BACKGROUND: The accuracy and efficacy of the Medtronic Diabetes (Northridge, CA) Real-Time (RT)-Continuous Glucose Monitoring (CGM) sensor were analyzed in 72 subjects with type 1 diabetes. METHODS: This was a retrospective analysis of 60,050 temporally paired data points (sensor and glucose meter values) obtained during the course of an outpatient ambulatory study evaluating the efficacy of a sensor-augmented pump system in adults and adolescents. Subjects uploaded sensor values and self-monitoring blood glucose data to the CareLink Clinical Application (Medtronic Diabetes) via the Internet, every 2 weeks during the course of the study. RESULTS: The overall percentage of sensor readings within +/-20% or +/-30% agreement of reference glucose readings was 75.6% and 86.8%, respectively. The highest rate of agreement occurred in the 240-400 mg/dL range, where 79.9% of sensor readings were within +/-20% of meter values and 91.5% of sensor readings were within 30% of meter values. The mean absolute relative difference for all subjects was 15.8%, and the median absolute relative difference was 10.9%. The bias was -2.13 mg/dL. Paired glucose measurements from the RT-CGM and meter demonstrated that 95.9% of paired points in the overall subject population fell in zones A and B of the Clarke Error Grid. Consensus Error Grid Analysis established that 99.2% of paired data points were in zones A and B. CONCLUSIONS: This study reports the accuracy of a continuous glucose sensor with a large number of paired data points (60,050). RT-CGM is safe and well tolerated and provides readings that are in close agreement with glucose meter values.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2007.0291

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 118

Metadata

Authors
Mastrototaro, J., Shin, J., Marcus, A., Sulur, G., & for the STAR 1 Clinical Trial Investigators
Venue
Diabetes Technology & Therapeutics
Published
2008-10
Volume
10
Issue
5
Pages
385-390
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W1963833218
Cited by
118
References
1
FWCI
8.28
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.