Chronic stability and selectivity of four-contact spiral nerve-cuff electrodes in stimulating the human femoral nerve
Open accessAbstract
This study describes the stability and selectivity of four-contact spiral nerve-cuff electrodes implanted bilaterally on distal branches of the femoral nerves of a human volunteer with spinal cord injury as part of a neuroprosthesis for standing and transfers. Stimulation charge threshold, the minimum charge required to elicit a visible muscle contraction, was consistent and low (mean threshold charge at 63 weeks post-implantation: 23.3 +/- 8.5 nC) for all nerve-cuff electrode contacts over 63 weeks after implantation, indicating a stable interface with the peripheral nervous system. The ability of individual nerve-cuff electrode contacts to selectively stimulate separate components of the femoral nerve to activate individual heads of the quadriceps was assessed with fine-wire intramuscular electromyography while measuring isometric twitch knee extension moment. Six of eight electrode contacts could selectively activate one head of the quadriceps while selectively excluding others to produce maximum twitch responses of between 3.8 and 8.1 N m. The relationship between isometric twitch and tetanic knee extension moment was quantified, and selective twitch muscle responses scaled to between 15 and 35 N m in tetanic response to pulse trains with similar stimulation parameters. These results suggest that this nerve-cuff electrode can be an effective and chronically stable tool for selectively stimulating distal nerve branches in the lower extremities for neuroprosthetic applications.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Helical and spiral extra-neural interfaces have been stable for decades in clinical applications.
“The helical (Agnew et al., 1989) and spiral (Naples et al., 1988) type interfaces have been proven to be stable for decades in clinical applications (Fisher et al., 2009; Polasek et al., 2009).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Fisher et al., 2009
The Abstract and Conclusion report stable stimulation thresholds for spiral nerve-cuff electrodes implanted in one human volunteer over 63 weeks. They do not establish stability for decades or for helical interfaces, so the citing sentence extends beyond this paper's findings. (LLM, conf=0.90)
Experimental — accuracy not yet measuredMotor units in their absolute refractory period, which lasts between 1.5 and 2.1 ms, will not respond to stimulation.
“Another method for quantifying selectivity takes advantage of the concept that motor units in their absolute refractory period, which lasts between 1.5 and 2.1 ms, will not respond to stimulation [1,3,6,11-15].”
From Optimization of selective stimulation parameters for multi-contact electrodes (2013) · cites this paper as [3]
If non-adjacent pairwise combinations of contacts were ignored and a limited data set was used to fit the Gompertz and third-order polynomial models, the method would require approximately 13 hours to completely optimize selective stimulation for a 100-contact Utah slant electrode array, as compared to 11.6 days otherwise.
“If non-adjacent pairwise combinations of contacts were ignored and a limited data set was used to fit the Gompertz and third-order polynomial models, which have three and ten parameters, respectively, the method described above would require approximately 13 hours to completely optimize selective stimulation for a 100-contact Utah slant electrode array, as compared to 11.6 days otherwise [3].”
