A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Open accessAbstract
OBJECTIVE: To evaluate the lifetime, response time, linearity, glucose range, and calibration stability of two different types of continuous glucose sensor implants in a dog model. RESEARCH DESIGN AND METHODS: Glucose sensors based on the enzyme electrode principle that are coupled to a radio transmitter were evaluated on the bench top, sterilized, and then implanted subcutaneously in nondiabetic mongrel dogs. A multichannel radio receiver and PC data processor were used to record the sensor glucose data. Initial early reliable sensor responsivity was recognized by a vigorous hyperglycemic excursion after an intramuscular injection of glucagon. Periodically the dogs were made temporarily diabetic by blocking pancreatic insulin secretion by subcutaneous injection of a synthetic somatostatin (octreotide). By using exogenous insulin injection followed by intravenous glucose infusion, glucose levels were manipulated through the entire clinical range of interest: 2.2-38.9 mmol/l (40-700 mg/dl). Every 5-10 min, reference blood glucose samples were obtained and run in our hospital clinical laboratory. The glucose sensor data was evaluated by linear least squares optimization and by the error grid method. RESULTS: Beginning as early as postimplant day 7, the in vivo performances of sensors were evaluated by using glucose infusion studies performed every 1-4 weeks. Bench-top and in vivo 90% response-time sensors were in the range of 4-7 min during sensor lifetime. Best-performing sensors from both types are summarized as follows. The earlier-stage technology was less linear with a dynamic range of no more than 22 mmol/l glucose, had a best-case recalibration interval of 18 days, and had a maximum lifetime of 94 days. The improved later-stage technology sensors, which were constructed with the addition of bioprotective and angiogenic membranes, were linear over the full extended range of clinical interest (2.2-38.9 mmol/l [40-700 mg/dl glucose]), had a best-case recalibration interval of 20 days, and had a maximum lifetime of >160 days. CONCLUSIONS: Stable clinically useful sensor performance was demonstrated as early as 7 days after implantation and for a sensor lifetime of 3-5 months. This type of subcutaneous glucose sensor appears to be promising as a continuous and painless long-term method for monitoring blood glucose. Specifically sensors with top-layer materials that stimulate angiogenesis at the sensor/tissue interface may have better dynamic measurement range, longer lifetimes, and better calibration stability than our previously reported sensors.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Sensors lacking the ePTFE membrane were functional for approximately 94 days, with a best-case recalibration interval of 18 days.
“sensors lacking the ePTFE membrane (i.e., having only an underlying layer of unspecified composition) were functional for approximately 94 d with a best-case recalibration interval of 18 d.156”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [156]
Biosensors with an ePTFE membrane having 1-10 μm pores implanted in non-diabetic canines were operable for more than 160 days, with a best-case recalibration interval of 20 days.
“Updike and coworkers implanted biosensors with an affixed expanded PTFE (ePTFE) membrane having 1–10 μm pores in non-diabetic canines and found they were operable for greater than 160 d with a best-case recalibration interval of 20 d.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [156]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Evaluation of a Subcutaneous Glucose Sensor out to 3 Months in a Dog Model
GILLIGAN, B. J., Shults, M. C., Rhodes, R. K., & Updike, S. J. (1994) · Diabetes Care · doi:10.2337/diacare.17.8.882 · Cited by 90 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707 · see citing works
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Concurrent delivery of dexamethasone and VEGF for localized inflammation control and angiogenesis
Patil, S. D., Papadmitrakopoulos, F., & Burgess, D. J. (2007) · Journal of Controlled Release · doi:10.1016/j.jconrel.2006.10.013 · Cited by 230 · see citing works
Koschwanez, H. E., & Reichert, W. M. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.034 · Cited by 181 · see citing works
Ward, W. K., Slobodzian, E. P., Tiekotter, K. L., & Wood, M. D. (2002) · Biomaterials · doi:10.1016/s0142-9612(02)00160-6 · Cited by 176 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500317 · Cited by 125 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500318 · Cited by 102 · see citing works
Feasibility of Continuous Long-Term Glucose Monitoring from a Subcutaneous Glucose Sensor in Humans
Gilligan, B. C., Shults, M., Rhodes, R. K., Jacobs, P. G., Brauker, J. H., Pintar, T. J., & Updike, S. J. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774198089 · Cited by 71 · see citing works
Self-citationKoschwanez, H. E., Yap, F. Y., Klitzman, B., & Reichert, W. M. (2008) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31824 · Cited by 69 · see citing works
A dexamethasone‐loaded PLGA microspheres/collagen scaffold composite for implantable glucose sensors
Ju, Y. M., Yu, B., West, L., Moussy, Y., & Moussy, F. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32512 · Cited by 44 · see citing works
Intravital microscopy evaluation of angiogenesis and its effects on glucose sensor performance
Koschwanez, H. E., Reichert, W. M., & Klitzman, B. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32630 · Cited by 15 · see citing works
Percutaneous Window Chamber Method for Chronic Intravital Microscopy of Sensor—Tissue Interactions
Koschwanez, H. E., Klitzman, B., & Reichert, W. M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200608 · Cited by 9 · see citing works
- Authors
- Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K.
- Affiliations
- University of Wisconsin–Madison (US)
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2000-02-01
- Volume
- 23
- Issue
- 2
- Pages
- 208-214
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2119535003
- Cited by
- 135
- References
- 1
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 1 · 2023: 1 · 2022: 1 · 2021: 1 · 2020: 3 · 2018: 5 · 2017: 1 · 2016: 7 · 2015: 2 · 2014: 3 · 2013: 5 · 2012: 2
- FWCI
- 5.56
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5r44dk048611-03)
- National Institute of Diabetes and Digestive and Kidney Diseases (2R44-DK-48611)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Surgery › Pancreatic function and diabetes
- Keywords
- subcutaneous glucose sensor, continuous glucose monitoring, enzyme electrodes, sensor lifetime, calibration stability, dynamic range, sensor response time, error grid analysis, sensor linearity, dogs, biomedical implants, diabetes management
- MeSH terms
- Animals, Blood Glucose, Calibration, Dogs, Glucagon, Humans, Monitoring, Physiologic, Prosthesis Design, Regression Analysis, Biosensing Techniques, Prostheses and Implants
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review