Power Profiles of Multifocal Contact Lenses and Their Interpretation
The DOI matches its Crossref record; 49 of the 49 references it cites were found in a registry or an index.
Abstract
Purpose: Many contact lens (CL) manufacturers produce simultaneous‐image lenses in which power varies either smoothly or discontinuously with zonal radius. We present in vitro measurements of some recent CLs and discuss how power profiles might be approximated in terms of nominal distance corrections, near additions, and on‐eye visual performance. Methods: Fully hydrated soft, simultaneous‐image CLs from four manufacturers (Air Optix AQUA, Alcon; PureVision multifocal, Bausch & Lomb; Acuvue OASYS for Presbyopia, Vistakon; Biofinity multifocal‐ “D” design, Cooper Vision) were measured with a Phase focus Lens Profiler (Phase Focus Ltd., Sheffield, UK) in a wet cell and powers were corrected to powers in air. All lenses had zero labeled power for distance. Results: Sagittal power profiles revealed that the “low” add PureVision and Air Optix lenses exhibit smooth (parabolic) profiles, corresponding to negative spherical aberration. The “mid” and “high” add PureVision and Air Optix lenses have bi‐aspheric designs, leading to different rates of power change for the central and peripheral portions. All OASYS lenses display a series of concentric zones, separated by abrupt discontinuities; individual profiles can be constrained between two parabolically decreasing curves, each giving a valid description of the power changes over alternate annular zones. Biofinity lenses have constant power over the central circular region of radius 1.5 mm, followed by an annular zone where the power increases approximately linearly, the gradient increasing with the add power, and finally an outer zone showing a slow, linear increase in power with a gradient being almost independent of the add power. Conclusions: The variation in power across the simultaneous‐image lenses produces enhanced depth of focus. The through‐focus nature of the image, which influences the “best focus” (distance correction) and the reading addition, will vary with several factors, including lens centration, the wearer's pupil diameter, and ocular aberrations, particularly spherical aberration; visual performance with some designs may show greater sensitivity to these factors.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“accuracy in these measurements is key to realizing the maximum benefits of FFLs [75].”
From Freeform vs. Aspheric Spectacle Lenses: A Comprehensive Review of Optical Performance, Clinical Outcomes, and Patient Considerations (2026) · cites this paper as [75]
Citations
Cites
High-Resolution Scanning X-ray Diffraction Microscopy
Thibault, P., Dierolf, M., Menzel, A., Bunk, O., David, C., & Pfeiffer, F. (2008) · Science · doi:10.1126/science.1158573 · Cited by 1,321
Monochromatic aberrations of the human eye in a large population
Porter, J., Guirao, A., Cox, I. G., & Williams, D. R. (2001) · Journal of the Optical Society of America A · doi:10.1364/josaa.18.001793 · Cited by 711
(1994) · PubMed · Cited by 688
A unified formula for light-adapted pupil size
Watson, A. B., & Yellott, J. I. (2012) · Journal of Vision · doi:10.1167/12.10.12 · Cited by 399
Age-related changes in monochromatic wave aberrations of the human eye.