From Optimization of selective stimulation parameters for multi-contact electrodes (2013) · cites this paper as [3]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Anatomical Guide for the Electromyographer: The Limbs and Trunk
Perotto, A., & Delagi, E. F. (1980) · Cited by 623 · see citing works
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · see citing works
Veraart, C., Raftopoulos, C., Mortimer, J. T., Delbeke, J., Pins, D., Michaux, G., Vanlierde, A., Parrini, S., & Wanet-Defalque, M.-C. (1998) · Brain Research · doi:10.1016/s0006-8993(98)00977-9 · Cited by 407 · see citing works
Anatomical Guide for the Electromyographer: The Limbs and Trunk
Buschbacher, R. M. (2007) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/phm.0b013e31803217b1 · Cited by 345 · see citing works
The influence of chair height on lower limb mechanics during rising
Rodosky, M. W., Andriacchi, T. P., & Andersson, G. B. J. (1989) · Journal of Orthopaedic Research® · doi:10.1002/jor.1100070215 · Cited by 337 · see citing works
Selective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
Smith, B., Tang, Z., Johnson, M. W., Pourmehdi, S., Gazdik, M. M., Buckett, J. R., & Peckham, P. H. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.664202 · Cited by 270 · see citing works
Vick, N. A. (1985) · JAMA · doi:10.1001/jama.1985.03350370136049 · Cited by 252 · see citing works
Selective stimulation of peripheral nerve fibers using dual intrafascicular electrodes
Yoshida, K., & Horch, K. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.243412 · Cited by 208 · see citing works
Polyimide cuff electrodes for peripheral nerve stimulation
Rodri´guez, F. J., Ceballos, D., Schu¨ttler, M., Valero, A., Valderrama, E., Stieglitz, T., & Navarro, X. (2000) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(00)00192-8 · Cited by 207 · see citing works
A nerve cuff technique for selective excitation of peripheral nerve trunk regions
Sweeney, J. D., Ksienski, D. A., & Mortimer, J. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55681 · Cited by 196 · see citing works
Methods for estimating isometric recruitment curves of electrically stimulated muscle
Durfee, W. K., & MacLean, K. E. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32097 · Cited by 155 · see citing works
Quantification of recruitment properties of multiple contact cuff electrodes
Grill, W. M., & Mortimer, J. T. (1996) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.506402 · Cited by 136 · see citing works
Standing-up of a healthy subject and a paraplegic patient
Bajd, T., Kralj, A. R., & Turk, R. (1982) · Journal of Biomechanics · doi:10.1016/0021-9290(82)90029-x · Cited by 124 · see citing works
Arborelius, U. P., Wretenberg, P. F., & LINDBERG, F. (1992) · Ergonomics · doi:10.1080/00140139208967399 · Cited by 123 · see citing works
An analysis of sit-to-stand movements
KOTAKE, T., Dohi, N., Kajiwara, T., Sumi, N., Koyama, Y., & Miura, T. (1993) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/0003-9993(93)90068-l · Cited by 113 · see citing works
Human nerve stimulation thresholds and selectivity using a multi-contact nerve cuff electrode
Polasek, K. H., Hoyen, H. A., Keith, M. W., & Tyler, D. J. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.891383 · Cited by 111 · see citing works
Self-citationTarler, M. D., & Mortimer, J. T. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828415 · Cited by 110 · see citing works
Grill, W. M., & Mortimer, J. T. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.736150 · Cited by 103 · see citing works
Clinical Applications of Electrical Stimulation After Spinal Cord Injury
Creasey, G. H., Ho, C., Triolo, R. J., Gater, D. R., DiMarco, A. F., Bogie, K. M., & Keith, M. W. (2004) · Journal of Spinal Cord Medicine · doi:10.1080/10790268.2004.11753774 · Cited by 102 · see citing works
Self-citationTriolo, R. J., Bieri, C., Uhlir, J., Kobetic, R., Scheiner, A., & Marsolais, E. B. (1996) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(96)90133-1 · Cited by 102 · see citing works
Self-citationDavis, J. A., Triolo, R. J., Uhlir, J. P., Bieri, C., Rohde, L., Lissy, D. A., & Kukke, S. N. (2001) · PubMed · Cited by 99 · see citing works
Self-citationMcDonnall, D., Clark, G. A., & Normann, R. A. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828425 · Cited by 98 · see citing works
Implanted stimulators for restoration of function in spinal cord injury
Bhadra, N., Kilgore, K. L., & Peckham, P. H. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00012-1 · Cited by 92 · see citing works
Standing After Spinal Cord Injury With Four-Contact Nerve-Cuff Electrodes for Quadriceps Stimulation
Fisher, L. E., Miller, M. E., Bailey, S. N., Davis Jr., J. A., Anderson, J. S., Rhode, L., Tyler, D. J., & Triolo, R. J. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.2003390 · Cited by 90 · see citing works
