Reduction in Duration of Hypoglycemia by Automatic Suspension of Insulin Delivery: The In-Clinic ASPIRE Study
Abstract
Background The efficacy of automatic suspension of insulin delivery in induced hypoglycemia among subjects with type 1 diabetes was evaluated. Subjects and methods In this randomized crossover study, subjects used a sensor-augmented insulin pump system with a low glucose suspend (LGS) feature that automatically stops insulin delivery for 2 h following a sensor glucose (SG) value ≤70 mg/dL. Subjects fasted overnight and exercised until their plasma glucose (measured with the YSI 2300 STAT Plus™ glucose and lactate analyzer [YSI Life Sciences, Yellow Springs, OH]) value reached ≤85 mg/dL on different occasions separated by washout periods lasting 3-10 days. Exercise sessions were done with the LGS feature turned on (LGS-On) or with continued insulin delivery regardless of SG value (LGS-Off). The order of LGS-On and LGS-Off sessions was randomly assigned. YSI glucose data were used to compare the duration and severity of hypoglycemia from successful LGS-On and LGS-Off sessions and to estimate the risk of rebound hyperglycemia after pump suspension. Results Fifty subjects attempted 134 sessions, 98 of which were successful. The mean±SD hypoglycemia duration was less during LGS-On than during LGS-Off sessions (138.5±76.68 vs. 170.7±75.91 min, P=0.006). During LGS-On compared with LGS-Off sessions, mean nadir YSI glucose was higher (59.5±5.72 vs. 57.6±5.69 mg/dL, P=0.015), as was mean end-observation YSI glucose (91.4±41.84 vs. 66.2±13.48 mg/dL, P 250 mg/dL. Conclusions Automatic s…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
When comparing the safety and efficacy of overnight closed-loop delivery of insulin with conventional insulin-pump therapy in adults with type 1 diabetes, closed-loop delivery improved overnight control of glycaemia and reduced the risk of nocturnal hypoglycaemia.
“when comparing the safety and efficacy of overnight closed-loop delivery of insulin with conventional insulin-pump therapy in adults with type 1 diabetes, closed-loop delivery improved overnight control of glycaemia and reduced the risk of nocturnal hypoglycaemia.98,99”
From Type 1 diabetes (2014) · cites this paper as [98]
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
Mechanisms of Hypoglycemia-Associated Autonomic Failure and Its Component Syndromes in Diabetes
Cryer, P. E. (2005) · Diabetes · doi:10.2337/diabetes.54.12.3592 · Cited by 402 · see citing works
Insulin Pump Therapy With Automated Insulin Suspension in Response to Hypoglycemia
Choudhary, P., Shin, J., Wang, Y., Evans, M. L., Hammond, P. J., Kerr, D., Shaw, J. A. M., Pickup, J. C., & Amiel, S. A. (2011) · Diabetes Care · doi:10.2337/dc10-2411 · Cited by 175 · see citing works
Self-citationPrevention of Hypoglycemia by Using Low Glucose Suspend Function in Sensor-Augmented Pump Therapy
Danne, T., Kordonouri, O., Holder, M., Haberland, H., Golembowski, S., Remus, K., Bläsig, S., Wadien, T., Zierow, S., Hartmann, R., et al. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0084 · Cited by 150 · see citing works
Duration of Nocturnal Hypoglycemia Before Seizures
Buckingham, B. A., Wilson, D. M., Lecher, T., Hanås, R., Kaiserman, K. B., & Cameron, F. J. (2008) · Diabetes Care · doi:10.2337/dc08-0863 · Cited by 111 · see citing works
Self-citationAgrawal, P. K., Welsh, J. B., Kannard, B., Askari, S., Yang, Q., & KAUFMAN, F. R. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500514 · Cited by 86 · see citing works
Self-citationLy, T. T., Hewitt, J., Davey, R. J., Lim, E. M., Davis, E. A., & Jones, T. W. (2011) · Diabetes Care · doi:10.2337/dc10-1042 · Cited by 81 · see citing works
Hypoglycemia Begets Hypoglycemia in IDDM
