Speech Perception as a Function of Electrical Stimulation Rate: Using the Nucleus 24 Cochlear Implant System
The DOI matches its Crossref record; 33 of the 33 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To investigate the effect of varying electrical stimulation rate on speech comprehension by cochlear implant users, while keeping the number of stimulated channels constant. DESIGN: Three average rates of electrical stimulation, 250, 807, and 1615 pulses per second per channel (pps/ch), were compared using a speech processing strategy that employed an electrode selection technique similar to that used in the Spectral Maxima Sound Processor strategy (McDermott, McKay, & Vandali, 1992; McDermott & Vandali, Reference Note 1; McKay, McDermott, Vandali, & Clark, 1991) and the Spectral Peak strategy (Skinner et al., 1994; Whitford et al., 1995). Speech perception tests with five users of the Nucleus 24 cochlear implant system were conducted over a 21-wk period. Subjects were given take-home experience with each rate condition. A repeated ABC evaluation protocol with alternating order was employed so as to account for learning effects and to minimize order effects. Perception of open-set monosyllabic words in quiet and open-set sentences at signal to noise ratios ranging from +20 to 0 dB, depending on the subject's ability, were tested. A comparative performance questionnaire was also administered. RESULTS: No statistical differences in group performance between the 250 and 807 pps/ch rates were observed in any of the speech perception tests. However, significantly poorer group performance was observed for the 1615 pps/ch rate for some tests due predominantly to the results of one subject. Analysis of individual scores showed considerable variation across subjects. For some subjects, one or more of the three rate conditions evaluated provided benefits on some speech perception tasks. The results of the comparative performance questionnaire indicated a preference for the 250 and 807 pps/ch rates over the 1615 pps/ch rate for most listening situations. CONCLUSIONS: For the speech processing strategy, implant system, and subjects evaluated in this study, the group results indicated that the use of electrical stimulation rates higher than 250 pps/ch (up to 1615 pps/ch) generally provided no significant improvement to speech comprehension. However, individual results indicated that perceptual benefits could be obtained by adjusting rate of stimulation optimally to suit each subject. Results from one subject, together with tinnitus problems arising from high-rate stimulation for another subject, indicated that high rates of stimulation may in fact be undesirable for some subjects.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
SCHEFFE, H. (1960) · Soil Science · doi:10.1097/00010694-196006000-00016 · Cited by 5,516
Speech Recognition with Primarily Temporal Cues
Shannon, R. V., Zeng, F.-G., Kamath, V., Wygonski, J., & Ekelid, M. (1995) · Science · doi:10.1126/science.270.5234.303 · Cited by 3,024
Better speech recognition with cochlear implants
Wilson, B. S., Finley, C. C., Lawson, D. T., Wolford, R. D., Eddington, D. K., & Rabinowitz, W. M. (1991) · Nature · doi:10.1038/352236a0 · Cited by 1,190
Electrical Stimulation of the Auditory Nerve in Man
SIMMONS, F. B. (1966) · Archives of Otolaryngology - Head and Neck Surgery · doi:10.1001/archotol.1966.00760030004003 · Cited by 613
Intonation, Perception and Language
Pollack, I., & Lieberman, P. (1968) · The American Journal of Psychology · doi:10.2307/1421071 · Cited by 594
Multichannel electrical stimulation of the auditory nerve in man. I. Basic psychophysics
Shannon, R. V. (1983) · Hearing Research · doi:10.1016/0378-5955(83)90077-1 · Cited by 364
Intonation and the perceptual separation of simultaneous voices
Brokx, J. P. L., & Nooteboom, S. G. (1982) · Journal of Phonetics · doi:10.1016/s0095-4470(19)30909-x · Cited by 364
Auditory Prostheses Research with Multiple Channel Intracochlear Stimulation in Man
Eddington, D. K., Dobelle, W. H., Mladejovsky, M. G., Brackmann, D. E., & Parkin, J. L. (1978) · Annals of Otology, Rhinology & Laryngology · doi:10.1177/00034894780870s602 · Cited by 334
Skinner, M. W., Clark, G. M., Whitford, L. A., Seligman, P., Staller, S. J., Shipp, D. B., Shallop, J. K., Everingham, C., Menapace, C., & Arndt, P. L. (1994) · PubMed · Cited by 322
Self-citationPitch perception by cochlear implant subjects
Townshend, B., Cotter, N., Van Compernolle, D., & White, R. L. (1987) · The Journal of the Acoustical Society of America · doi:10.1121/1.395554 · Cited by 306
Speech waveform envelope cues for consonant recognition
Van Tasell, D. J., Soli, S. D., Kirby, V. M., & Widin, G. P. (1987) · The Journal of the Acoustical Society of America · doi:10.1121/1.395251 · Cited by 292
Xu, J., Shepherd, R. K., Millard, R. E., & Clark, G. M. (1997) · Hearing Research · doi:10.1016/s0378-5955(96)00193-1 · Cited by 242
McDermott, H. J., McKay, C. M., & Vandali, A. E. (1992) · The Journal of the Acoustical Society of America · doi:10.1121/1.402826 · Cited by 218
Self-citationTemporal envelope and fine structure cues for speech intelligibility
Drullman, R. (1994) · The Journal of the Acoustical Society of America · doi:10.1121/1.408825 · Cited by 213
Enhancement of vowel coding for cochlear implants by addition of noise
Morse, R. P., & Evans, E. F. (1996) · Nature Medicine · doi:10.1038/nm0896-928 · Cited by 198
Temporal response patterns of single auditory nerve fibers elicited by periodic electrical stimuli
van den Honert, C., & Stypulkowski, P. H. (1987) · Hearing Research · doi:10.1016/0378-5955(87)90168-7 · Cited by 146
Tong, Y. C., Blamey, P. J., Dowell, R. C., & Clark, G. M. (1983) · The Journal of the Acoustical Society of America · doi:10.1121/1.389620 · Cited by 137
Phase-locking of auditory-nerve discharges to sinusoidal electric stimulation of the cochlea
Dynes, S., & Delgutte, B. (1992) · Hearing Research · doi:10.1016/0378-5955(92)90011-b · Cited by 137
Sifting vowels. Auditory pitch analysis and sound segregation.