Self-citationSubfascicle stimulation selectivity with the flat interface nerve electrode
Leventhal, D. K., & Durand, D. M. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1569266 · Cited by 81 · see citing works
Effects of regular use of neuromuscular electrical stimulation on tissue health
Bogie, K. M., & Triolo, R. J. (2003) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2003.11.0469 · Cited by 79 · see citing works
Self-citationSurgical Technique for Installing an Eight-Channel Neuroprosthesis for Standing
Davis, J. A., Triolo, R. J., Uhlir, J. P., Bhadra, N., Lissy, D. A., Nandurkar, S., & Marsolais, E. B. (2001) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200104000-00035 · Cited by 68 · see citing works
Self-citationRecruitment characteristics of nerve fascicles stimulated by a multigroove electrode
Koole, P., Holsheimer, J., Struijk, J. J., & Verloop, A. J. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559348 · Cited by 56 · see citing works
Kagaya, H., Shimada, Y., Ebata, K., Sato, M., Satô, K., YUKAWA, T., & Obinata, G. (1995) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(95)80556-7 · Cited by 56 · see citing works
Clinical experience with a helical bipolar stimulating lead
TARVER, W. B., GEORGE, R. E., MASCHINO, S. E., HOLDER, L. K., & WERNICKE, J. F. (1992) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1992.tb02933.x · Cited by 46 · see citing works
Uhlir, J. P., Triolo, R. J., Davis, J. A., & Bieri, C. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.827224 · Cited by 44 · see citing works
Self-citationDavis, R., Houdayer, T., Andrews, B., et al. (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06368.x · Cited by 37 · see citing works
Surface Anatomy for Clinical Needle Electromyography
hang-j-lee-joel-a-delisa, Lee, H. J., & DeLisa, J. A. (1999) · Cited by 22 · see citing works
Electrical Stimulation: A Reflection on Current Clinical Practices
Bertoti, D. B. (2000) · Assistive Technology · doi:10.1080/10400435.2000.10132007 · Cited by 19 · see citing works
Shimada, Y., Satô, K., Matsunaga, T., et al. (2001) · The Tohoku Journal of Experimental Medicine · doi:10.1620/tjem.193.221 · Cited by 19 · see citing works
Linear Summation of Torque Produced by Selective Activation of Two Motor Fascicles
Tarler, M. D., & Mortimer, J. T. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.891377 · Cited by 14 · see citing works
Surface Anatomy for Clinical Needle Electromyography
(2000) · Archives of Neurology · doi:10.1001/archneur.57.8.1228 · Cited by 4 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [1]Chae J Triolo R Kilgore K Creasey G DiMarco A Kirschblum S Neuromuscular Electrical Stimulation in Spinal Cord Injury Spinal Cord Medicine 2003 360 388 Lippincott Williams & Wilkins Philadelphia
Cited by
Sorted by citations
Schiefer, M., Tan, D., Sidek, S. M., & Tyler, D. J. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/1/016001 · Cited by 255 · see citing works
Self-citationTan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., & Tyler, D. J. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/2/026002 · Cited by 175 · see citing works
Self-citationA review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
Charkhkar, H., Shell, C. E., Marasco, P. D., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aac964 · Cited by 131 · see citing works
Self-citationAdaptive self-healing electronic epineurium for chronic bidirectional neural interfaces
Song, K.-I., Seo, H., Seong, D., Kim, S., Yu, K. J., Kim, Y.-C., Kim, J., Kwon, S. J., Han, H.-S., Youn, I., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-18025-3 · Cited by 120 · see citing works
“Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves”
Christie, B. P., Freeberg, M., Memberg, W. D., Pinault, G. J. C., Hoyen, H. A., Tyler, D. J., & Triolo, R. J. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0285-3 · Cited by 97 · see citing works
Self-citationNeural interfaces for somatosensory feedback: bringing life to a prosthesis
Tyler, D. J. (2015) · Current Opinion in Neurology · doi:10.1097/wco.0000000000000266 · Cited by 105 · see citing works
Self-citationThin film multi-electrode softening cuffs for selective neuromodulation
González-González, M. A., Kanneganti, A., Joshi-Imre, A., Hernandez-Reynoso, A. G., Bendale, G., Modi, R., Ecker, M., Khurram, A., Cogan, S. F., Voit, W. E., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34566-6 · Cited by 97 · see citing works
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88 · see citing works
“The helical (Agnew et al., 1989) and spiral (Naples et al., 1988) type interfaces have been proven to be stable for decades in clinical applications (Fisher et al., 2009; Polasek et al., 2009).”
