The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index; 6 of the 15 sources its claims cite were checked for support.
Filtering voluntary motion for pathological tremor compensation
Open accessAbstract
This paper describes an algorithm to estimate tremor and voluntary motion from measured motion data. Estimation is performed by means of an extended Kalman filter (EKF), which also estimates tremor parameters. Comparison of the proposed method with techniques described in the literature are conducted with two experimental data sets from tremor patients performing the same task, drawing a spiral. The presented algorithm may be directly applied in real-time pathological tremor compensation systems.
Overview
CoChat Summary
TL;DR
The paper presents an extended Kalman filter (EKF) that simultaneously estimates pathological tremor and voluntary motion from measured motion data while also estimating tremor parameters. In two patient data sets, the EKF is reported to be more robust across different sensors and motions than WFLC and BBF+WFLC, making it a candidate for real-time tremor compensation.
Key findings
- The proposed EKF estimates both tremor and voluntary motion, with tremor modeled as a harmonic signal and voluntary motion modeled as low-pass filtered white noise; it also recursively estimates tremor parameters.
- In an angular-rate-sensor data set from a Parkinson’s disease patient with tremor frequency around 7.5 Hz, the EKF and WFLC gave similar voluntary-motion estimates, but the WFLC estimate was noisier; the BBF+WFLC method did not effectively filter tremor with the applied gains.
- In a digitizing-tablet data set from an essential tremor patient with tremor frequency around 6.8 Hz, the WFLC did not correctly estimate tremor and propagated error into voluntary-motion estimation, while BBF+WFLC was excessively smoothed and delayed; the EKF reportedly recovered the correct drawing without tremor components.
- The EKF’s performance was described as accurate and insensitive to the two different data sets, whereas WFLC and BBF+WFLC were sensitive to voluntary-motion dynamics, amplitude ratio, and sampling frequency.
Methods
- Design: an EKF-based recursive estimator with a harmonic tremor model and a low-pass voluntary-motion model; the filter states combine tremor motion, tremor model parameters, and voluntary motion states.
- Data: two experimental data sets from two tremor patients performing the same spiral-drawing task, measured with an IDG300 angular rate sensor in one case and a Wacom Intuos3 digitizing tablet with 5,080 lpi resolution in the other.
- Analysis: results were compared with WFLC and BBF+WFLC on a visual basis; the same algorithm parameters were used for both data sets except for the initial frequency estimate. EKF parameters included H = 1, σ²_st = 5·10⁻⁸, σ²_ω = 5·10⁻⁴, σ²_{a_h,b_h} = 5·10⁻³, F = 2, and σ²_si = 5·10⁻²; the voluntary-motion low-pass cutoff was 2 Hz.
Limitations
- The paper states that a perfect reference voluntary motion is difficult or impossible to measure, so quantitative comparison was difficult and the comparison between methods was made on a visual basis.
- The paper states that the performance of the proposed algorithm with different parameters was not deeply evaluated, and that tracking of tremor frequency was not compared because it was mainly addressed in prior work.
- The paper notes the algorithm is more complex than the other methods, though it argues this is negligible with current computational units.
- Evident from the text: the evaluation uses only two patients and two data sets, with no quantitative error metrics, statistical tests, or ground-truth voluntary motion.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
3 Valid · 8 Not valid · 6 Not checked · 18 Own results and interpretations
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“From Fig. 4, it is shown that the result was excessively smoothed and delayed.”
What the sources say
- Disputedcites [2] [2] E. Rocon, J. M. Belda-Lois, A. F. Ruiz, M. Manto, J. C. Moreno, and J. L. Pons, 'Design and validation of a rehabilitation robotic exoskeleton for tremor assessment and suppression,' IEEE Transactions on Neural Systems and Rehabilitation Engineering , vol. 15, no. 3, pp. 367-378, 2007. - sourceExperimental — accuracy not yet measured
In “A Model of Tremor” (page unavailable), the paper describes its Benedict–Bordner/WFLC two-stage method as estimating motion with minimum time lag, rather than producing an excessively delayed result. Its Fig. 4 depicts the WOTAS exoskeleton, not a digitizing-tablet experiment. The paper does note that preliminary filtering for WFLC can introduce unwanted lag, but does not report the claimed result for the combined method. (LLM, conf=0.84)
Checked by a reasoning model — may be wrong.
plainSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Disputed
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The algorithm proposed in [2] was not able to filter effectively the tremor from intentional motion with the gains applied in our work.”
What the sources say
- Partly supportedcites [2] [2] E. Rocon, J. M. Belda-Lois, A. F. Ruiz, M. Manto, J. C. Moreno, and J. L. Pons, 'Design and validation of a rehabilitation robotic exoskeleton for tremor assessment and suppression,' IEEE Transactions on Neural Systems and Rehabilitation Engineering , vol. 15, no. 3, pp. 367-378, 2007. - sourceExperimental — accuracy not yet measured
Section A, “A Model of Tremor” (page unavailable), describes the proposed algorithm for separating voluntary and tremorous motion and reports that it performed efficiently on the patient data evaluated. The paper does not assess the gains used in the citing authors’ work, so it cannot support their claim that the algorithm failed to filter effectively under those gains. (LLM, conf=0.82)
plainSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Tremor frequency was around 7 . 5 Hz .”
plainSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The proposed method presented a better performance, particularly in terms of robustness to different data sets, being a good candidate for application in real pathological tremor compensation systems.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: unclear — It is unclear whether it could be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The insensitiveness of the proposed algorithm was greatly due to the assumption of Eq. (4), as the voluntary motion model, and the explicit consideration of process and measurement uncertainties.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
2 more claims
Create a free account to see all 35 claims and every source verdict.
