Accuracy, Utilization, and Effectiveness Comparisons of Different Continuous Glucose Monitoring Systems
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Accuracy and feature sets of continuous glucose monitoring (CGM) systems may influence device utilization and outcomes. We compared clinical trial accuracy and real-world utilization and effectiveness of two different CGM systems. MATERIALS AND METHODS: Separately conducted accuracy studies of a fifth-generation and a sixth-generation CGM system involved 50 and 159 adults, respectively. For between-system performance comparisons, propensity score methods were utilized to balance cohort characteristics. Real-world outcomes were assessed in 10,000 anonymized patients who had switched from the fifth-generation to the sixth-generation system and had used connected mobile devices to upload data from both systems, allowing pairwise comparisons of device utilization and glucose concentration distributions. RESULTS: Propensity score-adjusted mean absolute relative differences for the fifth- and sixth-generation systems were 9.0% and 9.9%, and the percentages of values within ±20%/20 mg/dL were 93.1% and 92.5%, respectively. The sixth-generation system, but not the fifth-generation system, met accuracy criteria for interoperable CGM systems. Both systems had high real-world utilization rates (93.8% and 95.3% in the fifth- and sixth-generation systems, respectively). Use of the sixth-generation system was associated with fewer glucose values <55 mg/dL (<3.1 mmol/L) (0.7% vs. 1.1%, P < 0.001) and more values 70-180 mg/dL (3.9-10.0 mmol/L) (57.3% vs. 56.0%, P < 0.001) than the fifth-generation system. CONCLUSIONS: CGM performance outcomes can be compared through the propensity score analysis of clinical trial data and pairwise comparisons of real-world data. The systems compared here had nearly equivalent accuracy and utilization rates. Longer term biochemical and psychosocial benefits observed with the fifth-generation system are also expected with the sixth-generation system.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185
Beck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635
Heinemann, L., Freckmann, G., Ehrmann, D., Faber-Heinemann, G., Guerra, S., Waldenmaier, D., & Hermanns, N. (2018) · The Lancet · doi:10.1016/s0140-6736(18)30297-6 · Cited by 522
Polonsky, W. H., Hessler, D., Ruedy, K. J., & Beck, R. W. (2017) · Diabetes Care · doi:10.2337/dc17-0133 · Cited by 346
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168 · cited 1×
“Accuracy of the G4 system with an advanced signal processing algorithm (‘‘Software 505’’) was studied in a clinical trial of adults by comparing CGM data with simultaneously collected ‘‘YSI’’ blood glucose values obtained with the YSI 2300 STAT Plus reference instrument (YSI Inc, Yellow Springs, OH).(17) Because the s…”
MethodFOSTER, N. C., MILLER, K., DIMEGLIO, L., MAAHS, D. M., TAMBORLANE, W. V., BERGENSTAL, R. M., CLEMENTS, M. A., RICKELS, M. R., SMITH, E., OLSON, B. A., et al. (2018) · Diabetes · doi:10.2337/db18-1689-p · Cited by 19 · cited 1×
“However, device utilization rates have risen markedly in recent years among both adults and children and have been attributed to improvements in accuracy, functionality, and comfort.(26,27) The absence of patients transitioning from G6 to G5 is another limitation of the study.”
Background
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“90,96,123,127,131,138,140,155,158,160 However, this cannot be recommended as the CGM value distribution is dependent on CGM accuracy and thus impairs the comparison of testing procedures among studies.”
BackgroundSadhu, A. R., Serrano, I. A., Xu, J., Nisar, T., Lucier, J., Pandya, A. R., & Patham, B. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964264 · Cited by 84
Faulds, E. R., Boutsicaris, A., Sumner, L., Jones, L., McNett, M., Smetana, K. S., May, C. C., Buschur, E., Exline, M. C., Ringel, M. D., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab409 · Cited by 71
Karakus, K. E., Akturk, H. K., Alonso, G. T., Snell-Bergeon, J. K., & Shah, V. N. (2023) · Diabetes Care · doi:10.2337/dc23-0495 · Cited by 52
Metadata
- Authors
- Welsh, J. B., Gao, P., Derdzinski, M., Puhr, S., Johnson, T. K., Walker, T. C., & Graham, C.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2019-03
- Volume
- 21
- Issue
- 3
- Pages
- 128-132
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2911901959
- Cited by
- 72
- References
- 6
- FWCI
- 8.26
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy, Utilization, and Effectiveness Comparisons of Different Continuous Glucose Monitoring Systems. https://app.cochat.ai/papers/10.1089/dia.2018.0374
- The paper
- Welsh, J. B., Gao, P., Derdzinski, M., Puhr, S., Johnson, T. K., Walker, T. C., & Graham, C. (2019). Accuracy, Utilization, and Effectiveness Comparisons of Different Continuous Glucose Monitoring Systems. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2018.0374
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.