Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges
Open accessAbstract
Continuous glucose monitoring (CGM) sensors are portable devices that allow measuring and visualizing the glucose concentration in real time almost continuously for several days and are provided with hypo/hyperglycemic alerts and glucose trend information. CGM sensors have revolutionized Type 1 diabetes (T1D) management, improving glucose control when used adjunctively to self-monitoring blood glucose systems. Furthermore, CGM devices have stimulated the development of applications that were impossible to create without a continuous-time glucose signal, e.g., real-time predictive alerts of hypo/hyperglycemic episodes based on the prediction of future glucose concentration, automatic basal insulin attenuation methods for hypoglycemia prevention, and the artificial pancreas. However, CGM sensors’ lack of accuracy and reliability limited their usability in the clinical practice, calling upon the academic community for the development of suitable signal processing methods to improve CGM performance. The aim of this paper is to review the past and present algorithmic challenges of CGM sensors, to show how they have been tackled by our research group, and to identify the possible future ones.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 The smart CGM sensor architecture, which consists of placing, in cascade to the output of a commercial CGM sensor, three software modules, able to work in real time, for denoising the random noise component, enhancing the accuracy, and predicting the future glucose concentration (adapted from [27]).](https://figures.cochat.ai/paper-figures/c1dcdb45c40842afa0dbc799604b1883/continuous-glucose-monitoring-sensors-past-present-and-future-algorithmic-figure-2.jpg)

Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The accuracy of CGM devices has improved significantly over time, with manufacturers of current devices reporting mean absolute relative deviation (MARD) proportions of 8.1% to 12.3%, compared with 5% to 10% for current BGM devices (and 22% for the first intermittently read interstitial glucose monitor brought to market in 2001) (190).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [190]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Bolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893
The UVA/PADOVA Type 1 Diabetes Simulator: New Features
Man, C. D., Micheletto, F., Lv, D., Breton, M., Kovatchev, B., & Cobelli, C. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813514502 · Cited by 791 · cited 1×
“The UVA/Padova T1D simulator is a large-scale maximal computer model of glucose, insulin and glucagon dynamics in patients with T1D, described by 13 differential equations and provided with three different populations: 100 adult, 100 pediatric and 100 adolescent virtual subjects [52,53].”
BackgroundIn Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760 · cited 1×
“The UVA/Padova T1D simulator is a large-scale maximal computer model of glucose, insulin and glucagon dynamics in patients with T1D, described by 13 differential equations and provided with three different populations: 100 adult, 100 pediatric and 100 adolescent virtual subjects [52,53].”
BackgroundThe Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685 · cited 1×
“The reduction/elimination of calibrations in CGM systems is further pressured by the recent advent of the Abbott Freestyle Libre (Abbott Diabetes Care, Alameda, CA, USA) [62,63].”
MethodSparacino, G., Zanderigo, F., Corazza, S., Maran, A., Facchinetti, A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889774 · Cited by 338 · cited 1×
“Finally, the algorithm we employed in the prediction module is a simple but effective predictor based on an autoregressive model of order one [40], whose key feature is the real-time estimation of the predictor’s parameters using a recursive least squares implementation, exploiting a forgetting…”
MethodSelf-citationPerformance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home Use
Gross, T. M., Bode, B. W., Einhorn, D., Kayne, D. M., Reed, J. H., White, N. H., & Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/152091500316737 · Cited by 303
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234 · cited 2×
“Only Kovatchev et al. [47] explicitly addressed this problem and tried to give some indications on this topic.”
BackgroundComing of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223
Georga, E. I., Protopappas, V. C., Ardigo, D., Marina, M., Zavaroni, I., Polyzos, D., & Fotiadis, D. I. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/titb.2012.2219876 · Cited by 216 · cited 1×
“…can be employed, e.g., those of Zhao et al. [9,12], Zecchin et al. [13,41], Turksoy et al. [10], Zarkogianni et al. [11], and Georga et al. [42,43] The algorithmically “smart” CGM sensor concept has been validated on a dataset consisting of 23 T1D individuals whose glucose concentrations…”
Method
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
“As a result, these and many other unsuccessful outcomes have left many wondering if it is possible to measure glucose in a way that doesn’t need any blood drawn out of the body.”
BackgroundMethodCappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477
Self-citationWearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442
Wearable Electrochemical Sensors for the Monitoring and Screening of Drugs
Teymourian, H., Parrilla, M., Sempionatto, J. R., Montiel, N. F., Barfidokht, A., Van Echelpoel, R., De Wael, K., & Wang, J. (2020) · ACS Sensors · doi:10.1021/acssensors.0c01318 · Cited by 423
Metadata
- Authors
- Facchinetti, A.
- Affiliations
- University of Padua (IT)
- Corresponding author
- Facchinetti, Andrea
- ORCID
- Facchinetti, Andrea
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2016-12-09
- Volume
- 16
- Issue
- 12
- Pages
- 2093
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC5191073
- OpenAlex
- W2560030487
- Cited by
- 176
- References
- 68
- Citations by year
- 2026: 14 · 2025: 16 · 2024: 8 · 2023: 17 · 2022: 17 · 2021: 16 · 2020: 22 · 2019: 25 · 2018: 19 · 2017: 22
- FWCI
- 3.24
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Padua Research Archive (University of Padova) (submittedVersion, cc-by)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- continuous glucose monitoring, type 1 diabetes, glucose prediction, hypoglycemia prediction, artificial pancreas, signal processing algorithms, sensor accuracy and reliability
- MeSH terms
- Clinical Decision-Making, Algorithms, Diabetes Mellitus, Type 1, Humans, Software, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges. https://app.cochat.ai/papers/10.3390/s16122093
- The paper
- Facchinetti, A. (2016). Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges. Sensors. https://doi.org/10.3390/s16122093
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.