Chronic stability and selectivity of four-contact spiral nerve-cuff electrodes in stimulating the human femoral nerve

Open access
Fisher, L E, Tyler, D J, Anderson, J S, Triolo, R J·2009·Journal of Neural Engineering

Abstract

This study describes the stability and selectivity of four-contact spiral nerve-cuff electrodes implanted bilaterally on distal branches of the femoral nerves of a human volunteer with spinal cord injury as part of a neuroprosthesis for standing and transfers. Stimulation charge threshold, the minimum charge required to elicit a visible muscle contraction, was consistent and low (mean threshold charge at 63 weeks post-implantation: 23.3 +/- 8.5 nC) for all nerve-cuff electrode contacts over 63 weeks after implantation, indicating a stable interface with the peripheral nervous system. The ability of individual nerve-cuff electrode contacts to selectively stimulate separate components of the femoral nerve to activate individual heads of the quadriceps was assessed with fine-wire intramuscular electromyography while measuring isometric twitch knee extension moment. Six of eight electrode contacts could selectively activate one head of the quadriceps while selectively excluding others to produce maximum twitch responses of between 3.8 and 8.1 N m. The relationship between isometric twitch and tetanic knee extension moment was quantified, and selective twitch muscle responses scaled to between 15 and 35 N m in tetanic response to pulse trains with similar stimulation parameters. These results suggest that this nerve-cuff electrode can be an effective and chronically stable tool for selectively stimulating distal nerve branches in the lower extremities for neuroprosthetic applications.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1741-2560/6/4/046010

Cites

38 of 38 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Helical and spiral extra-neural interfaces have been stable for decades in clinical applications.

    “The helical (Agnew et al., 1989) and spiral (Naples et al., 1988) type interfaces have been proven to be stable for decades in clinical applications (Fisher et al., 2009; Polasek et al., 2009).”

    From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Fisher et al., 2009

    The Abstract and Conclusion report stable stimulation thresholds for spiral nerve-cuff electrodes implanted in one human volunteer over 63 weeks. They do not establish stability for decades or for helical interfaces, so the citing sentence extends beyond this paper's findings. (LLM, conf=0.90)

    Experimental — accuracy not yet measured
  • Motor units in their absolute refractory period, which lasts between 1.5 and 2.1 ms, will not respond to stimulation.

    “Another method for quantifying selectivity takes advantage of the concept that motor units in their absolute refractory period, which lasts between 1.5 and 2.1 ms, will not respond to stimulation [1,3,6,11-15].”
  • If non-adjacent pairwise combinations of contacts were ignored and a limited data set was used to fit the Gompertz and third-order polynomial models, the method would require approximately 13 hours to completely optimize selective stimulation for a 100-contact Utah slant electrode array, as compared to 11.6 days otherwise.

    “If non-adjacent pairwise combinations of contacts were ignored and a limited data set was used to fit the Gompertz and third-order polynomial models, which have three and ten parameters, respectively, the method described above would require approximately 13 hours to completely optimize selective stimulation for a 100-contact Utah slant electrode array, as compared to 11.6 days otherwise [3].”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 1

  1. [1]Chae J Triolo R Kilgore K Creasey G DiMarco A Kirschblum S Neuromuscular Electrical Stimulation in Spinal Cord Injury Spinal Cord Medicine 2003 360 388 Lippincott Williams & Wilkins Philadelphia

Cited by

Sorted by citations

Authors
Fisher, L. E., Tyler, D. J., Anderson, J. S., & Triolo, R. J.
Affiliations
  • Case Western Reserve University (US)
  • Louis Stokes Cleveland VA Medical Center (US)
Venue
Journal of Neural Engineering
Published
2009-07-15
Volume
6
Issue
4
Pages
046010
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2054973590
Cited by
89
References
38
Citations by year
2026: 2 · 2025: 3 · 2024: 3 · 2023: 3 · 2022: 4 · 2021: 6 · 2020: 6 · 2019: 5 · 2018: 6 · 2017: 8 · 2016: 8 · 2015: 9 · 2014: 3 · 2013: 12 · 2012: 5
FWCI
1.52
Type
journal article
Language
en
Open access
Locations
  • Journal of Neural Engineering (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion)
Funders
  • Case Western Reserve University
  • Deutsches Krebsforschungszentrum
  • National Institutes of Health
  • National Institute of Biomedical Imaging and Bioengineering (R01EB001889, 5-R01-EB001889, T32-EB04314-08, T32EB004314)
  • National Center for Research Resources (M01RR000080)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
  • Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
Keywords
femoral nerve, spinal cord injury, neural prostheses, stimulation selectivity, quadriceps activation, isometric twitch, lower limb rehabilitation
MeSH terms
Adult, Analysis of Variance, Electric Stimulation, Electrodes, Implanted, Electromyography, Femoral Nerve, Humans, Knee, Male, Muscle Contraction, Spinal Cord Injuries, Muscle, Skeletal, Prostheses and Implants

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review