Scheffers, M. T. M. (1983) · University of Groningen research database (University of Groningen / Centre for Information Technology) · Cited by 134
Skinner, M. W., Holden, L. K., Holden, T. A., Dowell, R. C., Seligman, P., Brimacombe, J. A., & Beiter, A. L. (1991) · Ear and Hearing · doi:10.1097/00003446-199102000-00002 · Cited by 130
Temporal representations with cochlear implants.
Wilson, B. S., Finley, C. C., Lawson, D. T., & Zerbi, M. (1997) · PubMed · Cited by 130
Responses of Cat Auditory Nerve Fibers to Biphasic Electrical Current Pulses
Javel, E., Tong, Y. C., Shepherd, R. K., & Clark, G. M. (1987) · Annals of Otology, Rhinology & Laryngology · doi:10.1177/00034894870960s111 · Cited by 123
Parkins, C. W. (1989) · Hearing Research · doi:10.1016/0378-5955(89)90007-5 · Cited by 102
The perception of temporal modulations by cochlear implant patients
Busby, P. A., Tong, Y. C., & Clark, G. M. (1993) · The Journal of the Acoustical Society of America · doi:10.1121/1.408212 · Cited by 72
Tykocinski, M., Shepherd, R. K., & Clark, G. M. (1995) · Hearing Research · doi:10.1016/0378-5955(95)00108-g · Cited by 61 · cited 1×
“Neverthe-less it may be appropriate to proceed with some caution in this area (Tykocinski, Shepherd, & Clark, 1995).”
BackgroundFundamental Considerations in Designing Auditory Implants
Kiang, N. Y. S., Eddington, D. K., & Delgutte, B. (1979) · Acta Oto-Laryngologica · doi:10.3109/00016487909126408 · Cited by 61
Response Characteristics of Nerve Fibers to Patterned Electrical Stimulation
Hartmann, R., & Klinke, R. (1990) · Springer eBooks · doi:10.1007/978-1-4612-3256-8_10 · Cited by 60
A Functioning Multichannel Auditory Nerve StimulatorA Preliminary Report on Two Human Volunteers
Simmons, F. B., Mathews, R. G., Walker, M. G., & White, R. L. (1979) · Acta Oto-Laryngologica · doi:10.3109/00016487909126403 · Cited by 47
Will there be noise in their ears?
Moss, F., Chiou‐Tan, F. Y., & Klinke, F. R. (1996) · Nature Medicine · doi:10.1038/nm0896-860 · Cited by 31
Tykocinski, M., Shepherd, R. K., & Clark, G. M. (1997) · Hearing Research · doi:10.1016/s0378-5955(97)00117-2 · Cited by 30
Huang, C. Q., Shepherd, R. K., Seligman, P., & Clark, G. M. (1998) · Hearing Research · doi:10.1016/s0378-5955(97)00196-2 · Cited by 26
McKay, C. M., McDermott, H. J., Vandali, A. E., & Clark, G. M. (1991) · Minerva Access (University of Melbourne) · Cited by 24
Self-citationDirect current measurements in cochlear implants: an in vivo and in vitro study
Huang, C., Carter, P., Shepherd, R. K., Seligman, P., Tabor, B., & Clark, G. M. (2002) · International Conference on Bioelectromagnetism · doi:10.1109/icbem.1998.666422 · Cited by 6
Cited by
Cochlear Implants: System Design, Integration, and Evaluation
Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., & Feng, H. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008250 · Cited by 878
“Although there has been a trend for all manufacturers to push for higher stimulation rates, there is little or no scientific evidence suggesting that higher rate produces better performance [87]–[89].”
BackgroundMusic Perception with Cochlear Implants: A Review
McDermott, H. J. (2004) · Trends in Amplification · doi:10.1177/108471380400800203 · Cited by 549
Temporal considerations for stimulating spiral ganglion neurons with cochlear implants.
Boulet, J., White, M., & Bruce, I. C. (2016) · Journal of the Association for Research in Otolaryngology · doi:10.1007/s10162-015-0545-5 · Cited by 114
“…demonstrated no benefit (Friesen et al. 2005; Plant et al. 2002; Weber et al. 2007; Holden et al. 2002; Plant et al. 2007; Arora et al. 2009), while other research indicates that low to moderate stimulation rates, i.e., 250 to 500 pulses/s, work best (Balkany et al. 2007; Vandali et al. 2000).”
Background
Metadata
- Authors
- Vandali, A. E., Whitford, L. A., Plant, K. L., & Clark, A. G. M.
- Venue
- Ear and Hearing
- Published
- 2000-12
- Volume
- 21
- Issue
- 6
- Pages
- 608-624
- ISSN
- 0196-0202 (ISSN-L)
- OpenAlex
- W1991298420
- Cited by
- 308
- References
- 33
- FWCI
- 3.30
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.