Current assessment of pulse wave velocity: comprehensive review of validation studies
The DOI matches its Crossref record; 65 of the 65 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: Carotid-femoral pulse wave velocity (PWV) is considered the gold standard for arterial stiffness assessment in clinical practice. A large number of devices to measure PWV have been developed and validated. We reviewed different validation studies of PWV estimation techniques and assessed their conformity to the Artery Society Guidelines and the American Heart Association recommendations. METHODS: Pubmed and Medline (1995-2017) were searched to identify PWV validation studies. Of the 96 article retrieved, 26 met the inclusion criteria. RESULTS: Several devices had been developed and validated to noninvasively measure arterial stiffness, using applanation tonometry (SphygmoCor, PulsePen), piezoelectric mechanotransducers (Complior), cuff-based oscillometry (Arteriograph, Vicorder and Mobil-O-Graph), photodiode sensors (pOpmètre) and devices assessing brachial-ankle pulse wave velocity and cardiac-ankle PWV. Ultrasound technique and MRI remain confined to clinical research. Good agreement was found with the Artery Society Guidelines. Two studies (Complior, SphygmoCor Xcel) showed best adherence with the guidelines. In Arteriograph, MRI, ultrasound and SphygmoCor Xcel validation studies sample size was smaller than the minimum suggested by the guidelines. High discrepancies between devices were shown in distance estimation: in two studies (Arteriograph, Complior) path length was estimated in conformity to the guidelines. Transit time was calculated using the intersecting tangent method, but in two studies (Vicorder, pOpmètre) best agreement was found using the maximum of the second derivative. Six studies reached the accuracy level 'excellent' defined in the Artery guidelines. CONCLUSION: Method to assess transit time and path length need validation in larger populations. Further studies are required in different risk population to implement clinical applicability of every device.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Expert consensus document on arterial stiffness: methodological issues and clinical applications
Laurent, S., Cockcroft, J., Van Bortel, L., Boutouyrie, P., Giannattasio, C., Hayoz, D., Pannier, B., Vlachopoulos, C., Wilkinson, I., Struijker-Boudier, H., et al. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl254 · Cited by 6,077 · cited 1×
“PWV, by definition, results as d over transit time ratio (in m/s) [11].”
BackgroundMancia, G., Fagard, R., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T., Cifkova, R., De Backer, G., Dominiczak, A., et al. (2013) · Journal of Hypertension · doi:10.1097/01.hjh.0000431740.32696.cc · Cited by 5,655
Prediction of Cardiovascular Events and All-Cause Mortality With Arterial Stiffness
Vlachopoulos, C., Aznaouridis, K., & Stefanadis, C. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.10.061 · Cited by 4,165 · cited 1×
“Moreover, aortic stiffness has gained independent predictive value for fatal and nonfatal cardiovascular events in hypertensive patients [1,5,6], and could thus reclassify intermediate risk patients into higher or lower cardiovascular risk [1,7,8].”
BackgroundLaurent, S., Boutouyrie, P., Asmar, R., Gautier, I., Laloux, B., Guize, L., Ducimetiere, P., & Benetos, A. (2001) · Hypertension · doi:10.1161/01.hyp.37.5.1236 · Cited by 3,856 · cited 2×
“Complior has been used extensively in epidemiologic studies in Europe and formed much of the available data, demonstrating the significant and independent contribution of PWV to cardiovascular morbidity and mortality [5,29,30].”
BackgroundMethodImpact of Aortic Stiffness on Survival in End-Stage Renal Disease
Blacher, J., Guerin, A. P., Pannier, B., Marchais, S. J., Safar, M. E., & London, G. M. (1999) · Circulation · doi:10.1161/01.cir.99.18.2434 · Cited by 2,138 · cited 2×
“Both noninvasively and PWVINV measurements were considered as reference method [54].”
BackgroundMethodVan Bortel, L. M., Laurent, S., Boutouyrie, P., Chowienczyk, P., Cruickshank, J. K., De Backer, T., Filipovsky, J., Huybrechts, S., Mattace-Raso, F. U. S., Protogerou, A. D., et al. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834fa8b0 · Cited by 2,010 · cited 3×
“80% of carotid–femoral direct distance is the most accurate way to assess path length (the ‘80% method’) [4,60].”