Cryer, P. E. (1993) · Diabetes · doi:10.2337/diab.42.12.1691 · Cited by 67 · see citing works
Brazg, R., Bailey, T. S., Garg, S. K., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Nguyen, X., Shin, J., Welsh, J. B., & Lee, S. W. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500621 · Cited by 28 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Atkinson, M. A., Eisenbarth, G. S., & Michels, A. W. (2014) · The Lancet · doi:10.1016/s0140-6736(13)60591-7 · Cited by 2,713 · see citing works
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652 · see citing works
Self-citationLy, T. T., Nicholas, J. A., Retterath, A., Lim, E. M., Davis, E. A., & Jones, T. W. (2013) · JAMA · doi:10.1001/jama.2013.277818 · Cited by 411 · see citing works
Insulin delivery methods: Past, present and future
Shah, R., Shah, V. N., Patel, M., & Maahs, D. M. (2016) · International Journal of Pharmaceutical Investigation · doi:10.4103/2230-973x.176456 · Cited by 289 · see citing works
Grunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288 · see citing works
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197 · see citing works
Adolfsson, P., Riddell, M. C., Taplin, C. E., Davis, E. A., Fournier, P. A., Annan, F., Scaramuzza, A. E., Hasnani, D., & Hofer, S. E. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12755 · Cited by 191 · see citing works
Challenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 182 · see citing works
Pozzilli, P., Battelino, T., Danne, T., Hovorka, R., Jarosz-Chobot, P. A., & Renard, É. (2015) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2653 · Cited by 162 · see citing works
Buckingham, B. A., Raghinaru, D., Cameron, F., Bequette, B. W., Chase, H. P., Maahs, D. M., Slover, R. H., Wadwa, R. P., Wilson, D. M., Ly, T. T., et al. (2015) · Diabetes Care · doi:10.2337/dc14-3053 · Cited by 138 · see citing works
Self-citationBosi, E., Choudhary, P., de Valk, H. W., Lablanche, S., Castañeda, J., de Portu, S., Silva, J. D., Ré, R., Vorrink, L., Shin, J., et al. (2019) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(19)30150-0 · Cited by 123 · see citing works
Self-citationRoutine Sensor-Augmented Pump Therapy in Type 1 Diabetes: The INTERPRET Study
Nørgaard, K., Scaramuzza, A. E., Bratina, N., Lalic, N. M., Jarosz-Chobot, P. A., Kocsis, G., Jašinskienė, E., Block, C. E. M. D., Carrette, O., Castañeda, J., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0288 · Cited by 107 · see citing works
Real-Time Improvement of Continuous Glucose Monitoring Accuracy
Facchinetti, A., Sparacino, G., Guerra, S., Luijf, Y. M., DeVries, J. H., Mader, J. K., Ellmerer, M., Benesch, C., Heinemann, L., Bruttomesso, D., et al. (2012) · Diabetes Care · doi:10.2337/dc12-0736 · Cited by 106 · see citing works
Buckingham, B. A., Cameron, F., Calhoun, P. M., Maahs, D. M., Wilson, D. M., Chase, H. P., Bequette, B. W., Lum, J. W., Sibayan, J., Beck, R. W., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0040 · Cited by 103 · see citing works
Self-citationExercise in children and adolescents with diabetes
Robertson, K. J., Riddell, M. C., Guinhouya, B. C., Adolfsson, P., & Hanås, R. (2014) · Pediatric Diabetes · doi:10.1111/pedi.12176 · Cited by 99 · see citing works
Susceptibility of Interstitial Continuous Glucose Monitor Performance to Sleeping Position
Mensh, B. D., Wisniewski, N. A., Neil, B. M., & Burnett, D. R. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700408 · Cited by 97 · see citing works
Current perspectives on physical activity and exercise for youth with diabetes
Pivovarov, J. A., Taplin, C. E., & Riddell, M. C. (2015) · Pediatric Diabetes · doi:10.1111/pedi.12272 · Cited by 96 · see citing works
Danne, T., Tsioli, C., Kordonouri, O., Bläsig, S., Remus, K., Roy, A., Keenan, D. B., Lee, S. W., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0327 · Cited by 91 · see citing works