McLellan, J. S., Marcos, S. C., & Burns, S. A. (2001) · PubMed · Cited by 293
Superresolution imaging via ptychography
Maiden, A. M., Humphry, M. J., Zhang, F., & Rodenburg, J. M. (2011) · Journal of the Optical Society of America A · doi:10.1364/josaa.28.000604 · Cited by 256
Effect of Focus on the Visual Response to a Sinusoidally Modulated Spatial Stimulus*
Green, D. G., & Campbell, F. W. (1965) · Journal of the Optical Society of America · doi:10.1364/josa.55.001154 · Cited by 182
The effect of ocular aberrations on steady-state errors of accommodative response
Plainis, S., Ginis, H. S., & Pallikaris, A. (2005) · Journal of Vision · doi:10.1167/5.5.7 · Cited by 181
Self-citationOptical ptychography: a practical implementation with useful resolution
Maiden, A. M., Rodenburg, J. M., & Humphry, M. J. (2010) · Optics Letters · doi:10.1364/ol.35.002585 · Cited by 180
Applegate, R. A., Donnelly, W. J., Marsack, J. D., Koenig, D. E., & Pesudovs, K. (2007) · Journal of the Optical Society of America A · doi:10.1364/josaa.24.000578 · Cited by 163
Research works printed but not linked: 6
- Plainis S, 1996, Methods for verifying the power of soft contact lenses: a review, Optician, 212, 42
- Vogt A, 2011, Using power profiles to evaluate aspheric lenses, Contact Lens Spectrum, 26, 43
- Plainis S, 2011, Commenting on “Using power profiles to evaluate aspheric lenses”, Contact Lens Spectrum, 26, 15
- Atchison DA, 2012, Presbyopia: Origins, Effects and Treatment, 21
- Born M, 1993, Principles of Optics
- Charman WN, 1988, Retinal images with centred aspheric varifocal contact lenses, Int Contact Lens Clin, 15, 87
Cited by
Items 101–129 of 129
Rodríguez-López, V., Esteban-Ibañez, E., Gonzalez-Ramos, A. M., Marcos, S. C., & Martínez-Enríquez, E. (2025) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2025.102445 · Cited by 1
Mawhinney, A. J., Buckhurst, P. J., Craig Mcneile, A., Hall, S. D., & Buckhurst, H. D. (2025) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2025.102454 · Cited by 1
Tolerance to rotation of toric monofocal and bifocal intraocular lenses. A theoretical study
Águila‐Carrasco, A. J. D., Domínguez‐Vicent, A., Monsálvez-Romín, D., Esteve-Taboada, J. J., & Papadatou, E. (2017) · Optik · doi:10.1016/j.ijleo.2017.11.110 · Cited by 1
Liquid Crystal Spatial Light Modulators for Simulating Zonal Multifocal Lenses
Li, Y., Bradley, A., Xu, R., & Kollbaum, P. S. (2017) · Optometry and Vision Science · doi:10.1097/opx.0000000000001108 · Cited by 1
Power Profiles of an Extended Depth of Focus Contact Lens for Myopia Management
Noya-Padin, V., Mawhinney, A. J., Buckhurst, P. J., Pena‐Verdeal, H., Craig Mcneile, A., Hall, S. D., & Buckhurst, H. D. (2026) · Ophthalmic and Physiological Optics · doi:10.1007/s44402-026-00070-y · Cited by 1
Bausili, S. L. (2015) · RECERCAT (Consorci de Serveis Universitaris de Catalunya) · Cited by 1
Ocular Straylight and Artificial Lenses
Łabuz, G. (2017) · RePub (Erasmus University Rotterdam) · Cited by 1
Guía Clínica para la Adaptación de Lentes de Contacto Multifocales Tóricas Hidrofílicas
García Valencia, L. (2015) · UVaDOC UVaDOC University of Valladolid Documentary Repository (University of Valladolid) · Cited by 1
Optimisation des optiques bifocales destinées à la correction de la presbytie
Rio, D. (2016) · theses.fr (ABES) · doi:10.70675/ef2ee061ze047z4fdezb42bz923d838e8081 · Cited by 1
Ladopoulos, K., Pateras, E., & Ninos, G. (2026) · Cureus · doi:10.7759/cureus.104008
“…of benefit associated with an FFL can be attributed exclusively to measurement error in one or more of the fitting parameters (monocular PD, fitting height, vertex distance, pantoscopic tilt, face form) [74], and accuracy in these measurements is key to realizing the maximum benefits of FFLs [75].”