BackgroundMethodBoutouyrie, P., Tropeano, A. I., Asmar, R., Gautier, I., Benetos, A., Lacolley, P., & Laurent, S. (2002) · Hypertension · doi:10.1161/hy0102.099031 · Cited by 1,745 · cited 1×
“Complior has been used extensively in epidemiologic studies in Europe and formed much of the available data, demonstrating the significant and independent contribution of PWV to cardiovascular morbidity and mortality [5,29,30].”
MethodYAMASHINA, A., TOMIYAMA, H., TAKEDA, K., TSUDA, H., ARAI, T., HIROSE, K., KOJI, Y., HORI, S., & YAMAMOTO, Y. (2002) · Hypertension Research · doi:10.1291/hypres.25.359 · Cited by 1,623 · cited 9×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialCruickshank, K., Riste, L., Anderson, S. G., Wright, J. S., Dunn, G., & Gosling, R. G. (2002) · Circulation · doi:10.1161/01.cir.0000033824.02722.f7 · Cited by 1,531 · cited 2×
“With the ‘simultaneous technique’, two microphones are placed, respectively, to the left side of the neck (to insonate the left subclavian artery) and on the abdomen (to insonate the abdominal aorta above the bifurcation) and record simultaneously the pressure wave [51].”
BackgroundMethodTownsend, R. R., Wilkinson, I. B., Schiffrin, E. L., Avolio, A. P., Chirinos, J. A., Cockcroft, J. R., Heffernan, K. S., Lakatta, E. G., McEniery, C. M., Mitchell, G. F., et al. (2015) · Hypertension · doi:10.1161/hyp.0000000000000033 · Cited by 1,473 · cited 7×
“A first source of inaccuracy is body surface measurements, which may not represent the true aortic and arterial length [9,10].”
BackgroundMethodInfluentialThe velocity of pulse wave in man
Bramwell, J. C., & Hill, A. V. (1922) · Proceedings of the Royal Society of London. Series B, Containing Papers of a Biological Character · doi:10.1098/rspb.1922.0022 · Cited by 806 · cited 1×
“However, when compared with other ‘foot-to-foot’ algorithms in a one-dimensional arterial model, PWV obtained with the ITM shows poor agreement with the real PWV (analytically calculated by the Bramwell–Hill equation) [66].”
ResultPannier, B., Guérin, A. P., Marchais, S. J., Safar, M. E., & London, G. M. (2005) · Hypertension · doi:10.1161/01.hyp.0000159190.71253.c3 · Cited by 408 · cited 1×
“Complior has been used extensively in epidemiologic studies in Europe and formed much of the available data, demonstrating the significant and independent contribution of PWV to cardiovascular morbidity and mortality [5,29,30].”
MethodHorváth, I. G., Németh, Á., Lenkey, Z., Alessandri, N., Tufano, F., Kis, P., Gaszner, B., & Cziráki, A. (2010) · Journal of Hypertension · doi:10.1097/hjh.0b013e32833c8a1a · Cited by 374 · cited 5×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialAge-Associated Elongation of the Ascending Aorta in Adults
Sugawara, J., Hayashi, K., Yokoi, T., & Tanaka, H. (2008) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2008.06.010 · Cited by 327 · cited 2×
“80% of carotid–femoral direct distance is the most accurate way to assess path length (the ‘80% method’) [4,60].”
BackgroundMethodRisk prediction is improved by adding markers of subclinical organ damage to SCORE
Sehestedt, T., Jeppesen, J., Hansen, T. W., Wachtell, K., Ibsen, H., Torp-Petersen, C., Hildebrandt, P., & Olsen, M. H. (2010) · European Heart Journal · doi:10.1093/eurheartj/ehp546 · Cited by 315 · cited 1×
“Moreover, aortic stiffness has gained independent predictive value for fatal and nonfatal cardiovascular events in hypertensive patients [1,5,6], and could thus reclassify intermediate risk patients into higher or lower cardiovascular risk [1,7,8].”
