited by 252 · see citing works
Wearable temporary tattoo sensor for real-time trace metal monitoring in human sweat
Kim, J., de Araujo, W. R., Samek, I. A., Bandodkar, A. J., Jia, W., Brunetti, B., Paixão, T. R. L. C., & Wang, J. (2015) · Electrochemistry Communications · doi:10.1016/j.elecom.2014.11.024 · Cited by 245 · see citing works
Self-citationSekine, Y., Kim, S. B., Zhang, Y., Bandodkar, A. J., Xu, S., Choi, J., Irie, M., Ray, T. R., Kohli, P., Kozai, N., et al. (2018) · Lab on a Chip · doi:10.1039/c8lc00530c · Cited by 242 · see citing works
Bio-sensing textile based patch with integrated optical detection system for sweat monitoring
Morris, D., Coyle, S. M., Wu, Y., Lau, K. T., Wallace, G. G., & Diamond, D. (2009) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2009.02.032 · Cited by 234 · see citing works
Noninvasive glucose monitoring using saliva nano-biosensor
Zhang, W., Du, Y., & Wang, M. L. (2015) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2015.02.002 · Cited by 232 · see citing works
Portable biosensor for monitoring cortisol in low-volume perspired human sweat
Kinnamon, D., Ghanta, R., Lin, K.-C., Muthukumar, S., & Prasad, S. (2017) · Scientific Reports · doi:10.1038/s41598-017-13684-7 · Cited by 231 · see citing works
Tseng, P., Napier, B., Garbarini, L., Kaplan, D. L., & Omenetto, F. G. (2018) · Advanced Materials · doi:10.1002/adma.201703257 · Cited by 221 · see citing works
Two-Dimensional Photonic Crystal Chemical and Biomolecular Sensors
Cai, Z., Smith, N. L., Zhang, J.-T., & Asher, S. A. (2015) · Analytical Chemistry · doi:10.1021/ac504679n · Cited by 216 · see citing works
An update: salivary hormones and physical exercise
Gatti, R., & De Palo, E. F. (2011) · Scandinavian Journal of Medicine & Science in Sports · doi:10.1111/j.1600-0838.2010.01252.x · Cited by 213 · see citing works
Curto, V. F., Fay, C., Coyle, S. M., Byrne, R. J., O’Toole, C., Barry, C., Hughes, S. M., Moyna, N. M., Diamond, D., & Benito‐Lopez, F. (2012) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2012.06.048 · Cited by 211 · see citing works
Roda, A., Guardigli, M., Calabria, D., Calabretta, M. M., Cevenini, L., & Michelini, E. (2014) · The Analyst · doi:10.1039/c4an01612b · Cited by 206 · see citing works
Tear glucose levels in normal people and in diabetic patients.
Sen, D. K., & Sarin, G. S. (1980) · British Journal of Ophthalmology · doi:10.1136/bjo.64.9.693 · Cited by 203 · see citing works
Salivary Glucose Concentration and Excretion in Normal and Diabetic Subjects
Jurysta, C., Bulur, N., Oguzhan, B., Satman, I., Yilmaz, T. M., Malaisse, W. J., & Sener, A. (2009) · BioMed Research International · doi:10.1155/2009/430426 · Cited by 198 · see citing works
Yamaguchi, M., Deguchi, M., Wakasugi, J., Ono, S., Takai, N., Higashi, T., & Mizuno, Y. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2005.03.014 · Cited by 195 · see citing works
Wearable Biofuel Cells: A Review
Bandodkar, A. J., & Wang, J. (2016) · Electroanalysis · doi:10.1002/elan.201600019 · Cited by 193 · see citing works
Self-citationTextile Organic Electrochemical Transistors as a Platform for Wearable Biosensors
Gualandi, I., Marzocchi, M., Achilli, A., Cavedale, D., Bonfiglió, A., & Fraboni, B. (2016) · Scientific Reports · doi:10.1038/srep33637 · Cited by 193 · see citing works
Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus
Baca, J. T., Finegold, D. N., & Asher, S. A. (2007) · The Ocular Surface · doi:10.1016/s1542-0124(12)70094-0 · Cited by 187 · see citing works
Measurement of tear glucose levels with amperometric glucose biosensor/capillary tube configuration
Yan, Q., Peng, B., Su, G., Cohan, B. E., Major, T. C., & Meyerhoff, M. E. (2011) · Analytical Chemistry · doi:10.1021/ac201700c · Cited by 185 · see citing works
Privacy and Security in Mobile Health: A Research Agenda
Kotz, D., Gunter, C. A., Kumar, S., & Weiner, J. P. (2016) · Computer · doi:10.1109/mc.2016.185 · Cited by 185 · see citing works
Frew, J. E., & Hill, H. A. O. (1987) · Analytical Chemistry · doi:10.1021/ac00142a720 · Cited by 185 · see citing works
Investigation of the human tear film proteome using multiple proteomic approaches.
Green‐Church, K. B., Nichols, K. K., Kleinholz, N. M., Zhang, L., & Nichols, J. J. (2008) · PubMed Central · Cited by 184 · see citing works
Wearable textile biofuel cells for powering electronics
Jia, W., Wang, X., Imani, S., Bandodkar, A. J., Ramírez, J., Mercier, P. P., & Wang, J. (2014) · J. Mater. Chem. A · doi:10.1039/c4ta04796f · Cited by 184 · see citing works
Self-citationLee, Y., Howe, C., Mishra, S., Lee, D. S., Mahmood, M., Piper, M., Kim, Y., Tieu, K., Byun, H.-S., Coffey, J. P., et al. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1719573115 · Cited by 183 · see citing works
Yao, H., Liao, Y., Lingley, A. R., Afanasiev, A., Lähdesmäki, I., Otis, B. P., & Parviz, B. A. (2012) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/22/7/075007 · Cited by 182 · see citing works
Caldara, M., Colleoni, C., Guido, E., Re, V., & Rosace, G. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.08.073 · Cited by 181 · see citing works
Stevens, R. C., Soelberg, S. D., Near, S., & Furlong, C. E. (2008) · Analytical Chemistry · doi:10.1021/ac800892h · Cited by 179 · see citing works
Miniature Biofuel Cell as a Potential Power Source for Glucose-Sensing Contact Lenses
Falk, M., Andoralov, V., Silow, M., Toscano, M. D., & Shleev, S. (2013) · Analytical Chemistry · doi:10.1021/ac4006793 · Cited by 173 · see citing works
A paper strip based non-invasive glucose biosensor for salivary analysis
Soni, A., & Jha, S. K. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.09.042 · Cited by 173 · see citing works
Kim, S. B., Zhang, Y., Won, S. M., Bandodkar, A. J., Sekine, Y., Xue, Y., Koo, J., Harshman, S. W., Martin, J. A., Park, J. M., et al. (2018) · Small · doi:10.1002/smll.201703334 · Cited by 167 · see citing works
Self-citationThennadil, S. N., Rennert, J. L., Wenzel, B. J., Hazen, K. H., Ruchti, T. L., & Block, M. B. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/15209150152607132 · Cited by 166 · see citing works
Kownacka, A. E., Vegelyte, D., Joosse, M., Anton, N., Toebes, B. J., Lauko, J., Buzzacchera, I., Lipinska, K., Wilson, D. A., Geelhoed-Duijvestijn, N., et al. (2018) · Biomacromolecules · doi:10.1021/acs.biomac.8b01429 · Cited by 164 · see citing works
Glucose estimation in the salivary secretion of diabetes mellitus patients
Panda, A., Ramesh, & Oza (2012) · Diabetes Metabolic Syndrome and Obesity · doi:10.2147/dmso.s32112 · Cited by 158 · see citing works
初, 明. (2011) · Medical Entomology and Zoology · Cited by 157 · see citing works
The effect of collection technique on tear composition.
