Oscillometric sphygmomanometers: a critical appraisal of current technology
The DOI matches its Crossref record.
Abstract
AIM: Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy. METHODS: A block diagram is presented covering the components from cuff to algorithm. The oscillometric waveform is described, considering factors that can alter its shape. Methods used to assess accuracy, including the potential use of simulators, are described. RESULTS AND DISCUSSION: The block diagram focuses attention on cuff, amplifier, signal processing and algorithm. The importance of correct cuff size is emphasized. Accuracy can be affected by the extraction of the oscillometric pulses and the interpolation to compensate for higher deflation rates. Modern electronic amplifiers are assumed to be stable and do not drift, an assumption largely untested. Crucial to accuracy is the algorithm, but there is no standard algorithm and limited theoretical basis, leading to significant measurement errors in groups of patients, even by approved devices. The causes are not well understood, but differences in oscillometric waveform shape between patient groups have been observed and may explain auscultatory-oscillometric differences. The ability of theoretical models to explain the effects of arterial stiffness on BP measurements is discussed. Validation remains statistical though steps have been taken to improve it. CONCLUSION: The indirect nature of BP measurement poses particularly problems for ensuring accuracy. Critical assessment has done much to improve standards, but a solid theoretical understanding of the technique has not been formulated and further work is required.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Some limitations of the oscillometric method lead to errors in comparison with manual auscultatory measurement (e.g., up to 10–15%) [60] in BP estimations that are either systematic, random or associated with clinical characteristics of specific patient populations [61].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [61]
Citations
Cites
Still being checked.
Cited by
Flynn, J. T., Daniels, S. R., Hayman, L. L., Maahs, D. M., McCrindle, B. W., Mitsnefes, M., Zachariah, J. P., & Urbina, E. M. (2014) · Hypertension · doi:10.1161/hyp.0000000000000007 · Cited by 594
Whalin, M. K., Halenda, K. M., Haussen, D. C., Rebello, L. C., Frankel, M., Gershon, R. Y., & Nogueira, R. G. (2016) · American Journal of Neuroradiology · doi:10.3174/ajnr.a4992 · Cited by 135
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00693-x · Cited by 84
Picone, D. S., Schultz, M. G., Peng, X., Black, J. A., Dwyer, N., Roberts-Thomson, P., Chen, C.-H., Cheng, H.-M., Pucci, G., Wang, J.-G., et al. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10696 · Cited by 47
Perioperative hemodynamic monitoring
Cove, M. E., & Pinsky, M. R. (2012) · Best practice & research. Clinical anaesthesiology · doi:10.1016/j.bpa.2012.10.003 · Cited by 32
Metadata
- Authors
- Amoore, J. N.
- Affiliations
- University Hospital Crosshouse (GB)
- Corresponding author
- Amoore, John Nicholas
- Venue
- Blood Pressure Monitoring
- Published
- 2012-04
- Volume
- 17
- Issue
- 2
- Pages
- 80-88
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2001232250
- Semantic Scholar
- b2766faf1f374edd3c14fd0766299f8515a3825d
- Cited by
- 49
- References
- 0
- Citations by year
- 2026: 2 · 2024: 3 · 2023: 2 · 2022: 3 · 2021: 4 · 2019: 4 · 2018: 4 · 2017: 5 · 2016: 6 · 2015: 2 · 2014: 4 · 2013: 6 · 2012: 3
- FWCI
- 2.73
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Blood Pressure Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- blood pressure measurement accuracy, cuff size, measurement algorithm, arterial stiffness, measurement error
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Sphygmomanometers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Oscillometric sphygmomanometers: a critical appraisal of current technology. https://app.cochat.ai/papers/10.1097/mbp.0b013e32835026b0
- The paper
- Amoore, J. N. (2012). Oscillometric sphygmomanometers: a critical appraisal of current technology. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e32835026b0
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.