OBJECTIVE: Nocturnal hypoglycemia is a common problem with type 1 diabetes. In the home setting, we conducted a pilot study to evaluate the safety of a system consisting of an insulin pump and continuous glucose monitor communicating wirelessly with a bedside computer running an algorithm that temporarily suspends insulin delivery when hypoglycemia is predicted. RESEARCH DESIGN AND METHODS: After the run-in phase, a 21-night randomized trial was conducted in which each night was randomly assigned 2:1 to have either the predictive low-glucose suspend (PLGS) system active (intervention night) or inactive (control night). Three predictive algorithm versions were studied sequentially during the study for a total of 252 intervention and 123 control nights. The trial included 19 participants 18-56 years old with type 1 diabetes (hemoglobin A1c level of 6.0-7.7%) who were current users of the MiniMed Paradigm® REAL-Time Revel™ System and Sof-sensor® glucose sensor (Medtronic Diabetes, Northridge, CA). RESULTS: With the final algorithm, pump suspension occurred on 53% of 77 intervention nights. Mean morning glucose level was 144±48 mg/dL on the 77 intervention nights versus 133±57 mg/dL on the 37 control nights, with morning blood ketones >0.6 mmol/L following one intervention night. Overnight hypoglycemia was lower on intervention than control nights, with at least one value ≤70 mg/dL occurring on 16% versus 30% of nights, respectively, with the final algorithm. CONCLUSIONS: This study demonstrated that the PLGS system in the home setting is safe and feasible. The preliminary efficacy data appear promising with the final algorithm reducing nocturnal hypoglycemia by almost 50%.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2013.0040

Cites

6 of 6 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 6 references

Cited by

See all 102

Metadata

Authors
Buckingham, B. A., Cameron, F., Calhoun, P., Maahs, D. M., Wilson, D. M., Chase, H. P., Bequette, B. W., Lum, J., Sibayan, J., Beck, R. W., & Kollman, C.
Venue
Diabetes Technology & Therapeutics
Published
2013-08
Volume
15
Issue
8
Pages
622-627
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W1967760785
Cited by
102
References
6
FWCI
10.14
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Outpatient Safety Assessment of an In-Home Predictive Low-Glucose Suspend System with Type 1 Diabetes Subjects at Elevated Risk of Nocturnal Hypoglycemia. https://app.cochat.ai/papers/10.1089/dia.2013.0040
The paper
Buckingham, B. A., Cameron, F., Calhoun, P., Maahs, D. M., Wilson, D. M., Chase, H. P., Bequette, B. W., Lum, J., Sibayan, J., Beck, R. W., & Kollman, C. (2013). Outpatient Safety Assessment of an In-Home Predictive Low-Glucose Suspend System with Type 1 Diabetes Subjects at Elevated Risk of Nocturnal Hypoglycemia. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2013.0040

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.