CoChat

Lumbar Range of Motion: Reliability and Validity of the Inclinometer Technique in the Clinical Measurement of Trunk Flexibility

Saur, Petra M. M., Ensink, Franz-Bernhard M., Frese, Knut, Seeger, Dagmar, Hildebrandt, Jan · 1996 · Spine
Cited by 238

The DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.

Abstract

STUDY DESIGN: This study examines the reliability and validity of measuring lumbar range of motion with an inclinometer. OBJECTIVES: To find out whether a manual determination of the reference points for measuring lumbar range of motion is as reliable as radiologic determination for positioning the inclinometers, lumbar range of motion was determined in degrees by evaluating radiographs and by using the inclinometer technique of Loebl. SUMMARY OF BACKGROUND DATA: Reliability and validity of the inclinometer technique as a clinical measurement of trunk flexibility were investigated. Fifty-four patients participated in the study. METHODS: Lumbar range of motion measurements were taken with and without radiologic control of the T12 and S1 vertebrae as reference points for positioning of the inclinometers. An interrater correlation was done of the inclinometer techniques of a physician and a physiotherapist. Functional radiographs were investigated in a standing position. Lumbar range of motion measurements based on radiographs and those taken using the inclinometer alone were correlated to validate the inclinometer technique. RESULTS: Lumbar range of motion measurements taken with and without radiologic determination showed a very close correlation (r = 0.93; P < 0.001). Flexion alone also demonstrated a close correlation (r = 0.95; P < 0.001), whereas extension showed a somewhat smaller correlation (r = 0.82; P < 0.001). Total lumbar range of motion (r = 0.94; P < 0.001) and flexion (r = 0.88; P < 0.001) were closely related, as indicated by the interrater correlation, whereas extension (r = 0.42; P < 0.05) showed a lesser correlation. Correlation of the measurements taken radiographically and by inclinometer demonstrated an almost linear correlation for measurements of the total lumbar range of motion (r = 0.97; P < 0.001) and flexion (r = 0.98; P < 0.001), whereas extension (r = 0.75; P < 0.001) did not correlate as well. CONCLUSIONS: The noninvasive inclinometer technique proved to be highly reliable and valid, but the measurement technique for extension needs further refinement.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00007632-199606010-00011

Cites

15 of 15 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 15 references

Cited by

Metadata

Authors
Saur, P. M. M., Ensink, F.-B. M., Frese, K., Seeger, D., & Hildebrandt, J.
Venue
Spine
Published
1996-06
Volume
21
Issue
11
Pages
1332-1338
ISSN
0362-2436 (ISSN-L)
OpenAlex
W2082161122
Cited by
238
References
15
FWCI
2.70
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.