Ultra-high Resolution EELS Analysis and STEM Imaging at 20 keV
Open accessAbstract
Aberration-corrected scanning transmission electron microscopy (STEM) at primary energy (Eo) of 60 keV and below allows imaging and analysis of light elements largely without knock-on damage [1]. Even though other damage mechanisms exist, per unit of beam damage elastic and inelastic scattering cross-sections increase as Eo is decreased [2], driving interest in operation at the lowest Eo. Modern aberration correctors enable correction of geometric probe aberrations to high-enough angles to allow atomic-scale imaging at Eo=30 keV and below, assuming that chromatic aberration can be overcome too, either by Cc correction or monochromation [3]. The latter option brings the additional benefit of simultaneously enabling high- and ultra-high energy resolution EELS studies such as vibrational spectroscopy [4, 5]. The absolute energy resolution of the Nion U-HERMES system becomes smaller roughly as √Eo, from ∼12 meV at 200 kV to <3 meV for 30 keV and below [6]. Spectra acquired with <3 meV energy resolution reveal many new features, including acoustic phonons near the Г point in the Brillouin Zone [7] motivating operation at the lowest-possible Eo. However, the utility of a STEM is significantly diminished if it cannot form an atom-sized probe. Here we show that Eo = 20 keV is a viable choice for atomic-resolution imaging and ultra-high energy resolution analysis. Figure 1 shows a monochromated zero loss peak (ZLP) from the U-HERMES system at Eo = 20 keV. The full width at half of maximum (FWHM) is 2.6 meV. The same spectrum is also plotted on a log scale to see the tails of the distribution, showing that the FW at 10-4 of maximum is ∼110 meV. In vibrational EELS, weak loss features tend to be on the scale of 10-3 to 10-4 of the ZLP, meaning that weak loss features at ΔE around ∼110/2=55 meV should be easily observable. Figure 2 shows a high-angle annular dark field (HAADF) image of monolayer graphene acquired at Eo = 20 keV, with illumination semi-angle α/2=55 mrad. The energy selecting slit was set to a width of 100 meV to mitigate the loss of contrast due to chromatic aberration. The individual carbon atoms are well resolved, and the absolute modulation between the atoms and center of the 6-member carbon rings is ∼25%. Removing the slit and using the full ∼350 meV wide distribution from the cold field emission source (including tails) increases the beam current and average signal level by ∼4.3x, and decreases the modulation to ∼13%. In other words, the slit-in image gives twice better modulation at less than a quarter of the total beam dose, providing a significant benefit for both imaging and spectroscopy in terms of information gained per unit of dose. . Monochromated ZLP at Eo = 20 keV on a linear scale (left) showing full width at half maximum (FWHM) of 2.6 meV, and log scale (right) showing FW at 10–4 M of ∼110 meV. [3ms x 1,000 aligned spectra, α/2=β/2=10 mrad] (left) HAADF image of monolayer graphene with inset low-pass filtered section showing the location of a line profile (bottom right) to show the absolute contrast, and (upper right) Fourier transform of the image showing 1.07 Å contrast in 3 directions. Eo=20kV, monochromated to δE=100 meV, Ibeam=3 pA, α/2=55 mrad, β/2=86 mrad, aligned sum of 60 images, each 8 seconds and 512 x 512 scan pixels.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The advent of high-speed detectors has enabled time-resolved in situ electron microscopy, allowing precise stimulus control with outstanding spatial, temporal and spectral resolution20,21”
From Multimodal electron microscopy of halide perovskite interfacial dynamics (2026) · cites this paper as [20]
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 7
- Krivanek, 2010, Ultramicroscopy, 110, 935, 10.1016/j.ultramic.2010.02.007 · DOI
- Peet, 2019, Ultramicroscopy, 203, 125, 10.1016/j.ultramic.2019.02.007 · DOI
- Krivanek, 2018, Microscopy and Microanalysis, 24, 18, 10.1017/S1431927618000582 · DOI
- Krivanek, 2014, Nature, 514, 209, 10.1038/nature13870 · DOI
- Miyata, 2014, Microscopy, 63, 377, 10.1093/jmicro/dfu023 · DOI
- Dellby, 2020, Microsc. Microanal, 26, 1804, 10.1017/S1431927620019406 · DOI
- Plotkin-Swing, 2020, Ultramicroscopy, 217, 113067, 10.1016/j.ultramic.2020.113067 · DOI
Cited by
Sorted by citations
Multimodal electron microscopy of halide perovskite interfacial dynamics
Li, X., Gu, Q., Huang, W., Fairclough, S. M., Friend, R. H., Stranks, S. D., Liu, T., & Ducati, C. (2026) · Nature · doi:10.1038/s41586-026-10238-8 · Cited by 14
“The advent of high-speed detectors has enabled time-resolved in situ electron microscopy, allowing precise stimulus control with outstanding spatial, temporal and spectral resolution 20,21 .”
Metadata
- Authors
- Dellby, N., Nelson-Quillin, A. M., Krivanek, O. L., Hrncirik, P., Mittelberger, A., Plotkin-Swing, B., & Lovejoy, T. C.
- Affiliations
- Nion (United States) (US)
- Nion R&D , Kirkland, WA ,
- Venue
- Microscopy and Microanalysis
- Publisher
- Oxford University Press (OUP)
- Published
- 2023-07-22
- Volume
- 29
- Issue
- Supplement_1
- Pages
- 626-627
- ISSN
- 1431-9276 (ISSN-L)
- 1435-8115
- OpenAlex
- W4385072440
- Cited by
- 9
- References
- 0
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 2
- FWCI
- 0.80
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Microscopy and Microanalysis (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Structural Biology › Advanced Electron Microscopy Techniques and Applications
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Electron and X-Ray Spectroscopy Techniques
- Physical Sciences › Materials Science › General Materials Science › Advanced Materials Characterization Techniques
- Keywords
- low-voltage electron microscopy, aberration-corrected STEM, chromatic aberration, atomic resolution imaging, monolayer graphene, acoustic phonons, beam damage, light elements, HAADF imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Ultra-high Resolution EELS Analysis and STEM Imaging at 20 keV. https://app.cochat.ai/papers/10.1093/micmic/ozad067.305
- The paper
- Dellby, N., Nelson-Quillin, A. M., Krivanek, O. L., Hrncirik, P., Mittelberger, A., Plotkin-Swing, B., & Lovejoy, T. C. (2023). Ultra-high Resolution EELS Analysis and STEM Imaging at 20 keV. Microscopy and Microanalysis. https://doi.org/10.1093/micmic/ozad067.305