Quantification of human upper extremity nerves and fascicular anatomy
Brill, N. A., & Tyler, D. J. (2017) · Muscle & Nerve · doi:10.1002/mus.25534 · Cited by 64 · see citing works
Self-citationRaspopovic, S., Capogrosso, M., Badia, J., Navarro, X., & Micera, S. (2012) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2012.2189021 · Cited by 63 · see citing works
Schiefer, M. A., Freeberg, M., Pinault, G. J. C., Anderson, J., Hoyen, H., Tyler, D. J., & Triolo, R. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/5/056006 · Cited by 57 · see citing works
Self-citationCaravaca, A. S., Tsaava, T., Goldman, L., Silverman, H., Riggott, G., Chavan, S. S., Bouton, C., Tracey, K. J., Desimone, R., Boyden, E. S., et al. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa7a42 · Cited by 57 · see citing works
The design of and chronic tissue response to a composite nerve electrode with patterned stiffness
Freeberg, M. J., Stone, M. A., Triolo, R. J., & Tyler, D. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6632 · Cited by 54 · see citing works
Self-citationMaterial considerations for peripheral nerve interfacing
Kim, Y.-T., & Romero-Ortega, M. I. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.99 · Cited by 47 · see citing works
Optimization of selective stimulation parameters for multi-contact electrodes
Fisher, L. E., Tyler, D. J., & Triolo, R. J. (2013) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-10-25 · Cited by 45 · see citing works
Self-citationPhantom limb pain: peripheral neuromodulatory and neuroprosthetic approaches to treatment
Petersen, B. A., Nanivadekar, A. C., Chandrasekaran, S., & Fisher, L. E. (2019) · Muscle & Nerve · doi:10.1002/mus.26294 · Cited by 39 · see citing works
Self-citationIntraoperative evaluation of the spiral nerve cuff electrode on the femoral nerve trunk
Polasek, K. H., Schiefer, M. A., Pinault, G. C. J., Triolo, R. J., & Tyler, D. J. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/6/066005 · Cited by 33 · see citing works
Self-citationDelianides, C., Tyler, D., Pinault, G., Ansari, R., & Triolo, R. (2020) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.13110 · Cited by 20 · see citing works
Self-citationIncreased selectivity of clinical peripheral nerve interfaces
Tyler, D. J., Peterson, E. J., Brill, N., & White, K. (2011) · 2011 5th International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2011.5910536 · Cited by 10 · see citing works
Self-citationFrederick, R. A., Troyk, P. R., & Cogan, S. F. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac2bb8 · Cited by 7 · see citing works
- Authors
- Fisher, L. E., Tyler, D. J., Anderson, J. S., & Triolo, R. J.
- Affiliations
- Case Western Reserve University (US)
- Louis Stokes Cleveland VA Medical Center (US)
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2009-07-15
- Volume
- 6
- Issue
- 4
- Pages
- 046010
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC2928075
- OpenAlex
- W2054973590
- Cited by
- 89
- References
- 38
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 3 · 2023: 3 · 2022: 4 · 2021: 6 · 2020: 6 · 2019: 5 · 2018: 6 · 2017: 8 · 2016: 8 · 2015: 9 · 2014: 3 · 2013: 12 · 2012: 5
- FWCI
- 1.52
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Case Western Reserve University
- Deutsches Krebsforschungszentrum
- National Institutes of Health
- National Institute of Biomedical Imaging and Bioengineering (R01EB001889, 5-R01-EB001889, T32-EB04314-08, T32EB004314)
- National Center for Research Resources (M01RR000080)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- femoral nerve, spinal cord injury, neural prostheses, stimulation selectivity, quadriceps activation, isometric twitch, lower limb rehabilitation
- MeSH terms
- Adult, Analysis of Variance, Electric Stimulation, Electrodes, Implanted, Electromyography, Femoral Nerve, Humans, Knee, Male, Muscle Contraction, Spinal Cord Injuries, Muscle, Skeletal, Prostheses and Implants
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review