Accuracy of Blood Pressure Measurement Devices in Pregnancy: A Systematic Review of Validation Studies

Open access
Bello, Natalie A., Woolley, Jonathan J., Cleary, Kirsten Lawrence, Falzon, Louise, Alpert, Bruce S., Oparil, Suzanne, Cutter, Gary, Wapner, Ronald, Muntner, Paul, Tita, Alan T., Shimbo, Daichi · 2018 · Hypertension
Cited by 126

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

The accurate measurement of blood pressure (BP) in pregnancy is essential to guide medical decision making that affects both mother and fetus. The aim of this systematic review was to determine the accuracy of ambulatory, home, and clinic BP measurement devices in pregnant women. We searched Ovid MEDLINE, The Cochrane Library, EMBASE, CINAHL EBSCO, ClinicalTrials.gov, International Clinical Trials Registry Platform, and dabl from inception through August 3, 2017 for articles that assessed the validity of an upper arm BP measurement device against a mercury sphygmomanometer in pregnant women. Two independent investigators determined eligibility, extracted data, and adjudicated protocol violations. From 1798 potential articles identified, 41, that assessed 28 devices, met the inclusion criteria. Most articles (n=32) followed a standard or modified American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization, British Hypertension Society, or European Society of Hypertension validation protocol. Several articles described the results of validation studies performed on >1 device (n=7) or in >1 population of pregnant women (n=12), comprising 64 pairwise validity assessments. The device was validated in 61% (32 of 52) of studies which used a standard or modified protocol. Only 34% (11 of 32) of the studies wherein the device was successfully validated were performed without a protocol violation. Given the implications of inaccurate BP measurement in pregnant women, healthcare providers should be aware of and try to use the BP measurement devices which have been properly validated in this population.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/hypertensionaha.117.10295

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 7 references

Cited by

See all 126

Metadata

Authors
Bello, N. A., Woolley, J. J., Cleary, K. L., Falzon, L., Alpert, B. S., Oparil, S., Cutter, G., Wapner, R., Muntner, P., Tita, A. T., & Shimbo, D.
Venue
Hypertension
Published
2018-02
Volume
71
Issue
2
Pages
326-335
ISSN
  • 0194-911X (ISSN-L)
  • 1524-4563
OpenAlex
W2772190048
Cited by
126
References
7
FWCI
5.62
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accuracy of Blood Pressure Measurement Devices in Pregnancy: A Systematic Review of Validation Studies. https://app.cochat.ai/papers/10.1161/hypertensionaha.117.10295
The paper
Bello, N. A., Woolley, J. J., Cleary, K. L., Falzon, L., Alpert, B. S., Oparil, S., Cutter, G., Wapner, R., Muntner, P., Tita, A. T., & Shimbo, D. (2018). Accuracy of Blood Pressure Measurement Devices in Pregnancy: A Systematic Review of Validation Studies. Hypertension. https://doi.org/10.1161/hypertensionaha.117.10295

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.