Signal, Noise, and Contrast in Nuclear Magnetic Resonance (NMR) Imaging
The DOI matches its Crossref record.
Abstract
Calculations of the sensitivity of the saturation recovery and inversion recovery pulse sequences used in nuclear magnetic resonance imaging show the former to be superior in discriminating between tissues with the same proton density but different T1's. Two other pulse sequences, which are combinations of the above, have also been analyzed. These have lower T1 discrimination sensitivity, but other considerations, such as self-normalization, may still make them attractive. The calculations are only valid for selective excitation pulse sequences in which the selected slice profiles are approximately rectangular, and thus a sin(bt)/t radiofrequency excitation is desirable. In order to ensure that the saturation recovery sequence gives valid results for pulse repetition times comparable to or shorter than T2, it is necessary to destroy the coherence between pulse applications. For this purpose we use a series of "spoiler" gradient pulses between pulse trains. The saturation recovery pulse sequence also has the advantage that, by the correct choice of interpulse spacing, sensitivity close to the optimum T1 discrimination can be achieved over a wide range of T1 values. This has the potential advantage to the clinician of simplifying his choice of parameters for imaging.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
The intrinsic signal-to-noise ratio in NMR imaging
Edelstein, W. A., Glover, G. H., Hardy, C. J., & Redington, R. W. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030413 · Cited by 758
Self-citationSpoiling of transverse magnetization in steady‐state sequences
Zur, Y., Wood, M. L., & Neuringer, L. J. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910210210 · Cited by 306
Fundamental Physics of MR Imaging
Pooley, R. A. (2005) · RadioGraphics · doi:10.1148/rg.254055027 · Cited by 210
Relaxation times of breast tissue at 1.5T and 3T measured using IDEAL
Rakow‐Penner, R., Daniel, B., Yu, H., Sawyer‐Glover, A., & Glover, G. H. (2006) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20469 · Cited by 196
“According to Edelstein et al (19), the TR for saturation recovery imaging, and the TI for inversion recovery imaging should be set approximately equal to the T1 of the tissue of interest at a given field strength.”
MethodMagnetic resonance imaging of cavernous hemangioma of the liver: tissue-specific characterization
Stark, D., Felder, R., Wittenberg, J., Saini, S., Butch, R., White, M., Edelman, R., Mueller, P., Simeone, J., Cohen, A., et al. (1985) · American Journal of Roentgenology · doi:10.2214/ajr.145.2.213 · Cited by 183
Lori, N. F., Akbudak, E., Shimony, J. S., Cull, T. S., Snyder, A. Z., Guillory, R. K., & Conturo, T. E. (2002) · NMR in Biomedicine · doi:10.1002/nbm.779 · Cited by 164
“The MRI intensity derived from the modulus of the complex Fourier transformation can have bias due to noise.(64,65) To include this bias, the synthetic noiseless MR signal intensity in the previous section was placed entirely in the real channel, then independent noise fluctuations were added to the real and imaginary…”
Metadata
- Authors
- Edelstein, W. A., Bottomley, P. A., Hart, H. R., & Smith, L. S.
- Venue
- Journal of Computer Assisted Tomography
- Published
- 1983-06
- Volume
- 7
- Issue
- 3
- Pages
- 391-401
- ISSN
- 0363-8715 (ISSN-L)
- OpenAlex
- W2070818218
- Cited by
- 201
- References
- 0
- FWCI
- 23.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.