Real-Time Robust Heart Rate Estimation From Wrist-Type PPG Signals Using Multiple Reference Adaptive Noise Cancellation
The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Heart rate (HR) monitoring using photoplethysmographic (PPG) signals recorded from wearers' wrist greatly facilitates design of wearable devices and maximizes user experience. However, placing PPG sensors in wrist causes much stronger and complicated motion artifacts (MA) due to loose interface between sensors and skin. Therefore, developing robust HR estimation algorithms for wrist-type PPG signals has significant commercial values. In this paper, we propose a robust HR estimation algorithm for wrist-type PPG signals using multiple reference adaptive noise cancellation (ANC) technique-termed here as "MURAD." The main challenge of using ANC for MA reduction is to devise a qualified reference noise signal (RNS) to the adaptive filter. We propose a novel solution by using four RNSs, namely, the three-axis accelerometer data and the difference signal between the two PPG signals. For each RNS, we get a different version of the cleaned PPG signal. Then, a set of probable HR values is estimated using all of the cleaned PPG signals, and then, the value that is closest to the estimated HR of the previous time window is chosen to be the HR estimate of the current window. Then, some peak verification techniques are employed to ensure accurate HR estimations. The proposed technique gives lower average absolute error compared to state-of-the art methods. So, MURAD method provides a promising solution to the challenge of HR monitoring using PPG in wearable devices during severe MA conditions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Zhang, Z. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2406332 · Cited by 388 · cited 11×
“As for the testing datasets, MURAD provides mean absolute error of 2.24 BPM, whereas JOSS reported this error to be 3.21 BPM.”
BackgroundMethodResultInfluentialSkin photoplethysmography — a review
Kamal, A. A. R., Harness, J. B., Irving, G., & Mearns, A. J. (1989) · Computer Methods and Programs in Biomedicine · doi:10.1016/0169-2607(89)90159-4 · Cited by 328
Cited by
Heart Rate Estimation From Wrist-Worn Photoplethysmography: A Review
Biswas, D., Simoes-Capela, N., Van Hoof, C., & Van Helleputte, N. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2914166 · Cited by 258
Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths
Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., & Groenendaal, W. (2019) · Sensors · doi:10.3390/s19030673 · Cited by 140
“These properties will be exploited in the proposed algorithm framework in Section 4.”
BackgroundHeart rate tracking in photoplethysmography signals affected by motion artifacts: a review
Ismail, S., Akram, U., & Siddiqi, I. (2021) · EURASIP Journal on Advances in Signal Processing · doi:10.1186/s13634-020-00714-2 · Cited by 101
Xu, K., Jiang, X., Lin, S., Dai, C., & Chen, W. (2020) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2020.2988097 · Cited by 37
Mahloko, L., & Adebesin, F. (2020) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-45002-1_9 · Cited by 23
Metadata
- Authors
- Chowdhury, S. S., Hyder, R., Hafiz, M. S. B., & Haque, M. A.
- Venue
- IEEE Journal of Biomedical and Health Informatics
- Published
- 2018-03
- Volume
- 22
- Issue
- 2
- Pages
- 450-459
- ISSN
- 2168-2194 (ISSN-L)
- 2168-2208
- OpenAlex
- W2551912285
- Cited by
- 102
- References
- 3
- FWCI
- 3.80
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Real-Time Robust Heart Rate Estimation From Wrist-Type PPG Signals Using Multiple Reference Adaptive Noise Cancellation. https://app.cochat.ai/papers/10.1109/jbhi.2016.2632201
- The paper
- Chowdhury, S. S., Hyder, R., Hafiz, M. S. B., & Haque, M. A. (2018). Real-Time Robust Heart Rate Estimation From Wrist-Type PPG Signals Using Multiple Reference Adaptive Noise Cancellation. IEEE Journal of Biomedical and Health Informatics. https://doi.org/10.1109/jbhi.2016.2632201
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.