Adverse Tissue Reactions to Bioabsorbable Fixation Devices

B??stman, Ole M., Pihlajam??ki, Harri K.·2000·Clinical Orthopaedics and Related Research

Abstract

Among 2528 patients operated on using pins, rods, bolts, and screws made of polyglycolic acid or polylactic acid, 108 (4.3%) were affected by a clinically significant local inflammatory, sterile tissue reaction. The three most common indications for the use of these fixation devices were a displaced malleolar fracture, a chevron osteotomy for hallux valgus, and a displaced fracture of the radial head. In 107 patients, the reaction was elicited by a polyglycolic acid implant, and in one patient by a polylactic acid implant. The incidences were 5.3% (107 of 2037) and 0.2% (one of 491), respectively. The adverse tissue responses to polyglycolic acid were seen 11 weeks after the operation, on average, whereas the reaction to polylactic acid occurred 4.3 years after fixation of an ankle fracture. The mild reactions consisted of a painful erythematous papule of a few weeks' duration. Those of medium severity had a sinus that discharged remnants of the implant for up to 6 months. In the patients affected by severe reactions, extensive osteolytic lesions developed at the implant tracks. The histopathologic picture was that of a nonspecific foreign body reaction. In four patients with vigorous reactions, an arthrodesis of the wrist or ankle later was necessary because of severe osteoarthritis. Several markers of increased risk of the occurrence of a foreign body reaction were found. These included a poorly vascularized bone section such as scaphoid, use of a quinone dye as an additive in the polymer, and an implant geometry with large surface area (screw versus pin or rod). For polyglycolic acid implants, the risk of an adverse tissue response in a given clinical situation can be estimated from the findings of this study. For slow degrading polymers like polylactic acid, however, the ultimate biocompatibility still is unsettled, and additional clinical research with long followup is required.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003086-200002000-00026

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 476

Metadata

Authors
B??stman, O. M., & Pihlajam??ki, H. K.
Venue
Clinical Orthopaedics and Related Research
Published
2000-02
Volume
371
Issue
371
Pages
216-227
ISSN
0009-921X (ISSN-L)
OpenAlex
W127024613
Cited by
476
References
1
FWCI
11.33
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Adverse Tissue Reactions to Bioabsorbable Fixation Devices. https://app.cochat.ai/papers/10.1097/00003086-200002000-00026
The paper
B??stman, O. M., & Pihlajam??ki, H. K. (2000). Adverse Tissue Reactions to Bioabsorbable Fixation Devices. Clinical Orthopaedics and Related Research. https://doi.org/10.1097/00003086-200002000-00026

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.