Stuchell, R. N., Feldman, J. J., Farris, R. L., & Mandel, I. D. (1984) · PubMed · Cited by 155 · see citing works
Novartis signs up for Google smart lens
Senior, M. (2014) · Nature Biotechnology · doi:10.1038/nbt0914-856 · Cited by 152 · see citing works
Han, W., He, H., Zhang, L., Dong, C., Zeng, H. J., Dai, Y., Xing, L., Zhang, Y., & Xue, X. (2017) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.7b07990 · Cited by 151 · see citing works
An epidermal alkaline rechargeable Ag–Zn printable tattoo battery for wearable electronics
Berchmans, S., Bandodkar, A. J., Jia, W., Ramírez, J., Meng, Y. S., & Wang, J. (2014) · J. Mater. Chem. A · doi:10.1039/c4ta03256j · Cited by 149 · see citing works
Self-citationMunje, R. D., Muthukumar, S., Jagannath, B., & Prasad, S. (2017) · Scientific Reports · doi:10.1038/s41598-017-02133-0 · Cited by 149 · see citing works
Smartphone based non-invasive salivary glucose biosensor
Soni, A., & Jha, S. K. (2017) · Analytica Chimica Acta · doi:10.1016/j.aca.2017.10.003 · Cited by 143 · see citing works
Wearable electrochemical alcohol biosensors
Campbell, A. S., Kim, J., & Wang, J. (2018) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2018.05.014 · Cited by 140 · see citing works
Self-citationPankratov, D., González‐Arribas, E., Blum, Z., & Shleev, S. (2016) · Electroanalysis · doi:10.1002/elan.201501116 · Cited by 135 · see citing works
Wearable organic solar cells with high cyclic bending stability: Materials selection criteria
O’Connor, T. F., Zaretski, A. V., Savagatrup, S., Printz, A. D., Wilkes, C. D., Diaz, M. I., Sawyer, E. J., & Lipomi, D. J. (2016) · Solar Energy Materials and Solar Cells · doi:10.1016/j.solmat.2015.09.049 · Cited by 131 · see citing works
Noninvasively measuring blood glucose using saliva
Yamaguchi, M., Mitsumori, M., & Kano, Y. (1998) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.677170 · Cited by 129 · see citing works
Qiu, Y., Li, G., Hou, Y., Pan, Z., Li, H., Li, W., Liu, M., Ye, F., Yang, X., & Zhang, Y. (2015) · Chemistry of Materials · doi:10.1021/cm503784x · Cited by 125 · see citing works
Piezoelectric crystal biosensors
Ngeh-Ngwainbi, J., Suleiman, A. A., & Guilbault, G. G. (1990) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(90)80023-7 · Cited by 123 · see citing works
Zhang, Y., Zhao, Y., Cheng, X., Weng, W., Ren, J., Fang, X., Jiang, Y., Chen, P., Zhang, Z., Wang, Y., et al. (2015) · Angewandte Chemie International Edition · doi:10.1002/anie.201506142 · Cited by 117 · see citing works
New Developments in the Origins of the Homochirality of Biologically Relevant Molecules
Buschmann, H., Thede, R., & Heller, D. (2000) · Angewandte Chemie International Edition · doi:10.1002/1521-3773(20001117)39:22 · Cited by 111 · see citing works
Li, Q., Lu, X.-F., Xu, H., Tong, Y.-X., & Li, G.-R. (2014) · ACS Applied Materials & Interfaces · doi:10.1021/am405271q · Cited by 107 · see citing works
A realtime and continuous assessment of cortisol in ISF using electrochemical impedance spectroscopy
Venugopal, M., Arya, S. K., Chornokur, G., & Bhansali, S. (2011) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2011.04.028 · Cited by 100 · see citing works
Wearable Potentiometric Chloride Sweat Sensor: The Critical Role of the Salt Bridge
Choi, D.-H., Kim, J. S., Cutting, G. R., & Searson, P. C. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03391 · Cited by 100 · see citing works
Self-citationSialochemistry: A Diagnostic Tool?