Self-citationZhong, A., Choudhary, P., McMahon, C., Agrawal, P. K., Welsh, J. B., Cordero, T. L., & KAUFMAN, F. R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0216 · Cited by 77 · see citing works
Self-citationContinuous Glucose Monitoring: Evidence and Consensus Statement for Clinical Use
Liebl, A., Henrichs, H. R., Heinemann, L., Freckmann, G., Biermann, E., & Thomas, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700227 · Cited by 76 · see citing works
Prediction and prevention of hypoglycaemic events in type-1 diabetic patients using machine learning
Vehí, J., Contreras, I., Oviedo, S., Biagi, L., & Bertachi, A. (2020) · Health Informatics Journal · doi:10.1177/1460458219850682 · Cited by 76 · see citing works
Zaharieva, D. P., Messer, L. H., Paldus, B., O'Neal, D. N., Maahs, D. M., & Riddell, M. C. (2020) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2020.06.003 · Cited by 75 · see citing works
Closed-Loop System in the Management of Diabetes: Past, Present, and Future
Shah, V. N., Shoskes, A., Tawfik, B., & Garg, S. K. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0193 · Cited by 74 · see citing works
Self-citationA review of artificial pancreas technologies with an emphasis on bi‐hormonal therapy
Bakhtiani, P. A., Zhao, L., Youssef, J. E., Castle, J. R., & Ward, W. K. (2013) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12107 · Cited by 65 · see citing works
Improved Real-World Glycemic Control With Continuous Glucose Monitoring System Predictive Alerts
Abraham, S. B., Arunachalam, S., Zhong, A., Agrawal, P., Cohen, O., & McMahon, C. M. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819859334 · Cited by 63 · see citing works
Hypoglycemia Reduction and Changes in Hemoglobin A1c in the ASPIRE In-Home Study
Weiss, R., Garg, S. K., Bode, B. W., Bailey, T. S., Ahmann, A., Schultz, K. A., Welsh, J. B., Shin, J., & for the ASPIRE In-Home Study Group (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0306 · Cited by 56 · see citing works
Self-citationSensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes
Steineck, I. I. K., Ranjan, A. G., Nørgaard, K., & Schmidt, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672689 · Cited by 57 · see citing works
Defective Counterregulation and Hypoglycemia Unawareness in Diabetes
Reno, C. M., Litvin, M., Clark, A. L., & Fisher, S. J. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2012.11.005 · Cited by 53 · see citing works
Khalil, A., Beshyah, S., Awad, S. A., Benbarka, M., Haddad, M., Al-Hassan, D., Kahwatih, M., & Nagelkerke, N. J. D. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0061 · Cited by 53 · see citing works
Evaluation of a Predictive Low-Glucose Management System In-Clinic
Buckingham, B. A., Bailey, T. S., Christiansen, M. P., Garg, S. K., WEINZIMER, S. A., Bode, B. W., Anderson, S. M., Brazg, R., Ly, T. T., & KAUFMAN, F. R. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0319 · Cited by 52 · see citing works
Self-citationVettoretti, M., & Facchinetti, A. (2019) · BioMedical Engineering OnLine · doi:10.1186/s12938-019-0658-x · Cited by 52 · see citing works
Lane, J. E., Shivers, J. P., & Zisser, H. (2013) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0b013e32835edb9d · Cited by 50 · see citing works
The Impact of Accelerometer and Heart Rate Data on Hypoglycemia Mitigation in Type 1 Diabetes
Stenerson, M. B., Cameron, F., Wilson, D. M., Harris, B., Payne, S. R., Bequette, B. W., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813516208 · Cited by 48 · see citing works
Self-citationEmerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 48 · see citing works
Self-citationvan Beers, C. A. J., & DeVries, J. H. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816653411 · Cited by 40 · see citing works
