Diagnostic accuracy of a reagent strip for assessing urinary albumin excretion in the general population
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Albuminuria is a sensitive marker of renal derangement and has been included in a number of studies investigating chronic kidney diseases (CKDs). This study is aimed to evaluate the diagnostic performances of a strip for measuring the albumin/creatinine ratio (ACR) in the general population and to compare it with those found in a diabetic population. METHODS: Urine samples were obtained from 201 consecutive subjects enrolled in an epidemiological study and from 259 type 2 diabetic patients. Urine was tested for albumin and creatinine using the strip (Clinitek Microalbumin) and laboratory methods. A hundred samples were stored under various conditions to assess analyte stability. RESULTS: In the general population, the strip test reached a 90% sensitivity and 91% specificity, considering the laboratory method as the 'gold standard', sparing >80% of subjects the laboratory tests at the expense of a 1% false negative rate and an 8% false positive rate. Regarding sensitivity and specificity, the ACR test performs very similarly in the general population and in the diabetics. The stability study showed that storage at -20 degrees C induced a significant decrease in the albumin concentration with both methods, such that 5% of the samples were re-classified in the lower ACR class. Storage at -80 degrees C for up to 12 months did not affect the measurement with both methods. CONCLUSION: Clinitek Microalbumin strips can be used for screening purposes in the general population since they correctly classify a significant percentage of subjects, particularly those with a normal albuminuria. Storage at -80 degrees C does not affect strip results. Screening with the strip and confirming positive results with a wet chemistry method are an efficient strategy for detecting albuminuria in the general population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A comparative evaluation of various methods for microalbuminuria screening
Sarafidis, P. A., Riehle, J., Bogojevic, Z., Basta, E., Chugh, A., & Bakris, G. L. (2008) · American Journal of Nephrology · doi:10.1159/000111825 · Cited by 41
Meinhardt, U. (2003) · Journal of Diabetes and its Complications · doi:10.1016/s1056-8727(02)00180-0 · Cited by 38 · cited 1×
“[19] obtained a sensitivity of 89% and a specificity of 73%, with a PPV of 17% and an NPV of 99%.”
Background
Cited by
KDIGO 2024 clinical practice guideline for the evaluation and management of chronic kidney disease
Stevens, P. E., Ahmed, S. B., Carrero, J. J., Foster, B., Francis, A., Hall, R. K., Herrington, W. G., Hill, G., Inker, L. A., Kazancıoğlu, R., et al. (2024) · Kidney International · doi:10.1016/j.kint.2023.10.018 · Cited by 8,530
Chapter 1: Definition and classification of CKD
(2013) · Kidney International Supplements · doi:10.1038/kisup.2012.64 · Cited by 1,110
How should proteinuria be detected and measured?
Lamb, E. J., MacKenzie, F., & Stevens, P. E. (2009) · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · doi:10.1258/acb.2009.009007 · Cited by 271
McTaggart, M. P., Newall, R. G., Hirst, J. A., Bankhead, C. R., Lamb, E. J., Roberts, N. W., & Price, C. P. (2014) · Annals of Internal Medicine · doi:10.7326/m13-2331 · Cited by 39
Mejia, J. R., Fernandez-Chinguel, J. E., Dolores-Maldonado, G., Becerra-Chauca, N., Goicochea-Lugo, S., Herrera-Añazco, P., Zafra-Tanaka, J. H., & Taype-Rondan, A. (2021) · Heliyon · doi:10.1016/j.heliyon.2021.e08253 · Cited by 35
“…Kingdom Diabetic patients/ outpatient 252 Not defined 20.63 Clinitek 50 (Bayer Diagnostics, Tarrytown, USA) ACR Early morning Not defined Graziani (2009) [30] Italy General population and diabetic patients/primary care 460 Not defined 18.15 Clinitek Microalbumin (Siemens Medical…”
BackgroundInfluential
Metadata
- Authors
- Graziani, M. S., Gambaro, G., Mantovani, L., Sorio, A., Yabarek, T., Abaterusso, C., Lupo, A., & Rizzotti, P.
- Venue
- Nephrology Dialysis Transplantation
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-11-26
- Volume
- 24
- Issue
- 5
- Pages
- 1490-1494
- ISSN
- 0931-0509 (ISSN-L)
- 1460-2385
- OpenAlex
- W2058434703
- Cited by
- 54
- References
- 2
- FWCI
- 1.92
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Diagnostic accuracy of a reagent strip for assessing urinary albumin excretion in the general population. https://app.cochat.ai/papers/10.1093/ndt/gfn639
- The paper
- Graziani, M. S., Gambaro, G., Mantovani, L., Sorio, A., Yabarek, T., Abaterusso, C., Lupo, A., & Rizzotti, P. (2009). Diagnostic accuracy of a reagent strip for assessing urinary albumin excretion in the general population. Nephrology Dialysis Transplantation. https://doi.org/10.1093/ndt/gfn639
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.