Aguirre, A., Testa-Weintraub, L. A., Banderas, J. A., Haraszthy, G. G., Reddy, M. S., & Levine, M. J. (1993) · Critical Reviews in Oral Biology & Medicine · doi:10.1177/10454411930040031201 · Cited by 95 · see citing works
Disposable electrochemiluminescent biosensor for lactate determination in saliva
Ballesta Claver, J., Valencia Mirón, M. C., & Capitán-Vallvey, L. F. (2009) · The Analyst · doi:10.1039/b821922b · Cited by 93 · see citing works
Excretion of Urea, Sodium, Potassium and Chloride in Human Tears
Thaysen, J. H., & Thorn, N. A. (1954) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1954.178.1.160 · Cited by 93 · see citing works
Enzyme-Based Fiber Optic Sensor
Arnold, M. A. (1985) · Analytical Chemistry · doi:10.1021/ac50001a055 · Cited by 92 · see citing works
Xue, X., Qu, Z., Fu, Y., Yu, B., Xing, L., & Zhang, Y. (2016) · Nano Energy · doi:10.1016/j.nanoen.2016.05.021 · Cited by 92 · see citing works
Biomedical soft contact-lens sensor for in situ ocular biomonitoring of tear contents
Chu, M., Shirai, T., Takahashi, D., Arakawa, T., Kudo, H., Sano, K., Sawada, S.-I., Yano, K., Iwasaki, Y., Akiyoshi, K., et al. (2011) · Biomedical Microdevices · doi:10.1007/s10544-011-9530-x · Cited by 85 · see citing works
Noninvasive Method for Glucose Level Estimation by Saliva
Agrawal, R. P. (2013) · Journal of Diabetes & Metabolism · doi:10.4172/2155-6156.1000266 · Cited by 85 · see citing works
Initial Clinical Testing of a Holographic Non-Invasive Contact Lens Glucose Sensor
Domschke, A., March, W. F., Kabilan, S., & Lowe, C. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.89 · Cited by 81 · see citing works
Electrochemical bioaffinity sensors for salivary biomarkers detection
Campuzano, S., Yánez-Sedeño, P., & Pingarrón, J. M. (2017) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2016.10.002 · Cited by 81 · see citing works
Stretchable biofuel cell with enzyme-modified conductive textiles
Ogawa, Y., Takai, Y., Kato, Y., Kai, H., Miyake, T., & Nishizawa, M. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.063 · Cited by 76 · see citing works
Venugopal, M., Feuvrel, K. E., Mongin, D., Bambot, S., Faupel, M., Panangadan, A., Talukder, A., & Pidva, R. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912544 · Cited by 76 · see citing works
Open-source do-it-yourself multi-color fluorescence smartphone microscopy
Sung, Y., Campa, F., & Shih, W.-C. (2017) · Biomedical Optics Express · doi:10.1364/boe.8.005075 · Cited by 76 · see citing works
Recent trends in the development of diagnostic tools for diabetes mellitus using patient saliva
Viswanath, B., Choi, C. S., Lee, K., & Kim, S. (2017) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2017.01.011 · Cited by 75 · see citing works
Handbook of Large-Scale Distributed Computing in Smart Healthcare
Abbas, A., Zomaya, A. Y., & Khan, S. U. (2017) · Scalable computing and communications · doi:10.1007/978-3-319-58280-1 · Cited by 74 · see citing works
A new approach to the assessment of anaerobic metabolism: measurement of lactate in saliva.
Segura, R., Javierre, C., Ventura, J. L., Lizarraga, M. A., Campos, B., & Garrido, E. (1996) · British Journal of Sports Medicine · doi:10.1136/bjsm.30.4.305 · Cited by 71 · see citing works
Cavitas Sensors: Contact Lens Type Sensors & Mouthguard Sensors
Mitsubayashi, K., & Arakawa, T. (2016) · Electroanalysis · doi:10.1002/elan.201600083 · Cited by 71 · see citing works
Disposable electrochemical biosensors
Hilditch, P. I., & Green, M. J. (1991) · The Analyst · doi:10.1039/an9911601217 · Cited by 69 · see citing works
Accuracy of Noninvasive Glucose Sensing Based on Near-Infrared Spectroscopy
Liu, J., Liu, R., & Xu, K. (2015) · Applied Spectroscopy · doi:10.1366/14-07728 · Cited by 63 · see citing works
A Wearable Device for Monitoring Sweat Rates via Image Analysis
Matzeu, G., Fay, C., Vaillant, A., Coyle, S. M., & Diamond, D. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2477676 · Cited by 63 · see citing works
A paper-based skin patch for the diagnostic screening of cystic fibrosis
Mu, X., Xin, X., Fan, C., Li, X., Tian, X., Xu, K., & Zheng, Z. J. (2015) · Chemical Communications · doi:10.1039/c5cc00717h · Cited by 59 · see citing works
Glucose in the saliva of the non-diabetic and the diabetic patient
Campbell, M. J. A. (1965) · Archives of Oral Biology · doi:10.1016/0003-9969(65)90020-8 · Cited by 58 · see citing works
Smartphone-based point-of-care testing of salivary α-amylase for personal psychological measurement.
Zhang, L., Yang, W.-T., Yang, Y.-K., Liu, H., & Gu, Z.-Z. (2015) · In Analysis · doi:10.1039/c5an01664a · Cited by 57 · see citing works
Zhang, K., & Yang, Y. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201703331 · Cited by 57 · see citing works
Materials, Devices and Systems of Soft Bioelectronics for Precision Therapy
Wu, H., Gao, W. F., & Yin, Z. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700017 · Cited by 56 · see citing works
Effects of a 30-km race upon salivary lactate correlation with blood lactate
Santos, R. V. T., Almeida, A. L. R., Caperuto, E. C., Martins, E., & Costa Rosa, L. F. B. P. (2006) · Comparative Biochemistry and Physiology Part B: Biochemistry and Molecular Biology · doi:10.1016/j.cbpb.2006.07.001 · Cited by 51 · see citing works
Sakaguchi, K., Hirota, Y., Hashimoto, N., Ogawa, W., Hamaguchi, T., Matsuo, T., Miyagawa, J.-I., Namba, M., Sato, T., Okada, S., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700313 · Cited by 50 · see citing works
Self‐Healing Inks for Autonomous Repair of Printable Electrochemical Devices
Bandodkar, A. J., Mohan, A. M. V., Silva‐López, C., Ramírez, J., & Wang, J. (2015) · Advanced Electronic Materials · doi:10.1002/aelm.201500289 · Cited by 47 · see citing works
Self-citation磁気レオロジカルエラストマーの周波数,振幅,磁気依存粘弾性のモデリングと実験的検証
Fu-Long, X., Xian-Xu, B., & Qian, L. (2016) · Smart Materials and Structures · Cited by 47 · see citing works
In vivo evaluation of a chip based near infrared sensor for continuous glucose monitoring
Ben Mohammadi, L., Klotzbuecher, T., Sigloch, S., Welzel, K., Göddel, M., Pieber, T. R., & Schaupp, L. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.09.043 · Cited by 46 · see citing works
A disposable biosensor for the determination of alpha-amylase in human saliva
Mahosenaho, M., Caprio, F., Micheli, L., Sesay, A. M., Palleschi, G., & Virtanen, V. (2010) · Microchimica Acta · doi:10.1007/s00604-010-0360-y · Cited by 45 · see citing works
Universal Principles for Ultra Low Power and Energy Efficient Design
Sarpeshkar, R. (2012) · IEEE Transactions on Circuits and Systems II: Express Briefs · doi:10.1109/tcsii.2012.2188451 · Cited by 40 · see citing works
Effect of garment design on piezoelectricity harvesting from joint movement
Yang, J.-H., Cho, H.-S., Park, S.-H., Song, S.-H., Yun, K.-S., & Lee, J. H. (2016) · Smart Materials and Structures · doi:10.1088/0964-1726/25/3/035012 · Cited by 40 · see citing works
Enzyme-Based Fiber Optic Sensor for Glucose Determination
Abdel‐Latif, M. S., Suleiman, A. A., & Guilbault, G. G. (1988) · Analytical Letters · doi:10.1080/00032718808071922 · Cited by 34 · see citing works
Diurnal variation in human tear enzymes
Fullard, R. J., & Carney, L. G. (1984) · Experimental Eye Research · doi:10.1016/0014-4835(84)90134-9 · Cited by 32 · see citing works
Cyclopentane-modified PNA improves the sensitivity of nanoparticle-based scanometric DNA detection
Pokorski, J. K., Nam, J., Vega, R. A., Mirkin, C. A., & Appella, D. H. (2005) · Chemical Communications · doi:10.1039/b418383e · Cited by 30 · see citing works
Noninvasive Photonic-Crystal Material for Sensing Glucose in Tears
Khalil, O. S. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2004.042978 · Cited by 30 · see citing works
A dual microscale glucose sensor on a contact lens, tested in conditions mimicking the eye
Yao, H., Afanasiev, A., Lähdesmäki, I., & Parviz, B. A. (2011) · IEEE International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2011.5734353 · Cited by 29 · see citing works
Use of a new sample-collection device (Macroduct) in anion analysis of human sweat.