Agrawal, P. K., Zhong, A., Welsh, J. B., Shah, R. V., & KAUFMAN, F. R. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0257 · Cited by 40 · see citing works
Self-citationAbraham, M. B., Davey, R. J., O'Grady, M. J., Ly, T. T., Paramalingam, N., Fournier, P. A., Roy, A., Grosman, B., Kurtz, N., Fairchild, J. M., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0141 · Cited by 40 · see citing works
Social Determinants of Health, Goals and Outcomes in High-Risk Children With Type 1 Diabetes
Hershey, J. A., Morone, J. F., Lipman, T. H., & Hawkes, C. P. (2021) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2021.02.005 · Cited by 40 · see citing works
Elbarbary, N. S. (2016) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2781 · Cited by 37 · see citing works
A Peek Under the Hood: Explaining the MiniMed™ 780G Algorithm with Meal Detection Technology
Grosman, B., Roy, A., LINTEREUR, L. J., Turksoy, K., Benedetti, A., Cordero, T. L., Vigersky, R. A., MCVEAN, J. J. F., Rhinehart, A. S., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0446 · Cited by 36 · see citing works
Technology to Optimize Pediatric Diabetes Management and Outcomes
Markowitz, J. T., Harrington, K. R., & Laffel, L. M. (2013) · Current Diabetes Reports · doi:10.1007/s11892-013-0419-3 · Cited by 35 · see citing works
Sherr, J. L., Palau Collazo, M., Cengiz, E., Michaud, C., Carria, L., Steffen, A. T., Weyman, K., Zgorski, M., Tichy, E., Tamborlane, W. V., et al. (2014) · Diabetes Care · doi:10.2337/dc13-1608 · Cited by 35 · see citing works
Abraham, M. B., de Bock, M. I., Paramalingam, N., O'Grady, M. J., Ly, T. T., George, C., Roy, A., Spital, G., Karula, S., Heels, K., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0364 · Cited by 34 · see citing works
PAULEY, M. E., Tommerdahl, K. L., Snell‐Bergeon, J. K., & Forlenza, G. P. (2022) · Current Cardiology Reports · doi:10.1007/s11886-022-01799-x · Cited by 36 · see citing works
Hanazaki, K., Munekage, M., Kitagawa, H., Yatabe, T., Munekage, E., Shiga, M., Maeda, H., & Namikawa, T. (2016) · Journal of Artificial Organs · doi:10.1007/s10047-016-0904-y · Cited by 33 · see citing works
The Impact of Accelerometer Use in Exercise-Associated Hypoglycemia Prevention in Type 1 Diabetes
Stenerson, M. B., Cameron, F., Payne, S. R., Payne, S. L., Ly, T. T., Wilson, D. M., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814551045 · Cited by 31 · see citing works
Self-citationUtility of Continuous Glucose Monitoring in Type 1 and Type 2 Diabetes
Toschi, E., & Wolpert, H. A. (2016) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2016.06.003 · Cited by 32 · see citing works
Katz, M. L., Giani, E., & Laffel, L. M. (2015) · Current Diabetes Reports · doi:10.1007/s11892-015-0692-4 · Cited by 31 · see citing works
Wood, M. A., Shulman, D. I., Forlenza, G. P., Bode, B. W., Pinhas‐Hamiel, O., Buckingham, B. A., Kaiserman, K. B., Liljenquist, D. R., Bailey, T. S., Shin, J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0209 · Cited by 30 · see citing works
Self-citationHypoglycemia Begets Hypoglycemia: The Order Effect in the ASPIRE In-Clinic Study
Garg, S. K., Brazg, R., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0219 · Cited by 28 · see citing works
Self-citationChen, E., King, F., Kohn, M. A., Spanakis, E. K., Breton, M. D., & Klonoff, D. C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0119 · Cited by 28 · see citing works
Self-citationWake, A. D. (2022) · World Journal of Diabetes · doi:10.4239/wjd.v13.i3.161 · Cited by 30 · see citing works
Bakhtiani, P. A., Youssef, J. E., Duell, A. K., Branigan, D. L., Jacobs, P. G., Lasarev, M. R., Castle, J. R., & Ward, W. K. (2014) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2014.09.001 · Cited by 25 · see citing works
