Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT
The DOI matches its Crossref record; 17 of the 17 references it cites were found in a registry or an index.
Abstract
Background Two devices for self-measurement of blood pressure at the brachial artery—the Omron M5-I and the Omron 705IT—were evaluated according to the international protocol of the European Society of Hypertension. Design The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 supplementary subjects are included making a total number of 33 subjects on which the final validation is performed. Methods For each subject, four blood pressure (BP) measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with three measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 differences were classified into categories (≤5, ≤10, ≤15 mmHg). The number of differences in each category was compared to the number required by the international protocol. An individual analysis was then done to determine for each subject the number of comparisons ≤5 mmHg. At least 22 of the 33 subjects should have two of their three comparisons ≤5 mmHg. Results The two tested devices passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were −0.9±5.8 and −0.8±4.8 mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP) respectively for the Omron M5-I device and −0.2±4.5 and −2.0±4.8 mmHg for the Omron 705IT device. Readings for the two devices differing by less than 5, 10 and 15 mmHg for systolic and diastolic values fulfil the recommendation criteria of the international protocol as well as the individual analysis. Conclusions The Omron M-5I and the Omron 705IT devices pass the validation recommendations of the international protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
O??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
OʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 744
Self-citationIMPROVEMENT OF MEDICATION COMPLIANCE IN UNCONTROLLED HYPERTENSION
Haynes, R. B., Gibson, E., Hackett, B., Sackett, D., Taylor, D. W., Roberts, R., & Johnson, A. (1976) · The Lancet · doi:10.1016/s0140-6736(76)91737-2 · Cited by 513
Recommendations for the use of home (self) and ambulatory blood pressure monitoring
Pickering, T. (1996) · American Journal of Hypertension · doi:10.1016/0895-7061(95)00341-x · Cited by 472
O'Brien, E. T., Mee, F. I., Atkins, N., & Thomas (1996) · PubMed · Cited by 409
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Validation of the Omron HEM-907 device for blood pressure measurement
El Assaad, M. A., Topouchian, J. A., Darné, B. M., & Asmar, R. G. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200208000-00006 · Cited by 279
Self-citationAsmar, R., & Zanchetti, A. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018050-00001 · Cited by 213
Self-citationWhite, W. B., & Anwar, Y. A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200104000-00007 · Cited by 213
Appel, L. J., & Stason, W. B. (1993) · Annals of Internal Medicine · doi:10.7326/0003-4819-118-11-199306010-00008 · Cited by 206
Cited by
Celis-Morales, C. A., Perez-Bravo, F., Ibañez, L., Salas, C., Bailey, M. E. S., & Gill, J. M. R. (2012) · PLoS ONE · doi:10.1371/journal.pone.0036345 · Cited by 486
Whyte, L. J., Gill, J. M. R., & Cathcart, A. J. (2010) · Metabolism · doi:10.1016/j.metabol.2010.01.002 · Cited by 424
“Blood pressure was measured using an automated blood pressure monitor (Omron HEM705 CP; Omron Healthcare, Milton Keynes, United Kingdom) that has been validated according to the European Society of Hypertension International Protocol [27].”
MethodRajzer, M. W., Wojciechowska, W., Klocek, M., Palka, I., Brzozowska-Kiszka, M., & Kawecka-Jaszcz, K. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e32830a4a25 · Cited by 308
Reference Values of Pulse Wave Velocity in Healthy Children and Teenagers
Reusz, G. S., Cseprekal, O., Temmar, M., Kis, E., Cherif, A. B., Thaleb, A., Fekete, A., Szabó, A. J., Benetos, A., & Salvi, P. (2010) · Hypertension · doi:10.1161/hypertensionaha.110.152686 · Cited by 302
Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A. (2008) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3282cb57b6 · Cited by 259
Home Blood Pressure Measurement
Verberk, W. J., Kroon, A. A., Kessels, A. G. H., & de Leeuw, P. W. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.05.058 · Cited by 212
Left ventricular strain and strain rate in a general population
Kuznetsova, T., Herbots, L., Richart, T., D'hooge, J., Thijs, L., Fagard, R. H., Herregods, M.-C., & Staessen, J. A. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehn280 · Cited by 211
Stergiou, G. S., Yiannes, N. G., & Rarra, V. C. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000209074.38331.16 · Cited by 192
Martínez Vizcaíno, V, Salcedo Aguilar, F., Franquelo Gutiérrez, R., Solera Martínez, M., Sánchez López, M., Serrano Martínez, S., López García, E., & Rodríguez Artalejo, F. (2008) · International Journal of Obesity · doi:10.1038/sj.ijo.0803738 · Cited by 189
Coleman, A., Freeman, P., Steel, S., & Shennan, A. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000189788.05736.5f · Cited by 179
Asmar, R. (2011) · Vascular Health and Risk Management · doi:10.2147/vhrm.s27193 · Cited by 163
Self-citationTopouchian, J. A., El Assaad, M. A., Orobinskaia, L. V., El Feghali, R. N., & Asmar, R. G. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000209078.17246.34 · Cited by 158
Self-citationHome blood pressure normalcy in children and adolescents: the Arsakeion School study
Stergiou, G. S., Yiannes, N. G., Rarra, V. C., & Panagiotakos, D. B. (2007) · Journal of Hypertension · doi:10.1097/hjh.0b013e328122d3fc · Cited by 119
Orthostatic hypotension in very old individuals living in nursing homes: the PARTAGE study
Valbusa, F., Labat, C., Salvi, P., Vivian, M. E., Hanon, O., & Benetos, A. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834d3d73 · Cited by 91
Engbers, L. H., van Poppel, M. N. M., & van Mechelen, W. (2007) · Preventive Medicine · doi:10.1016/j.ypmed.2006.11.005 · Cited by 41
Blood Pressure and Cognitive Function: The Role of Central Aortic and Brachial Pressures
Pase, M. P., Stough, C., Grima, N. A., Harris, E., Macpherson, H., Scholey, A. B., & Pipingas, A. (2013) · Psychological Science · doi:10.1177/0956797613488602 · Cited by 38
Metadata
- Authors
- El Assaad, M. A., Topouchian, J. A., & Asmar, R. G.
- Venue
- Blood Pressure Monitoring
- Published
- 2003-06
- Volume
- 8
- Issue
- 3
- Pages
- 127-133
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2099393780
- Cited by
- 224
- References
- 17
- FWCI
- 4.82
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.