Cole, D. E. C., & Boucher, M. J. (1986) · Clinical Chemistry · doi:10.1093/clinchem/32.7.1375 · Cited by 25 · see citing works
A neural algorithm for the non-uniform and adaptive sampling of biomedical data
Mesin, L. (2016) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2016.02.004 · Cited by 22 · see citing works
A CMOS integrated low-voltage low-power time-controlled interface for chemical resistive sensors
De Marcellis, A., Depari, A., Ferri, G., Flammini, A., & Sisinni, E. (2013) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2012.09.104 · Cited by 20 · see citing works
Catecholamine Content of the Lacrimal Fluid of Healthy People and Glaucoma Patients
Zubareva, T. V., & Kiseleva, Z. M. (2010) · Ophthalmologica · doi:10.1159/000308678 · Cited by 20 · see citing works
Salivary Biomarkers-A Review of Powerful Diagnostic tool
Journals, I., Goswami, D. Y., Mishra, R., Agrawal, A., & Agrawal, L. A. (2015) · Figshare · doi:10.6084/m9.figshare.1384292 · Cited by 16 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Singh, M. Non invasive diagnostic tool for pathological conditions salivary biomarkers. Int. J. Pharm. Biol. Arch. 5, 1–12 (2014).
- Imani, S., Mercier, P.P., Bandodkar, A.J., Kim, J. & Wang, J. in 2016 IEEE International Symposium on Circuits and Systems (ISCAS) 1122–1125 (IEEE, 2016).
- Dimitriou, T. & Ioannis, K. in First International Symposium on Applied Sciences on Biomedical and Communication Technologies, 2008 ISABEL’08 1–5 (IEEE, 2008).
Cited by
Sorted by citations
Photothermal Nanomaterials: A Powerful Light-to-Heat Converter
Cui, X., Ruan, Q., Zhuo, X., Xia, X., Hu, J., Fu, R., Li, Y., Wang, J., & Xu, H. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00159 · Cited by 1,985 · see citing works
Self-citationYang, J. C., Mun, J., Kwon, S. Y., Park, S., Bao, Z., & Park, S. (2019) · Advanced Materials · doi:10.1002/adma.201904765 · Cited by 1,641 · see citing works
Technology Roadmap for Flexible Sensors
Luo, Y., Abidian, M. R., Ahn, J., Akinwande, D., Andrews, A. M., Antonietti, M., Bao, Z., Magnus Berggren, A., Berkey, C. A., Bettinger, C. J., et al. (2023) · ACS Nano · doi:10.1021/acsnano.2c12606 · Cited by 1,419 · see citing works
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306 · see citing works
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270 · see citing works
Smart Textiles for Electricity Generation
Chen, G., Li, Y., Bick, M., & Chen, J. (2020) · Chemical Reviews · doi:10.1021/acs.chemrev.9b00821 · Cited by 933 · see citing works
Device integration of electrochemical biosensors
Wu, J., Liu, H., Chen, W., Ma, B., & Ju, H. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00032-w · Cited by 931 · see citing works
A wearable electrochemical biosensor for the monitoring of metabolites and nutrients
Wang, M., Yang, Y., Min, J., Song, Y., Tu, J., Mukasa, D., Ye, C., Xu, C., Heflin, N., McCune, J. S., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00916-z · Cited by 920 · see citing works
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 890 · see citing works
Self-citationSmart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
Wearable and flexible electrochemical sensors for sweat analysis: a review
Gao, F., Liu, C., Zhang, L., Liu, T., Wang, Z., Song, Z., Cai, H., Fang, Z., Chen, J., Wang, J., et al. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00443-6 · Cited by 813 · see citing works
Self-citationSmart textiles for personalized healthcare
Libanori, A., Chen, G., Zhao, X., Zhou, Y., & Chen, J. (2022) · Nature Electronics · doi:10.1038/s41928-022-00723-z · Cited by 756 · see citing works
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. F. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 728 · see citing works
Advances in healthcare wearable devices
Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A., & Asghar, W. (2021) · npj Flexible Electronics · doi:10.1038/s41528-021-00107-x · Cited by 714 · see citing works
“, saliva, blood, urine, sweat, etc.(6,7).”