Automatic Artificial Pancreas Systems Using an Intelligent Multiple-Model PID Strategy
Batmani, Y., Khodakaramzadeh, S., & Moradi, P. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3116376 · Cited by 26 · see citing works
Shah, V. N., Moser, E. G., Blau, A., Dhingra, M., & Garg, S. K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0228 · Cited by 25 · see citing works
Self-citationAlotaibi, A., Khalifah, R. A. A., & McAssey, K. (2020) · Pediatric Diabetes · doi:10.1111/pedi.13088 · Cited by 26 · see citing works
Continuous Glucose Monitoring Adherence
de Bock, M. I., Cooper, M. N., Retterath, A. J., Nicholas, J. A., Ly, T. T., Jones, T. W., & Davis, E. A. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816633484 · Cited by 22 · see citing works
Insulin pump therapy in youth with type 1 diabetes: toward closed-loop systems
Tauschmann, M., & Hovorka, R. (2014) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2014.910192 · Cited by 21 · see citing works
Do youth with type 1 diabetes exercise safely? A focus on patient practices and glycemic outcomes
Roberts, A. J. C., Yi‐Frazier, J. P., Aitken, K. E., Mitrovich, C., Pascual, M., & Taplin, C. E. (2016) · Pediatric Diabetes · doi:10.1111/pedi.12402 · Cited by 21 · see citing works
Development and Future of Automated Insulin Delivery (AID) Systems
Garg, S. K., & MCVEAN, J. J. F. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0467 · Cited by 23 · see citing works
Self-citationThe role of continuous glucose monitoring in the care of children with type 1 diabetes
Larson, N. S., & Pinsker, J. E. (2013) · International Journal of Pediatric Endocrinology · doi:10.1186/1687-9856-2013-8 · Cited by 21 · see citing works
The Artificial Pancreas: Is it Important to Understand How the β Cell Controls Blood Glucose?
Steil, G. M., & Grodsky, G. M. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700528 · Cited by 20 · see citing works
Abraham, M. B., Nicholas, J. A., Ly, T. T., Roby, H. C., Paramalingam, N., Fairchild, J. M., King, B. R., Ambler, G. R., Cameron, F. J., Davis, E. A., et al. (2016) · BMJ Open · doi:10.1136/bmjopen-2016-011589 · Cited by 18 · see citing works
ARX model for interstitial glucose prediction during and after physical activities
Ugalde, H. M. R., Garnotel, M., Doron, M., Jallon, P., Charpentier, G., Franc, S., Huneker, E., Simon, C., & Bonnet, S. (2019) · Control Engineering Practice · doi:10.1016/j.conengprac.2019.07.013 · Cited by 17 · see citing works
Davis, T., Salahi, A., Welsh, J. B., & Bailey, T. S. (2015) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12542 · Cited by 15 · see citing works
Self-citationWolpert, H. A., Kavanagh, M., Atakov-Castillo, A., & Steil, G. M. (2015) · Diabetic Medicine · doi:10.1111/dme.12823 · Cited by 14 · see citing works
Weiss, R., Garg, S. K., Bergenstal, R. M., Klonoff, D. C., Bode, B. W., Bailey, T. S., Thrasher, J. R., Schwartz, F., Welsh, J. B., KAUFMAN, F. R., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815586014 · Cited by 13 · see citing works
Self-citationVillafuerte, B., Frías, M. M., Martín, M. B. R., Valverde, R. Y., Gómez, M. Á. Á., & Castellanos, R. B. (2017) · Endocrinología Diabetes y Nutrición · doi:10.1016/j.endinu.2017.02.008 · Cited by 13 · see citing works
Automated Insulin Delivery—The Light at the End of the Tunnel
Turksoy, K., Frantz, N., Quinn, L., Dumin, M., Kilkus, J., Hibner, B., Çinar, A., & Littlejohn, E. (2017) · Jornal de Pediatria · doi:10.1016/j.jpeds.2017.02.055 · Cited by 8 · see citing works
Klonoff, D. C., Bergenstal, R. M., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A., Welsh, J. B., & Lee, S. W. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700424 · Cited by 12 · see citing works