- It may state two separate claims as one. 10
- The quote only partly states this claim. 1
- Probably not a substantial claim. 21
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“A similar approach has been recently described in [27].”
From Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data (2010) · cites this paper as [27]
Citations
Cites
Optimal State Estimation: Kalman, H∞, and Nonlinear Approaches
Simon, D. (2006) · doi:10.1002/0470045345 · Cited by 4,875
Optimal state estimation: Kalman, H [infinity], and nonlinear approaches
(2007) · Choice Reviews Online · doi:10.5860/choice.44-3334 · Cited by 2,108
Design and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression
Rocon, E., Belda-Lois, J. M., Ruiz, A. F., Manto, M., Moreno, J. C., & Pons, J. L. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903917 · Cited by 349
Adaptive cancelling of physiological tremor for improved precision in microsurgery
Riviere, C. N., Rader, R. S., & Thakor, N. V. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.686791 · Cited by 260 · cited 8×
“H, for instance, was set to 1, since it was the value used in [6].”
BackgroundMethodResultInfluentialAttenuation of pathological tremors by functional electrical stimulation I: Method
Prochazka, A., Elek, J., & Javidan, M. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368521 · Cited by 150
Bandlimited Multiple Fourier Linear Combiner for Real-time Tremor Compensation
Veluvolu, K. C., Tan, U. X., Latt, W. T., Shee, C. Y., & Ang, W. T. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4352922 · Cited by 83
Optimal digital filtering for tremor suppression
Gonzalez, J. G., Heredia, E. A., Rahman, T., Barner, K. E., & Arce, G. R. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.841338 · Cited by 56 · cited 1×
“[5] presented the application of digital”
Online pathological tremor characterization using extended Kalman filtering
Bó, A. P. L., Poignet, P., Widjaja, F., & Ang, W. T. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649516 · Cited by 23
Self-citation
Research works printed but not linked: 2
- elble, 1990, Tremor
- bo?, 2008, Online pathological tremor characterization with Kalman filtering, the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 1753
Cited by
Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data
Gallego, J., Rocon, E., Roa, J. O., Moreno, J., & Pons, J. L. (2010) · Sensors · doi:10.3390/s100302129 · Cited by 107
“A similar approach has been recently described in [27].”
Zhang, Z., Chu, B., Liu, Y., Li, Z., & Owens, D. H. (2022) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2022.3150301 · Cited by 37
“Extended-Kalman-filter-based control [12], proportional–integral (PI) control [13], neural oscillator control [14], and repetitive control (RC) [15] were also applied to improve the performance of FES-based tremor suppression.”
MethodTremor attenuation using FES-based joint stiffness control
Bó, A. P. L., & Poignet, P. (2010) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2010.5509560 · Cited by 26
“Here, we briefly present a method which was originally presented in our previous work [6], where a further presentation of this problem may be found.”
BackgroundMethodInfluentialSelf-citationOn the use of FES to attenuate tremor by modulating joint impedance
Bó, A. P. L., Coste, C., Poignet, P., Gény, C., & Fattal, C. (2011) · IEEE Conference on Decision and Control (CDC) · doi:10.1109/cdc.2011.6161352 · Cited by 8
“Here, we briefly present a method which was originally presented in our previous works [10], [11], where a deeper presentation of this problem may be found.”
MethodSelf-citationVision-Based Control of a Handheld Micromanipulator for Robot-Assisted Retinal Surgery
Becker, B. C. (2012) · KiltHub Repository · doi:10.1184/r1/6724241 · Cited by 3
Metadata
- Authors
- Bo, A. P. L., Poignet, P., & Geny, C.
- Affiliations
- Centre National de la Recherche Scientifique (FR)
- ORCID
- Poignet, Philippe
- Venue
- 2009 IEEE/RSJ International Conference on Intelligent Robots and Systems
- Publisher
- IEEE
- Published
- 2009-10
- Pages
- 55-60
- ISSN
- 2153-0858 (ISSN-L)
- OpenAlex
- W2108772193
- Cited by
- 9
- References
- 8
- Citations by year
- 2022: 1 · 2019: 1 · 2012: 1
- FWCI
- 4.33
- Type
- conference paper
- Language
- en
- Open access
- Licence
- other-oa
- Locations
- IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (publishedVersion)
- HAL (submittedVersion, other-oa)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Keywords
- pathological tremor, tremor compensation, extended Kalman filter, motor control
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-29, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-29 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Filtering voluntary motion for pathological tremor compensation. https://app.cochat.ai/papers/10.1109/iros.2009.5353972
- The paper
- Bo, A. P. L., Poignet, P., & Geny, C. (2009). Filtering voluntary motion for pathological tremor compensation. 2009 IEEE/RSJ International Conference on Intelligent Robots and Systems. https://doi.org/10.1109/iros.2009.5353972
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.