CoChat

Surgical Technique for Installing an Eight-Channel Neuroprosthesis for Standing

Davis, John A., Triolo, Ronald J., Uhlir, James P., Bhadra, Niloy, Lissy, Dawn A., Nandurkar, Sanjay, Marsolais, E. B. · 2001 · Clinical Orthopaedics and Related Research
Cited by 68

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

A standardized surgical procedure to implant an eight-channel functional neuromuscular stimulation system in the lower extremities for standing, exercise, and transfers for individuals with spinal cord injury has been developed. The implanted components include: (1) one eight-channel receiver-stimulator, (2) epimysial electrodes, (3) intramuscular electrodes, and (4) inline connectors. The development process included identifying the target muscle set for electrode placement and the corresponding surgical approaches, determining the stages of the surgical procedure, and assessing the effectiveness and stability of the implanted neuroprosthesis. The bilateral muscle set consists of the vastus lateralis, the gluteus maximus, the semimembranosus, and the erector spinae. Surgical approaches to the nerve entry points were developed through a series of cadaveric studies and intraoperative tests. Electrode placement is related to bony landmarks and based on standard orthopaedic approaches. The components of the neuroprosthesis are installed in one surgical session, with three stages. This procedure has been applied successfully in seven individuals, resulting in strong, isolated stimulated contractions adequate to raise and lower the body, maintain standing with a walker, and perform pivot transfers. The standardized surgical procedure is repeatable and teachable and will be used in upcoming multicenter clinical trials of the implanted neuroprosthesis.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003086-200104000-00035

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 68

Metadata

Authors
Davis, J. A., Triolo, R. J., Uhlir, J. P., Bhadra, N., Lissy, D. A., Nandurkar, S., & Marsolais, E. B.
Venue
Clinical Orthopaedics and Related Research
Published
2001-04
Volume
385
Issue
385
Pages
237-252
ISSN
0009-921X (ISSN-L)
OpenAlex
W2035743367
Cited by
68
References
1
FWCI
2.08
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Surgical Technique for Installing an Eight-Channel Neuroprosthesis for Standing. https://app.cochat.ai/papers/10.1097/00003086-200104000-00035
The paper
Davis, J. A., Triolo, R. J., Uhlir, J. P., Bhadra, N., Lissy, D. A., Nandurkar, S., & Marsolais, E. B. (2001). Surgical Technique for Installing an Eight-Channel Neuroprosthesis for Standing. Clinical Orthopaedics and Related Research. https://doi.org/10.1097/00003086-200104000-00035

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.