Self-citationPetruželková, L., Picková, K., Šumnı́k, Z., Šoupal, J., & Obermannová, B. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0459 · Cited by 12 · see citing works
Roze, S., Smith‐Palmer, J., de Portu, S., Saltik, A. Z. O., Akgül, T., & Deyneli, O. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0198 · Cited by 12 · see citing works
Gargallo-Fernández, M., Escalada, J., Peralta, F. G., Moreno, P. R., Martínez, A. M., Botella-Serrano, M., Pérez, C. T., & Fernández, J. L. (2015) · Endocrinología y Nutrición · doi:10.1016/j.endonu.2015.02.004 · Cited by 11 · see citing works
Marçal, D. F. D. S., Alexandrino, E. G., Cortez, L. E. R., & Bennemann, R. M. (2018) · journal of physical education · doi:10.4025/jphyseduc.v29i1.2917 · Cited by 11 · see citing works
Roze, S., Smith‐Palmer, J., Portu, S. D., Delbaere, A., Brouwer, B. D., & de Valk, H. W. (2019) · ClinicoEconomics and Outcomes Research · doi:10.2147/ceor.s186298 · Cited by 11 · see citing works
Predictive Low-Glucose Suspend to Prevent Hypoglycemia
Pinsker, J. E., & Dassau, E. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0064 · Cited by 10 · see citing works
Saiman, Y., & Mahmud, N. (2021) · European Journal of Gastroenterology & Hepatology · doi:10.1097/meg.0000000000002218 · Cited by 9 · see citing works
Modern clinical management helps reducing the impact of type 1 diabetes in children
Scaramuzza, A. E., & Zuccotti, G. (2015) · Pharmacological Research · doi:10.1016/j.phrs.2015.03.001 · Cited by 9 · see citing works
Hypoglycemia Prevention Using Low Glucose Suspend Systems
Bequette, B. W. (2014) · Lecture notes in bioengineering · doi:10.1007/978-3-642-54464-4_3 · Cited by 9 · see citing works
JENDLE, J. H., Adolfsson, P., & Pollock, N. W. (2020) · Diving and Hyperbaric Medicine Journal · doi:10.28920/dhm50.2.135-143 · Cited by 9 · see citing works
Continuous glucose monitoring in pregnant women with type 1 diabetes
Garg, S. K., & Polsky, S. (2017) · The Lancet · doi:10.1016/s0140-6736(17)32449-2 · Cited by 8 · see citing works
Self-citationSperling, M. A., Wolfsdorf, J. I., Menon, R. K., Tamborlane, W. V., Maahs, D. M., Battelino, T., & Phillip, M. (2021) · Elsevier eBooks · doi:10.1016/b978-0-323-62520-3.00021-x · Cited by 9 · see citing works
Kurnaz, E., Aycan, Z., Yıldırım, N., & Çetinkaya, S. Ç. (2017) · The Turkish Journal of Pediatrics · doi:10.24953/turkjped.2017.04.013 · Cited by 8 · see citing works
Swarm hybrid optimization for a piecewise model fitting applied to a glucose model
Acedo, L., Botella, M., Cortés, J. ‐., Hidalgo, J. I., Maqueda, E., & Villanueva, R. J. (2018) · Journal of Systems and Information Technology · doi:10.1108/jsit-10-2017-0103 · Cited by 8 · see citing works
Use of continuous glucose monitoring for sport in type 1 diabetes
Abdulrahman, A., Manhas, J., Linane, H., Gurney, M. E., Fitzgerald, C., & O'Sullivan, E. (2018) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2018-000432 · Cited by 8 · see citing works
The Future of Glucose Monitoring
Garg, S. K. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0421 · Cited by 8 · see citing works
Self-citationClosed-loop insulin delivery in adults with type 1 diabetes
Kumareswaran, K. (2012) · Apollo (University of Cambridge) · doi:10.17863/cam.16644 · Cited by 7 · see citing works
Nutritional Management of Athletes with Type 1 Diabetes: A Narrative Review
Cavallo, M., Fano, M. D., Barana, L., Dozzani, I., Bianchini, E., Pellegrino, M., Cisternino, L., Migliarelli, S., Giulietti, C., Pippi, R., et al. (2024) · Nutrients · doi:10.3390/nu16060907 · Cited by 6 · see citing works
Registration and Real-Life Studies on Automated Insulin Delivery Systems