BackgroundRajzer, M. W., Wojciechowska, W., Klocek, M., Palka, I., Brzozowska-Kiszka, M., & Kawecka-Jaszcz, K. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e32830a4a25 · Cited by 308 · cited 6×
“Similar findings were described by other authors when comparing the Complior SP and SphygmoCor devices [26,27].”
BackgroundResultInfluentialSalvi, P., Lio, G., Labat, C., Ricci, E., Pannier, B., & Benetos, A. (2004) · Journal of Hypertension · doi:10.1097/00004872-200412000-00010 · Cited by 295 · cited 6×
“Other recurrent discrepancies regarded the BMI of included patients with only two validation study considering BMI more than 30 kg/m(2) among the exclusion criteria [17,28] and sample size, that was lower than the recommended minimum of 83 individuals in nearly 40% of the studies [18,24,26,37,38,41, 52,53,55,56,58,59].”
BackgroundMethodInfluentialEvaluation of Carotid-Femoral Pulse Wave Velocity: Influence of Timing Algorithm and Heart Rate
Millasseau, S. C., Stewart, A. D., Patel, S. J., Redwood, S. R., & Chowienczyk, P. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000154229.97341.d2 · Cited by 289 · cited 11×
“Pulse transit time is determined, according to the Artery Society guidelines [9], using the ITM, the less influenced by changes in wave shape or wave reflection [13,14].”
BackgroundMethodInfluentialHametner, B., Wassertheurer, S., Kropf, J., Mayer, C., Eber, B., & Weber, T. (2013) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283614168 · Cited by 284 · cited 11×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodInfluentialWeber, T., Ammer, M., Rammer, M., Adji, A., O'Rourke, M. F., Wassertheurer, S., Rosenkranz, S., & Eber, B. (2009) · Journal of Hypertension · doi:10.1097/hjh.0b013e32832cb04e · Cited by 262 · cited 7×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialLarge Artery Stiffness Assessment Using SphygmoCor Technology
Butlin, M., & Qasem, A. (2016) · Pulse · doi:10.1159/000452448 · Cited by 244
Carotid and Aortic Stiffness: Determinants of Discrepancies
Paini, A., Boutouyrie, P., Calvet, D., Tropeano, A.-I., Laloux, B., & Laurent, S. (2006) · Hypertension · doi:10.1161/01.hyp.0000202052.25238.68 · Cited by 244 · cited 1×
“In animals, carotid-PWV is 2–3m/s higher than in the aortic arch [63], and in humans differences between aortic and carotid stiffness are higher in diabetic and hypertensive patients [64].”
BackgroundRegional pulse-wave velocity in the arterial tree.
McDonald, D. A. (1968) · Journal of Applied Physiology · doi:10.1152/jappl.1968.24.1.73 · Cited by 222 · cited 1×
“In animals, carotid-PWV is 2–3m/s higher than in the aortic arch [63], and in humans differences between aortic and carotid stiffness are higher in diabetic and hypertensive patients [64].”
BackgroundAge-associated changes in regional aortic pulse wave velocity
Rogers, W. J., Hu, Y.-L., Coast, D., Vido, D. A., Kramer, C. M., Pyeritz, R. E., & Reichek, N. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(01)01504-2 · Cited by 219 · cited 9×
“[56], compared PWVMRI measured in three aortic regions (proximal, midthoracic, abdominal) with tonometric PWV: when PWVMRI was averaged over the three locations, good agreement was found with non-PWVINV, although the proximal segment showed the best relationship between aging and increasing arterial stiffness.”
BackgroundMethodResultInfluentialVerbeke, F., Van Biesen, W., Honkanen, E., Wikström, B., Jensen, P. B., Krzesinski, J.-M., Rasmussen, M., Vanholder, R., & Rensma, P. L. (2011) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.05120610 · Cited by 218 · cited 1×
“The SphygmoCor device has been used in the AngloCardiff Collaborative Study to assess the correlation between cardiovascular risk factors and PWV [20] and the Chronic Renal Insufficiency Cohort study of chronic kidney disease [21], as well as other cohorts and intervention studies [22,23].”
