The Effect of Local Subcutaneous Delivery of Vascular Endothelial Growth Factor on the Function of a Chronically Implanted Amperometric Glucose Sensor
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
The foreign body capsule that forms around implanted devices such as glucose sensors is hypovascular and has limited permeability to glucose. Such a capsule may function better if well vascularized. We hypothesized that capsular vascularization achieved by local release of vascular endothelial growth factor (VEGF) would lead to enhanced function. Amperometric glucose sensor array disks, each with four indicating electrodes, were implanted into rats. Animals received local subcutaneous infusions of VEGF(165) via osmotic pumps at a location on the sensor face 2 mm from one of the electrodes ("near units"). "Intermediate" electrode units were 15 mm, and "distant" units were 22 mm, from the VEGF source. Every 2 weeks, a glucose infusion was given to assess sensor function by telemetry. Near units demonstrated a lower lag duration (delay after blood glucose) than intermediate and distant units. The mean absolute relative difference for near units was less than for distant units. The percentage of data pairs in the A region of the Clarke error grid of the near sensing units was greater than that of the distant units. Values for the functional measures for saline controls fell between near and distant VEGF values. Glucose sensor function was found to be more favorable in units immediately adjacent to the VEGF infusion port. The most likely cause for this finding is increased neovessel growth in the surrounding foreign body capsule. Slow release of angiogenic growth factors may be a potential method for chronically enhancing the function of a subcutaneously implanted biosensor.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K. (2000) · Diabetes Care · doi:10.2337/diacare.23.2.208 · Cited by 135
Cited by
Onuki, Y., Bhardwaj, U., Papadimitrakopoulos, F., & Burgess, D. J. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200610 · Cited by 418
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Ward, W. K. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200504 · Cited by 325
Self-citationThe Foreign Body Response Demystified
Chandorkar, Y., K, R., & Basu, B. (2019) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.8b00252 · Cited by 201
All charged up about implanted biomaterials
Grainger, D. W. (2013) · Nature Biotechnology · doi:10.1038/nbt.2600 · Cited by 186
Metadata
- Authors
- Ward, W. K., Wood, M. D., Casey, H. M., Quinn, M. J., & Federiuk, I. F.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2004-04
- Volume
- 6
- Issue
- 2
- Pages
- 137-145
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2068615272
- Cited by
- 39
- References
- 1
- FWCI
- 2.62
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Effect of Local Subcutaneous Delivery of Vascular Endothelial Growth Factor on the Function of a Chronically Implanted Amperometric Glucose Sensor. https://app.cochat.ai/papers/10.1089/152091504773731320
- The paper
- Ward, W. K., Wood, M. D., Casey, H. M., Quinn, M. J., & Federiuk, I. F. (2004). The Effect of Local Subcutaneous Delivery of Vascular Endothelial Growth Factor on the Function of a Chronically Implanted Amperometric Glucose Sensor. Diabetes Technology & Therapeutics. https://doi.org/10.1089/152091504773731320
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.