Glucose Sensors and the Alternate Site Testing-Like Phenomenon: Relationship Between Rapid Blood Glucose Changes and Glucose Sensor Signals

Koschinsky, Theodor, Jungheim, Karsten, Heinemann, Lutz · 2003 · Diabetes Technology & Therapeutics
Cited by 64

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Abstract

In the last few years blood glucose meters have been developed allowing glucose measurements in capillary blood samples collected at sites other than the fingertips. The main reason for establishing this so-called alternate site testing (AST) was to sample blood from locations with reduced pain perception. It is well known that capillary blood glucose is closely correlated to systemic (i.e., arterial) glucose levels and that under steady-state conditions, glucose values measured in blood samples collected from alternate sites are virtually identical to those collected from the fingertip. However, during rapid changes in blood glucose levels, glucose concentrations in capillary blood samples from the fingertips can differ considerably in both domains (time and concentration) from those determined in capillary blood from alternate sites (i.e., the so-called AST phenomenon). Such differences can have serious clinical consequences (e.g., risky delays in hypoglycemia detection). There is evidence that all skin sites exhibiting a reduced blood flow (in comparison with the fingertip) within the superficial skin layers are prone to this AST phenomenon. Nearly all glucose sensors having been developed so far or being currently under development measure glucose levels at alternate sites and also in another compartment [e.g., interstitial fluid (ISF)] than blood. So, in principle they might be prone to an AST-like phenomenon (i.e., rapid changes in systemic blood glucose levels may also result in delayed ISF glucose readings). Our knowledge about the impact of an AST-like phenomenon on the performance of glucose monitoring systems is presently very limited. Glucose kinetics in the different compartments during dynamic systemic blood glucose changes have not been fully elucidated yet. If an AST-like phenomenon plays a role with glucose sensors should therefore be studied. Depending on the measurement technology used for the individual type of glucose monitoring system probably this phenomenon has a variable impact on the results obtained.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/152091503322527030

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

See all 64

Metadata

Authors
Koschinsky, T., Jungheim, K., & Heinemann, L.
Venue
Diabetes Technology & Therapeutics
Published
2003-10
Volume
5
Issue
5
Pages
829-842
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2178956461
Cited by
64
References
4
FWCI
3.37
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Glucose Sensors and the Alternate Site Testing-Like Phenomenon: Relationship Between Rapid Blood Glucose Changes and Glucose Sensor Signals. https://app.cochat.ai/papers/10.1089/152091503322527030
The paper
Koschinsky, T., Jungheim, K., & Heinemann, L. (2003). Glucose Sensors and the Alternate Site Testing-Like Phenomenon: Relationship Between Rapid Blood Glucose Changes and Glucose Sensor Signals. Diabetes Technology & Therapeutics. https://doi.org/10.1089/152091503322527030

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.