MethodAortic PWV in Chronic Kidney Disease: A CRIC Ancillary Study
Townsend, R. R., Wimmer, N. J., Chirinos, J. A., Parsa, A., Weir, M., Perumal, K., Lash, J. P., Chen, J., Steigerwalt, S. P., Flack, J., et al. (2010) · American Journal of Hypertension · doi:10.1038/ajh.2009.240 · Cited by 218 · cited 1×
“The SphygmoCor device has been used in the AngloCardiff Collaborative Study to assess the correlation between cardiovascular risk factors and PWV [20] and the Chronic Renal Insufficiency Cohort study of chronic kidney disease [21], as well as other cohorts and intervention studies [22,23].”
MethodMcEniery, C. M., Spratt, M., Munnery, M., Yarnell, J., Lowe, G. D., Rumley, A., Gallacher, J., Ben-Shlomo, Y., Cockcroft, J. R., & Wilkinson, I. B. (2010) · Hypertension · doi:10.1161/hypertensionaha.110.150896 · Cited by 211 · cited 1×
“The SphygmoCor device has been used in the AngloCardiff Collaborative Study to assess the correlation between cardiovascular risk factors and PWV [20] and the Chronic Renal Insufficiency Cohort study of chronic kidney disease [21], as well as other cohorts and intervention studies [22,23].”
MethodClinical Significance and Reproducibility of New Arterial Distensibility Index
Kubozono, T., Miyata, M., Ueyama, K., Nagaki, A., Otsuji, Y., Kusano, K., Kubozono, O., & Tei, C. (2007) · Circulation Journal · doi:10.1253/circj.71.89 · Cited by 208 · cited 1×
“PWV is estimated by dividing the distance from the aortic valve to the ankle cuff by the sum of the difference between the time the pulse waves are transmitted to the brachial and the ankle cuffs, and the time difference between the second heart sound on the phonocardiogram and the notch of the brachial waves [45].”
MethodValidation and reproducibility of aortic pulse wave velocity as assessed with velocity‐encoded MRI
Grotenhuis, H. B., Westenberg, J. J. M., Steendijk, P., van der Geest, R. J., Ottenkamp, J., Bax, J. J., Jukema, J. W., & de Roos, A. (2009) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21886 · Cited by 204 · cited 10×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialHwang, M.-H., Yoo, J.-K., Kim, H.-K., Hwang, C.-L., Mackay, K., Hemstreet, O., Nichols, W. W., & Christou, D. D. (2014) · Journal of Human Hypertension · doi:10.1038/jhh.2013.144 · Cited by 202 · cited 16×
“On the other hand, three studies (SphygmoCor Xcel [18], Arteriograph [38] and Doppler ultrasound [53]) showed a small sample size even though the reference standard was non-PWVINV assessment.”
BackgroundMethodResultInfluentialWilkinson, I. B., McEniery, C. M., Schillaci, G., Boutouyrie, P., Segers, P., Donald, A., & Chowienczyk, P. J. (2010) · Artery Research · doi:10.1016/j.artres.2010.03.001 · Cited by 202 · cited 9×
“The Artery society guidelines [9] suggest that in every validation study two investigators should be employed (alternating the devices between individuals) to assess inter-operator variability.”
BackgroundMethodInfluentialValidity and repeatability of the Vicorder apparatus: a comparison with the SphygmoCor device
Hickson, S. S., Butlin, M., Broad, J., Avolio, A. P., Wilkinson, I. B., & McEniery, C. M. (2009) · Hypertension Research · doi:10.1038/hr.2009.154 · Cited by 189 · cited 15×
“Moreover, Obeid et al. [50] proved that the best correlations relating ftPWV and cf-PWV, were obtained with the MSDA.”
