Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data
Open accessAbstract
This paper presents a two stage algorithm for real-time estimation of instantaneous tremor parameters from gyroscope recordings. Gyroscopes possess the advantage of providing directly joint rotational speed, overcoming the limitations of traditional tremor recording based on accelerometers. The proposed algorithm first extracts tremor patterns from raw angular data, and afterwards estimates its instantaneous amplitude and frequency. Real-time separation of voluntary and tremorous motion relies on their different frequency contents, whereas tremor modelling is based on an adaptive LMS algorithm and a Kalman filter. Tremor parameters will be employed to drive a neuroprosthesis for tremor suppression based on biomechanical loading.
Overview
CoChat Summary
## TL;DR
The paper proposes a two-stage real-time method for estimating pathological tremor amplitude and frequency from gyroscope data by first removing voluntary motion and then modeling the remaining tremor. The authors report that the best setup is a Critically Dampened Filter for voluntary motion tracking plus a WFLC-KF cascade for tremor estimation, with average amplitude error of 0.001 ± 0.002 rad/s.
## Key findings
- The authors report that pathological tremor generally lies in the 3–12 Hz band, while voluntary motion is mostly 0–2 Hz, and use this separation to justify their two-stage design.
- For voluntary motion tracking, the Critically Dampened Filter (CDF) with θ = 0.990 had the least KTE and was selected as the best stage-1 filter.
- For tremor modeling, the best performer was a cascade of WFLC + Kalman Filter, which the authors say produced the least FMSEd for all tasks; they state FMSEd was at least five times smaller than with the BMFLC.
- The final system achieved an average tremor amplitude estimation error of 0.001 ± 0.002 rad/s and frequency estimates that agreed with spectrograms.
## Methods
- The study recorded wrist flexion-extension using two solid-state gyroscopes placed on the distal forearm and hand; data were sampled at 1 kHz.
- The sample included 5 patients with tremor-related pathologies: 4 men and 1 woman, aged 48 to 74 years (average 63), recruited at Hôpital Erasme, Brussels.
- Patients performed four tasks: arms outstretched (30 s), finger to nose (30 s), rest (30 s), and pouring 20 cl of water.
- Main analysis: stage 1 compared BBF, CDF, and KF using KTE; stage 2 compared WFLC, BMFLC, and a WFLC-KF cascade using FMSEd.
## Limitations
- The paper states that the patient sample was small: 5 patients only.
- It also states that patients had different pathologies that cause tremor, but the text provided does not specify the diagnosis mix in the body beyond referring to Table 1, so generalizability across tremor types is limited by the evidence shown here.
- Evident from the text: the evaluation is based on a controlled set of short clinical/functional tasks, so the results may not extend to unconstrained daily-life movement; this is not explicitly stated as a limitation in the paper.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures








Verification
Claims the paper makes
4 Valid · 0 Not valid · 41 Not checked · 46 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“In [9], the authors have validated, both functionally and clinically, the concept of tremor suppression by means of a wearable robot that applies biomechanical loads in the upper limb.”
§ Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data > Introduction
What the sources say
- Supportedcites [9] Rocon, 2007, Design and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression, IEEE Trans. Neural Syst. Rehab. Eng source
The abstract reports clinical trials of WOTAS, a wearable exoskeleton that suppresses tremor through biomechanical loading. The Conclusion states that evaluation with 10 users validated the concept of tremor suppression through wearable robotics, and the Discussion reports reductions in tremor power. (LLM, conf=0.94)
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The data we collected yields that pathological tremors occur in a frequency band higher than voluntary motion, Figure 3, which is in agreement with the literature.”
§ Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data > Real-Time Estimation of Instantaneous Tremor Parameters
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Robust, precise and low delay estimation of tremor parameters is achieved: average amplitude estimation error is 0.001 ± 0.002 rad/s, and frequency estimation agrees with spectrograms.”
§ Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data > Introduction
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Robust, precise and low delay estimation of tremor parameters is achieved: average amplitude estimation error is 0.001 ± 0.002 rad/s, and frequency estimation agrees with spectrograms.”
§ Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data > Introduction
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This is demonstrated because performance in terms of FMSEd degrades the most during finger to nose and water glass tests, Table 3.”
§ Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data > Discussion
boostedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
40 more claims
Create a free account to see all 91 claims and every source verdict.
- Probably not a substantial claim. 24
- It may state two separate claims as one. 28
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In contrast, the Kalman filter (KF) computes the optimal solution by minimizing the covariance of the a posteriori estimation error [28] without any a priori assumption.”
From Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device (2022) · cites this paper as [28]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This is a relevant issue to face during the experimental validation of these technologies because of the high fluctuations in tremorous movements (44, 87), which even could be caused by subjective factors such as anxiety, distraction, or surprise (44, 112, 115, 116).”
From A Review on Wearable Technologies for Tremor Suppression (2021) · cites this paper as [112]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“voluntary [24,25] or artificially elicited [26] muscle contractions affect the severity of the tremor, by generating forces on the musculoskeletal system.”
From A neuroprosthesis for tremor management through the control of muscle co-contraction (2013) · cites this paper as [26]
The cited source says what the claim attributes to it.
In the citing paper
“Pathological tremor (patients with essential tremor and Parkinson disease) tend to display tremor with a dominant frequency [5–8].”