A wearable biosensing system with in-sensor adaptive machine learning for hand gesture recognition
Moin, A., Zhou, A., Rahimi, A., Menon, A., Benatti, S., Alexandrov, G., Tamakloe, S., Ting, J., Yamamoto, N. A. D., Khan, Y., et al. (2020) · Nature Electronics · doi:10.1038/s41928-020-00510-8 · Cited by 685 · see citing works
Lu, P., Ruan, D., Huang, M., Tian, M., Zhu, K., Gan, Z., & Xiao, Z. (2024) · Signal Transduction and Targeted Therapy · doi:10.1038/s41392-024-01852-x · Cited by 726 · see citing works
Multifunctional materials for implantable and wearable photonic healthcare devices
Lee, G., Moon, H., Kim, H., Lee, G. H., Kwon, W., Yoo, S., Myung, D. J., Yun, S. H., Bao, Z., & Hahn, S. K. (2020) · Nature Reviews Materials · doi:10.1038/s41578-019-0167-3 · Cited by 673 · see citing works
Hydrogel interfaces for merging humans and machines
Yuk, H., Wu, J., & Zhao, X. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00483-4 · Cited by 683 · see citing works
Dai, Y., Somoza, R. A., Wang, L., Welter, J. F., Li, Y., Caplan, A. I., & Liu, C.-C. (2019) · Angewandte Chemie International Edition · doi:10.1002/anie.201910772 · Cited by 660 · see citing works
Shi, Q., Dong, B., He, T., Sun, Z., Zhu, J., Zhang, Z., & Lee, C. (2020) · InfoMat · doi:10.1002/inf2.12122 · Cited by 667 · see citing works
Yu, Y., Nassar, J., Xu, C., Min, J., Yang, Y., Dai, A., Doshi, R., Huang, A., Song, Y., Gehlhar, R., et al. (2020) · Science Robotics · doi:10.1126/scirobotics.aaz7946 · Cited by 655 · see citing works
An epidermal patch for the simultaneous monitoring of haemodynamic and metabolic biomarkers
Sempionatto, J. R., Lin, M., Yin, L., De la paz, E., Pei, K., Sonsa-ard, T., de Loyola Silva, A. N., Khorshed, A. A., Zhang, F., Tostado, N., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00685-1 · Cited by 658 · see citing works
Self-citationWireless battery-free wearable sweat sensor powered by human motion
Song, Y., Min, J., Yu, Y., Wang, H., Yang, Y., Zhang, H., & Gao, W. (2020) · Science Advances · doi:10.1126/sciadv.aay9842 · Cited by 652 · see citing works
Tehrani, F., Teymourian, H., Wuerstle, B. L., Kavner, J., Patel, R., Furmidge, A., Aghavali, R., Hosseini-Toudeshki, H., Brown, C., Zhang, F., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00887-1 · Cited by 651 · see citing works
“Lastly, the stability of the microneedle sensor against mechanical wear and tear, particularly the sensor chemistry, was evaluated by comparing the amperometric response of the glucose sensors in the presence of 5, 10 and 15 mM glucose in synthetic ISF solution under three conditions: (1) before application to the ski…”
BackgroundMethodInfluentialSelf-citationArtificial multimodal receptors based on ion relaxation dynamics
You, I., Mackanic, D. G., Matsuhisa, N., Kang, J., Kwon, J., Beker, L., Mun, J., Suh, W., Kim, T. Y., Tok, J. B.-H., et al. (2020) · Science · doi:10.1126/science.aba5132 · Cited by 632 · see citing works
Functional Fibers and Fabrics for Soft Robotics, Wearables, and Human–Robot Interface
Xiong, J., Chen, J., & Lee, P. S. (2020) · Advanced Materials · doi:10.1002/adma.202002640 · Cited by 640 · see citing works
Wearable chemical sensors for biomarker discovery in the omics era
Sempionatto, J. R., Lasalde-Ramírez, J. A., Mahato, K., Wang, J., & Gao, W. (2022) · Nature Reviews Chemistry · doi:10.1038/s41570-022-00439-w · Cited by 613 · see citing works
Biosensors applications in medical field: A brief review
Haleem, A., Javaid, M., Singh, R. P., Suman, R., & Rab, S. (2021) · Sensors International · doi:10.1016/j.sintl.2021.100100 · Cited by 602 · see citing works
Evolution of Wearable Devices with Real-Time Disease Monitoring for Personalized Healthcare
Guk, K., Han, G., Lim, J., Jeong, K., Kang, T., Lim, E.-K., & Jung, J. (2019) · Nanomaterials · doi:10.3390/nano9060813 · Cited by 624 · see citing works
Scalable production of high-performing woven lithium-ion fibre batteries
He, J., Lu, C., Jiang, H., Han, F., Shi, X., Wu, J., Wang, L., Chen, T., Wang, J., Zhang, Y., et al. (2021) · Nature · doi:10.1038/s41586-021-03772-0 · Cited by 561 · see citing works
Self-citationThe emergence of AI-based wearable sensors for digital health technology: a review
Shajari, S., Kuruvinashetti, K., Komeili, A., & Sundararaj, U. (2023) · Sensors · doi:10.3390/s23239498 · Cited by 535 · see citing works
“Tears contain protein, peptides, lipids, metabolites, and electrolytes [126].”
BackgroundInfluentialMultifunctional conductive hydrogel-based flexible wearable sensors
Wang, L., Xu, T., & Zhang, X. (2021) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116130 · Cited by 507 · see citing works
Digital health for aging populations
Chen, C., Ding, S., & Wang, J. (2023) · Nature Medicine · doi:10.1038/s41591-023-02391-8 · Cited by 525 · see citing works
“These include continuous monitoring of dynamically changing glucose levels in patients with diabetes, potassium ions and the stress hormone cortisol in individuals with cardiac disease or the Parkinson’s disease drug l-DOPA (also known as levodopa) 4 .”