BackgroundMethodResultInfluentialLuzardo, L., Lujambio, I., Sottolano, M., da Rosa, A., Thijs, L., Noboa, O. A., Staessen, J. A., & Boggia, J. G. (2012) · Hypertension Research · doi:10.1038/hr.2012.78 · Cited by 168 · cited 4×
“Noninvasive validation was performed in 14 studies [17,18,24,27,28,34,38,39,48,50,53,56,58], with SphygmoCor, Complior, PulsePen or general tonometry as reference standards (Table 3).”
BackgroundMethodInfluentialComparative study of methodologies for pulse wave velocity estimation
Salvi, P., Magnani, E., Valbusa, F., Agnoletti, D., Alecu, C., Joly, L., & Benetos, A. (2008) · Journal of Human Hypertension · doi:10.1038/jhh.2008.42 · Cited by 160 · cited 1×
“It is based on the intersection of the horizontal line crossing the lowest point of the pressure waveform with the interpolant line of the early protosystolic phase [24,25].”
BackgroundTurin, T. C., Kita, Y., Rumana, N., Takashima, N., Kadota, A., Matsui, K., Sugihara, H., Morita, Y., Nakamura, Y., Miura, K., et al. (2010) · Hypertension Research · doi:10.1038/hr.2010.103 · Cited by 160 · cited 1×
“In prospective observational studies ba-PWV proved its independent predictive value in loss of kidney function [42], CVD [43] and all cause death [44].”
BackgroundDoppler ultrasound in the measurement of pulse wave velocity: agreement with the Complior method
Martínez, J. C., Torguet, P., García-Gil, M., Garcia, I., Martín, N., Guasch, B., Faur, D., & Vallés, M. (2011) · Cardiovascular Ultrasound · doi:10.1186/1476-7120-9-13 · Cited by 147
Sarafidis, P. A., Loutradis, C., Καρπέτας, Α., Tzanis, G. S., Piperidou, A., Koutroumpas, G., Raptis, V., Syrgkanis, C., Liakopoulos, V., Efstratiadis, G., et al. (2017) · Hypertension · doi:10.1161/hypertensionaha.117.09023 · Cited by 139 · cited 1×
“aPWV measured with Mobil-O-Graph independently predicted mortality in 135 patients with chronic kidney disease [35]; in a recent study, ambulatory PWV has been associated with cardiovascular events risk and all-cause mortality in 170 hemodialysis patients [36].”
MethodWesterhof, B. E., van den Wijngaard, J. P. H. M., Murgo, J. P., & Westerhof, N. (2008) · Hypertension · doi:10.1161/hypertensionaha.108.116525 · Cited by 135 · cited 1×
“Using a transfer function, central pressure curves are obtained and processed using the ARCSolver algorithm (Austrian Institute of Technology, Vienna, Austria) [16,31]; various parameters from pulse wave analysis to wave separation analysis are incorporated in a mathematical model, that combines age, central pressure…”
MethodWeber, T., Wassertheurer, S., Hametner, B., Parragh, S., & Eber, B. (2015) · Journal of Hypertension · doi:10.1097/hjh.0000000000000518 · Cited by 131 · cited 9×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialMeasurement of Local Pulse Wave Velocity: Effects of Signal Processing on Precision
Hermeling, E., Reesink, K. D., Reneman, R. S., & Hoeks, A. P. G. (2007) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2006.11.018 · Cited by 124 · cited 9×
“Pulse transit time is determined, according to the Artery Society guidelines [9], using the ITM, the less influenced by changes in wave shape or wave reflection [13,14].”
MethodInfluentialAortic pulse wave velocity predicts mortality in chronic kidney disease stages 2–4
Baumann, M., Wassertheurer, S., Suttmann, Y., Burkhardt, K., & Heemann, U. W. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000113 · Cited by 120 · cited 2×
“Mobil-O-Graph showed independent prognostic value in a small study in patients with chronic kidney disease [35], whereas outcome studies with Arteriograph are, currently, absent.”
