Automated Overnight Closed-Loop Control Using a Proportional-Integral-Derivative Algorithm with Insulin Feedback in Children and Adolescents with Type 1 Diabetes at Diabetes Camp
The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: This study determined the feasibility and efficacy of an automated proportional-integral-derivative with insulin feedback (PID-IFB) controller in overnight closed-loop (OCL) control of children and adolescents with type 1 diabetes over multiple days in a diabetes camp setting. RESEARCH DESIGN AND METHODS: The Medtronic (Northridge, CA) Android™ (Google, Mountain View, CA)-based PID-IFB system consists of the Medtronic Minimed Revel™ 2.0 pump and Enlite™ sensor, a control algorithm residing on an Android phone, a translator, and remote monitoring capabilities. An inpatient study was completed for 16 participants to determine feasibility. For the camp study, subjects with type 1 diabetes were randomized to either OCL or sensor-augmented pump therapy (control conditions) per night for up to 6 nights at diabetes camp. RESULTS: During the camp study, 21 subjects completed 50 OCL nights and 52 control nights. Based on intention to treat, the median time spent in range, from 70 to 150 mg/dL, was greater during OCL at 66.4% (n = 55) versus 50.6% (n = 52) during the control period (P = 0.004). A per-protocol analysis allowed for assessment of algorithm performance with the median percentage time in range, 70-150 mg/dL, being 75.5% (n = 37) for OCL versus 47.6% (n = 32) for the control period (P < 0.001). There was less time spent in the hypoglycemic ranges <60 mg/dL and <70 mg/dL during OCL compared with the control period (P = 0.003 and P < 0.001, respectively). CONCLUSIONS: The PID-IFB controller is effective in improving time spent in range as well as reducing nocturnal hypoglycemia during the overnight period in children and adolescents with type 1 diabetes in a diabetes camp setting.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Home Use of an Artificial Beta Cell in Type 1 Diabetes
Thabit, H., Tauschmann, M., Allen, J. M., Leelarathna, L., Hartnell, S., Wilinska, M. E., Acerini, C. L., Dellweg, S., Benesch, C., Heinemann, L., et al. (2015) · New England Journal of Medicine · doi:10.1056/nejmoa1509351 · Cited by 450
Fully Integrated Artificial Pancreas in Type 1 Diabetes
Breton, M., Farret, A., Bruttomesso, D., Anderson, S., Magni, L., Patek, S., Dalla Man, C., Place, J., Demartini, S., Del Favero, S., et al. (2012) · Diabetes · doi:10.2337/db11-1445 · Cited by 315
Davis, E. A., Keating, B., Byrne, G. C., Russell, M., & Jones, T. W. (1997) · Diabetes Care · doi:10.2337/diacare.20.1.22 · Cited by 273 · cited 1×
“14 The Medtronic (Northridge, CA) Android (cid:2) (Google, Mountain View, CA)-based system using a proportional-integral-derivative with insulin feedback (PID-IFB) algorithm is an investigational device designed for physician-supervised OCL control in patients with type 1 diabetes.”
Hovorka, R., Kumareswaran, K., Harris, J., Allen, J. M., Elleri, D., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Dunger, D. B., et al. (2011) · BMJ · doi:10.1136/bmj.d1855 · Cited by 244 · cited 1×
“8 Hovorka and co-workers 1,2,21,22 have completed several landmark studies examining closed-loop control in adolescents and children.”
Kovatchev, B., Cobelli, C., Renard, E., Anderson, S., Breton, M., Patek, S., Clarke, W., Bruttomesso, D., Maran, A., Costa, S., et al. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400611 · Cited by 200
Cited by
Weisman, A., Bai, J.-W., Cardinez, M., Kramer, C. K., & Perkins, B. A. (2017) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(17)30167-5 · Cited by 463
Bekiari, E., Kitsios, K., Thabit, H., Tauschmann, M., Athanasiadou, E., Karagiannis, T., Haidich, A.-B., Hovorka, R., & Tsapas, A. (2018) · BMJ · doi:10.1136/bmj.k1310 · Cited by 437
Carlson, A. L., Sherr, J. L., Shulman, D. I., Garg, S. K., Pop-Busui, R., Bode, B. W., Lilenquist, D. R., Brazg, R. L., Kaiserman, K. B., Kipnes, M. S., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0319 · Cited by 230
Self-citationOne year clinical experience of the first commercial hybrid closed-loop system
Lal, R. A., Basina, M., Maahs, D. M., Hood, K., Buckingham, B., & Wilson, D. M. (2019) · Diabetes Care · doi:10.2337/dc19-0855 · Cited by 215
Self-citationISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Metadata
- Authors
- Ly, T. T., Keenan, D. B., Roy, A., Han, J., Grosman, B., Cantwell, M., Kurtz, N., von Eyben, R., Clinton, P., Wilson, D. M., & Buckingham, B. A.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2016-06
- Volume
- 18
- Issue
- 6
- Pages
- 377-384
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2407185352
- Cited by
- 51
- References
- 5
- FWCI
- 7.79
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Automated Overnight Closed-Loop Control Using a Proportional-Integral-Derivative Algorithm with Insulin Feedback in Children and Adolescents with Type 1 Diabetes at Diabetes Camp. https://app.cochat.ai/papers/10.1089/dia.2015.0431
- The paper
- Ly, T. T., Keenan, D. B., Roy, A., Han, J., Grosman, B., Cantwell, M., Kurtz, N., von Eyben, R., Clinton, P., Wilson, D. M., & Buckingham, B. A. (2016). Automated Overnight Closed-Loop Control Using a Proportional-Integral-Derivative Algorithm with Insulin Feedback in Children and Adolescents with Type 1 Diabetes at Diabetes Camp. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2015.0431
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.