Mączyńska, E., Lyu, H. C., Karnowski, K., Grulkowski, I., & Wojtkowski, M. (2015) · Frontiers in Optics · doi:10.1364/fio.2015.jtu4a.33
Soldevila, L. G. (2014) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Pauné, J. (2015) · UPCommons institutional repository (Universitat Politècnica de Catalunya) · doi:10.5821/dissertation-2117-95987
Visual simulation of different optical designs
Papadatou, E. (2016) · Repository of Digital Objects for Teaching Research and Culture (University of Valencia)
New orthokeratology lens design for presbyopia
Travé‐Huarte, S. (2017) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Contact lens prescribing : USA vs. the world
Morgan, P. B., Efron, N., & Woods, C. A. (2012) · Deakin Research Online (Deakin University)
Using automated photo-ophthalmic device (APOD) technology in optometry practice
Barnard, S. (2017) · Optician · doi:10.12968/opti.2017.1.184
Cilvēka fizioloģijas un uztveres aktuālie pētījumi 2020. gadā : rakstu krājums
Ikaunieks, G., Krūmiņa, G., Švede, A., & Panke, K. (2020) · Latvijas Universitāte eBooks · doi:10.22364/cfuap.2020
Presbyopia and multifocal contact lenses 1: Attitudes and beliefs
Dumbleton, K., Laughton, D., & Palombi, J. (2021) · Optician · doi:10.12968/opti.2021.8.8702
On customized soft contact lenses
Faria‐Ribeiro, M. Á., & González‐Méijome, J. M. (2025) · Elsevier eBooks · doi:10.1016/b978-0-323-87929-3.00018-9
Specialty contact lenses for presbyopia: Multifocals
Barnett, M., & Johnson, L. (2025) · Elsevier eBooks · doi:10.1016/b978-0-323-87929-3.00023-2
Teren, D., & ALIO, J. L. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-13983-3.00023-7
Impact of ocular aberrations and pupil diameter on vision with four daily multifocal contact lenses
Plaza, E. M., Molina‐Martín, A., Mena-Guevara, K. J., Sáiz, D. D. F., Gómez‐Vicente, V., & Pinero, D. P. (2025) · Clinical and Experimental Optometry · doi:10.1080/08164622.2025.2565457
“The observed benefits may stem from two mechanisms: (1) enhanced adaptation to environmental light changes, and (2) modulation of the effective optical power profile of MCLs 5,6 as the pupil constricts and dilates.”
Evaluation Method for Resin Mold Using Reflective Wavefront Sensor
Tatsumi, K., Saeki, K., Kubota, S., Kaneda, Y., Uno, K., Ohnuma, K., & Shiina, T. (2025) · Sensors · doi:10.3390/s25216682
Influência do diâmetro pupilar no rendimento visual e ótico das lentes de contacto multifocais
Isla-Paradelo, L. (2014) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Valoración de los protocolos de adaptación de tres tipos de lentes de contacto multifocales
Giménez Ruiz, A. (2017) · UPCommons institutional repository (Universitat Politècnica de Catalunya)
Pérez‐prados, R. (2017) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Plainis, S. (2026) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2026.102702
Self-citationOptical Principles and Characteristics of Presbyopia-correcting Multifocal Contact Lenses
Park, H.-S., Kwon, N. E., & Bang, S. P. (2026) · Annals of Optometry and Contact Lens · doi:10.52725/aocl.2026.25.3.109
Metadata
- Authors
- Plainis, S., Atchison, D. A., & Charman, W. N.
- Affiliations
- University of Crete (GR)
- Corresponding author
- Plainis, Sotiris
- Venue
- Optometry and Vision Science
- Publisher
- Wiley
- Published
- 2013-10
- Volume
- 90
- Issue
- 10
- Pages
- 1066-1077
- ISSN
- 1040-5488 (ISSN-L)
- 1538-9235
- OpenAlex
- W2110300024
- Cited by
- 115
- References
- 48
- Citations by year
- 2026: 3 · 2025: 11 · 2024: 10 · 2023: 7 · 2022: 3 · 2021: 8 · 2020: 6 · 2019: 11 · 2018: 9 · 2017: 19 · 2016: 16 · 2015: 9 · 2014: 2 · 2013: 1
- FWCI
- 6.80
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- onlinelibrary.wiley.com/termsAndConditions#vor (vor)
- doi.wiley.com/10.1002/tdm_license_1.1 (tdm)
- Locations
- Optometry and Vision Science (publishedVersion)
- PubMed (publishedVersion)
- Research Explorer (The University of Manchester) (submittedVersion)
- Topics
- Health Sciences › Medicine › Ophthalmology › Ophthalmology and Visual Impairment Studies
- Health Sciences › Medicine › Ophthalmology › Ocular Surface and Contact Lens
- Health Sciences › Medicine › Ophthalmology › Intraocular Surgery and Lenses
- Keywords
- multifocal contact lenses, power profile, spherical aberration, depth of focus, near addition, distance correction, lens centration, pupil size, ocular aberrations
- MeSH terms
- Contact Lenses, Equipment Design, Humans, Presbyopia, Optics and Photonics, In Vitro Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.