BackgroundMethodButlin, M., Qasem, A., Battista, F., Bozec, E., McEniery, C. M., Millet-Amaury, E., Pucci, G., Wilkinson, I. B., Schillaci, G., Boutouyrie, P., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363c789 · Cited by 115
Tomiyama, H., Tanaka, H., Hashimoto, H., Matsumoto, C., Odaira, M., Yamada, J., Yoshida, M., Shiina, K., Nagata, M., & Yamashina, A. (2010) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2010.05.033 · Cited by 114 · cited 1×
“In prospective observational studies ba-PWV proved its independent predictive value in loss of kidney function [42], CVD [43] and all cause death [44].”
BackgroundPapaioannou, T. G., Argyris, A., Protogerou, A. D., Vrachatis, D. A., Nasothimiou, E., Sfikakis, P. P., Stergiou, G. S., & Stefanadis, C. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2013.08.079 · Cited by 114 · cited 1×
“Using a transfer function, central pressure curves are obtained and processed using the ARCSolver algorithm (Austrian Institute of Technology, Vienna, Austria) [16,31]; various parameters from pulse wave analysis to wave separation analysis are incorporated in a mathematical model, that combines age, central pressure…”
MethodAortic Stiffness and Central Wave Reflections Predict Outcome in Renal Transplant Recipients
Verbeke, F., Maréchal, C., Laecke, S. V., Biesen, W. V., Devuyst, O., Bortel, L. M. V., Jadoul, M. Y., & Vanholder, R. (2011) · Hypertension · doi:10.1161/hypertensionaha.111.176594 · Cited by 112 · cited 1×
“The SphygmoCor device has been used in the AngloCardiff Collaborative Study to assess the correlation between cardiovascular risk factors and PWV [20] and the Chronic Renal Insufficiency Cohort study of chronic kidney disease [21], as well as other cohorts and intervention studies [22,23].”
MethodRing, M., Eriksson, M. J., Zierath, J. R., & Caidahl, K. (2014) · Hypertension Research · doi:10.1038/hr.2014.115 · Cited by 104 · cited 9×
“On the other hand, three studies (SphygmoCor Xcel [18], Arteriograph [38] and Doppler ultrasound [53]) showed a small sample size even though the reference standard was non-PWVINV assessment.”
BackgroundMethodResultInfluentialMatsuoka, O., Otsuka, K., Murakami, S., Hotta, N., Yamanaka, G., Kubo, Y., Yamanaka, T., Shinagawa, M., Nunoda, S., Nishimura, Y., et al. (2005) · Biomedicine & Pharmacotherapy · doi:10.1016/s0753-3322(05)80008-3 · Cited by 92 · cited 4×
“Keywords: accuracy, arterial stiffness, pulse wave velocity Abbreviations: aPWV, aortic pulse wave velocity; baPWV, brachial-ankle pulse wave velocity; cf-PWV, carotid– femoral pulse wave velocity; cf-TT, carotid–femoral transit time; CVD, cardiovascular disease; ft-PTT, finger–toe pulse transit time; ft-PWV, finger–t…”
BackgroundInfluentialWestenberg, J. J. M., van Poelgeest, E. P., Paul Steendijk, A., Grotenhuis, H. B., Jukema, J. W., & Roos, A. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-2 · Cited by 90 · cited 7×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialSehestedt, T. B., Jeppesen, J., Hansen, T. W., Rasmussen, S. L., Wachtell, K., Ibsen, H., Torp-Pedersen, C. T., & Olsen, M. H. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e328356c579 · Cited by 87 · cited 1×
“Moreover, aortic stiffness has gained independent predictive value for fatal and nonfatal cardiovascular events in hypertensive patients [1,5,6], and could thus reclassify intermediate risk patients into higher or lower cardiovascular risk [1,7,8].”
BackgroundVardoulis, O., Papaioannou, T. G., & Stergiopulos, N. (2013) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00963.2012 · Cited by 85 · cited 1×
“Currently, several algorithms are used, resulting in a 5– 15% difference in PWV measurements [67].”