BackgroundSelf-citationFundamentals, Applications, and Future Directions of Bioelectrocatalysis
Chen, H., Simoska, O., Lim, K., Grattieri, M., Yuan, M., Dong, F., Lee, Y. S., Beaver, K., Weliwatte, N. S., Gaffney, E. M., et al. (2020) · Chemical Reviews · doi:10.1021/acs.chemrev.0c00472 · Cited by 501 · see citing works
Shrivastava, S., Trung, T. Q., & Lee, N. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00319c · Cited by 495 · see citing works
Torrente-Rodríguez, R. M., Lukas, H., Tu, J., Min, J., Yang, Y., Xu, C., Rossiter, H. B., & Gao, W. F. (2020) · Matter · doi:10.1016/j.matt.2020.09.027 · Cited by 489 · see citing works
Wireless smart contact lens for diabetic diagnosis and therapy
Keum, D. H., Kim, S.-K., Koo, J., Lee, G.-H., Jeon, C., Mok, J. W., Mun, B. H., Lee, K. J., Kamrani, E., Joo, C.-K., et al. (2020) · Science Advances · doi:10.1126/sciadv.aba3252 · Cited by 483 · see citing works
Shao, Y., Ying, Y. B., & Ping, J. F. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00587k · Cited by 480 · see citing works
Investigation of Cortisol Dynamics in Human Sweat Using a Graphene-Based Wireless mHealth System
Torrente-Rodríguez, R. M., Tu, J., Yang, Y., Min, J., Wang, M., Song, Y., Yu, Y., Xu, C., Ye, C., IsHak, W. W., et al. (2020) · Matter · doi:10.1016/j.matt.2020.01.021 · Cited by 479 · see citing works
A wearable aptamer nanobiosensor for non-invasive female hormone monitoring
Ye, C., Wang, M., Min, J., Tay, R. Y., Lukas, H., Sempionatto, J. R., Li, J., Xu, C., & Gao, W. F. (2023) · Nature Nanotechnology · doi:10.1038/s41565-023-01513-0 · Cited by 469 · see citing works
“In addition, it could also provide skin temperature information continuously via a built-in temperature sensor in the AFE.”
BackgroundZhang, Z., Wen, F., Sun, Z., Guo, X., He, T., & Lee, C. (2022) · Advanced Intelligent Systems · doi:10.1002/aisy.202100228 · Cited by 474 · see citing works
Flexible Wearable Sensors for Cardiovascular Health Monitoring
Chen, S., Qi, J., Fan, S., Qiao, Z., Yeo, J. C., & Lim, C. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100116 · Cited by 453 · see citing works
Dong, K., Peng, X., An, J., Wang, A. C., Luo, J., Sun, B., Wang, J., & Wang, Z. L. (2020) · Nature Communications · doi:10.1038/s41467-020-16642-6 · Cited by 443 · see citing works
Self-citationPiezoelectric nanogenerators for personalized healthcare
Deng, W., Zhou, Y., Libanori, A., Chen, G., Yang, W., & Chen, J. (2022) · Chemical Society Reviews · doi:10.1039/d1cs00858g · Cited by 442 · see citing works
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442 · see citing works
Self-citationSensors for daily life: A review
Javaid, M., Haleem, A., Rab, S., Singh, R. P., & Suman, R. (2021) · Sensors International · doi:10.1016/j.sintl.2021.100121 · Cited by 454 · see citing works
Wearable Electrochemical Sensors for the Monitoring and Screening of Drugs
Teymourian, H., Parrilla, M., Sempionatto, J. R., Montiel, N. F., Barfidokht, A., Echelpoel, R. V., Wael, K. D., & Wang, J. (2020) · ACS Sensors · doi:10.1021/acssensors.0c01318 · Cited by 431 · see citing works
Self-citationNanocomposite hydrogel-based strain and pressure sensors: a review
Sun, X., Yao, F., & Li, J. (2020) · Journal of Materials Chemistry A · doi:10.1039/d0ta06965e · Cited by 422 · see citing works
Biosensor nanoengineering: Design, operation, and implementation for biomolecular analysis
Purohit, B., Vernekar, P. R., Shetti, N. P., & Chandra, P. (2020) · Sensors International · doi:10.1016/j.sintl.2020.100040 · Cited by 425 · see citing works
Skin-inspired soft bioelectronic materials, devices and systems
Zhao, C., Park, J., Root, S. E., & Bao, Z. (2024) · Nature Reviews Bioengineering · doi:10.1038/s44222-024-00194-1 · Cited by 451 · see citing works
“Flexible electrochemical biosensor arrays have been integrated to monitor metabolites (such as glucose and lactose) and ions (such as K + and Cl − ) in real time 2,3,73 (Fig.”
InfluentialMaterials-Driven Soft Wearable Bioelectronics for Connected Healthcare
Gong, S. P., Yan, L., Yin, J., Levin, A., & Cheng, W. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00502 · Cited by 424 · see citing works
Hybrid multimodal wearable sensors for comprehensive health monitoring
Mahato, K., Saha, T., Ding, S., Sandhu, S. S., Chang, A., & Wang, J. (2024) · Nature Electronics · doi:10.1038/s41928-024-01247-4 · Cited by 435 · see citing works
“Modern wearable sensors are an alternative to periodic clinical assessments and can offer continuous non-invasive monitoring of a wide range of physical and biochemical indicators 3 .”
Self-citationWearable aptamer-field-effect transistor sensing system for noninvasive cortisol monitoring
Wang, B., Zhao, C., Wang, Z., Yang, K.-A., Cheng, X., Liu, W., Yu, W., Lin, S., Zhao, Y., Cheung, K. M., et al. (2022) · Science Advances · doi:10.1126/sciadv.abk0967 · Cited by 411 · see citing works
“The components are (1) MCU, (2) ADC, (3) potentiostat chip, (4) digital-to-analog converter, and (5) bluetooth.”
A physicochemical-sensing electronic skin for stress response monitoring
Xu, C., Song, Y., Sempionatto, J. R., Solomon, S. A., Yu, Y., Nyein, H. Y. Y., Tay, R. Y., Li, J., Heng, W., Min, J., et al. (2024) · Nature Electronics · doi:10.1038/s41928-023-01116-6 · Cited by 420 · see citing works
Liu, L., Guo, X., & Lee, C. (2021) · Nano Energy · doi:10.1016/j.nanoen.2021.106304 · Cited by 410 · see citing works
“With the increasing demand of humans monitoring their physical conditions, external devices with wearable and portable designs of biomechanical energy harvesters, have played an increasingly important role in constructing self-powered or self-sustainable systems in smart healthcare [374 – 380].”