From Estimation of physiological tremor from accelerometers for real-time applications (2011) · cites this paper as [7]
Citations
Cites
Adaptive noise cancelling: Principles and applications
Widrow, B., Glover, J. R., McCool, J. M., Kaunitz, J., Williams, C. S., Hearn, R. H., Zeidler, J. R., Eugene Dong, Jr., & Goodlin, R. C. (1975) · Proceedings of the IEEE · doi:10.1109/proc.1975.10036 · Cited by 3,939
Estimation and Tracking: Principles, Techniques, and Software
Bar‐Shalom, Y., & Li, X. (1993) · Medical Entomology and Zoology · Cited by 2,203
Consensus Statement of the Movement Disorder Society on Tremor
Deuschl, G., Bain, P., & Brin, M. (2008) · Movement Disorders · doi:10.1002/mds.870131303 · Cited by 1,981
Mayagoitia, R. E., Nene, A. V., & Veltink, P. H. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00231-7 · Cited by 655
Measuring orientation of human body segments using miniature gyroscopes and accelerometers
Luinge, H. J., & Veltink, P. H. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345966 · Cited by 647
Roetenberg, D., Luinge, H. J., Baten, C. T. M., & Veltink, P. H. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847353 · Cited by 588
Wearable Robots: Biomechatronic Exoskeletons
(2008) · Wiley eBooks · doi:10.1002/9780470987667 · Cited by 517
Deuschl, G., Raethjen, J., Lindemann, M., & Krack, P. (2001) · Muscle & Nerve · doi:10.1002/mus.1063 · Cited by 508 · cited 5×
“Tremor is defined as a rhythmic oscillation of a body part [1].”
BackgroundInfluentialA practical gait analysis system using gyroscopes
Tong, K., & Granat, M. H. (1999) · Medical Engineering & Physics · doi:10.1016/s1350-4533(99)00030-2 · Cited by 460
Kolawole, M. O. (2002) · Radar Systems, Peak Detection and Tracking · doi:10.1016/b978-075065773-0/50016-9 · Cited by 453
Research works printed but not linked: 3
- Elble, R.J., and Koller, W.C. (1990). Tremor, The John Hopkins Univesity Press.
- Riviere, C.N. (Adaptive Suppression of Tremor for Improved Human Machine Control, 1981). Adaptive Suppression of Tremor for Improved Human Machine Control, PhD Thesis;.
- Rocon, E. (Reducción Activa de Temblor Patológico de Miembro Superior Mediante Exoesqueletos Robóticos, 2006). Reducción Activa de Temblor Patológico de Miembro Superior Mediante Exoesqueletos Robóticos, PhD Thesis;.
Cited by
Estimation of physiological tremor from accelerometers for real-time applications
Veluvolu, K. C., & Ang, W. T. (2011) · Sensors · doi:10.3390/s110303020 · Cited by 117
Towards remote monitoring of Parkinson’s disease tremor using wearable motion capture systems
Delrobaei, M., Memar, S., Pieterman, M., Stratton, T. W., McIsaac, K., & Jog, M. (2017) · Journal of the Neurological Sciences · doi:10.1016/j.jns.2017.11.004 · Cited by 120
Using Portable Transducers to Measure Tremor Severity
Elble, R. J., & McNames, J. (2016) · Tremor and Other Hyperkinetic Movements · doi:10.5334/tohm.320 · Cited by 95
Inertial sensors versus standard systems in gait analysis: a systematic review and meta-analysis
Petraglia, F., Scarcella, L., Pedrazzi, G., Brancato, L., Puers, R., & Costantino, C. (2019) · European Journal of Physical and Rehabilitation Medicine · doi:10.23736/s1973-9087.18.05306-6 · Cited by 107
Smartwatch for the analysis of rest tremor in patients with Parkinson's disease
Blanco, R. L., Velasco, M. A., Méndez-Guerrero, A., Romero, J. P., del Castillo, M. D., Serrano, J. I., Rocón, E., & Benito‐León, J. (2019) · Journal of the Neurological Sciences · doi:10.1016/j.jns.2019.04.011 · Cited by 95
Self-citation
Metadata
- Authors
- Gallego, J., Rocon, E., Roa, J. O., Moreno, J., & Pons, J. L.
- Affiliations
- Consejo Superior de Investigaciones Científicas (ES)
- Bioengineering Group, Consejo Superior de Investigaciones Científicas, CSIC, Ctra. Campo Real, km 0.2 La Poveda, 28500, Arganda del Rey, Spain
- Corresponding author
- Gallego, Juan
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2010-03-16
- Volume
- 10
- Issue
- 3
- Pages
- 2129-2149
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC3264472
- OpenAlex
- W2123467994
- Semantic Scholar
- 2d7696bbf96c68ab7179e0da6464f15709525163
- Cited by
- 107
- References
- 42
- Citations by year
- 2025: 6 · 2024: 5 · 2023: 5 · 2022: 2 · 2021: 13 · 2020: 10 · 2019: 9 · 2018: 8 · 2017: 6 · 2016: 10 · 2015: 2 · 2014: 6 · 2013: 12 · 2012: 6
- FWCI
- 4.84
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Keywords
- pathological tremor, gyroscope data, real-time estimation, adaptive LMS algorithm, Kalman filter, tremor suppression, neural prostheses, biomechanical loading
- MeSH terms
- Aged, Algorithms, Biomechanical Phenomena, Clothing, Female, Humans, Male, Middle Aged, Pattern Recognition, Automated, Rotation, Tremor, Wrist, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-09-29, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-09-29 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data. https://app.cochat.ai/papers/10.3390/s100302129
- The paper
- Gallego, J., Rocon, E., Roa, J. O., Moreno, J., & Pons, J. L. (2010). Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data. Sensors. https://doi.org/10.3390/s100302129
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.