MethodDogui, A., Redheuil, A. B., Lefort, M., Cesare, A. D., Kachenoura, N., Herment, A., & Mousseaux, É. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22570 · Cited by 84 · cited 8×
“A three-dimensional volumetric acquisition (performed by manual annotation of three centrelines), on the other hand, is more accurate due to the effects of out-of-plane curvature [58,60], but also challenging and time consuming [61].”
BackgroundMethodResultInfluentialDistance measurements for the assessment of carotid to femoral pulse wave velocity
Vermeersch, S. J., Rietzschel, E. R., Buyzere, M. D., Bortel, L. M. V., Gillebert, T. C., Verdonck, P., Laurent, S., Segers, P., & Boutouyrie, P. (2009) · Journal of Hypertension · doi:10.1097/hjh.0b013e3283313a8a · Cited by 84 · cited 1×
“SphygmoCor uses a tonometric sensor placed on a pen-like support acquiring pressure waves first at the carotid site and then at the femoral site [12].”
MethodGrillo, A., Parati, G., Rovina, M., Moretti, F., Salvi, L., Gao, L., Baldi, C., Sorropago, G., Faini, A., Millasseau, S., et al. (2017) · American Journal of Hypertension · doi:10.1093/ajh/hpx140 · Cited by 73 · cited 3×
“[62] corroborate these disparities, proving that the influence of transit time in PWV assessment is higher at high levels of arterial stiffness, and differences in sampling rate may contribute to this trend [62].”
BackgroundObeid, H., Khettab, H., Marais, L., Hallab, M., Laurent, S., & Boutouyrie, P. (2017) · Journal of Hypertension · doi:10.1097/hjh.0000000000001371 · Cited by 64 · cited 6×
“Noninvasive validation was performed in 14 studies [17,18,24,27,28,34,38,39,48,50,53,56,58], with SphygmoCor, Complior, PulsePen or general tonometry as reference standards (Table 3).”
BackgroundMethodResultInfluentialAlivon, M., Phuong, T. V.-D., Vignon, V., Bozec, E., Khettab, H., Hanon, O., Briet, M. B., Halimi, J., Hallab, M., Plichart, M., et al. (2015) · Archives of cardiovascular diseases · doi:10.1016/j.acvd.2014.12.003 · Cited by 63 · cited 9×
“Noninvasive validation was performed in 14 studies [17,18,24,27,28,34,38,39,48,50,53,56,58], with SphygmoCor, Complior, PulsePen or general tonometry as reference standards (Table 3).”
BackgroundMethodResultInfluentialDavies, J. M., Bailey, M. A., Griffin, K. J., & Scott, D. J. A. (2012) · Vascular · doi:10.1258/vasc.2011.ra0054 · Cited by 61 · cited 8×
“Moreover, Obeid et al. [50] proved that the best correlations relating ftPWV and cf-PWV, were obtained with the MSDA.”
BackgroundMethodInfluentialArterial Stiffness: From Physiology to Clinical Implications
Milan, A., Tosello, F., Fabbri, A., Vairo, A., Leone, D., Chiarlo, M., Covella, M., & Veglio, F. (2011) · High Blood Pressure & Cardiovascular Prevention · doi:10.2165/11588020-000000000-00000 · Cited by 58
Self-citationAscending aortic dilatation, arterial stiffness and cardiac organ damage in essential hypertension
Milan, A., Tosello, F., Naso, D., Avenatti, E., Leone, D., Magnino, C., & Veglio, F. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835aa588 · Cited by 58
Self-citationvan Leeuwen-Segarceanu, E. M., Tromp, W. F., Bos, W. J. W., Vogels, O. J. M., Groothoff, J. W., & van der Lee, J. H. (2010) · Journal of Hypertension · doi:10.1097/hjh.0b013e32833a8b83 · Cited by 57 · cited 1×
“Using different path lengths may lead to different results even when the same device is evaluated [39,40,65].”
BackgroundStea, F., Bozec, E., Millasseau, S., Khettab, H., Boutouyrie, P., & Laurent, S. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000091 · Cited by 56 · cited 17×
“Pulse transit time is determined, according to the Artery Society guidelines [9], using the ITM, the less influenced by changes in wave shape or wave reflection [13,14].”