Aktürk, H. K., Brown, S. A., Beck, R. W., Sherr, J. L., Bergenstal, R. M., Kovatchev, B. P., Vigersky, R. A., Breton, M. D., Hovorka, R., Klonoff, D. C., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251334993 · Cited by 6 · see citing works
Self-citationRegistration and Real-Life Studies on Automated Insulin Delivery Systems
Aktürk, H. K., Brown, S. A., Beck, R. W., Sherr, J. L., Bergenstal, R. M., Kovatchev, B. P., Vigersky, R. A., Breton, M. D., Hovorka, R., Klonoff, D. C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0130 · Cited by 6 · see citing works
Self-citationVillafuerte, B., Frías, M. M., Martín, M. B. R., Valverde, R. Y., Gómez, M. Á. Á., & Castellanos, R. B. (2017) · Endocrinología Diabetes y Nutrición (English ed ) · doi:10.1016/j.endien.2017.02.009 · Cited by 6 · see citing works
Experience With the Enlite Sensor in a Multicenter Pediatric Study
Evans, K., Richardson, C., Landry, A., Muileboom, J., Cormack, L., & Lawson, M. L. (2014) · The Science of Diabetes Self-Management and Care · doi:10.1177/0145721714560589 · Cited by 6 · see citing works
Exercise for People with Type 1 Diabetes
GALLEN, I. W. (2014) · Medicine and sport science/Medicine and sport · doi:10.1159/000357344 · Cited by 5 · see citing works
Continuous Glucose Monitoring in 2012
Bode, B. W., & Battelino, T. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.1502 · Cited by 5 · see citing works
Islet cell transplantation: current status in the UK
Choudhary, P., PARROTT, N. R., Birtles, L., & Rutter, M. K. (2012) · Practical Diabetes · doi:10.1002/pdi.1707 · Cited by 5 · see citing works
Yafi, M., Shah, A., & Vélez, K. (2023) · Translational Pediatrics · doi:10.21037/tp-23-145 · Cited by 5 · see citing works
De la bomba de insulina y el monitoreo continuo de glucosa al páncreas artificial
Apablaza, P., Néstor Soto, I., & Ethel Codner, D. (2017) · Revista médica de Chile · doi:10.4067/s0034-98872017000500011 · Cited by 5 · see citing works
Refining the Closed Loop in the Data Age: Research-to-Practice Transitions in Diabetes Technology
Forlenza, G. P., Sankaranarayanan, S., & Maahs, D. M. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0055 · Cited by 5 · see citing works
Gargallo-Fernández, M., Martín, J. E. S., Peralta, F. G., Moreno, P. R., Martínez, A. M., Botella-Serrano, M., Pérez, C. T., & Fernández, J. L. (2016) · PubMed · doi:10.1016/j.endonu.2016.03.002 · Cited by 5 · see citing works
Shah, V. N., Rewers, A. H., & Garg, S. K. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00013-2 · Cited by 4 · see citing works
Self-citation
- Authors
- Garg, S., Brazg, R. L., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & Kaufman, F. R.
- Affiliations
- Barbara Davis Center for Childhood Diabetes, Aurora, Colorado.
- Rainier Clinical Research Center (US)
- AMCR Institute (US)
- Stanford Medicine (US)
- Mills Peninsula Health Services (US)
- Medtronic Inc., Northridge, California.
- Corresponding author
- Garg, Satish
- ORCID
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2012-03
- Volume
- 14
- Issue
- 3
- Pages
- 205-209
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W1978849955
- Cited by
- 138
- References
- 9
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 4 · 2023: 4 · 2022: 5 · 2021: 7 · 2020: 6 · 2019: 10 · 2018: 8 · 2017: 11 · 2016: 16 · 2015: 17 · 2014: 16 · 2013: 21 · 2012: 10
- FWCI
- 16.43
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- type 1 diabetes, hypoglycemia, low glucose suspend, sensor-augmented pump, exercise-induced hypoglycemia, rebound hyperglycemia, hypoglycemia duration, hypoglycemia severity, continuous glucose monitoring
- MeSH terms
- Adolescent, Adult, Blood Glucose, Diabetes Mellitus, Type 1, Female, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged, Treatment Outcome, Cross-Over Studies, Monitoring, Ambulatory, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review