BackgroundNaturally sourced hydrogels: emerging fundamental materials for next-generation healthcare sensing
Wang, Z., Wei, H. Z., Huang, Y., Wei, Y., & Chen, J. (2023) · Chemical Society Reviews · doi:10.1039/d2cs00813k · Cited by 400 · see citing works
Artificial intelligence biosensors: Challenges and prospects
Jin, X., Liu, C., Xu, T., Su, L., & Zhang, X. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112412 · Cited by 399 · see citing works
Shirzaei Sani, E., Xu, C., Wang, C., Song, Y., Min, J., Tu, J., Solomon, S. A., Li, J., Banks, J. L., Armstrong, D. G., et al. (2023) · Science Advances · doi:10.1126/sciadv.adf7388 · Cited by 405 · see citing works
A wireless patch for the monitoring of C-reactive protein in sweat
Tu, J., Min, J., Song, Y., Xu, C., Li, J., Moore, J., Hanson, J., Hu, E., Parimon, T., Wang, T.-Y., et al. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01059-5 · Cited by 395 · see citing works
Organic field-effect transistor-based flexible sensors
Yuvaraja, S., Nawaz, A., Liu, Q., Dubal, D. P., Surya, S. G., Salama, K. N., & Sonar, P. M. (2020) · Chemical Society Reviews · doi:10.1039/c9cs00811j · Cited by 388 · see citing works
Reviews of wearable healthcare systems: Materials, devices and system integration
Lou, Z., Wang, L., Jiang, K., Wei, Z., & Shen, G. (2019) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2019.100523 · Cited by 389 · see citing works
Opportunities and Challenges for Biosensors and Nanoscale Analytical Tools for Pandemics: COVID-19
Bhalla, N., Pan, Y., Yang, Z., & Payam, A. F. (2020) · ACS Nano · doi:10.1021/acsnano.0c04421 · Cited by 385 · see citing works
An autonomous wearable biosensor powered by a perovskite solar cell
Min, J., Demchyshyn, S., Sempionatto, J. R., Song, Y., Hailegnaw, B., Xu, C., Yang, Y., Solomon, S. A., Putz, C., Lehner, L. E., et al. (2023) · Nature Electronics · doi:10.1038/s41928-023-00996-y · Cited by 377 · see citing works
“More specifically, the wearable device’s electronic system consists of (1) an energy-harvesting power management integrated circuit (PMIC) that efficiently boosts, converts and manages the output from the quasi-2D FPSC, (2) a compact programmable system-on-chip (PSoC) Bluetooth low-energy (BLE) module that integrates…”
Connected healthcare: Improving patient care using digital health technologies
Awad, A., Trenfield, S. J., Pollard, T. D., Ong, J. J., Elbadawi, M., McCoubrey, L. E., Goyanes, A., Gaisford, S., & Basit, A. W. (2021) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2021.113958 · Cited by 405 · see citing works
All-printed soft human-machine interface for robotic physicochemical sensing
Yu, Y., Li, J., Solomon, S. A., Min, J., Tu, J., Guo, W. H., Xu, C., Song, Y., & Gao, W. F. (2022) · Science Robotics · doi:10.1126/scirobotics.abn0495 · Cited by 367 · see citing works
Age of Flexible Electronics: Emerging Trends in Soft Multifunctional Sensors
Lee, J., Cho, K., & Kim, J. (2024) · Advanced Materials · doi:10.1002/adma.202310505 · Cited by 367 · see citing works
Self-citationYin, L., Cao, M., Kim, K. N., Lin, M., Moon, J., Sempionatto, J. R., Yu, J., Liu, R., Wicker, C., Trifonov, A., et al. (2022) · Nature Electronics · doi:10.1038/s41928-022-00843-6 · Cited by 370 · see citing works
Self-citationFrom Diagnosis to Treatment: Recent Advances in Patient-Friendly Biosensors and Implantable Devices
Li, P., Lee, G., Kim, S. Y., Kwon, S. Y., Kim, H., & Park, S. (2021) · ACS Nano · doi:10.1021/acsnano.0c06688 · Cited by 362 · see citing works
Wearable Biosensors: An Alternative and Practical Approach in Healthcare and Disease Monitoring
Sharma, A., Badea, M., Tiwari, S., & Marty, J. (2021) · Molecules · doi:10.3390/molecules26030748 · Cited by 373 · see citing works
“Recent advancements in wearable contact lens biosensors indicate the use of smartphones for the detection of the analyte of interest in tear fluids [10,193].”
BackgroundInfluentialGoud, K. Y., Moonla, C., Mishra, R. K., Yu, C., Narayan, R. J., Litvan, I., & Wang, J. (2019) · ACS Sensors · doi:10.1021/acssensors.9b01127 · Cited by 358 · see citing works
Self-citationThe Era of Digital Health: A Review of Portable and Wearable Affinity Biosensors
Tu, J., Torrente-Rodríguez, R. M., Wang, M., & Gao, W. F. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201906713 · Cited by 359 · see citing works
Lou, J., Jiao, Y., Yang, R., Huang, Y., Xu, X., Zhang, L., Ma, Z., Yu, Y., Peng, W., Yuan, Y., et al. (2022) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2209218119 · Cited by 356 · see citing works
Liang, C., Ruan, K., Zhang, Y., & Gu, J. (2020) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.0c04482 · Cited by 347 · see citing works
Soft wearable devices for deep-tissue sensing
Lin, M., Hu, H., Zhou, S., & Xu, S. H. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00427-y · Cited by 349 · see citing works
Hydrogels and Their Role in Biosensing Applications
Herrmann, A., Haag, R., & Schedler, U. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100062 · Cited by 339 · see citing works
Graphene-Based Sensors for Human Health Monitoring
Huang, H., Su, S., Wu, N., Wan, H., Wan, S., Bi, H., & Sun, L. (2019) · Frontiers in Chemistry · doi:10.3389/fchem.2019.00399 · Cited by 336 · see citing works
Wireless battery-free body sensor networks using near-field-enabled clothing
Lin, R., Kim, H.-J., Achavananthadith, S., Kurt, S. A., Tan, S. C. C., Yao, H., Tee, B. C. K., Lee, J. K. W., & Ho, J. S. (2020) · Nature Communications · doi:10.1038/s41467-020-14311-2 · Cited by 331 · see citing works
Wearable Soft Technologies for Haptic Sensing and Feedback
Yin, J., Hinchet, R., Shea, H. R., & Majidi, C. (2020) · Advanced Functional Materials · doi:10.1002/adfm.202007428 · Cited by 335 · see citing works
A Wearable Cellulose Acetate-Coated Mouthguard Biosensor for In Vivo Salivary Glucose Measurement
Arakawa, T., Tomoto, K., Nitta, H., Toma, K., Takeuchi, S., Sekita, T., Minakuchi, S., & Mitsubayashi, K. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c01201 · Cited by 320 · see citing works