BackgroundMethodResultInfluentialKato, A., Takita, T., Furuhashi, M., Maruyama, Y., Miyajima, H., & Kumagai, H. (2012) · Therapeutic Apheresis and Dialysis · doi:10.1111/j.1744-9987.2012.01058.x · Cited by 54 · cited 1×
“This device has been used in Japanese individuals either as a predictor of cardiovascular outcomes (although ba-PWV was superior) [46], and cognitive impairment [47].”
MethodYamamoto, N., Yamanaka, G., Ishikawa, M., Takasugi, E., Murakami, S., Yamanaka, T., Ishine, M., Matsubayashi, K., Hanafusa, T., & Otsuka, K. (2009) · Dementia and Geriatric Cognitive Disorders · doi:10.1159/000235642 · Cited by 45 · cited 1×
“This device has been used in Japanese individuals either as a predictor of cardiovascular outcomes (although ba-PWV was superior) [46], and cognitive impairment [47].”
MethodStyczyński, G., Rdzanek, A., Pietrasik, A., Kochman, J., Huczek, Z., Sobieraj, P., Gaciong, Z. A., & Szmigielski, C. A. (2016) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2016.07.013 · Cited by 38 · cited 10×
“PWVINV assessment was compared with SphygmoCor [15,16], Mobil-O-Graph [32] (and the ARCSolver algorithm [16]), Arteriograph [37], ba-PWV [41], Doppler ultrasound [52] and MRI [55,59].”
BackgroundMethodResultInfluentialvan Engelen, A., Vieira, M. S., Rafiq, I., Cecelja, M., Schneider, T., de Bliek, H., Figueroa, C. A., Hussain, T., Botnar, R. M., & Alastruey, J. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-017-0341-y · Cited by 17 · cited 2×
“For these reasons automated three centrelines extraction method has been developed and validated recently [61].”
MethodMorales, M. S., Cuffaro, P. E., Barochiner, J., Rada, M. A., Alfie, J., Aparicio, L. S., Marín, M. J., Galarza, C. R., & Waisman, G. D. (2015) · Artery Research · doi:10.1016/j.artres.2015.03.001 · Cited by 15 · cited 17×
“Pulse transit time is determined, according to the Artery Society guidelines [9], using the ITM, the less influenced by changes in wave shape or wave reflection [13,14].”
BackgroundMethodInfluential
Cited by
How to Measure Arterial Stiffness in Humans
Segers, P., Rietzschel, E. R., & Chirinos, J. A. (2020) · Arteriosclerosis, Thrombosis, and Vascular Biology · doi:10.1161/atvbaha.119.313132 · Cited by 248
Arterial stiffness and vascular aging: mechanisms, prevention, and therapy
Herzog, M. J., Müller, P., Lechner, K., Stiebler, M., Arndt, P., Kunz, M., Ahrens, D., Schmeißer, A., Schreiber, S., & Braun-Dullaeus, R. C. (2025) · Signal Transduction and Targeted Therapy · doi:10.1038/s41392-025-02346-0 · Cited by 189
“28,156 The latter is clinically very well validated and one of the most reliable devices for measuring PWV. 156,331 The simultaneous version uses the foot-to-foot method, which illustrates the beginning of the pressure curve rise at the end of diastole as a foot.”
Ning, R., Li, Y., Du, Z., Li, T., Sun, Q., Lin, L., Xu, Q., Duan, J., & Sun, Z. (2021) · Redox Biology · doi:10.1016/j.redox.2021.102113 · Cited by 96
“PWV is the gold standard used for evaluating arterial stiffness [37].”
Metadata
- Authors
- Milan, A., Zocaro, G., Leone, D., Tosello, F., Buraioli, I., Schiavone, D., & Veglio, F.
- Venue
- Journal of Hypertension
- Published
- 2019-08
- Volume
- 37
- Issue
- 8
- Pages
- 1547-1557
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2922842315
- Cited by
- 236
- References
- 65
- FWCI
- 13.86
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.