Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: The Vicorder is a new brachial cuff-based device that estimates central blood pressure (cBP) using a brachial-to-aortic transfer function. The aim of this study was to evaluate cBP estimated by the Vicorder. METHODS: During cardiac angiography, cBP estimated by the Vicorder and the SphygmoCor was evaluated against simultaneous invasive cBP in 50 patients. The two devices were also compared noninvasively in a separate group of 90 healthy individuals. RESULTS: Central SBP (cSBP) obtained with each device satisfied the American Association for the Advancement of Medical Instrumentation accuracy criteria when peripheral waveforms were calibrated to invasive mean arterial pressure (MAP)/DBP: estimated - invasive cSBP difference, -4.0 ± 7.4 mmHg, Vicorder, P < 0.001; -1.4 ± 7.9 mmHg, SphygmoCor, P = 0.21. When oscillometric brachial SBP/DBP was used for peripheral waveform calibration, cSBP was underestimated by Vicorder (Δ -6.4 ± 7.4 mmHg, P < 0.001 versus invasive) and more so by SphygmoCor (Δ -11.9 ± 7.2 mmHg, P < 0.001 versus invasive). Conversely, cSBP was more closely estimated by SphygmoCor when waveforms were calibrated to brachial MAP/DBP (Δ -2.8 ± 9.4 mmHg, P = 0.04 versus invasive). In the noninvasive study, Vicorder cSBP correlated well with SphygmoCor cSBP when SphygmoCor waveforms were calibrated to brachial MAP/DBP (121 ± 16 versus 121 ± 17 mmHg, P = 0.2) but not when brachial SBP/DBP was used for calibration (115 ± 19 mmHg, P < 0.001). CONCLUSION: The Vicorder and SphygmoCor devices provide reliable estimates of cSBP when calibrated to invasive pressure. When calibrated to brachial BP, both devices underestimated cSBP, although this was attenuated when SphygmoCor was calibrated to brachial MAP/DBP. Vicorder may be a simple alternative to tonometry-based methods for noninvasive assessment of cBP.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785 · cited 1×
“The overall strength of agreement between estimated and measured cBP values was also quantified according to Lin’s concordance correlation coefficient [23], whereas the Pitman–Morgan test was used to test the difference between correlated variances [24].”
MethodMcDonald's Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles
Vlachopoulos, C., O'Rourke, M., & Nichols, W. W. (2011) · doi:10.1201/b13568 · Cited by 1,613
Vlachopoulos, C., Aznaouridis, K., O'Rourke, M. F., Safar, M. E., Baou, K., & Stefanadis, C. (2010) · European Heart Journal · doi:10.1093/eurheartj/ehq024 · Cited by 1,363 · cited 2×
“(cSBP) and PP (cPP), and the differences with the corresponding brachial values, named SBP and PP amplification, may have prognostic relevance and refine individual cardiovascular risk stratification [10–13].”
BackgroundChen, C.-H., Nevo, E., Fetics, B., Pak, P. H., Yin, F. C. P., Maughan, W. L., & Kass, D. A. (1997) · Circulation · doi:10.1161/01.cir.95.7.1827 · Cited by 1,307
Pauca, A. L., O’Rourke, M. F., & Kon, N. D. (2001) · Hypertension · doi:10.1161/hy1001.096106 · Cited by 1,179
KARAMANOGLU, M., O'ROURKE, M. F., AVOLIO, A. P., & KELLY, R. P. (1993) · European Heart Journal · doi:10.1093/eurheartj/14.2.160 · Cited by 873 · cited 1×
“We showed that the brachial cuff-based approach used by the Vicorder device provided results generally in line with those obtained with the SphygmoCor device, which has been demonstrated to estimate cSBP with high accuracy compared with invasively measured cSBP [6,8,9] when radial waveforms are calibrated to invasive…”
Cited by
Vlachopoulos, C., Xaplanteris, P., Aboyans, V., Brodmann, M., Cífková, R., Cosentino, F., De Carlo, M., Gallino, A., Landmesser, U., Laurent, S., et al. (2015) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2015.05.007 · Cited by 716
Self-citationCentral blood pressure: current evidence and clinical importance
McEniery, C. M., Cockcroft, J. R., Roman, M. J., Franklin, S. S., & Wilkinson, I. B. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht565 · Cited by 678
Self-citationBissell, M. M., Hess, A. T., Biasiolli, L., Glaze, S. J., Loudon, M., Pitcher, A., Davis, A., Prendergast, B., Markl, M., Barker, A. J., et al. (2013) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.113.000528 · Cited by 380
Jennings, A., Berendsen, A. M., de Groot, L. C. P. G. M., Feskens, E. J. M., Brzozowska, A., Sicinska, E., Pietruszka, B., Meunier, N., Caumon, E., Malpuech-Brugère, C., et al. (2019) · Hypertension · doi:10.1161/hypertensionaha.118.12259 · Cited by 172
Cicero, A. F. G., Fogacci, F., Desideri, G., Grandi, E., Rizzoli, E., D’Addato, S., & Borghi, C. (2019) · Nutrients · doi:10.3390/nu11112674 · Cited by 37
Metadata
- Authors
- Pucci, G., Cheriyan, J., Hubsch, A., Hickson, S. S., Gajendragadkar, P. R., Watson, T., O'Sullivan, M., Woodcock-Smith, J., Schillaci, G., Wilkinson, I. B., & McEniery, C. M.
- Venue
- Journal of Hypertension
- Published
- 2013-01
- Volume
- 31
- Issue
- 1
- Pages
- 77-85
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2006645451
- Cited by
- 128
- References
- 7
- FWCI
- 6.14
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure. https://app.cochat.ai/papers/10.1097/hjh.0b013e32835a8eb1
- The paper
- Pucci, G., Cheriyan, J., Hubsch, A., Hickson, S. S., Gajendragadkar, P. R., Watson, T., O'Sullivan, M., Woodcock-Smith, J., Schillaci, G., Wilkinson, I. B., & McEniery, C. M. (2013). Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure. Journal of Hypertension. https://doi.org/10.1097/hjh.0b013e32835a8eb1
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.