Kumar, P., Gupta, G. P., & Tripathi, R. (2020) · Computer Communications · doi:10.1016/j.comcom.2020.12.003 · Cited by 317 · see citing works
Zhang, Z., Wang, Z., Zhang, C., Yao, Z., Zhang, S., Wang, R., Tian, Z., Han, J., Chang, C., Lou, J., et al. (2024) · Advanced Materials · doi:10.1002/adma.202308453 · Cited by 317 · see citing works
Technology evolution from self-powered sensors to AIoT enabled smart homes
Dong, B., Shi, Q., Yang, Y., Wen, F., Zhang, Z., & Lee, C. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.105414 · Cited by 316 · see citing works
Sypabekova, M., Hagemann, A., Rho, D., & Kim, S. (2022) · Biosensors · doi:10.3390/bios13010036 · Cited by 313 · see citing works
Functional Fiber Materials to Smart Fiber Devices
Chen, C., Feng, J., Li, J., Guo, Y., Shi, X., & Peng, H. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00192 · Cited by 312 · see citing works
Two-Dimensional Field-Effect Transistor Sensors: The Road toward Commercialization
Dai, C., Liu, Y., & Wei, D. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00924 · Cited by 309 · see citing works
Applied body-fluid analysis by wearable devices
Brasier, N., Wang, J., Gao, W. F., Sempionatto, J. R., Dincer, C., Ates, H. C., Güder, F., Olenik, S., Schauwecker, I., Schaffarczyk, D., et al. (2024) · Nature · doi:10.1038/s41586-024-08249-4 · Cited by 317 · see citing works
Self-citationConstruction of Bio‐Piezoelectric Platforms: From Structures and Synthesis to Applications
Xu, Q., Gao, X., Zhao, S., Liu, Y., Zhang, D., Zhou, K., Khanbareh, H., Chen, W., Zhang, Y., & Bowen, C. (2021) · Advanced Materials · doi:10.1002/adma.202008452 · Cited by 301 · see citing works
Review—An Overview on Recent Progress in Screen-Printed Electroanalytical (Bio)Sensors
Singh, S., Wang, J., & Cinti, S. (2022) · ECS Sensors Plus · doi:10.1149/2754-2726/ac70e2 · Cited by 300 · see citing works
Self-citationA self-sustainable wearable multi-modular E-textile bioenergy microgrid system
Yin, L., Kim, K. N., Lv, J., Tehrani, F., Lin, M., Lin, Z., Moon, J.-M., Ma, J., Yu, J., Xu, S. H., et al. (2021) · Nature Communications · doi:10.1038/s41467-021-21701-7 · Cited by 296 · see citing works
Self-citationXiao, J., Liu, Y., Su, L., Zhao, D., Zhao, L., & Zhang, X. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b03110 · Cited by 294 · see citing works
Lee, I., Probst, D., Klonoff, D. C., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 291 · see citing works
An ultrasensitive CRISPR/Cas12a based electrochemical biosensor for Listeria monocytogenes detection
Li, F., Ye, Q., Chen, M., Zhou, B., Zhang, J., Pang, R., Xue, L., Wang, J., Zeng, H., Wu, S., et al. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113073 · Cited by 291 · see citing works
Self-citationApplications, challenges, and needs for employing synthetic biology beyond the lab
Brooks, S. M., & Alper, H. S. (2021) · Nature Communications · doi:10.1038/s41467-021-21740-0 · Cited by 292 · see citing works
Wearable plasmonic paper–based microfluidics for continuous sweat analysis
Mogera, U., Guo, H., Namkoong, M., Rahman, M. S., Nguyen, N. T., & Tian, L. (2022) · Science Advances · doi:10.1126/sciadv.abn1736 · Cited by 288 · see citing works
Wearable electrochemical biosensors in North America
Min, J., Sempionatto, J. R., Teymourian, H., Wang, J., & Gao, W. F. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112750 · Cited by 280 · see citing works
“However, due to innovative technological and scientific advances in diverse areas, ranging from nano/micro-fabrication methods, materials chemistry, flexible electronics, to bioelectronics, along with the advent of digital communication technology and wireless sensor networks, the field of wearable chemical sensors ha…”
BackgroundSelf-citationRoy, A., Zenker, S., Jain, S., Afshari, R., Oz, Y., Zheng, Y., & Annabi, N. (2024) · Advanced Materials · doi:10.1002/adma.202404225 · Cited by 285 · see citing works
Teymourian, H., Moonla, C., Tehrani, F., Vargas, E., Aghavali, R., Barfidokht, A., Tangkuaram, T., Mercier, P. P., Dassau, E., & Wang, J. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b05109 · Cited by 278 · see citing works
Self-citation3D-printed PEDOT:PSS for soft robotics
Li, J., Cao, J., Lu, B., & Gu, G. (2023) · Nature Reviews Materials · doi:10.1038/s41578-023-00587-5 · Cited by 277 · see citing works
“6a).”
Continuous monitoring of deep-tissue haemodynamics with stretchable ultrasonic phased arrays
Wang, C., Qi, B., Lin, M., Zhang, Z., Makihata, M., Liu, B., Zhou, S., Huang, Y.-H., Hu, H., Gu, Y., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00763-4 · Cited by 274 · see citing works
Reshaping healthcare with wearable biosensors
Smith, A. A., Li, R., & Tse, Z. T. H. (2023) · Scientific Reports · doi:10.1038/s41598-022-26951-z · Cited by 283 · see citing works
- Authors
- Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J.
- Affiliations
- University of California San Diego (US)
- Corresponding author
- Wang, Joseph
- Venue
- Nature Biotechnology
- Published
- 2019-02-25
- Volume
- 37
- Issue
- 4
- Pages
- 389-406
- ISSN
- 0733-222X (ISSN-L)
- 1087-0156
- 1546-1696
- PMC
- PMC8183422
- OpenAlex
- W2917386416
- Cited by
- 3,373
- References
- 196
- Citations by year
- 2026: 424 · 2025: 544 · 2024: 557 · 2023: 511 · 2022: 472 · 2021: 418 · 2020: 340 · 2019: 106
- FWCI
- 26.94
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Nature Biotechnology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute on Alcohol Abuse and Alcoholism (T32AA013525)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Keywords
- wearable biosensors, biological fluids, sweat, saliva, tears, interstitial fluid, electrochemical biosensor, optical biosensors, multiplexed biosensing, flexible materials, real-time physiological monitoring, wearable healthcare
- MeSH terms
- Wearable Electronic Devices, Biotechnology, Body Fluids, Computer Systems, Humans, Iontophoresis, Monitoring, Physiologic, Saliva, Sweat, Tears, Biosensing Techniques, Lab-On-A-Chip Devices
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review